Osstem Abutment System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. · Link Abutment for CEREC The Link Abutment for CEREC is titanium alloy abutments placed onto OSSTEM dental implants to provide support for customized prosthetic restorations. Link Abutment for CEREC is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. All digitally designed copings and/or crowns for use with the Link abutment for CEREC is to be scanned using Sirona CEREC AC or CEREC AF or CEREC AI, designed using Sirona inLab software (Version 3.65) or Sirona CEREC Software (Version 4.2) and manufactured using a Sirona CEREC or inLab MC X or MC XL milling unit. CAD/CAM manufacturing/milling occurs at dental laboratories per the design limitations of the Sirona CEREC.
Device Story
System comprises various titanium alloy and Co-Cr-Mo dental abutments and components; used to connect dental implants to prosthetic restorations (crowns, bridges, overdentures). Components include transfer, angled, temporary, stud, and port abutments, plus cylinders and screws. Link Abutment for CEREC integrates with Sirona CAD/CAM systems for customized restorations. Used in dental clinics by clinicians; CAD/CAM manufacturing occurs in dental laboratories. Provides structural support for prosthetics; enables path adjustment for implants; maintains esthetics during temporary phases. Benefits patients by restoring chewing function and dental esthetics.
Clinical Evidence
No clinical studies submitted. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993), steam sterilization validation (ISO 17665-1), mechanical fatigue testing (ISO 14801), and retention testing for dental attachments.
Technological Characteristics
Materials: Titanium Alloy (ASTM F136), Titanium Gr. 3 (ASTM F67), Co-Cr-Mo Alloy, POM, Nylon, PC. Components include various abutments, cylinders, and screws. Connectivity: Compatible with Sirona CEREC CAD/CAM systems. Sterilization: Provided non-sterile; validated for steam sterilization (ISO 17665-1).
Indications for Use
Indicated for use with dental implants to support prosthetic restorations (crowns, bridges, overdentures). Link Abutment for CEREC indicated for screw-retained or cement-retained single tooth and bridge restorations.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
July 12, 2019
Osstem Implant Co., Ltd. % Peter Lee QA/RA Manager Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, Pennsylvania 19030
Re: K182091
Trade/Device Name: Osstem Abutment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: June 12, 2019 Received: June 13, 2019
Dear Peter Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The logo includes the company name in bold, black font, with the word "OSSTEM" also appearing in orange next to the company name. The contact information includes the company's address, telephone number (+82 51 850 2500), fax number (+82 51 861 4693), and website (www.osstem.com).
510(k) Number: K182091
Device Name: Osstem Abutment System
### Indication for Use:
Osstem Abutment System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.
### · Link Abutment for CEREC
The Link Abutment for CEREC is titanium alloy abutments placed onto OSSTEM dental implants to provide support for customized prosthetic restorations. Link Abutment for CEREC is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. All digitally designed copings and/or crowns for use with the Link abutment for CEREC is to be scanned using Sirona CEREC AC or CEREC AF or CEREC AI, designed using Sirona inLab software (Version 3.65) or Sirona CEREC Software (Version 4.2) and manufactured using a Sirona CEREC or inLab MC X or MC XL milling unit. CAD/CAM manufacturing/milling occurs at dental laboratories per the design limitations of the Sirona CEREC.
X Prescription Use (21 CFR 801 Subpart D) Over-The Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
# 510(k) Summary
# K182091
Date: July 12, 2019
### 1. Administrative Information
| Submitter | OSSTEM IMPLANT Co., Ltd. |
|---------------|-------------------------------------------------------------------|
| - Address | 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, 612-070, Korea |
| - Contact | Ms. Jungmin Yoo |
| - Phone | +82-51-850-2575 |
| Correspondent | HIOSSEN Inc. |
| - Address | 85 Ben Fairless Dr. Fairless Hills, PA 19030 |
| - Contact | Mr. Peter Lee |
| - Phone | 267-759-7031 |
# 2. Device Name and Classification
| Trade or (Propreitary) Name | Osstem Abutment System |
|-----------------------------|------------------------------------|
| Common or Usual Name | Dental Abutment |
| Classification Name | Endosseous dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Device Classification | Class II |
| Product Code | NHA |
# 3. Predicate Device
| Primary Predicate | |
|----------------------|------------------------------------------------------------|
| K161689 | OSSTEM Implant System - Abutment, Osstem Implant Co., Ltd. |
| Reference Predicates | |
| K160670 | ET US SS Prosthetic System |
| K160519 | Link Abutment for CEREC, Osstem Implant Co., Ltd. |
| K150295 | LOCATOR RTx, Zest Anchors, Inc. |
| K140507 | Hiossen Prosthetic System, Osstem Implant Co., Ltd. |
| K132067 | Multi Angled Abutment System, Osstem Implant Co., Ltd. |
| K120847 | ET/SS Implant System, Osstem Implant Co., Ltd. |
| K080594 | MS System (Narrow Ridge), Osstem Implant Co., Ltd. |
| K063861 | GS System, Osstem Implant Co., Ltd. |
| K062030 | US System, Osstem Implant Co., Ltd. |
{5}------------------------------------------------
### 4. Indication for Use
Osstem Abutment System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.
### · Link Abutment for CEREC
The Link Abutment for CEREC is titanium alloy abutments placed onto OSSTEM dental implants to provide support for customized prosthetic restorations. Link Abutment for CEREC is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. All digitally designed copings and/or crowns for use with the Link abutment for CEREC is to be scanned using Sirona CEREC AC or CEREC AF or CEREC AI, designed using Sirona inLab software (Version 3.65) or Sirona CEREC Software (Version 4.2) and manufactured using a Sirona CEREC or inLab MC X or MC XL milling unit. CAD/CAM manufacturing/milling occurs at dental laboratories per the design limitations of the Sirona CEREC.
### 5. Device Description
| Manufacturer | Model Name | Connection | Diameter (mm) |
|--------------------------|---------------|---------------|----------------------------------------------------------------------------------------------------------------------------|
| Osstem Implant Co., Ltd. | TS SA Fixture | Internal Hex | 3.2, 3.5, 3.75, 3.77, 4.2, 4.25, 4.4, 4.6, 4.63,<br>4.65, 4.8, 4.9, 5.05, 5.08, 5.1, 5.25, 5.92, 5.95,<br>6, 6.2, 6.8, 7.1 |
| | SS SA Fixture | Internal Octa | 3.75, 4.1, 4.25, 4.45, 4.6, 4.9, 5, 5.05, 5.92,<br>5.95, 5.96, 6, 6.8, 6.93 |
| | US SA Fixture | External Hex | 3.6, 4.2, 5.1, 5.2 |
| | MS SA Implant | Narrow Ridge | 2.5, 2.9 |
Osstem Abutment System is compatible with the following implant systems.
Osstem Abutment System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures.
Osstem Abutment System is similar to other commercially available products based on the intended use, technology used, claims, material composition employed and performance characteristics.
Osstem Abutment System is substantially equivalent in design, function and intended use to the predicate devices as above.
| Device | Content | |
|---------------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Transfer Abutment | Description Transfer Abutment is used for prosthetic restoration. It | |
| | | is used for making general cement-type prosthesis. |
| | Material | Titanium Alloy (Ti-6Al-4V, ASTM F136) |
| | Diameter (mm) | 4.0, 4.6, 5.0, 6.0, 7.0 |
| | Post Height (mm) | 4.0, 5.5, 7.0 |
| | Description | Angled Abutment is used for prosthetic restoration. It is used for making general cement-type prosthesis. It is used when a prosthetic's path adjustment is necessary |
| Angled Abutment | Material | Titanium Alloy (Ti-6Al-4V, ASTM F136) |
| | Diameter (mm) | 4.0, 4.5, 5.0, 6.0 |
| | Post Height (mm) | 8 |
| | Angulation | 17° |
| Link Abutment for Cerec | Description | Link Abutment for Cerec is used for customized prosthetic restoration. It is indicated for screw-retained single tooth or cement-retained single tooth and bridges restorations. It is compatible with the Sirona CEREC MC X and MC XL prosthetic milling system. |
| | Material | Titanium Alloy (Ti-6Al-4V, ASTM F136) |
| | Diameter (mm) | 4.5 |
| | Post Height (mm) | 4.7 |
| Temporary Abutment | Description | Temporary Abutment is used for prosthetic restoration. It is used temporarily to maintain esthetic appearance until final prosthesis is made. |
| | Material | Titanium Gr.3 (ASTM F67) |
| | Diameter (mm) | 4.0, 4.5 |
| | G/H (mm) | 1.0, 3.0 |
| Multi Angled Abutment | Description | If a few numbers of fixtures were implanted in mandibular bone for making full denture, some of fixture path should be leaned. Multi Angled Abutment is used to adjust path of prosthesis. |
| | Material | Titanium Alloy (Ti-6Al-4V, ASTM F136) |
| | Diameter (mm) | 4.9 |
| | Height (mm) | 5, 5.1, 5.5, 5.6, 6, 6.1, 6.5, 6.6, 7.5, 7.6 |
| | Angulation | 17°, 30° |
| Multi NP-Cast Cylinder | Description | Multi NP-Cast Cylinder is used for prosthetic restoration. It is used for making screw-retained type prosthesis by casting with non-precious metal alloy; and used with Multi Abutment together. It creates framework of the final prosthesis to be fixed on top of the abutment. |
| | Material | Co-Cr-Mo Alloy (Cylinder Body) + POM (Cylinder Sleeve) |
| | Diameter (mm) | 5.0 |
| | Length (mm) | 7.3 |
| Multi Combination Cylinder | Description | Multi Combination Cylinder is used for prosthetic restoration. It is used for making combination-retained type prosthesis with using Multi Abutment together. It |
| | | tope of the abutment. |
| | Material | Titanium Gr. 3 (ASTM F67) |
| | Diameter (mm) | 5.0 |
| | Length (mm) | 7.3 |
| Convertible Angled Cylinder | Description | Convertible Angled Cylinder is used for prosthetic<br>restoration. It is used for making combination-retaiend<br>type prosthesis by using with Convertible Abutment<br>together. It is used when path adjustment is necessary<br>at 17° axial angle. |
| | Material | Titanium Gr. 3 (ASTM F67) |
| | Diameter (mm) | 4.2, 5.0, 6.3 |
| | Length (mm) | 7.8 |
| | Angulation | 17° |
| | Description | Stud Abutment is used for prosthetic restoration. It is<br>used for making stud type overdenture prosthetics. |
| Stud Abutment | Material | Titanium Alloy (Ti-6Al-4V, ASTM F136) |
| | Diameter (mm) | 3.5 |
| | Head length (mm) | 2.5 |
| O-ring | Description | O-ring is inserted into retainer or retainer cap and<br>serves as a buffer for abutments/implants and denture<br>fixation. |
| | Material | NBR (Acrylonitrile & Butadiene Polymer) |
| | Diameter (mm) | 3.5 |
| | Description | O-ring Retainer Cap is used for making stud-type<br>overdenture. It is inserted and fixed into denture; and is<br>connected with abutment/implants. |
| O-ring Retainer Cap | Material | Titanium Gr. 3 (ASTM F67) |
| | Diameter (mm) | 3.95 |
| | Height (mm) | 2.9 |
| | Description | Port Abutment is used for prosthetic restoration. It is<br>for implant retained overdenture at maxilla/mandible in<br>case of the patient has no teeth. |
| Port Abutment | Material | Titanium Alloy (Ti-6Al-4V, ASTM F136) |
| | Diameter (mm) | 3.5, 3.7, 4.1, 4.8, 5.1 |
| | G/H (mm) | 1, 2, 3, 4, 5, 6, 7 |
| | Description | Port Angled Abutment is used for prosthetic<br>restoration. It is for implant retained overdenture<br>needed of path compensation at maxilla/mandible in<br>case of the patient has no teeth. |
| Port Angled Abutment | Material | Titanium Alloy (Ti-6Al-4V, ASTM F136) |
| | Diameter (mm) | 4.63, 4.66, 4.68, 4.72, 4.73, 4.74, 4.77, 4.82 |
| | Height (mm) | 6.5, 6.8, 7.5, 7.6 |
| | Angulation | 10°, 17°, 30° |
| Port Angled Abutment Head | Description | Port Angled Abutment Head is used for prosthetic<br>restoration. It is a part that is used to connect with Port<br>Angled Abutment. |
| | Material | Titanium Alloy (Ti-6Al-4V, ASTM F136) |
| | Diameter (mm) | 4.6 |
| | Height (mm) | 3.7 |
| Port Male, Port Extended Male | Description | Port Male and Port Extended Male is used for<br>prosthetic restoration. It is used to be inserted between<br>Port Abutment/Port Angled Abutment and Port Male<br>Cap and takes a role to maintain retention of<br>overdenture. |
| | Material | Nylon |
| | Diameter (mm) | 4.75 |
| | Height (mm) | 1.8 |
| Port Male Cap | Description | Port Male Cap is used for prosthetic restoration. It is<br>used to fix Port Male or Port Extended Male by<br>inserted into the denture. |
| | Material | Titanium Alloy (Ti-6Al-4V, ASTM F136) |
| | Diameter (mm) | 5.5 |
| | Height (mm) | 2.25 |
| Port Male Kit | Description | Port Male Kit is used for prosthetic restoration by<br>providing accessories that are used with Port Abutment<br>or Port Angled Abutment, and it is included of Port<br>Male Cap, Port Provisional Male, Port Spacer, and 3<br>different types of Port Male as a set. |
| | Set Configuration | Port Male Cap + Port Provisional Male + Port Spacer +<br>Port Male (3ea) |
| Abutment Screw | Description | Abutment Screw is used to connect an abutment to the<br>fixture. |
| | Material | Titanium Alloy (Ti-6Al-4V, ASTM F136) |
| | Diameter (mm) | 2, 2.05 |
| | Length (mm) | 7.5, 9.6 |
| Cylinder Screw | Description | Cylinder Screw is used to connect a cylinder to the<br>abutment. |
| | Material | Titanium Alloy (Ti-6Al-4V, ASTM F136) |
| | Diameter (mm) | 2.2, 2.5 |
| | Length (mm) | 4.35, 4.9 |
| Esthetic-low Temporary Cylinder | Description | Esthetic-low Temporary Cylinder is used for prosthetic<br>restoration. It is used for making temporary prosthesis<br>before loading final prosthesis. It is used by connected<br>with Multi Abutment, US Multi Angled Abutment or<br>Esthetic-low Abutment to make overdenture and bridge<br>as multiple cases. |
| | Material | Titanium Gr. 3 (ASTM F67) |
| | Diameter (mm) | 4.8, 5.5 |
| | Length (mm) | 12 |
| Temporary Cap (Narrow Ridge) | Description | Temporary Cap is used for prosthetic restoration<br>temporarily. It is a component used to protect upper<br>structure while final prosthesis is made. |
| | Material | PC (Poly Carbonate) |
| Diameter (mm) | 4 | |
| Length (mm) | 9.6 | |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is written in orange, with the superscript "6" next to it. Below the word "OSSTEM" is the word "IMPLANT" in gray. The logo is simple and modern.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is written in orange, with the superscript symbol for a registered trademark next to the "M". Below "OSSTEM" is the word "IMPLANT" in gray. The logo is simple and modern.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is written in orange, with a small superscript "6" next to the "M". Below "OSSTEM", the word "IMPLANT" is written in gray. The logo is simple and modern.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is written in orange, with a small superscript "6" next to it. Below the word "OSSTEM", the word "IMPLANT" is written in gray. The logo is simple and modern.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
#### 6. Substantial Equivalence Discussion
These subject devices are cleared in past 510(k) submissions but submitted to change their identifiers without modifications of dimensions or shape.
Link Abutment for Cerec, K160519 Multi Angled Abutment, K132067 Abutment Screw, K132067
The Indications for Use Statements are compared in the tables below. It can be seen that the Indications for Use Statements of the Subject and Primary Predicate devices are identical, with the exception of the addition to the Subject Indications for Use Statements language regarding the Link Abutment for CEREC. This language is identical to the Indications for Use Statement of the Reference Device being used for the Substantial Equivalence comparison for the Link Abutment for CEREC. All other reference devices have minor changes in wording, but nothing that affects the intended use or safety and effectiveness of the subject device system.
| | Osstem Abutment System | OSSTEM Implant System -<br>Abutment | Link Abutment for CEREC |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) No. | Proposed | Predicated (K161689) | Referenced (K160519) |
| Manufacturer | Osstem Implant Co., Ltd. | Osstem Implant Co., Ltd. | Osstem Implant Co., Ltd. |
| Indications for<br>Use Statement | Osstem Abutment System is<br>intended for use with a dental<br>implant to provide support<br>for prosthetic restorations<br>such as crowns, bridges, or<br>overdentures.<br>• Link Abutment for CEREC<br>The Link Abutment for<br>CEREC is titanium alloy<br>abutments placed onto<br>OSSTEM dental implants to<br>provide support for<br>customized prosthetic<br>restorations. Link Abutment<br>for CEREC is indicated for<br>screw-retained single tooth or<br>cement-retained single tooth<br>and bridge restorations. All<br>digitally designed copings<br>and/or crowns for use with<br>the Link abutment for<br>CEREC is to be scanned | The OSSTEM Implant<br>System - Abutment is<br>intended for use with a dental<br>implant to provide support<br>for prosthetic restorations<br>such as crowns, bridges, or<br>overdentures. | The Link Abutment for<br>CEREC is titanium alloy<br>abutments placed onto<br>HIOSSEN dental implants to<br>provide support for<br>customized prosthetic<br>restorations. Link Abutment<br>for CEREC is indicated for<br>screw-retained single tooth or<br>cement-retained single tooth<br>and bridge restorations. All<br>digitally designed copings<br>and/or crowns for use with<br>the Link abutment for<br>CEREC is to be scanned<br>using Sirona CEREC AC or<br>CEREC AF or CEREC AI,<br>designed using Sirona inLab<br>software (Version 3.65) or<br>Sirona CEREC Software<br>(Version 4.2) and<br>manufactured using a Sirona<br>CEREC or inLab MC X or |
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray. There is a small superscript "6" next to the "M" in "OSSTEM."
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
| using Sirona CEREC AC or<br>CEREC AF or CEREC AI,<br>designed using Sirona inLab<br>software (Version 3.65) or<br>Sirona CEREC Software<br>(Version 4.2) and<br>manufactured using a Sirona<br>CEREC or inLab MC X or<br>MC XL milling unit.<br>CAD/CAM<br>manufacturing/milling occurs<br>at dental laboratories per the<br>design limitations of the<br>Sirona CEREC. | MC XL milling unit.<br>CAD/CAM<br>manufacturing/milling occurs<br>at dental laboratories per the<br>design limitations of the<br>Sirona CEREC. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
Except for them, the subject devices are substantially equivalent in indications and design principles to the predicate devices as shown below.
| | Transfer Abutment | Transfer Abutment | Remark |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k) No. | Proposed | Predicated (K161689) | - |
| Manufacturer | Osstem Implant Co., Ltd. | Osstem Implant Co., Ltd. | Identical |
| Design | Image: Transfer Abutment | Image: Transfer Abutment | Identical |
| Indications for Use<br>Statement | The Osstem Abutment System<br>is intended for use with a<br>dental implant to provide<br>support for prosthetic<br>restorations such as crowns,<br>bridges, or overdentures. | The OSSTEM Implant System<br>- Abutment is intended for use<br>with a dental implant to<br>provide support for prosthetic<br>restorations such as crowns,<br>bridges, or overdentures. | Identical |
| Principle of<br>Operation | Using making for general<br>cement-type prosthesis. | Using making for general<br>cement-type prosthesis. | Identical |
| Material | Titanium Alloy<br>(Ti-6Al-4V, ASTM F136) | Titanium Alloy<br>(Ti-6Al-4V, ASTM F136) | Identical |
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray. The logo is simple and modern.
| | D(Ø) | G/H | Post | | D(Ø) | G/H | Post | | |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|--------------|--|------|------------------|--------------|--------------------------------------------------|--|
| Dimension (mm) | 4 | 1, 2, 3, 4,<br>5, 6, 7 | 5.5, 7 | | | | | -. Addition of Ø4<br>(New, smallest<br>diameter) | |
| | 4.6 | 6, 7 | 5.5, 7 | | 4.6 | 1, 2, 3, 4,<br>5 | 5.5, 7 | -. Addition of Ø4.6<br>(G/H 6, 7mm) | |
| | 5 | 6, 7 | 4, 5.5,<br>7 | | 5 | 1, 2, 3, 4,<br>5 | 4, 5.5,<br>7 | -. Addition of Ø5<br>(G/H 6, 7mm) | |
| | 6 | 6, 7 | 4, 5.5,<br>7 | | 6 | 1, 2, 3, 4,<br>5 | 4, 5.5,<br>7 | -. Addition of Ø6<br>(G/H 6, 7mm) | |
| | 7 | 1, 2, 3, 4,<br>5, 6, 7 | 4 | | | | | -. Addition of Ø7<br>(Post 4mm) | |
| | 7 | 6, 7 | 5.5 | | 7 | 1, 2, 3, 4,<br>5 | 5.5 | -. Addition of Ø7<br>(G/H 6, 7mm) | |
| S.E. | Similarities<br>Proposed Transfer Abutment has same design, function and indications for use<br>statement; and is made with same material with same manufacturing method by same<br>manufacturer and is generally used for cement-retained restoration compared to that of<br>the predicated Transfer Abutment (K161689).<br>Differences<br>The proposed Transfer Abutment has diameter of 4.0 to 7.0mm that has smallest<br>diameter among the proposed and the predicated Transfer Abutment (K161689).<br>However, the proposed abutment is straight type; therefore, we do not consider<br>additional fatigue testing.<br>.. Except for its size of diameter, proposed Transfer Abutment and the predicated<br>Transfer Abutment have common in design, function, indication for use, material,<br>manufacturing process, manufacturer, etc.; therefore, the proposed Transfer Abutment is<br>substantially equivalent to the predicated Transfer Abutment (K161689). | | | | | | | | |
| | Angled Abutment | Angled Abutment | Remark |
|----------------------------------|----------------------------------------------------------|-----------------------------------------------------------|----------------------------------------------------|
| 510(k) No. | Proposed | Predicated (K120847) | - |
| Manufacturer | Osstem Implant Co., Ltd. | Osstem Implant Co., Ltd. | Identical |
| Design | Image: Angled abutment with TiN Coating | Image: Angled abutment with TiN Coating | Identical except for<br>its TiN coating<br>section |
| Indications for Use<br>Statement | The Osstem Abutment System<br>is intended for use with a | The abutment is intended for<br>use with a dental implant | Identical |
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image contains the logo for Osstem Implant. The word "OSSTEM" is in orange, with a small superscript "6" next to it. Below "OSSTEM" is the word "IMPLANT" in gray. The logo is simple and clean, with a focus on the company name.
| | dental implant to provide<br>support for prosthetic<br>restorations such as crowns,<br>bridges, or overdentures. | fixture to provide support for<br>prosthetic restorations such as<br>crowns, bridges, or<br>overdenture. | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Principle of<br>Operation | Using making general cement-<br>type prosthesis when a<br>prosthetic's path adjustment is<br>necessary. | Using making general cement-<br>type prosthesis when a<br>prosthetic's path adjustment is<br>necessary. | Identical |
| Material | Titanium Alloy<br>(Ti-6Al-4V, ASTM F136)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.