The Elos Accurate® Hybrid Base™ is intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Hybrid Base™ is used as an interface between a dental implant and a zirconia superstructure and will be attached to the implant using the included prosthetic screw and attached to the zirconia superstructure by cementing.
Device Story
Elos Accurate® Hybrid Base is a titanium base interfacing between dental implants and zirconia superstructures; supports single-unit (engaging) or multi-unit (non-engaging) restorations. Device consists of titanium base and prosthetic screw; zirconia superstructure is cemented to base. Used in dental clinics; operated by dentists/prosthodontists. Digitally designed superstructures must be manufactured at FDA-registered Elos Medtech-approved milling facilities. Provides stable interface for definitive prosthetic restorations; benefits patient by restoring chewing function.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing: biocompatibility (ISO 10993-5 cytotoxicity) and dynamic compression-bending fatigue testing (ISO 14801). Engineering/dimensional analysis confirmed compatibility with OEM implant bodies, abutments, and screws.
Technological Characteristics
Materials: Ti-6Al-4V alloy (ASTM/ISO compliant), Zirconia (ISO 13356). Interface: Internal connection. Form factor: Two-piece abutment (titanium base + zirconia superstructure). Surface: Anodized or non-anodized; Medicarb (DLC) coating on screws. Sterilization: Validated per ISO 17665-1/2. Connectivity: N/A (mechanical device).
Indications for Use
Indicated for patients requiring single or multiple tooth prosthetic restorations supported by dental implants. Compatible with specific Nobel Biocare, Straumann, and Astra Tech implant systems. Contraindications include patients where the device cannot be properly attached to the implant or superstructure.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 10, 2020
Elos Medtech Pinol A/S % Floyd Larson President PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K191919
Trade/Device Name: Elos Accurate® Hybrid Base Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: January 6, 2020 Received: January 7, 2020
Dear Floyd Larson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name
Elos Accurate® Hybrid Base
K191919
#### Indications for Use (Describe)
The Elos Accurate® Hybrid Base™ is intended for attaching to dental implants in order to single or multiple tooth prostheto restorations. The Hybrid Base" is used as an interface between a dental implant and a zirconia superstructure and will be attached to the included prosthetic screw and attached to the zirconia superstructure by cementing.
The Elos Accurate® Hybrid Base™ is compatible with the implant systems listed in Table 1.
| Table 1. | | |
|--------------------------------|------------------------|----------------------------|
| Implant Platform compatibility | Platform diameter [mm] | Implant Body diameter [mm] |
| Nobel Replace NP | 3.5 | 3.5 |
| Nobel Replace RP | 4.3 | 4.3 |
| Nobel Replace WP | 5 | 5 |
| Nobel Replace 6.0 | 6 | 6 |
| Nobel CC 3.0 | 3 | 3 |
| Nobel CC NP | 3.5 | 3.5 & 3.75 |
| Nobel CC RP | 3.9 | 4.3 & 5 |
| Nobel CC WP | 5.1 | 5.5 |
| Straumann Bone Level NC | 3.3 | 3.3 |
| Straumann Bone Level RC | 4.1 & 4.8 | 4.1 & 4.8 |
| Astra Tech 3.0 | 3 | 3 |
| Astra Tech 3.5/4.0 | 3.5 & 4 | 3.5 & 4 |
| Astra Tech 4.5/5.0 | 4.5 & 5 | 4.5 & 5 |
| Astra Tech EV 3.0 | 3 | 3 |
| Astra Tech EV 3.6 | 3.6 | 3.6 |
| Astra Tech EV 4.2 | 4.2 | 3.6 & 4.2 |
| Astra Tech EV 4.8 | 4.8 | 4.2 & 4.8 |
| Astra Tech EV 5.4 | 5.4 | 5.4 |
All digitally designed zirconia superstructures for use "Hybrid Base" are only intended to be sent and manufactured at an FDA registered Elos Medtech approved milling facility.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# K191919
# Elos Medtech Pinol A/S Elos Accurate® Hybrid Base™
February 10, 2020
## ADMINISTRATIVE INFORMATION
| Manufacturer Name | Elos Medtech Pinol A/S<br>Engvej 33<br>DK-3330 Gørløse, Denmark<br>Telephone +45 48 21 64 69 | | | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Official Contact | Tina Friis Poulsen, Head of Compliance | | | |
| Representative/Consultant | Floyd G. Larson, MS, MBA<br>Kevin A. Thomas, PhD<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 858-792-1235<br>Fax: +1 858-792-1236<br>Email: flarson@paxmed.com<br>kthomas@paxmed.com | | | |
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Elos Accurate® Hybrid Base™ |
|------------------------|------------------------------------|
| Common Name | Dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Division | DHT1B: Division of Dental Devices |
## PREDICATE DEVICE INFORMATION
Primary Predicate K173908, DESS Dental Smart Solutions, Terrats Medical SL
Reference Devices K151621, BioHorizons CAD/CAM Abutments, BioHorizons Implant Systems, Inc. K171799, Elos Accurate® Customized Abutment, Elos Medtech Pinol A/S K183518, Preat Abutments, Preat Corporation K080396, OsseoSpeed Narrow, Dentsply Sirona K053384, Fixture MicroThread OsseoSpeed, Dentsply Sirona K120414, OsseoSpeed Plus, Dentsply Sirona K102436, NobelActive 3.0, Nobel Biocare AB
{5}------------------------------------------------
K071370. NobelActive Internal Connection Implant. Nobel Biocare AB K133731, NobelActive Wide Platform (WP), Nobel Biocare AB K023113, Replace TiUnite Endoessous Implant, Nobel Biocare AB K062129, P.0004 Implants (Straumann Dental Implant System), Institut Straumann K130436, Multilink Hybrid Abutment Cement, Ivoclar Vivadent AG
# INDICATIONS FOR USE
The Elos Accurate® Hybrid Base" is intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Hybrid Base" is used as an interface between a dental implant and a zirconia superstructure and will be attached to the implant using the included prosthetic screw and attached to the zirconia superstructure by cementing.
| Table 1. | | |
|--------------------------------|------------------------------|----------------------------------|
| Implant Platform compatibility | Platform<br>diameter<br>[mm] | Implant Body<br>diameter<br>[mm] |
| Nobel Replace NP | 3.5 | 3.5 |
| Nobel Replace RP | 4.3 | 4.3 |
| Nobel Replace WP | 5 | 5 |
| Nobel Replace 6.0 | 6 | 6 |
| Nobel CC 3.0 | 3 | 3 |
| Nobel CC NP | 3.5 | 3.5 & 3.75 |
| Nobel CC RP | 3.9 | 4.3 & 5 |
| Nobel CC WP | 5.1 | 5.5 |
| Straumann Bone Level NC | 3.3 | 3.3 |
| Straumann Bone Level RC | 4.1 & 4.8 | 4.1 & 4.8 |
| Astra Tech 3.0 | 3 | 3 |
| Astra Tech 3.5/4.0 | 3.5 & 4 | 3.5 & 4 |
| Astra Tech 4.5/5.0 | 4.5 & 5 | 4.5 & 5 |
| Astra Tech EV 3.0 | 3 | 3 |
| Astra Tech EV 3.6 | 3.6 | 3.6 |
| Astra Tech EV 4.2 | 4.2 | 3.6 & 4.2 |
| Astra Tech EV 4.8 | 4.8 | 4.2 & 4.8 |
| Astra Tech EV 5.4 | 5.4 | 5.4 |
The Elos Accurate® Hybrid Base™ is compatible with the implant systems listed in Table 1.
All digitally designed zirconia superstructures for use with the Elos Accurate® Hybrid Base™ are only intended to be sent and manufactured at an FDA registered Elos Medtech approved milling facility.
# DEVICE DESCRIPTION
The Elos Accurate Hybrid Base is a titanium base designed to interface with a dental implant and to support a patient-specific ceramic superstructure or a multi-unit direct restoration cemented to the base. The base and the superstructure form a patient-specific abutment that will support a definitive restoration, either a single crown or a multi-unit restoration. The Elos Accurate Hybrid Base Engaging is intended for single-unit restorations and Elos Accurate Hybrid Base Nonengaging is intended for multi-unit restorations. Alternatively, a definitive multi-unit restoration may be cemented directly to the Elos Accurate Hybrid Base.
{6}------------------------------------------------
Manufacture of the final finished device will be at an Elos Medtech-approved milling facility that is registered with FDA as a medical device manufacturer and is qualified as a contract manufacturer to Elos Medtech.
This submission includes a two-piece abutment (titanium base and zirconia superstructure) and abutment screws compatible with OEM implants from Dentsply Implants (Astra Tech TX, Astra Tech EV), Nobel Biocare (Nobel Active/Conical Connection and Nobel Replace) and Institut Straumann (Straumann Bone Level NC and RC).
Abutments compatible with Astra Tech implants are available in engaging (anti-rotational) designs, and all others are available in both engaging and non-engaging designs.
## PERFORMANCE DATA
Non-clinical data submitted to demonstrate substantial equivalence include: biocompatibility evaluation and confirmatory cytotoxicity testing according to ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity, and dynamic compression-bending testing according to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants. No clinical data were included in this submission.
Substantial equivalence with regard to compatibility with OEM components is supported by engineering and dimensional analysis of OEM implant bodies, OEM abutments and OEM abutment fixation screws to confirm compatibility.
Substantial equivalence with regard to mechanical performance is supported by dynamic testing according to ISO 14801 Dentistry – Implants – Dynamic fatigue test for endosseous dental implants.
The coatings used on selected screws are identical to coatings on previously cleared devices. The Medicarb coating is identical to the Medicarb (DLC) coating on reference device K171799 and the Nobel DLC coating is identical to the DLC coating on compatible device K071370. Sterilization validation of the reprocessing instructions were conducted in accordance to ISO 17665-1 and ISO 17665-2.
Minor differences in the designs, dimensions, sizes, or compatible OEM implant lines among the subject device, the primary predicate device, and the reference devices do not affect substantial equivalence. These minor differences do not impact a determination of substantial equivalence because these differences are related to the compatible OEM implant designs or are mitigated by the mechanical performance testing.
{7}------------------------------------------------
# EQUIVALENCE TO MARKETED DEVICE
Overall, the subject device has the following similarities to the predicate devices:
- has the same intended use,
- o uses the same operating principle,
- o incorporates the same basic design,
- . incorporates the same or very similar materials, and
- has similar packaging and is sterilized using the same materials and processes. .
The basis for the belief of Elos Medtech Pinol A/S that the subject device is substantially equivalent to the predicate devices is summarized in the following Table of Substantial Equivalence.
{8}------------------------------------------------
| Subject Device | Primary Predicate | Reference Device | Reference Device | Reference Device | | | | | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|---------------------|-------------------------------------------|-------------------------------|-----------------------------------------|
| | K173908 | K151621 | K171799 | K183518 | | | | | | | |
| Elos Accurate® Hybrid BaseTM | DESS Dental Smart Solutions | BioHorizons CAD/CAM Abutments | Elos Accurate® Customized Abutment, | Preat Abutments | | | | | | | |
| | | BioHorizons Implant Systems, Inc. | Elos Medtech Pinol A/S | Preat Corporation | | | | | | | |
| Elos Medtech Pinol A/S | Terrats Medical SL | | | | | | | | | | |
| The Elos Accurate® Hybrid BaseTM is intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Hybrid BaseTM is used as an interface between a dental implant and a zirconia superstructure and will be attached to the implant using the included prosthetic screw and attached to the zirconia superstructure by cementing. | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.<br>All digitally designed custom abutments for use with AurumTM Abutment or Pre-milled Blank are to be sent to a Terrats Medical validated milling center for manufacture. | BioHorizons CAD/CAM Abutments are dental abutments placed onto a dental implant to provide support for dental prosthetic restorations. The abutments include: 1) Titanium abutment blanks with a pre-machined implant connection where the upper portion may be custom-milled in accordance with a patient-specific design using CAD/CAM techniques; and 2) Titanium bases with a pre-machined implant connection upon which a CAD/CAM designed superstructure may be fitted to complete a two-piece dental abutment. The abutments include an abutment screw for fixation to the underlying implant. The abutments may be used for single-unit (single-tooth) or multiple-unit (bridges and bars) restorations and are compatible for use with BioHorizons Internal and Tapered Internal implant systems and Zimmer® Dental Screw-Vent® and Tapered Screw-Vent® implants with 3.5mm, 4.5mm and 5.7mm internal hex-connection mating platform diameters.<br>All digitally designed abutments and/or copings for use with BioHorizons CAD/CAM Abutments are intended to be sent to a BioHorizons-validated milling center for manufacture. BioHorizons abutments designed using CAD/CAM techniques must fulfill the BioHorizons allowable range of design parameters. | The Elos Accurate® Customized Abutments are intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to a dental implant using the included Elos Prosthetic screw.<br>The Elos Accurate® Customized Abutments are compatible with the following implant systems: | Preat Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit on multi-unit prosthetic restorations. The Titanium Base abutments consists of two major parts. Specifically, the titanium base and mesostructured components make up a two-piece abutment.<br>All digitally designed custom abutments, superstructures, and/or hybrid crowns for use with Titanium Base or Titanium Blank are to be sent to a Preat validated milling center for manufacture. | | | | | | | |
| The Elos Accurate® Hybrid BaseTM is compatible with the implant systems listed in Table 1. | Compatible Implant Systems | | | Compatible Implant Systems | | | | | | | |
| Table 1. | | | | | | | | | | | |
| Implant Platform<br>compatibility | Platform<br>diameter<br>[mm] | Implant<br>Body<br>diameter<br>[mm] | Implant System<br>Compatibility | Implant Body | Implant<br>Platform | Ref. No. | Platform<br>compatibility | Implant<br>diameter | Compatible Implant System | Implant Body<br>Diameter (mm) | Implant<br>Platform<br>Diameter<br>(mm) |
| Nobel Replace NP | 3.5 | 3.5 | 3i Certain® | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 | AB-BRA411213-US | Nobel Biocare®/<br>Brånemark® RP | 3.75 mm<br>& 4 mm | 3i OSSEOTITE® Certain® | 3.25 | 3.4 |
| Nobel Replace RP | 4.3 | 4.3 | 3i OSSEOTITE® | 3.25, 3.75,<br>4.0, 5.0 | 3.4, 4.1, 5.0 | AB-BRA351213-US | Nobel Biocare®/<br>Brånemark® RP | 3.3 mm | | 4.0 | 4.1 |
| Nobel Replace WP | 5 | 5 | OsseoSpeedTM | 3.5, 4.0, 5.0 | 3.5/4.0, 4.5/5.0 | AB-BRA511213-US | Nobel Biocare®/<br>Brånemark® RP | 5 mm | | 5.0 | 5.0 |
| Nobel Replace 6.0 | 6 | 6 | FRIADENT XIVE | 3.4, 3.8, 4.5 | 3.4, 3.8, 4.5 | All digitally designed Elos Accurate® Customized<br>Abutments are intended to be manufactured at an Elos<br>Medtech approved milling facility. | | | | 6.0 | 6.0 |
| Nobel CC 3.0 | 3 | 3 | NobelActive® | 3.5, 4.3, 5.0 | NP, RP | | | | Astra Tech OsseoSpeedTM | 3.0 | 3.0 |
| Nobel CC NP | 3.5 | 3.5 & 3.75 | NobelReplace<br>Conical | 3.5, 4.3, 5.0 | NP, RP | | | | | 3.5, 4.0 | 3.5/4.0 |
| Nobel CC RP | 3.9 | 4.3 & 5 | Nobel Replace<br>Trilobe | 3.5, 4.3, 5.0 | NP, RP, WP | | | | | 4.5, 5.0 | 4.5/5.0 |
| Nobel CC WP | 5.1 | 5.5 | Brånemark | 3.5, 3.75/4.0,<br>5.0 | NP, RP, WP | | | | BioHorizons Tapered Internal | 3.0 | 3.0 |
| Straumann Bone Level NC | 3.3 | 3.3 | Straumann® Bone<br>Level | 3.3, 4.1, 4.8 | NC, RC | | | | | 3.5 | 3.5 |
| Straumann Bone Level RC…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.