K173908 · Terrats Medical SL · NHA · Jun 15, 2018 · Dental
Device Facts
Record ID
K173908
Device Name
DESS Dental Smart Solutions
Applicant
Terrats Medical SL
Product Code
NHA · Dental
Decision Date
Jun 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with Aurum™ Abutment or Pre-milled Blank are to be sent to a Terrats Medical validated milling center for manufacture.
Device Story
DESS Dental Smart Solutions are dental implant abutments (Aurum Base, Pre-milled Blanks, CoCr Abutments) used to support prosthetic restorations (single or multi-unit). The system includes titanium alloy and cobalt-chromium alloy components. Clinicians use these abutments with compatible endosseous dental implants. Digitally designed custom abutments are manufactured via CAD/CAM at a Terrats Medical validated milling center. The device provides a mechanical interface between the implant and the final prosthesis (cement or screw-retained). By providing a stable connection, the device enables the restoration of missing teeth, improving patient function and aesthetics.
Clinical Evidence
No clinical data were included in this submission. Substantial equivalence is supported by non-clinical bench testing, including sterilization validation (ISO 17665-1/2) and biological evaluation/cytotoxicity testing (ISO 10993-1/5).
Technological Characteristics
Materials: Titanium alloy (ASTM F136) and Cobalt-Chromium alloy (ASTM 1537). Design: CAD/CAM custom abutments, pre-milled blanks, and cast-to abutments. Interface: Internal and external hex/conical connections compatible with various OEM implant platforms (3.3mm to 6.5mm). Sterilization: Non-sterile, validated for terminal sterilization (SAL 10^-6).
Indications for Use
Indicated for use in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations in partially or completely edentulous patients.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
June 15, 2018
Terrats Medical SL % Linda Schulz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K173908
Trade/Device Name: DESS Dental Smart Solutions Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: May 16, 2018 Received: May 17, 2018
# Dear Linda Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Tina Kiang, Ph.D. for Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
## 510(k) Number (if known)
#### K173908
#### Device Name
#### DESS Dental Smart Solutions
#### Indications for Use (Describe)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with Aurum™ Abutment or Pre-milled Blank are to be sent to a Terrats Medical validated milling center for manufacture.
| Compatible Implant Systems | | |
|---------------------------------|----------------------|------------------|
| Implant System<br>Compatibility | Implant Body | Implant Platform |
| 3i Certain® | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| 3i OSSEOTITE® | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| OsseoSpeed™ | 3.5, 4.0, 5.0 | 3.5/4.0, 4.5/5.0 |
| FRIADENT XiVE | 3.4, 3.8, 4.5 | 3.4, 3.8, 4.5 |
| NobelActive® | 3.5, 4.3, 5.0 | NP, RP |
| NobelReplace® Conical | 3.5, 4.3, 5.0 | NP, RP |
| NobelReplace® Trilobe | 3.5, 4.3, 5.0 | NP, RP, WP |
| Brånemark | 3.5, 3.75/4.0, 5.0 | NP, RP, WP |
| Straumann® Bone Level | 3.3, 4.1, 4.8 | NC, RC |
| Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN, WN |
| Tapered Screw-Vent® | 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 |
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Terrats Medical SL DESS Dental Smart Solutions
# June 15, 2018
# ADMINISTRATIVE INFORMATION
| Manufacturer Name | Terrats Medical SL<br>Carrer Mogoda 75-99<br>Barberà del Vallès 08210<br>Barcelona, Spain | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--|
| Telephone | +34 93 564 60 06 | |
| Fax | +34 93 564 73 17 | |
| Official Contact | Roger Terrats, COO | |
| Representative/Consultant | Linda Schulz, BSDH, RDH<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 | |
| Telephone: | +1-858-792-1235 | |
| Fax: | +1-858-792-1236 | |
| Email: | LSchulz@paxmed.com<br>FLarson@paxmed.com | |
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | DESS Dental Smart Solutions |
|----------------------------|------------------------------------|
| Common Name | Dental implant abutment |
| Classification Name | Endosseous dental implant abutment |
| Classification Regulations | 21 CFR 872.3630, Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
# PREDICATE DEVICE INFORMATION
Primary Predicate K170588, DESS Dental Smart Solutions, Terrats Medical SL
| Reference Device | | |
|------------------|---------------------------------|-----------------|
| K092341 | Low Profile Abutment | Biomet 3i, Inc. |
| K150669 | Neoss TiBase and CoCr Abutments | Neoss Ltd. |
| K120414 | OsseoSpeed™ Plus | Astra Tech AB |
| K160784 | CAM Titanium Blanks | Altatec GmbH |
{4}------------------------------------------------
| K063341 | 3i OSSEOTITE Certain® Dental Implants | Implant Innovations, Inc. |
|---------|---------------------------------------------------------------------------|------------------------------|
| K063286 | OSSEOTITE® Dental Implants | Implant Innovations, Inc. |
| K101732 | OsseoSpeed™ | Astra Tech AB |
| K073075 | FRIADENT Implant Systems | DENTSPLY International, Inc. |
| K142260 | NobelActive® | Nobel Biocare AB |
| K073142 | NobelReplace Hexagonal Implants | Nobel Biocare AB |
| K050705 | TiUnite® Implants | Nobel Biocare AB |
| K050406 | NOBELSPEEDY™ Implants | Nobel Biocare USA LLC |
| K022562 | Various Brånemark System Implants -<br>Immediate Function Indication | Nobel Biocare AB |
| K140878 | Straumann® Bone Level Tapered Implants | Straumann USA, LLC |
| K062129 | P.004 Implants | Institut Straumann AG |
| K130222 | Straumann® Dental Implant System SLActive and<br>Roxolid Product Families | Straumann USA, LLC |
| K112160 | Tapered Screw-Vent® X Implant | Zimmer Dental, Incorporated |
#### Compatible Implant Systems
#### INDICATIONS FOR USE
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with Aurum™ Abutment or Pre-milled Blank are to be sent to a Terrats Medical validated milling center for manufacture.
| Compatible Implant Systems | | |
|------------------------------|----------------------|------------------|
| Implant System Compatibility | Implant Body | Implant Platform |
| 3i Certain® | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| 3i OSSEOTITE® | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| OsseoSpeed™ | 3.5, 4.0, 5.0 | 3.5/4.0, 4.5/5.0 |
| FRIADENT XiVE | 3.4, 3.8, 4.5 | 3.4, 3.8, 4.5 |
| NobelActive® | 3.5, 4.3, 5.0 | NP, RP |
| NobelReplace Conical | 3.5, 4.3, 5.0 | NP, RP |
| Nobel Replace Trilobe | 3.5, 4.3, 5.0 | NP, RP, WP |
| Brånemark | 3.5, 3.75/4.0, 5.0 | NP, RP, WP |
| Straumann® Bone Level | 3.3, 4.1, 4.8 | NC, RC |
| Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN, WN |
| Tapered Screw-Vent® | 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 |
Compatible Implant Systems
#### DEVICE DESCRIPTION
DESS Dental Smart Solutions subject devices include four abutment design types (Aurum Base, Pre-milled Blank, CoCr Pre-milled Blank, CoCr Abutment) and one screw type (Aurum Base Screw). Abutments are provided in ten abutment connections compatible with eleven implant platform diameters range from 3.3 mm to 6.5 mm. Corresponding implant body diameters range from 3.25 mm to 6.0 mm. All abutments are provided non-sterile. The following table outlines the body and platform diameters by abutment design and corresponding compatible implant line.
{5}------------------------------------------------
## Summary of Abutment Designs
| | CAD / CAM | | | | Cast-To | | | OEM<br>Implant System | Connection |
|----------------------------|------------------------------------------------------------|--------------------------------------------------------|-----------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------|------------------------------------------------|----------------------------|------------------------------------|------------|
| DESS<br>Abutment<br>System | AURUM base™<br>Non-Engaging<br>Titanium Alloy<br>ASTM F136 | AURUM base™<br>Engaging<br>Titanium Alloy<br>ASTM F136 | CoCr Pre-milled<br>Blank Engaging<br>CoCr ASTM 1537 | Ti Pre-milled<br>Blank Engaging<br>Titanium Alloy<br>ASTM E136 | CoCr Abutment<br>Non-Engaging<br>CoCr<br>ASTM 1537 | CoCr Abutment<br>Engaging<br>CoCr<br>ASTM 1537 | | | |
| Internal Hex<br>"Click" | NP<br>RP<br>WP | NP<br>RP<br>WP | NP<br>RP<br>WP | | NP<br>RP<br>WP | NP<br>RP<br>WP | 3i Certain® | 3.4 (NP)<br>4.1 (RP)<br>5.0 (WP) | Internal |
| External Hex<br>USA | NP<br>RP<br>WP | NP<br>RP<br>WP | NP<br>RP<br>WP | | NP<br>RP<br>WP | NP<br>RP<br>WP | 3i<br>OSSEOTITE® | 3.4 (NP)<br>4.1 (RP)<br>5.0 (WP) | External |
| Internal Hex<br>Conic | RP<br>WP | RP<br>WP | RP<br>WP | | RP<br>WP | RP<br>WP | OsseoSpeed™ | 3.5/4.0<br>(RP)<br>4.5/5.0<br>(WP) | Internal |
| Internal Hex<br>FD | NP<br>RP<br>WP | NP<br>RP<br>WP | NP<br>RP<br>WP | NP<br>RP<br>WP | NP<br>RP<br>WP | NP<br>RP<br>WP | FRIADENT<br>XiVE® | 3.4 (NP)<br>3.8 (RP)<br>4.5 (WP) | Internal |
| Active Hex | NP<br>RP | NP<br>RP | NP<br>RP | | NP<br>RP | NP<br>RP | NobelActive® | 3.5 (NP)<br>3.9 (RP) | Internal |
| | | | | | | | NobelReplace®<br>Conical | 3.5 (NP)<br>3.9 (RP) | Internal |
| Tri-Lobe | NP<br>RP<br>WP | NP<br>RP<br>WP | NP<br>RP<br>WP | | NP<br>RP<br>WP | NP<br>RP<br>WP | NobelReplace®<br>Trilobe | 3.5 (NP)<br>4.3 (RP)<br>5.0 (WP) | Internal |
| External Hex<br>Universal | NP<br>RP | NP<br>RP | NP<br>RP<br>WP | NP | NP<br>RP<br>WP | NP<br>RP<br>WP | Brånemark<br>System® | 3.5 (NP)<br>4.1 (RP)<br>5.1 (WP) | External |
| Conical BL | NC<br>RC | NC<br>RC | NC<br>RC | | NC<br>RC | NC<br>RC | Straumann®<br>Bone Level | 3.3 (NC)<br>4.1 (RC) | Internal |
| Octagon | RN<br>WN | RN<br>WN | RN<br>WN | | RN<br>WN | RN<br>WN | Straumann®<br>Tissue Level | 4.8 (RN)<br>6.5 (WN) | Internal |
| Internal Hex<br>USA | NP<br>RP<br>WP | NP<br>RP<br>WP | NP<br>RP<br>WP | | NP<br>RP<br>WP | NP<br>RP<br>WP | Tapered Screw-<br>Vent | 3.5 (NP)<br>4.5 (RP)<br>5.7 (WP) | Internal |
Aurum Abutment is a titanium abutment that can be used for a direct multi-unit restoration or to support a zirconia superstructure plus a single-unit or multi-unit restoration. Design parameters for the zirconia superstructure are a minimum wall thickness of 0.4 mm, a minimum post height for single-unit restorations of 4.0 mm, and a maximum gingival height of 6.0 mm. All zirconia superstructures are for straight abutments only.
Pre-milled Blank abutments are cylindrical abutments designed for custom abutment fabrication by a CAD/CAM process. All patient-specific abutment fabrication is by prescription on the order of the clinician. Pre-milled Blanks are made of cobalt-chromium alloy or titanium alloy. Design parameters for the Pre-Milled Blanks are, a minimum wall thickness of 0.45 mm, a minimum post height for single-unit restorations of 4.0 mm, a maximum gingival height of 6.0 mm, and a maximum total abutment height of 19 mm. All Pre-Milled Blanks are for straight abutments only.
{6}------------------------------------------------
CoCr Abutments are designed as a cast-to abutment for support of a single-unit or a multi-unit restoration. They are made of cobalt-chromium alloy. Design parameters for CoCr Abutments are, a minimum wall thickness of 0.4 mm, a minimum post height for single-unit restorations of 4.0 mm, and a maximum gingival height of 6.0 mm. All CoCr Abutments are for straight abutments only.
## PERFORMANCE DATA
Non-clinical testing data submitted to demonstrate substantial equivalence included: sterilization to an SAL of 106 according to ISO 17665-1 and ISO 17665-2 to ensure sterilization of the final finished device; biological evaluation according to ISO 10993-1 and cytotoxicity testing to ISO 10993-5 to demonstrate that all devices are non-cytotoxic, and compatibility analysis by reference to K170588.
No clinical data were included in this submission.
# EQUIVALENCE TO MARKETED DEVICE
The subject device is substantially equivalent in indications and design principles to the predicate devices shown above. Below are summary tables comparing the Indications for Use and the technological characteristics of the subject device and the predicate devices.
| Indications for Use Statement | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|-----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Subject<br>Device | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| DESS Dental<br>Smart<br>Solutions<br>Terrats<br>Medical SL | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the<br>maxillary or mandibular arch to provide support for prosthetic restorations.<br>All digitally designed custom abutments for use with Aurum Base or Pre-Milled Blank are to be sent to a Terrats Medical<br>validated milling center for manufacture. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Compatible Implant Systems | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | Implant System Compatibility | Implant Body Implant Platform 3i Certain® 3.25, 4.0, 5.0 3.4, 4.1, 5.0 3i OSSEOTITE® 3.25, 3.75, 4.0, 5.0 3.4, 4.1, 5.0 OsseoSpeed™ 3.5, 4.0, 5.0 3.5/4.0, 4.5/5.0 FRIADENT XIVE 3.4, 3.8, 4.5 3.4, 3.8, 4.5 NobelActive® 3.5, 4.3, 5.0 NP, RP NobelReplace Conical 3.5, 4.3, 5.0 NP, RP Nobel Replace Trilobe 3.5, 4.3, 5.0 NP, RP, WP Brånemark 3.5, 3.75/4.0, 5.0 NP, RP, WP Straumann® Bone Level 3.3, 4.1, 4.8 NC, RC Straumann® Tissue Level 3.3, 4.1, 4.8 RN, WN Tapered Screw-Vent® 3.7, 4.1, 4.7, 6.0 3.5, 4.5, 5.7 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Primary<br>Predicate<br>Devices | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| K170588<br>DESS Dental<br>Smart<br>Solutions<br>Terrats<br>Medical SL | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary<br>or mandibular arch to provide support for prosthetic restorations.<br>All digitally designed custom abutments for use with TiBase or Pre-milled Blank are to be sent to a Terrats Medical validated<br>milling center for manufacture. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Compatible Implant Systems | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | Implant System Compatibility | Implant Diameter (mm) Platform Diameter (mm) 3i Certain® 3.25, 4.0, 5.0 3.4, 4.1, 5.0 3i OSSEOTITE® 3.25, 3.75, 4.0, 5.0 3.4, 4.1, 5.0 OsseoSpeed™ 3.5, 4.0, 5.0 3.5/4.0, 4.5/5.0 FRIADENT XIVE 3.4, 3.8, 4.5 3.4, 3.8, 4.5 NobelActive® 3.5, 4.3, 5.0 NP, RP NobelReplace Conical 3.5, 4.3, 5.0 NP, RP Nobel Replace Trilobe 3.5, 4.3, 5.0 NP, RP, WP Brånemark 3.5, 3.75/4.0, 5.0 NP, RP, WP Straumann® Bone Level 3.3, 4.1, 4.8 NC, RC Straumann® Tissue Level 3.3, 4.1, 4.8 RN, WN Tapered Screw-Vent® 3.7, 4.1, 4.7, 6.0 3.5, 4.5, 5.7 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
#### Comparison of Indications for Use Statements
{7}------------------------------------------------
| Reference<br>Devices | |
|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K092341<br>Low Profile<br>Abutment<br>Biomet 3i, Inc. | BIOMET <i>3i</i> Low Profile Abutments are intended for use as accessories to endosseous dental implants to support a prosthetic<br>device in a partially or completely edentulous patient. A dental abutment is intended for use to support single and multiple tooth<br>prosthesis, in the mandible or maxilla. The prosthesis is screw retained to the abutment. |
| K150669<br>Neoss TiBase<br>and CoCr<br>Abutments<br>Neoss Ltd. | Neoss TiBase:<br>Neoss Abutments are designed to be connected to the Neoss Implants and intended for use as an aid in prosthetic rehabilitation.<br>The Neoss TiBase is compatible with the Sirona Dental System inCoris ZI Meso L. All digitally designed copings and/or crowns<br>for use with the Neoss TiBase Abutments are to be designed and milled using the Sirona Dental CAD/CAM System.<br>Neoss CoCr Abutments:<br>Neoss abutments are designed to be connected to the Neoss Implants and intended for use as an aid in prosthetic rehabilitation. |
| K120414<br>OsseoSpeed™<br>Plus<br>Astra Tech<br>AB | Implants:<br>The Astra Tech Dental Implants are intended for both one- and two-stage surgical procedures in the following situations and with<br>the following clinical protocols:<br>replacing single and multiple missing teeth in the mandible and maxilla, immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge, especially indicated for use in soft bone applications where implants with other implant surface treatments may be less effective, immediate loading in all indications, except in single tooth situations on implants shorter than 8 mm, or in soft bone (type IV) where implant stability may be difficult to obtain and immediate loading may not be appropriate. The intended use for OsseoSpeed™ Plus 3.0S is limited to replacement of maxillary lateral incisors and mandibular incisors. |
| | Abutments:<br>Astra Tech Implant System Plus abutments are intended to be used in conjunction with Astra Tech Implant System Plus in fully<br>edentulous or partially edentulous maxillary and/or mandibular arches to provide support for crowns, bridges or overdentures.<br>Atlantis Abutments: |
| | The Atlantis™ Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or<br>completely edentulous; patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla.<br>The prosthesis can be cemented, screw retained or friction fit to the abutment. The abutment screw is intended to secure the<br>abutment to the endosseous implant. |
| | The Atlantis™ Crown Abutment in Zirconia is intended for use with an endosseous; implant to function as a substructure that<br>also serves as the final restoration, in partially or completely edentulous; patients. The prosthesis is screw retained. The abutment<br>screw is intended to secure the crown abutment to the endosseous implant. |
| K160784<br>CAM<br>Titanium<br>Blanks Altatec<br>GmbH | CAM Titanium Blanks are intended for the fabrication of abutments and healing caps on CAMLOG® SCREW LINE and<br>CAMLOG® ROOT-LINE implants in the maxilla and mandible. Digitally designed abutments fabricated with CAD/CAM<br>techniques for use with CAM Titanium Blanks are intended to be sent to a CAMLOG validated milling center for manufacture. |
| | CAM Titanium Blanks are intended for the fabrication of abutments and healing caps on CONELOG® SCREW-LINE implants<br>in the maxilla and mandible. Digitally designed abutments fabricated with CAD/CAM techniques for use with CAM Titanium<br>Blanks are intended to be sent to a CAMLOG validated milling center for manufacture. |
| | CAM Titanium Blanks are intended for the fabrication of abutments and healing caps/gingiva former on iSy® implants in the<br>maxilla and mandible. Digitally designed abutments fabricated with CAD/CAM techniques for use with CAM Titanium Blanks<br>are intended to be sent to a CAMLOG validated milling center for manufacture. |
{8}------------------------------------------------
| Comparison | Subject<br>Device | Primary<br>Predicate<br>Device | Reference Devices | | | |
|------------------------------------------------|------------------------------------------------------------------------------------|---------------------------------------------------------------|--------------------------------------------------|-------------------------------------------------------|-------------------------------------------------|---------------------------------------------|
| | K170588 | K092341 | K150669 | K120414 | K160784 | |
| | DESS<br>Dental<br>Smart<br>Solutions<br>Terrats<br>Medical SL | DESS<br>Dental<br>Smart<br>Solutions<br>Terrats<br>Medical SL | Low<br>Profile<br>Abutment<br>Biomet 3i,<br>Inc. | Neoss<br>TiBase and<br>CoCr<br>Abutment<br>Neoss Ltd. | OsseoSpeed<br>Plus<br>Astra Tech<br>AB | CAM<br>Titanium<br>Blank<br>Altatec<br>GmbH |
| Design | | | | | | |
| Abutment Design | CAD/CAM<br>Blank<br>CAD/CAM<br>TiBase<br>Castable<br>Abutment<br>Aurum<br>Abutment | CAD/CAM<br>Blank<br>CAD/CAM<br>TiBase<br>Abutment | Titanium<br>Abutment | CAD/CAM<br>TiBase<br>Castable<br>Abutment | Castable<br>Abutment | CAD/CAM<br>Blank |
| Prosthesis<br>Attachment | Cement-<br>retained<br>Screw-<br>retained | Cement-<br>retained<br>Screw-<br>retained | Screw-<br>retained | Cement-<br>retained<br>Screw-<br>retained | Cement-<br>retained<br>Screw-<br>retained | Cement-<br>retained |
| Restoration | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-<br>unit, Multi-<br>unit |
| Abutment/Implant<br>Platform<br>Diameter* (mm) | 3.4 - 6.5 | 3.4 - 6.5 | 3.4 - 5.0 | 4.1 | 3.0 - 5.4 | 3.3 - 6.0 |
| Abutment Angle | 0° | 0° | 0°- 30° | 0°- 20° | 0°- 30° | 0°- 30° |
| Abutment/<br>Implant Interface | Internal<br>and<br>External | Internal<br>and<br>External | Internal<br>and<br>External | Internal | Internal | Internal |
| Material | | | | | | |
| Abutment | Titanium<br>Alloy<br>CoCr,<br>Zirconia | Titanium<br>Alloy<br>Zirconia | Titanium<br>Alloy | Titanium<br>Alloy,<br>CoCr,<br>Zirconia | Titanium<br>Alloy<br>Zirconia,<br>Gold,<br>PEEK | Titanium<br>Alloy |
| Screw | Titanium<br>Alloy | Titanium<br>Alloy | Titanium<br>Alloy | Titanium<br>Alloy | Titanium<br>Alloy | Titanium<br>Alloy |
# Comparison of Technological Characteristics
*See Summary of Abutment Designs (p.3) for individual OEM platform diameters
{9}------------------------------------------------
DESS Dental Smart Solutions subject device abutments are substantially equivalent in intended use, design, function, material, size, and Indications for Use to DESS Dental Smart Solutions abutments (K170588). The only change in the Indications for Use statement is the name of the abutment. The Aurum Abutment is substantially equivalent to the TiBase abutment in K170588. Design parameters for minimum wall thickness, maximum gingival height and no angulation correction remain the same as the primary predicate K170588. The minimum post height for single-unit restorations is now 4.0 mm. Titanium and CoCr Pre-milled Blanks are substantially equivalent to the titanium Pre-milled Blanks in K170588. Design parameters for minimum wall thickness, minimum post height for single-unit restorations, maximum gingival height and no angulation correction remain the same as the primary predicate K170588. The CoCr Abutment is substantially equivalent in design and function to OsseoSpeed Plus abutments (K120414) and in material and function to Neoss CoCr abutments (K150669). Design parameters for minimum wall thickness, minimum post height for single-unit restorations, maximum gingival height and no angulation correction remain the same as the primary predicate K170588 and reference device K160784 and are substantially equivalent in intended use, design, function, material, size and final design parameters. The Aurum Abutment has a substantially equivalent titanium post height to the titanium post height of the Low Profile Abutment in K092341. Both are used for single-unit and multi-unit restorations. When used for a single-unit restoration the Aurum Abutment and the Low Profile Abutment are to be used with an additional castable component to create a minimum post height of 4 mm). All are intended for use with endosseous dental implants in the maxilla and mandible to provide prosthetic support. Digital files for subject device DESS Dental Smart Solutions abutments and DESS Dental Smart Solutions abutments cleared in K170588 are to be sent to a validated milling center for manufacture. Implant/abutment interface compatibility for the subject device is substantially equivalent to DESS Dental Smart Solutions abutments (K170588) and to the compatible implant system predicates listed above.
## CONCLUSION
The subject device and the predicate devices have intended use, have similar technological characteristics, and are made of similar materials. The subject device and predicate devices encompass the same range of physical dimensions, including diameter and design of the abutments. The subject and predicate devices are packaged in similar materials and are to be sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.