The C Scope Visualization system is intended to visualize an interior cavity of the body.
Device Story
C Scope Visualization System is an arthroscope for minimally-invasive procedures. Input: CMOS camera sensor in a single-use rigid scope captures visual data; optical fiber provides illumination. Transformation: Scope connects to a reusable handpiece; handpiece connects via USB to a tablet running ImageClear software for image/video processing, display, and data management. Output: Live video and captured images on the tablet screen. Used in hospitals or doctor's offices by trained physicians. The system aids in visualizing anatomical structures during arthroscopy. Benefits include real-time visualization for diagnostic or operative procedures. The scope includes an integral drape for a sterile gripping surface.
Clinical Evidence
No clinical studies were required to demonstrate substantial equivalence. Evidence is based on bench-only testing, including physical integrity, functionality, mechanical durability, electrical safety (IEC 60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), biocompatibility (ISO 10993 series), and sterilization validation (ISO 11135-1). Human factors validation was performed via two formative studies and one summative study.
Technological Characteristics
System includes a single-use rigid scope (CMOS sensor, optical fiber illumination), reusable handpiece, and tablet. Materials: medical-grade plastics and stainless steel. Connectivity: USB. Sterilization: Ethylene Oxide (EO) for scope kit. Software: ImageClear for image processing. Electrical safety: IEC 60601-1, IEC 60601-2-18. EMC: IEC 60601-1-2. Applied part: Type BF.
Indications for Use
Indicated for use by a trained physician to provide illumination and visualization in arthroscopic procedures of an interior cavity of the body through a surgical opening.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
January 12, 2021
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
CIT Ortho, LLC % Prithul Bom Most Responsible Person Regulatory Technology Services 1000 Westgate Drive, Suite #510k Saint Paul. Minnesota 55114
Re: K203526/S001
Trade/Device Name: C Scope™ Visualization System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: November 11, 2020 Received: January 8, 2021
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203526
Device Name C Scope Visualization System
Indications for Use (Describe)
The C Scope Visualization System is indicated to be used by a trained physician to provide illumination and visualization in arthroscopic procedures of an interior cavity of the body through a surgical opening.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for CIT Ortho. The logo features a stylized image of a knee joint inside of a circle on the left. To the right of the image is the text "CIT Ortho", with "CIT" in black and "Ortho" in blue.
### 510(k) SUMMARY
| Applicant: | CIT Ortho, LLC<br>26202 Detroit Rd. Ste. 340<br>Westlake, Ohio 44145 | |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Megan Lecavalier<br>Chief Operating Officer<br>Phone: 440-772-1421<br>Fax: 440-772-1422<br>Email: mlecavalier@clearimg.com | |
| Date Summary Prepared: | December 28, 2020 | |
| Trade Name: | C Scope™ Visualization System | |
| Common Name: | Arthroscope | |
| Classification Name: | Arthroscope, Class II | |
| Product Code: | HRX | |
| Regulatory Class: | Class II | |
| Regulation Number: | 21 CFR §888.1100 | |
| Predicate Device: | NeedleCam HD Visualization System [K143705]<br>BioVision Technologies. Golden, CO | |
| | Subject Device | Predicate Device |
| Feature | C Scope Visualization System | NeedleCam HD Visualization System |
| FDA Classification | Class II | Class II |
| Product Code(s) | HRX | HRX/GCJ |
| Regulation Number | 21 CFR §888.1100 | 21 CFR §888.1100/ 876.1500 |
| Regulation Name | Arthroscope | Arthroscope/ Laparoscope, General & Plastic Surgery |
| Indications for Use / Intended Use | Intended Use<br>The C Scope Visualization System is<br>intended to visualize an interior<br>cavity of the body. | Intended Use<br>The Needle Cam HD™ Visualization System is<br>intended to visualize an<br>interior cavity of the body. |
| | Indications for use<br>The C Scope Visualization System is<br>indicated to be used by a trained<br>physician to provide illumination<br>and visualization in arthroscopic<br>procedures of an interior cavity of<br>the body through a surgical<br>opening. | Indications for Use<br>Needle Cam HD™ Visualization<br>System is indicated to be used<br>by a trained physician to<br>provide illumination and<br>visualization of an interior<br>cavity of the body through a<br>natural or surgical opening in<br>diagnostic and operative<br>arthroscopic and endoscopic<br>procedures. Examples of<br>surgical use include but are not<br>limited to procedures on the<br>knee, shoulder, ankle, elbow,<br>wrist, shoulder,<br>temporomandibular joint<br>(TMJ), spinal, ophthalmic, ENT,<br>and cervix. |
| Where Used | Hospital / Doctor's Office | Hospital / Doctor's office |
| Device Description | The C Scope Visualization System<br>Includes the following components:<br>Tablet with custom software for<br>image/video processing and<br>display, capture and USB data | NeedleCam HD Visualization<br>System includes the following<br>components:<br>An Image Capture Box with<br>custom software for<br>image/video processing<br>display, capture, and USB |
| Feature | Subject Device<br>C Scope Visualization System | Predicate Device<br>NeedleCam HD Visualization System |
| | management, and external power supply Reusable handpiece with LED light source. Disposable single-use sterile kit including a semi-rigid scope with integral drape, and supplemental instruments including a cannula, trocar/obturator, and syringe and tubing for flushing. | data management, and external power supply.<br>Video Outputs for external monitor Reusable Camera-handpiece with an LED light source A disposable single-use sterile kit including a semi-rigid fiberoptic scope with integral drape, and supplemental instruments including a cannula, trocar, obturator, and cannula plug. |
| Single Use/Reusable | Tablet and Handpiece: Reusable.<br>Scope and instruments: Sterile, Single Use. | Image capture box and camera handpiece: Reusable.<br>Scopes & instruments: Sterile, Single Use |
| Sterilization | Scope kit - Ethylene Oxide (EO) sterilization | Scope kit - Ethylene Oxide (EO) sterilization |
| Sterility, How Supplied | Scope Kit - sterile in PETG tray w/ sealed LLDPE header pouch | Scope Kit – sterile in PETG tray w/ sealed Tyvek lid. |
| | Tablet, Handpiece - non-sterile in cardboard packaging with protective padding | Camera Handpiece and Image Capture Box- non-sterile in cardboard packaging with protective padding |
| Image Acquisition | Image acquisition is achieved through the camera, which is a CMOS sensor located in the scope. | Image acquisition is achieved through the camera, which is a CCD sensor located in the handpiece. |
| Connectivity | The scope w/ camera connects to the handpiece with a quick release connector. The handpiece (with illumination source) and cable connects to the Tablet through USB connection. | The scope connects to the camera handpiece with a quick release connection. The handpiece (with illumination source) cable connects to the Image Capture Box through a custom electrical connection. |
| Image Display(s) | All-in-one Tablet | External display connection |
| Data Storage | Images and video automatically stored to Tablet.<br>Images and video can be exported to USB device. | Image and video storage to USB device. |
| Type of Light Source | LED | LED |
| Feature | Subject Device | Predicate Device |
| | C Scope Visualization System | NeedleCam HD Visualization System |
| Endoscope coupler | Built-in | Built-in |
| Materials | Commonly used medical-grade plastics and stainless steel | Commonly used medical-grade plastics and stainless steel |
### Device Description:
The C Scope Visualization System is an arthroscope that allows users to illuminate and visualize patient anatomy during minimally-invasive arthroscopic procedures. The system consists of a single-use rigid Scope, a reusable Handpiece and a Tablet for viewing the images. The Scope contains a CMOS camera chip to visualize and an optical fiber to illuminate anatomical structures. The Scope includes an integral drape which covers the Handpiece to provide a sterile gripping surface for the user. The Scope is packaged in a sterile C Scope Disposable Scope Kit containing single use items syringe, tubing, cannula, and trocar to aid in the procedure. The Handpiece provides control of the direction and depth of the Scope and allows the user to capture images and videos. The reusable tablet Viewing Tablet is loaded with custom ImageClear software which provides image and video display and data management.
### Indications for Use:
The C Scope Visualization System is indicated to be used by a trained physician to provide illumination and visualization in arthroscopic procedures of an interior cavity of the body through a surgical opening.
### Intended Use:
The C Scope Visualization system is intended to visualize an interior cavity of the body.
### Technology Characteristics:
The technological characteristics of the C Scope Visualization System are similar to its predicate device –
{4}------------------------------------------------
NeedleCam HD Visualization System (K143705). The indications and contraindications listed in this submission are in congruence between these devices. The C Scope Visualization system utilizes a disposable arthroscope, included in a sterile kit with supplemental instruments. The disposable scope connects to a reusable handpiece to display a live image on a viewing tablet with custom software for image processing and capture.
# Comparison to Predicate Device:
The fundamental scientific technology of the C Scope Visualization System and the previously cleared predicate device, NeedleCam HD Visualization System [K143705], are the same. The table below shows relevant similarities and differences.
Table 1- Predicate Comparison
{5}------------------------------------------------
{6}------------------------------------------------
Based on the evaluation of the performance characteristics, construction, and indications of use, CIT Ortho has concluded that the C Scope Visualization System is substantially equivalent to the predicate device listed in this submission.
# Performance Data Summary:
Determination of substantial equivalence is based on an assessment of non-clinical performance test data. The C Scope Visualization System has successfully completed the following relevant performance testing to demonstrate substantial equivalence. Testing was performed to evaluate physical integrity, functionality, and performance of the system.
The results of these tests provide reasonable assurance that the C Scope Visualization System has been designed and tested to assure conformance to the requirements for its intended use and indications for use. No new safety or performance issues were raised during the testing; therefore, this device is considered to be substantially equivalent to the predicate device.
# Environmental and Packaging
The sterile packaging for the Disposable Scope Kit of the C Scope Visualization system was verified for transportation and environmental conditions according to ASTM D4169-16 DC-13 and ASTM D4332-14. Seal formation for the sterile LLDPE header pouch was validated per ISO 11607-2 2019 (E).
# Mechanical and Durability
Testing was performed to verify the mechanical integrity and durability of the C Scope Visualization system. Testing also included ingress protection testing per IEC/EN 60529:2013 and luer testing per ISO 80369-7:2016. These tests demonstrated that the system met all product requirements.
# Human Factors
Human factors validations were performed per Guidance for Industry and FDA Staff "Applying Human Factors and Usability Engineering to Medical Devices." Two formative studies and one summative validation study were conducted utilizing the complete C Scope Visualization System in a productionequivalent state.
# Electrical Safety and Electromagnetic Compatibility
Electrical safety and Electromagnetic Compatibility (EMC) testing were conducted on the C Scope Visualization System consisting of the Tablet, Handpiece, and Disposable Scope. The system complies with the IEC 60601-1:2012 and IEC 60601-2-18:2009 standards for safety with the Disposable Scope being a type BF applied part. The system complies with the IEC 60601-1-2:2014 standard for EMC. Immunity of the system to RFID readers was verified using AIM 7351731.
# Software Validation and Verification Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care
CIT Ortho C Scope Visualization System Page 5- 4
{7}------------------------------------------------
provider. Software validation was conducted according to AAMI/IEC 62304:2006/A1:2016 and FDA's Guidance for Industry and FDA Staff "General Principles of Software Validation."
### Biocompatibility Testing
The patient-contacting components of the device are categorized as Externally Communicating Device, Tissue and Bone, Limited Contact (< 24 hours). Per ISO 10993-1:2018, the following testing was conducted:
- ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity .
- ISO 10993-10:2013 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
- . ISO 10993-11:2018 Biological evaluation of medical devices Part 11: Tests for systemic toxicity
- ISO 10993-17:2009 Biological evaluation of medical devices Part 17: Establishment of allowable limits for leachable substances
- ISO 10993-18:2020 Biological evaluation of medical devices Part 18: Chemical characterization of ● medical device materials within a risk management process
Documentation for the biocompatibility testing was provided as recommended by Guidance for Industry and FDA Staff, "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.'"
# Sterilization and Shelf Life
The C Scope Visualization System Disposable Scope Kit sterilization process, using 100% Ethylene Oxide (EO), has been validated in accordance with ISO 11135-1:2014 to achieve a SAL of 10 °. EO and ECH residuals were below the limits specified in ISO 10993-7:2008 Amd 1:2019 (E). Both baseline and accelerated shelf-life testing were conducted demonstrating the device will perform as intended to support the proposed shelf-life. Accelerated aging testing was carried out according to ASTM F1980:2016.
Reprocessing of the reusable C Scope Handpiece was validated through a Manual Cleaning Validation, including Simulated Use, and an Intermediate Level Disinfection Validation. These studies were conducted in accordance with AAMI TIR12:2010, AAMI TIR30:2011 (R2016), and "Guidance for Industry and FDA Staff – Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
# Animal Testing
No animal studies were required to demonstrate substantial equivalence.
# Clinical Testing
No clinical studies were required to demonstrate substantial equivalence.
# Conclusion
CIT Ortho concludes through a review of the benchtop assessments, the comparison of the device classification, intended use, operating principle, technological characteristics, sterility, and biocompatibility that the C Scope Visualization System is substantially equivalent to the predicate device and is as safe, as effective, and performs as well as or better than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.