K151621 · BioHorizons Implant Systems, Inc. · NHA · Jan 13, 2016 · Dental
Device Facts
Record ID
K151621
Device Name
BioHorizons CAD/CAM Abutments
Applicant
BioHorizons Implant Systems, Inc.
Product Code
NHA · Dental
Decision Date
Jan 13, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BioHorizons CAD/CAM Abutments are dental abutments placed onto a dental implant to provide support for dental prosthetic restorations. The abutments include: 1) Titanium abutment blanks with a pre-machined implant connection where the upper portion may be custom-milled in accordance with a patient-specific design using CAD/CAM techniques; and 2) Titanium bases with a pre-machined implant connection upon which a CAD/CAM designed superstructure may be fitted to complete a two-piece dental abutments include an abutment screw for fixation to the underlying implant. The abutments may be used for single-tooth) or multiple-unit (bridges and bars) restorations and are compatible for use with BioHorizons Internal and Tapered Internal implant systems and Zimmer® Dental Screw-Vent® and Tapered Screw-Vent® implants with 3.5mm, 4.5mm and 5.7mm internal hex-connection mating platform diameters. All digitally designed abutments and/or copings for use with BioHorizons CAD/CAM Abutments are intended to be sent to a BioHorizons-validated milling center for manufacture. BioHorizons abutments designed using CAD/CAM techniques must fulfill the BioHorizons allowable range of design parameters.
Device Story
Device consists of titanium abutment blanks and titanium bases for custom-milled dental restorations; includes abutment screw for implant fixation. Used in dental clinics; operated by dental professionals. Workflow: clinician creates custom design using 3Shape library/software; design file (.stl) converted to numerical control (.nc) file; uploaded to BioHorizons-validated milling center for fabrication. Output is a patient-specific abutment supporting prosthetic restorations. Benefits include customized fit for functional/aesthetic dental reconstruction. Compatible with specific BioHorizons and Zimmer implant systems.
Clinical Evidence
Bench testing only. Dynamic mechanical fatigue testing performed per ISO 14801 and Class II Special Controls Guidance. Compatibility testing verified mating platform geometry against Zimmer implant subset. Steam sterilization validated per AAMI/ANSI/ISO 17665-1:2006 (SAL 10^-6). Software validation confirmed design parameter adherence.
Indicated for patients requiring dental prosthetic restorations (single-tooth or multiple-unit bridges/bars) supported by BioHorizons Internal/Tapered Internal or Zimmer Screw-Vent/Tapered Screw-Vent implants (3.5mm, 4.5mm, 5.7mm platforms).
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Core 3D Abutment System for Digital Prosthetic Solutions (K122295)
BioHorizons Tapered Internal Implant System (K071638)
BioHorizons Tapered Internal Implant System (K143022)
Zimmer Dental Screw-Vent and Tapered Screw-Vent Systems (K013227)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird in flight, composed of three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 13, 2016
BioHorizons Implant Systems, Inc. Mr. Michael Davis Director, Regulatory Affairs 2300 Riverchase Center Birmingham, Alabama 35244
Re: K151621
Trade/Device Name: BioHorizons CAD/CAM Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: December 11, 2015 Received: December 14, 2015
Dear Mr. Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use Statement (as required by ODE for all 510(k)s received after January 1, 1996)
510(k) Number: K151621
Device Name: BioHorizons CAD/CAM Abutments
Indications for Use:
BioHorizons CAD/CAM Abutments are dental abutments placed onto a dental implant to provide support for dental prosthetic restorations. The abutments include: 1) Titanium abutment blanks with a pre-machined implant connection where the upper portion may be custom-milled in accordance with a patient-specific design using CAD/CAM techniques; and 2) Titanium bases with a pre-machined implant connection upon which a CAD/CAM designed superstructure may be fitted to complete a two-piece dental abutments include an abutment screw for fixation to the underlying implant. The abutments may be used for single-tooth) or multiple-unit (bridges and bars) restorations and are compatible for use with BioHorizons Internal and Tapered Internal implant systems and Zimmer® Dental Screw-Vent® and Tapered Screw-Vent® implants with 3.5mm, 4.5mm and 5.7mm internal hex-connection mating platform diameters.
All digitally designed abutments and/or copings for use with BioHorizons CAD/CAM Abutments are intended to be sent to a BioHorizons-validated milling center for manufacture. BioHorizons abutments designed using CAD/CAM techniques must fulfill the BioHorizons allowable range of design parameters.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 510(k) Summary 21 CFR 807.92
#### Submitter's Name & Address
| Manufacturer: | BioHorizons Implant Systems, Inc.<br>2300 Riverchase Center<br>Birmingham, AL 35244<br>Phone (205) 967-7880<br>Fax (205) 870-0304 |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Official contact: | Michael Davis, Director, Regulatory Affair |
| Date prepared: | January 13, 2016 |
#### Name of the Device
| Trade Name: | BioHorizons CAD/CAM Abutments |
|------------------------|------------------------------------|
| Common or Usual Name: | Dental implant abutment |
| Classification Name: | Endosseous dental implant abutment |
| Classification Number: | Class II (21 CFR 872.3630) NHA |
#### Predicate Devices
Primary Predicate Device:
- 1. K122295, Core 3D Abutment System for Digital Prosthetic Solutions, October 24, 2013.
Reference Predicate Devices:
- 1. K111935, NT-Trading Ti-Base Abutment, 2-CONnect Abutment, February 17, 2012.
- 2. K103691, BioHorizons Abutments for Zimmer, November 3, 2011.
- 3. K121787, BioHorizons Tapered Internal Plus Implants, September 5, 2012.
- 4. K073268, BioHorizons Internal Implant System, February 8, 2008.
- 5. K071638, BioHorizons Tapered Internal Implant System, October 10, 2007.
- 6. K143022, BioHorizons Tapered Internal Implant System, January 8, 2015.
- 7. K013227, Zimmer® Dental (formerly Sulzer Dental) Screw-Vent® and Tapered Screw-Vent® Systems, November 19, 2001.
#### Device Description
BioHorizons CAD/CAM Abutments are dental implant final restorative abutments supplied in platform diameters of 3.0mm, 3.5mm, 4.5mm and 5.7mm. The abutments are intended to provide support for dental prosthetic restorations. Each abutment includes an abutment screw for fixation to the underlying implant. Abutment material is titanium alloy as specified in ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy (UNS R56401) for Surgical Implant Applications.
Select abutments are further processed by applying patterns of micro-machined grooves or channels, known as Laser-Lok, to a specified region of the abutment margin. The abutments are provided non-sterile, and they are packaged using materials known in the industry to be appropriate for medical device packaging.
#### Indications for Use
BioHorizons CAD/CAM Abutments are dental abutments placed onto a dental implant to provide support for dental prosthetic restorations. The abutments include: 1) Titanium abutment
{4}------------------------------------------------
blanks with a pre-machined implant connection where the upper portion may be custom-milled in accordance with a patient-specific design using CAD/CAM techniques; and 2) Titanium bases with a pre-machined implant connection upon which a CAD/CAM designed superstructure may be fitted to complete a two-piece dental abutments include an abutment screw for fixation to the underlying implant. The abutments may be used for single-unit (single-tooth) or multiple-unit (bridges and bars) restorations and are compatible for use with BioHorizons Internal and Tapered Internal implant systems and Zimmer® Dental Screw-Vent® and Tapered Screw-Vent® implants with 3.5mm, 4.5mm and 5.7mm internal hex-connection mating platform diameters.
All digitally designed abutments and/or copings for use with BioHorizons CAD/CAM Abutments are intended to be sent to a BioHorizons-validated milling center for manufacture. BioHorizons abutments designed using CAD/CAM techniques must fulfill the BioHorizons allowable range of design parameters.
### Technological Characteristics
The fundamental scientific technology of the BioHorizons CAD/CAM Abutments that are the subject of this 510(k) is substantially equivalent to the existing abutments of the referenced predicate devices. (Refer to the following Summary Table of Substantial Equivalence for comparison of the subject device to the declared predicate devices.) Select abutment models are further processed by applying Laser-Lok to a specified region of the abutment margin.
Both the subject device and predicate devices include similar design features such as abutment base material, surface treatment, implant-abutment connection tailored to the implant system(s) with which the devices are intended to be used, anti-rotational geometry, prosthetic platform sizes and mode of retention which thereby demonstrates substantial equivalence in their respective designs. Any specific differences related to prosthetic platform sizes between the subject device and the predicate devices do not raise questions of safety or effectiveness based on the performance testing conducted on the worst-case manufactured dimensions which supports the determination that the subject devices are appropriate for their intended use and do not render the devices not substantially equivalent. The features, similarities and differences are further summarized in tabular format in Table 1. Summary Table of Substantial Equivalence following in this section. The Zimmer compatible platforms are identical to the declared predicate, K103691. The BioHorizons 3.0mm platform compatible implant is identical to the declared predicate device cleared in K143022. The compatible platform implants are substantially equivalent to the declared predicates.
All materials, suppliers and packaging methods remain the same as those utilized for the predicate BioHorizons Abutments for Zimmer (K103691). Additionally, the Laser-Lok feature is substantially equivalent to that cleared for the BioHorizons Abutments for Zimmer (K103691). The BioHorizons CAD/CAM Abutments are substantially equivalent to the features of the predicate abutment devices which could affect safety or effectiveness because of the similarities in design, materials and intended use.
## Abutment Manufacturing
A custom abutment design file is created by the customer using a 3Shape abutment library. The abutment design file is converted to a stereolithography (.stl) file using 3Shape software. The .stl file is converted to a numerical control (.nc) file using an appropriate software application (e.g.
{5}------------------------------------------------
SUM3D). After the custom abutment design is uploaded, the milling machine software executes the necessary commands to mill the final abutment. The digitally designed file is to be sent to a registered and listed BioHorizons contract manufacturer for milling. BioHorizons CAD/CAM Abutments are compatible with commercially available dental CAD/CAM systems, such as 3Shape.
## Summary of Testing
Dynamic mechanical fatigue testing was performed in accordance with the Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments, May 12, 2004 and ISO 14801. Results of the testing demonstrate that the fully milled, worst-case 30° angled configurations of both the 3.0mm and 3.5mm prosthetic platform abutments and the titanium bases with bonded zirconia superstructure perform in accordance with their intended use.
Compatibility testing was performed on a representative subset of Zimmer® Screw-Vent® and Tapered Screw-Vent® implants. The subset included the following Zimmer® item numbers: TSV4B8, TSV4B10, TSVWH10, TSVWB10, TSVWB11, TSVWH11, TSVT6B10, TSV6H11, TSV6H13 and TSV6H16. This testing verifies compatibility of BioHorizons Abutments for Zimmer® with all Zimmer® Screw-Vent® and Tapered Screw-Vent® items based on equivalent mating platform geometry.
Steam sterilization validation testing was performed new in accordance with AAMI/ANSVISO 17665-1:2006, Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices. Test results demonstrate a sterility assurance level (SAL) of 10°.
Finally, software intended use validation testing of the software systems utilized in the manufacture of the CAD/CAM abutments was performed to ensure that the program design limitations prevent the user from milling abutments that do not fulfill the BioHorizons design criteria.
#### Conclusion
The data presented in this submission demonstrates that the proposed devices are substantially equivalent with respect to performance and intended use. The proposed devices perform as well as the legally marketed predicate devices. Furthermore, the proposed devices do not pose any new or increased risks as compared to the legally marketed predicate devices.
{6}------------------------------------------------
# Table 1. Summary Table of Substantial Equivalence
| | Subject Device | Primary Predicate Device | Reference Predicate Devices | | | | Compatible Implant Devices | | | | |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| | BioHorizons Implant Systems,<br>Inc.<br>BioHorizons CAD/CAM<br>Abutments | Core 3D Abutment System for<br>Digital Prosthetic Solutions<br>K122295 | NT-Trading Ti-Base Abutment, 2-<br>CONnect Abutment<br>K111935 | BioHorizons Implant Systems,<br>Inc.<br>BioHorizons Abutments for<br>Zimmer K103691 | | Subject Device | BioHorizons Implant Systems, Inc.<br>BioHorizons CAD/CAM<br>Abutments | BioHorizons Implant Systems, Inc.<br>Tapered Internal Plus Implants<br>K121787 | BioHorizons Implant Systems, Inc.<br>Internal Implant System<br>K073268 | BioHorizons Implant Systems, Inc.<br>Tapered Internal Implant System<br>K071638 & K143022 | Zimmer® Dental (formerly<br>Sulzer Dental) Screw-Vent® and<br>Tapered Screw-Vent® Systems<br>K013227 |
| Classification/<br>Product Code | 872.3630/NHA | 872.3630/NHA | 872.3630/NHA | 872.3630/NHA | Design | | | | | | |
| Intended Use | BioHorizons CAD/CAM Abutments are<br>dental abutments placed onto a dental<br>implant to provide support for dental<br>prosthetic restorations. The abutments<br>include: 1) Titanium abutment blanks<br>with a pre-machined implant connection<br>where the upper portion may be custom-<br>milled in accordance with a patient-<br>specific design using CAD/CAM<br>techniques; and 2) Titanium bases with a<br>pre-machined implant connection upon<br>which a CAD/CAM designed<br>superstructure may be fitted to complete<br>a two-piece dental abutment. The<br>abutments include an abutment screw for<br>fixation to the underlying implant. The<br>abutments may be used for single-unit<br>(single-tooth) or multiple-unit (bridges<br>and bars) restorations and are compatible<br>for use with BioHorizons Internal and<br>Tapered Internal implant systems and<br>Zimmer® Dental Screw-Vent® and<br>Tapered Screw-Vent® implants with<br>3.5mm, 4.5mm and 5.7mm internal hex-<br>connection mating platform diameters.<br>All digitally designed abutments and/or<br>copings for use with BioHorizons<br>CAD/CAM Abutments are intended to<br>be sent to a BioHorizons-validated<br>milling center for manufacture.<br>BioHorizons abutments designed using<br>CAD/CAM techniques must fulfill the<br>BioHorizons allowable range of design<br>parameters. | The CORE 3D abutment system for<br>digital prosthetic solutions are dental<br>abutments placed into a dental implant to<br>provide support for dental prosthetic<br>restorations. The abutments include:<br>• Titanium Bases to be attached to the<br>underlying implant and upon which a<br>CAD/CAM designed superstructure<br>may be fitted to complete a two-<br>piece dental abutment;<br>• Titanium Abutment Blanks with a<br>pre-machined implant connection<br>where the upper portion may be<br>custom-milled in accordance with a<br>patient-specific design using<br>CAD/CAM techniques:<br>• Abutment Screws to permanently fix<br>the abutments to the underlying<br>implant.<br>Core 3D abutments are intended for use<br>to support single-tooth (unit) and<br>multiple-tooth (bridges and bars)<br>prostheses, in the mandible or maxilla for<br>functional and aesthetic restorations.<br>Core 3D abutments designed using<br>CAD/CAM techniques must fulfill the<br>Core 3D allowable range of design<br>specifications and be provided as straight<br>abutments only.<br>Core 3D abutments and are compatible<br>for use with the following dental<br>implants:<br>• Nobel Biocare® Branemark<br>System™ (K022562, K934825)<br>• Zimmer® Tapered Screwvent®<br>(K013227, K061410, K072589) | Ti-Base for individual Zirconium Abutments: The<br>devices covered by this submission are abutments<br>which are placed into a dental implant to provide<br>support for a prosthetic restoration.<br>The Ti-Base for individual Zirconium Abutments are<br>intended for use to support a prosthetic device in a<br>partially or completely edentulous patient. It is<br>intended for use to support single and multiple tooth<br>prosthesis, in the mandible or maxilla. The prosthesis<br>can be cement retained to the abutment. The<br>abutment screw is intended to secure the abutment to<br>the endosseous implant.<br>The Ti-Base abutments are indicated for use with the<br>following implant systems:<br>• Nobel Biocare® Replace Select®<br>• Nobel Biocare No be !Active™<br>• Biomet 3i® Osseotite®<br>• Biomet 3i® Osseotite® Certain®<br>• Nobel Biocare Branemark®<br>• Straumann® synOcta®<br>• Straumann® Bone Level®<br>• Zimmer® Tapered Screw-vent®<br>• Astra Tech OsseoSpeed®<br>• Dentsply-Friadent® Frialit®<br>2-CONnect Abutment for Bridges and Bars: 2-<br>CONnect Abutment for Bridges and Bars is indicated<br>for use to provide support for prosthetic restorations<br>such as bars and bridges. The 2-CONnect abutments<br>can be used in multiple tooth restorations. The 2-<br>CONnect abutment can be used together with<br>cemented bridges and bar constructions for functional<br>and aesthetical reconstruction.<br>The 2-CONnect abutments are indicated for use with<br>the following implant systems:<br>• Nobel Biocare® Replace Select®<br>• Straumann® synOcta® | BioHorizons Abutments for Zimmer are<br>abutments that include healing abutments<br>for contouring tissue and final restorative<br>abutments to support a prosthesis. The<br>abutments may be used for a single or<br>multiple unit restoration and are<br>compatible for use with BioHorizons<br>Internal and Tapered Internal implant<br>systems and Zimmer® Dental Screw-<br>Vent® and Tapered Screw-Vent®<br>implants with 3.5mm, 4.5mm and 5.7mm<br>internal hex-connection mating platform<br>diameters. | Implant Body<br>Diameter | | N/A | 3.8mm, 4.6mm, 5.8mm | 3.5mm, 4.0mm, 5.0mm, 6.0mm | 3.4mm, 3.8mm, 4.6mm, 5.8mm | 3.3mm, 3.7mm, 4.1mm, 4.7mm,<br>6.0mm |
| | Subject Device | Primary Predicate Device | Reference Predicate Devices | | Implant length | | N/A | 7.5mm (except 3.8mm body), 9mm,<br>10.5mm, 12mm, 15mm | 9mm, 10.5mm, 12mm, 15mm | 7.5mm (except 3.8mm body), 9mm,<br>10.5mm, 12mm, 15mm, 18mm<br>(except 5.8mm body) | 8mm, 10mm, 11.5mm, 13mm,<br>16mm |
| | BioHorizons Implant Systems,<br>Inc.<br>BioHorizons CAD/CAM<br>Abutments | Core 3D Abutment System for<br>Digital Prosthetic Solutions<br>K122295 | NT-Trading Ti-Base Abutment, 2-<br>CONnect Abutment<br>K111935 | BioHorizons Implant Systems,<br>Inc.<br>BioHorizons Abutments for<br>Zimmer K103691 | External<br>thread | | N/A | Buttress | Square | Buttress | V-shaped (Triple-Lead Thread) |
| Design | | | | | Surface | | All abutments (except 3.0mm<br>platform): Anodized on the hex<br>connection<br>Select abutments: TiN coated; Laser-<br>Lok in specified zone between<br>prosthetic margin and the implant-<br>abutment junction (IAJ) | Implant - RBT<br>Collar - Laser-Lok | Implant - RBT or HA<br>Collar - Laser-Lok or RBT | Implant - RBT or HA<br>Collar - Laser-Lok or RBT | Implant - MTX Microtexture or<br>HA<br>Collar - Machined |
| Surface | All abutments (except 3.0mm<br>platform): Anodized on the hex<br>connection<br>Select abutments: TiN coated;<br>Laser-Lok in specified zone<br>between prosthetic margin and<br>the implant-abutment junction<br>(IAJ) | Not known | Not known | All abutments: Anodized on the<br>hex connection<br>Select abutments: TiN coated;<br>Laser-Lok in specified zone<br>between prosthetic margin and the<br>implant-abutment junction (IAJ) | Connection | | Internal Hex | Internal Hex | Internal Hex | Internal Hex | Internal Hex |
| Laser-Lok<br>Zone | Groove size: 8µm width and<br>depth<br>Min. height (vertical dimension):<br>0.5mm | N/A | N/A | Groove size: 8µm width and depth<br>Min. height (vertical dimension):<br>0.5mm | Internal<br>thread | | N/A | Spiralock UNF 1-72 | Spiralock UNF 1-72 | Spiralock UNF 1-72 | UNF 1-72 |
| Connection | Internal Hex | Varies, External and Internal<br>dependent on implant system<br>compatibility | Varies, dependent on implant system<br>compatibility | Internal Hex | Prosthetic<br>platform | | 3.0mm, 3.5mm, 4.5mm, 5.7mm | 3.0mm, 3.5mm, 4.5mm | 3.5mm, 4.5mm, 5.7mm | 3.0mm, 3.5mm, 4.5mm, 5.7mm | 3.5mm, 4.5mm, 5.7mm |
| Prosthetic<br>platform | 3.0mm, 3.5mm, 4.5mm, 5.7mm | 3.5mm, 4.1mm, 4.5mm, 5.1mm,<br>5.7mm | 3.5mm (minimum), 6.5mm (maximum)…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.