K200817 · Truabutment, Inc. · DZE · Oct 7, 2020 · Dental
Device Facts
Record ID
K200817
Device Name
URIS OMNI Narrow System & Prosthetic
Applicant
Truabutment, Inc.
Product Code
DZE · Dental
Decision Date
Oct 7, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
URIS OMNI Narrow System is indicated for use in the treatment of missing maxillary lateral incisors or the mandbular central and lateral incisors, in support of single or multiple-unit restorations including: cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. The URIS OMNI Prosthetic abutments are intended for use with URIS OMNI dental implants to provide support for prosthetic restorations such as crowns, bridges, or over-dentures. All digitally designed abutments and/or coping for use with URIS OMNI Prosthetic abutments are intended to be sent to a TruAbutment-validated milling center for manufacture.
Device Story
Dental implant system comprising fixtures and prosthetic abutments; fixtures made of unalloyed titanium (ASTM F67) with SLA surface treatment; prosthetic components made of titanium alloy (ASTM F136). Inputs include patient clinical data for CAD/CAM design of custom abutments (URIS DS, URIS Base). Device transforms inputs via 3Shape CAD/CAM software into patient-specific designs manufactured at validated milling centers. Used in dental clinics by clinicians to support crowns, bridges, or overdentures. Output provides structural support for prosthetic restorations, restoring chewing function. Benefits include functional and esthetic rehabilitation of edentulous sites.
Clinical Evidence
No clinical data included. Evidence consists of bench testing: biocompatibility (ISO 10993-1), fatigue testing (ISO 14801:2016), sterilization validation (ISO 11137, ISO 17665), shelf-life testing (ASTM F1980), and material analysis (SEM/EDS).
Indicated for treatment of missing maxillary lateral incisors or mandibular central and lateral incisors in partially or fully edentulous patients. Supports single or multiple-unit restorations (cemented, screw-retained, overdenture) and fixed bridgework. Intended for delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 7, 2020
TruAbutment Inc. Chris Choi Director 17742 Cowan Irvine, California 92614
Re: K200817
Trade/Device Name: URIS OMNI Narrow System & Prosthetic Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: June 5. 2020 Received: September 8, 2020
Dear Chris Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200817
Device Name URIS OMNI Narrow System & Prosthetic
#### Indications for Use (Describe)
URIS OMNI Narrow System is indicated for use in the treatment of missing maxillary lateral incisors or the mandbular central and lateral incisors, in support of single or multiple-unit restorations including: cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading.
The URIS OMNI Prosthetic abutments are intended for use with URIS OMNI dental implants to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
All digitally designed abutments and/or coping for use with URIS OMNI Prosthetic abutments are intended to be sent to a TruAbutment-validated milling center for manufacture.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|---------------------------------------------|
| <div><span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirement of 21 CFR part 807.92
#### Submitter:
TruAbutment Inc. Eunjin Jang 17742 Cowan, Irvine CA 92614 USA Email: eunjin.jang@truabutment.com
Phone: 1-714-956-1488
#### Device Information:
Device Name: URIS OMNI Narrow System & Prosthetic Classification Name: Endosseous Dental Implant Classification: Class II Primary Product Code: DZE Secondary Product Code: NHA Regulation number: 21 CFR 872.3640 Date Prepared: 10/06/2020
### Predicate Device
- Primary Predicate Device:
- URIS OMNI System (K172100) by TruAbutment Korea Co., Ltd. -
- Reference Devices:
- -Nobel Active 3.0 (K102436) by Nobel Biocare.
- OSSTEM Implant System (K161689) by Osstem Implant Co., Ltd . -
- -Hiossen Prosthetic System (K140507) by Osstem Implant Co., Ltd .
- Osstem Abutment System (K182091) by Osstem Implant Co., Ltd . -
- Multi Angled Abutment (K123755) by Osstem Implant Co., Ltd . -
- -SMART builder System (K120951) by Osstem Implant Co., Ltd.
- AnyOne Internal Implant System (K123988) by Megagen Implant Co., Ltd. -
- ET US SS Prosthetic System (K160670) by Osstem Implant Co., Ltd. -
- Preat Abutments (K183518) by Preat Corporation. -
- Oneplant Dental Implant System (K081748) by WARANTEC Implant Co., Ltd. -
- InCoris Zi (K123664) by Sirona Dental Systems GmbH. -
- -RelyX Unicem 2Automix (K100756) by 3M ESPE
- TruAbutment DS (K183106) by TruAbutment Korea Co., Ltd. -
#### Official Correspondent
TruAbutment Inc. Chris Choi 17742 Cowan, Irvine CA 92614 USA Email: chris.choi@truabutment.com Phone: 1-714-956-1488
{4}------------------------------------------------
# Device Description
URIS OMNI Narrow System fixtures are dental implants made of Unalloyed Titanium, grade 4 (ASTM F67) intended for use in the treatment of missing maxillary lateral incisors or the mandibular central and lateral incisors. The surface is SLA (Sandblasted, Large grit and Acid etched) treated and is provided sterile. It consists of two implant lines, the OMNI Straight and the OMNI Tapered, with corresponding cover screws, healing abutments and prosthetic abutments. The OMNI Tapered implant has a tapered wall with a single thread design. The OMNI Straight implant has straight wall with smaller threading at the coronal end, and bigger threading at the apical end. Both implant lines have Narrow (Ø 3.15 mm) platform sizes. Both implant lines share the following diameters and lengths.
Ø 3.15 x 10, 11.5, 13, 14.5mm (L)
URIS Prosthetic System is made of titanium alloy (Ti-6Al-4V ELI) intended for use as an aid in prosthetic restoration. It consists of Ball Abutment, Retainer, T LOC Straight Abutment, T Loc Titanium Cap, Multi-Unit Straight Abutment, Multi-Unit Angled Abutment, Multi-Unit Healing Cap, Multi-Unit Ti Cylinder, Multi-unit temporary cylinder, Multi-Unit Cylinder screw, URIS DS, URIS Base. No additional angulation is to be included in the when using a coping or cylinder (i.e., Multi-unit Ti Cylinder, Multi-unit Temporary Cylinder, Multi-unit Base) with any of the Multi-unit Abutments.
| Device Component | Diameters (Ø) | Lengths | Angulation |
|-------------------------------------|---------------|---------------------------------------------------|------------|
| OMNI Straight Fixture | 3.15mm | 10~14.5mm | |
| OMNI Tapered Fixture | 3.15mm | 10~14.5mm | |
| Ball Abutment | 3.5mm | Cuff Height: 1.0~6.0mm | |
| Retainer Cap | 5.1mm | 3.9mm | |
| Retainer | 5.1mm | 2.1mm | |
| T LOC Straight Abutment | 3.8mm | Cuff Height: 1.0~6.0mm | |
| T LOC Titanium Cap | 5.4mm | 2.3mm | |
| Multi-Unit Straight Abutment | 5.0mm | Cuff Height: 1.0mm~6.0mm | |
| Multi-Unit Angled Abutment | | Cuff Height: 3.0mm~5.0mm | 17° |
| | | Cuff Height: 4.0mm~6.0mm | 29.5° |
| Multi-Unit Healing Cap | 5.1mm | 4.5mm | |
| Multi-Unit Ti Cylinder | 5.0mm | 5.0mm | |
| Multi-unit temporary cylinder | 5.0mm | 12mm | |
| Multi-Unit Base | 5.0mm | 4.35/7.35mm | |
| Multi-unit cylinder screw | 1.6mm | 3.3mm | |
| URIS DS | Ø3.8~ Ø 5.5mm | 6~11mm | 0~25° |
| URIS Base – Titanium Base Component | 4.0mm/4.3mm | Cuff Height: 1.0/2.0mm<br>Post Height : 3.5/5.5mm | |
Cover screw and healing abutment are anodized in yellow or green.
Fixtures and cover screw are provided sterile and other prosthetics are provided non-sterile. All non-sterile products must be sterilized by end users before use.
{5}------------------------------------------------
URIS Base consists of a two-piece abutment, where the titanium base is a pre-manufactured abutment that will be used to support a CAD/CAM designed superstructure (the second part of the two-piece abutment) that composes the final abutment. URIS Base is made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications. It is compatible with the following systems:
## Dental Implants
- URIS OMNI System Implants (K172100) 3.5, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5mm
- . URIS OMNI Narrow System Implants (Proposed) 3.15 mm
### Raw material blanks
- · InCoris Zi (ZrO2) by Sirona Dental Systems GmbH, L size blanks, cleared under K123664.
### Cement
- RelyX Unicem 2Automix by 3M ESPE, cleared under K100756.
All zirconia superstructure that composes the final abutment must be designed and milled through the 3 shape CAD/CAM System, by TruAbutment validated milling center, according to the prosthetic planning and patient clinical situation. The superstructure is cemented to the TruBase S in the lab. Use "RelyX Unicem 2Automix" as an adhesive extraorally to connect.
URIS Base is provided non-sterile therefore must be sterilized after the cementation of the zirconia superstructure on the URIS Base.
Design Limitation for Zirconia superstructure
| Design parameter | Design Limit |
|------------------------------------------------------------------------------------------------------|---------------|
| Minimum and Maximum abutment angle | 0~15° |
| Minimum and Maximum Cuff Height | 0.5~5 mm |
| Minimum and Maximum diameter at<br>abutment/implant interface | ∅5.0mm~∅8.0mm |
| Minimum Thickness | 0.4 mm |
| Minimum and Maximum length of<br>abutment post (length above the abutment<br>collar/gingival height) | 4 ~6 mm |
{6}------------------------------------------------
URIS DS abutment as a patient matched titanium abutment compatible with both URIS OMNI System (K172100) and URIS OMNI Narrow System (subject).
### Design Limitation for URIS DS
| Design parameter | Subject System<br>(K200817)<br>Design Limit | URIS OMNI System<br>(K172100)<br>Design Limit |
|------------------------------------------------------------------------------------------------------|---------------------------------------------|-----------------------------------------------|
| Minimum and Maximum Gingival<br>Height | 0.5~4mm | 0.5~4mm |
| Minimum and Maximum diameter at<br>abutment/implant interface | $Ø$ 3.8~ $Ø$ 5.5 | $Ø$ 3.8~ $Ø$ 5.5 |
| Minimum and Maximum length of<br>abutment | 6~11mm | 6~11mm |
| Minimum and Maximum length of<br>abutment post (length above the<br>abutment collar/gingival height) | 4~8mm | 4~8mm |
| Minimum wall thickness at<br>abutment/implant interface | 0.4mm | 0.4mm |
| Minimum and Maximum abutment<br>angle | 0~25° | 0~25° |
#### Indication for Use
URIS OMNI Narrow System is indicated for use in the treatment of missing maxillary lateral incisors or the mandibular central and lateral incisors, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading.
The URIS OMNI Prosthetic abutments are intended for use with URIS OMNI dental implants to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
All digitally designed abutments and/or coping for use with URIS OMNI Prosthetic abutments are intended to be sent to a TruAbutment-validated milling center for manufacture
## Summary of Technological Characteristics
The subject device is substantially equivalent to the currently cleared devices. They are substantially equivalent in intended use, material, design, dimension, connection, functions and surface treatments. Comparison demonstrating Substantial Equivalence follows at the end of this section.
{7}------------------------------------------------
# URIS Fixture
| | Subject Device | Primary<br>Predicate Device | Reference Devices |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K<br>Number | K200817 | K172100 | K102436 |
| Device<br>Name | URIS OMNI Narrow<br>System & Prosthetic | URIS OMNI System | Nobel Active 3.0 |
| Manufact<br>urer | TruAbutment Korea Co.,<br>Ltd | TruAbutment Korea Co.,<br>Ltd | Nobel Biocare |
| Indication<br>for Use | URIS OMNI Narrow<br>System is indicated for use<br>in the treatment of missing<br>maxillary lateral incisors<br>or the mandibular central<br>and lateral incisors, in<br>support of single or<br>multiple-unit restorations<br>including; cemented<br>retained, screw retained,<br>or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is intended<br>for delayed loading.<br>The URIS OMNI<br>Prosthetic abutments are<br>intended for use with<br>URIS OMNI dental<br>implants to provide<br>support for prosthetic<br>restorations such as<br>crowns, bridges, or over-<br>dentures.<br>All digitally designed<br>abutments and/or coping<br>for use with URIS OMNI<br>Prosthetic abutments are<br>intended to be sent to a<br>TruAbutment-validated<br>milling center for<br>manufacture | URIS OMNI System is<br>indicated for use in<br>partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is intended<br>for delayed loading. | The NobelActive<br>3.0mm implant is<br>indicated for use in the<br>treatment of missing<br>maxillary lateral incisors<br>or the mandibular<br>central and lateral<br>incisors to support<br>prosthetic devices, such as<br>artificial teeth,in order to<br>restore chewing<br>function in partially<br>edentulous patients. The<br>NobelActive 3.0<br>implants may be put into<br>immediate function<br>provided that stability<br>requirements detailed in<br>the manual are satisfied. |
| Design | Image: Subject Device Design | Image: Primary Predicate Device Design | Image: Reference Devices Design |
| | Subject Device | Primary<br>Predicate Device | Reference Devices |
| Structure | - Internal Hex- connected<br>- Submerged Fixture | - Internal Hex- connected<br>- Submerged Fixture | -Internal Hex<br>- Bone Level Implant |
| Body<br>Diameter<br>(D) | 3.15mm | 3.5, 4.0, 4.5, 5.0, 5.5, 6.0,<br>6.5 mm | 3.0mm |
| Length<br>(mm) | 10, 11.5, 13, 14.5mm | 7, 8.5, 10, 11.5, 13,<br>14.5mm | 10, 11.5, 13, 15mm |
| Material<br>of Fixture | CP Ti Grade 4 (ASTM<br>F67) | CP Ti Grade 4 (ASTM<br>F67) | Pure Titanium Grade 4<br>(ASTM F67) |
| Surface | Sand-blasted, Large grit,<br>Acid-etched<br>(S.L.A) | Sand-blasted, Large grit,<br>Acid-etched<br>(S.L.A) | TiUnite |
| Sterilizatio<br>n | Gamma Sterilization | Gamma Sterilization | Radiation Sterile |
| Shelf Life | 5years | 5years | N/A |
| Implant<br>Body<br>Features | Threaded | Threaded | Threaded |
| Product<br>Code | DZE | DZE | DZE |
| SE | The subject device is substantially equivalent in indications and design principles to<br>the primary predicate device and the reference devices listed above. Provided tables<br>are comparing the Indications for Use Statements and the technological<br>characteristics of the subject device, the primary predicate device, and the reference<br>devices.<br>The Indications for Use Statement (IFUS) for subject device abutment is<br>substantially equivalent in intended use to the primary predicate device K172100,<br>and the reference devices K102436. All are intended for use with endosseous dental<br>implants in the maxilla and mandible to provide functional and esthetic rehabilitation<br>of the edentulous maxilla and mandible. The minor differences between the IFUS for<br>the subject device and the primary predicate include:<br>the subject device IFUS includes the term "URIS OMNI Narrow System &<br>Prosthetic is indicated for use in the treatment of missing maxillary lateral incisors or<br>the mandibular central and lateral incisors" and the primary predicate IFUS does not.<br>The other minor differences are related to the specific device names, validated<br>milling centers, and the compatible OEM implant lines. None of these minor<br>differences impact substantial equivalence because both IFUS express equivalent<br>intended use to facilitate dental prosthetic restorations, and the indications are<br>expressed equivalently using different specific wording. | | |
{8}------------------------------------------------
{9}------------------------------------------------
# URIS Abutments
| | Subject Device | Predicate Devices | |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Part Name | Ball Abutment | Stud Abutment | |
| Design | | | |
| Applicant | TruAbutment Korea Co., Ltd | Osstem Implant Co., Ltd. | |
| Trade Name | URIS OMNI Narrow System &<br>Prosthetic | OSSTEM Implant System | |
| 510(K) No. | K200817 | K161689 | |
| Classification<br>Name | Endosseous Dental Implant<br>Abutments(872.3630) | Endosseous Dental Implant<br>Abutments(872.3630) | |
| Product Code | NHA | NHA | |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | |
| Indication<br>For Use/<br>Intended Use | Ball Abutment is intended for use in<br>conjunction with the fixture in<br>partially or fully edentulous<br>mandibles and maxillae, in support of<br>multiple-unit cement retained<br>restorations…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.