SMARTBUILDER SYSTEM

K120951 · Osstem Implant Co., Ltd. · JEY · Oct 3, 2012 · Dental

Device Facts

Record IDK120951
Device NameSMARTBUILDER SYSTEM
ApplicantOsstem Implant Co., Ltd.
Product CodeJEY · Dental
Decision DateOct 3, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4760
Device ClassClass 2
AttributesTherapeutic

Indications for Use

SMARTbuilder System is a metal (Non-resorbable membrane) device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites.

Device Story

SMARTbuilder System is a non-absorbable, 3D pre-formed titanium membrane system used in dental surgery. It stabilizes and supports bone grafts in dento-alveolar bony defect sites. The system includes a titanium membrane, healing abutments, spacers, and cover caps. It is used in the oral cavity by dental professionals. The device is supplied sterile and is intended for use with specific OSSTEM dental implant systems (GS3 and US.SS.GS). By providing a stable scaffold for bone graft material, the device facilitates bone regeneration in defect areas, aiding in successful dental implant placement.

Clinical Evidence

No clinical studies are submitted. Biocompatibility is supported by the long-term, widespread use of titanium alloys in dental implants.

Technological Characteristics

Materials: Pure Titanium Grade 2 (ASTM F67), Pure Titanium Grade 4 (ASTM F67), and Titanium alloy Ti 6Al 4V (ASTM F136). Form factor: 3D pre-formed membrane, healing abutments, spacers, and cover caps. Sterilization: Supplied sterile. Connectivity: None (mechanical device).

Indications for Use

Indicated for patients requiring stabilization and support of bone grafts in dento-alveolar bony defect sites in conjunction with dental implants.

Regulatory Classification

Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 120951 ОСТ 3 2012 - ## 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: August 28, 2012 1. Company and Correspondent making the submission: - Submitter's Name : - Address : OSSTEM Implant Co., Ltd. #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804, Republic of Korea - Contact : 2. Device : Trade or (Proprietary) Name : Classification Name : SMARTbuilder System Bone Plate 21CFR872.4760 Class II JEY, NHA Mr. Hee Kwon Son 3. Predicate Device: The Neo Titanium mesh, CTi-mem, Neobiotech Co., Ltd., K111761 The SQ IS SYSTEM, Neobiotech Co., Ltd., K090825 ## 4. Description: Customized 3D Pre-Formed Titanium Membrane. SMARTBuilder is the non-absorbable membrane that is made of titanium metal to stabilize and support of bone graft after bone transplantation at the area having autogenous bone deficiency in the oral cavity. | | | Dimensions (mm) | |-----------|------------------|------------------------------------------------------------------------------------| | Image: BW | Image: BD and BL | BW(Buccal width): 8.0, 10.0<br>BD(Buccal Distance): 5.5<br>BL(Buccal Length): 7, 9 | QS-QI-505-3(Rev.0) Letter(8.5 X 11in) {1}------------------------------------------------ Image /page/1/Figure/0 description: The image shows diagrams of dental implant components with associated measurements. The first two components have a buccal width of 9.0 to 12.0, a buccal distance of 5.5, and a buccal length of 7 to 9. Other components have height measurements ranging from 0.5 to 3.0 and diameters from 3.3 to 5.03, while another has a height of 1.52 and a diameter of 4.0. OSSTEM Implant Co., Ltd The SMARTbuilder is made of pure titanium metal and supplied sterile. The SMARTbuilder system is used with GS3 System (K091208) of OSSTEM Implant Co., Ltd. and US.SS.GS System (K073247) of OSSTEM Implant Co., Ltd. The SMARTbuilder is substantially equivalent in design, function and intended use to the Neo Titanium mesh, CTi-mem (K111761) of Neobiotech Co., Ltd Height, Healing abutment and Cover cap of The SMARTbuilder system are substantially equivalent in design, function and intended use to the SQ IS SYSTEM (K090825) of Neobiotech Co., Ltd QS-QI-505-3(Rev.0) Letter(8.5 X 11in) {2}------------------------------------------------ OSSTEM® MPLANT OSSTEM Implant Co., Ltd. #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com | - Substantial Equivalence Matrix | | | |----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------| | | SMARTbuilder | Predicate devices<br>CTi-mem (K111761) | | Design | Image: SMARTbuilder design | Image: Cti-mem design | | Intended<br>use | SMARTbuilder is a metal device intended<br>for use with a dental implant to stabilize and<br>support of bone graft in dento-alveolar bony<br>defect sites. | For Stabilization and support bone grafts in<br>dento-alveolar bony defect sites | | Material of<br>Fixture | Pure Titanium Grade 2 (ASTM F67) | Pure Titanium Grade 2 (ASTM F67) | | Width (D) | 8, 9, 10, 12 | 12 | | Sterilization | Sterilie | Sterilie | | Shelf life | 5 years | | | SE | The SMARTbuilder has same material and indication for use and similar design and<br>technological characteristics as the predicate device, such as the CTi-mem | | | - Substantial Equivalence Matrix | | | | |----------------------------------|--|--|--| | | | | | | | SMARTbuilder System<br>Healing Abutment | Predicate devices | | |------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|--| | | | SQ IS SYSTEM (K090825)<br>CTi Healing Abutment | | | Design | Image: SMARTbuilder System Healing Abutment | Image: SQ IS SYSTEM (K090825) CTi Healing Abutment | | | Intended<br>use | SMARTbuilder is a metal device intended<br>for use with a dental implant to stabilize and<br>support of bone graft in dento-alveolar bony<br>defect sites. | For Stabilization and support bone grafts in<br>dento-alveolar bony defect sites | | | Material of<br>Fixture | Pure Titanium Grade 4 (ASTM F67) | Pure Titanium (ASTM F67) | | | Diameter | 4.0~7 | 4.0 | | | Height | 3.35~4.22 | 3.0, 4.0 | | | Sterilization | Sterilie | Sterilie | | | Shelf life | 5years | | | | SE | The Healing Abutment has same material and indication for use and similar design and<br>technological characteristics as the predicate device, such as the CTi Healing Abutment | | | QS-QI-505-3(Rev.0) Letter(8.5 X 11in) September 1997 - 1992 - 1992 - 1991 - 1991 - 19 {3}------------------------------------------------ OSSTEM® IMPLANT OSSTEM Implant Co., Ltd. #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com | | SMARTbuilder System<br>Height | Predicate devices<br>SQ IS SYSTEM (K090825)<br>CTi Spacer | | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|--| | Design | | | | | Intended<br>use | SMARTbuilder is a metal device intended for<br>use with a dental implant to stabilize and<br>support of bone graft in dento-alveolar bony<br>defect sites. | For Stabilization and support bone grafts in<br>dento-alveolar bony defect sites | | | Material<br>of<br>Fixture | Titanium alloy Ti 6Al 4V (ASTM F 136) | Pure Titanium (ASTM F67) | | | Steriliza<br>tion | Sterilie | Sterilie | | | Shelf life | 5years | | | | SE | The Height has same material and indication for use and similar design and<br>technological characteristics as the predicate device, such as the CTi Spacer | | | | | SMARTbuilder<br>Cover Cap | Predicate devices | | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|--| | | Image: SMARTbuilder Cover Cap | SQ IS SYSTEM (K090825)<br>CTi Cover Cap | | | Design | Image: SMARTbuilder Cover Cap | Image: CTi Cover Cap | | | Intended<br>use | SMARTbuilder is a metal device intended for<br>use with a dental implant to stabilize and<br>support of bone graft in dento-alveolar bony<br>defect sites. | For Stabilization and support bone grafts in<br>dento-alveolar bony defect sites | | | Material<br>of Fixture | Pure Titanium Grade 4 (ASTM F67) | Pure Titanium (ASTM F67) | | | Diameter | 4.0 | 4.0 | | | Height | 1.52 | 1.5 | | | Sterilizati<br>on | Sterilie | Sterilie | | | Shelf life | 5years | - | | | S E | The Cover Cap has same material and indication for use and similar design and<br>technological characteristics as the predicate device, such as the CTi Cover Cap | | | {4}------------------------------------------------ OSSTEM Implant Co., Ltd. #507-8 Geoie3-Dong Yeonie-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com - 5. Indication for use : The SMARTbuilder system is a metal device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites. - 6. Review : The SMARTbuilder system has the same material and indication for use and similar design and technological characteristics as the predicate device. - 7. Summary of nonclinical testing Biocompatibility evaluation for SMARTbuilder System is not considered because SMARTbuilder System, which is made with titanium alloy, has been generally and widely used as a dental material such as implant for a long time - 8. Summary of clinical testing No clinical studies are submitted - 9. Conclusion : Based on the information provided in this premarket notification Osstem Implant Co., Ltd. concludes that the SMARTbuilder system is substantially equivalent to the predicate devices as described herein {5}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/5/Picture/1 description: The image shows a logo for the Department of Health & Human Services, USA. The logo consists of a circular arrangement of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, represented by thick, stylized lines. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OCT 3 2012 Osstem Implant Company, Limited C/O Mr. Patrick Lim Hiossen, Incorporated 85 Ben Fairless Drive Fairless Hills, Pennsylvania 19030 Re: K120951 Trade/Device Name: SMARTbuilder System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY, NHA Dated: August 28, 2012 Received: September 12, 2012 Dear Mr. Lim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In . addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2- Mr. Lim Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reportung (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, R. Staley Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ OSSTEM Implant Co., Ltd. #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com # 510(k) Number K_12095 1 Device Name : SMARTbuilder system Indication for use : SMARTbuilder System is a metal (Non-resorbable membrane) device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites. Prescription Use X (Per 21CFR801 Subpart D) OR Over-The-Counter Use (Per 21 CFR807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Russo (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K12095
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...