The Voyant System is indicated to provide minimally invasive access, visualization, illumination, and magnification of the surgical area of the spine.
Device Story
Voyant System is a sterile, single-use minimally invasive spine surgery access and visualization device. It consists of a tubular retractor (Voyant Sheath) and an external camera (Voyant Imager) with an LED light source. The Imager attaches to the Sheath and connects via HDMI to an Image Control Box (ICB). The ICB powers the Imager, processes video data, and outputs a live feed to an external HD monitor in the operating room. Surgeons use the system to visualize the surgical site during spinal procedures. The device provides magnification and illumination to assist in surgical decision-making and procedural execution. The system is intended for use by physicians in an operating room setting.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical performance verification testing, including design verification, software validation, electrical safety (IEC 60601-1), and electromagnetic compatibility (IEC 60601-1-2).
Technological Characteristics
System comprises a tubular retractor (Parylene coated 6061-T6 Aluminum, medical grade adhesive) and an external CMOS camera (1/4" sensor, 3280x2464 resolution, 1080p output). Illumination via integrated LED. Connectivity via HDMI to an external Image Control Box. Sterilization of single-use components via Ethylene Oxide (EtO) to 10^-6 SAL. Complies with IEC 60601-1, IEC 60601-1-2, IEC 62304, and EN ISO 14971.
Indications for Use
Indicated for patients undergoing minimally invasive spine surgery requiring access, visualization, and magnification of the surgical area.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
July 11, 2019
Viseon, Inc. Cora Sim Regulatory Affairs Manager 13900 Alton Parkway, Suite 125 Irvine, California 92618
Re: K191579
Trade/Device Name: Voyant System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: June 13, 2019 Received: June 14, 2019
Dear Cora Sim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191579
Device Name Voyant System
Indications for Use (Describe)
The Voyant System is indicated to provide minimally invasive access, visualization, illumination, and magnification of the surgical area of the spine.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for VISEON INC. The logo consists of a stylized V shape in teal and blue, followed by the word "VISEON" in blue, block letters. Below the word "VISEON" is the word "INC." in smaller, blue letters.
## Voyant System 510(k) Summary - K191579
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
#### A. Submitted by:
Cora Sim Regulatory Affairs Manager Viseon, Inc. 13900 Alton Parkway, Suite 125 Irvine, CA 92618 Telephone: (949) 662-3959 ext. 106 Facsimile: (949) 581-4761
Date Prepared: July 1, 2019
#### B. Device Name:
| Trade or Proprietary Name: | Voyant System |
|----------------------------|-------------------|
| Common or Usual Name: | Arthroscope |
| Classification Name: | Arthroscope |
| Device Class: | Class II |
| Classification: | 21 CFR § 888.1100 |
| Product Code: | HRX |
#### C. Predicate Device:
The subject Voyant System is substantially equivalent to the predicate device, Voyant System (K181885).
#### D. Device Description:
The Voyant System is composed of the Voyant Sheath and Voyant Imager, and Accessories which includes an Image Control Box (ICB) and connecting cables. The Voyant Sheath and Voyant Imager are supplied sterile by ethylene oxide (EtO) while the ICB and connecting cables are supplied non-sterile and are The Voyant System is used in minimally invasive spine surgery to provide access and reusable. visualization of the surgical area of the spine.
The Voyant System is a sterile, single-use access and visualization device consisting of a Sheath for access to the spine and an Imager for illumination and visualization of the surgical area. The Sheaths are available in multiple diameters and lengths to accommodate a variety of patient anatomies and spinal surgical procedures. Likewise, the Voyant Imager is available in corresponding diameters to attach to the Voyant Sheath. The Voyant Imager connects to the Image Control Box (ICB) via an HDMI cable and live video image is displayed to a connected external HD monitor in the operating room. The ICB functions to power the Voyant Imager, process video data from the Voyant Imager to display live video image to the connected third-party HD Monitor.
The purpose of this submission is to implement design changes to the predicate Voyant System (K181885) to create a product line extension which includes the Voyant Imager, and Accessories. These modifications to the Voyant System do not change the indications for use of the device, nor do they change the fundamental scientific technology of the device. Specifically, the following design and labeling modifications are described in this submission:
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo on the left and the word "Spec" on the right. The logo consists of a blue letter N and the word "INC." written below it. The word "Spec" is written in a serif font and is positioned to the right of the logo.
- Design modifications to the Voyant Sheath including additional diameters and lengths ●
- Material change of the patient contacting components of the Voyant System
- Design modifications to the Voyant Imager .
- Design modifications to the Image Control Box (ICB) .
- Labeling changes to include updated images and instructions based upon the modified subject device ● and additional warning based upon the improved performance of the subject device.
The models for the subject device are listed in the table below:
### Voyant System Models:
| | Model<br>Number | Package Contents** | | |
|---------------------------------------|-----------------|--------------------------|---------------|---------|
| | | Voyant Sheath | Voyant Imager | |
| Sterile, Single-use,<br>Voyant System | VSY-1640 | 16mm OD | 40mm length | 16mm OD |
| | VSY-1650 | | 50mm length | |
| | VSY-1660 | | 60mm length | |
| | VSY-1670 | | 70mm length | |
| | VSY-1680 | | 80mm length | |
| | VSY-1840 | 18mm OD | 40mm length | 18mm OD |
| | VSY-1850 | | 50mm length | |
| | VSY-1860 | | 60mm length | |
| | VSY-1870 | | 70mm length | |
| | VSY-1880 | | 80mm length | |
| | VSY-2040 | 20mm OD | 40mm length | 20mm OD |
| | VSY-2050 | | 50mm length | |
| | VSY-2060 | | 60mm length | |
| VSY-2070 | 70mm length | | | |
| VSY-2080 | 80mm length | | | |
| VSY-2240 | 22mm OD | 40mm length | | 22mm OD |
| VSY-2250 | | 50mm length | | |
| VSY-2260 | | 60mm length | | |
| VSY-2270 | | 70mm length | | |
| VSY-2280 | | 80mm length | | |
| VSY-2640 | 26mm OD | 40mm length | | 26mm OD |
| VSY-2650 | | 50mm length | | |
| VSY-2660 | | 60mm length | | |
| VSY-2670 | | 70mm length | | |
| VSY-2680 | | 80mm length | | |
| Non-sterile, Reusable,<br>Accessories | ICB-02*** | Voyant Image Control Box | | |
** One Voyant Sheath of specified diameter (OD) and length and one Voyant Imager of corresponding diameter (OD) are packaged together
***VLBL08-002 Operator's Manual
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for VISEON INC. The logo features a stylized "V" shape in teal and blue, followed by the word "VISEON" in a blue, sans-serif font. Below the word "VISEON" is the word "INC." in a smaller, sans-serif font.
#### E. Indications for Use:
The Voyant System is indicated to provide minimally invasive access, visualization, and magnification of the surgical area of the spine.
#### F. Comparison of Technological Characteristics with the Predicate Device:
As was established in this submission, the subject Voyant System is substantially equivalent to the predicate Voyant System (K181885) previously cleared by FDA for commercial distribution in the United States. There have been no design changes to the devices previously cleared in the predicate 510(k). The subject device has been shown to be substantially equivalent technological characteristics to its predicate device through comparison in areas including design, labeling/intended use, principles of operation, materials, and function.
| Feature | PREDICATE DEVICE<br>Voyant System | SUBJECT DEVICE<br>Voyant System |
|--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K181885 | K191579 |
| Product Code | HRX | Same |
| Indications for Use | The Voyant System is indicated to provide<br>minimally invasive access, visualization,<br>illumination, and magnification of the<br>surgical area of the spine. | Same |
| Intended Use | Visualization of surgical area. | Same |
| Design | The Voyant System is an access and<br>visualization system designed to provide<br>illumination and magnification of the<br>surgical area during minimally invasive<br>spine surgery.<br>The system consists of the Voyant Sheath<br>and the Voyant Imager and Accessories.<br>The Voyant Sheath is a tubular retractor<br>used to provide access to the surgical area.<br>The Voyant Imager is an external camera<br>with LED light source that attaches to the<br>Voyant Sheath. Accessories to the Voyant<br>Imager include an Image Control Box (ICB)<br>with connecting cables.<br>The Image Control Box is a portable<br>accessory used to control the image output<br>of the Voyant Imager, which is displayed on<br>a connected operating room HD monitor. | Same |
| Patient Contacting<br>Materials (Tubular<br>Retractor) | Anodized 6061-T6 Aluminum and medical<br>grade plastic and adhesive | Parylene coated 6061-T6<br>Aluminum and medical grade<br>adhesive |
| Feature | PREDICATE DEVICE | SUBJECT DEVICE |
| | Voyant System | Voyant System |
| Labeling (Single<br>use/Reusable) | Voyant Sheath and Voyant Imager:<br>Sterile, Single use<br>Image Control Box and Connecting<br>Cables: Non-sterile, Reusable | Same |
| Surgical Site Access<br>Device | Tubular retractors | Same |
| Image Acquisition | Image acquisition is achieved through an<br>integrated camera external to the surgical<br>opening | Same |
| Image Processing | The image is digitally processed. | Same |
| Image Display | External monitor connection. | Same |
| Illumination | Illumination is achieved via direct<br>transmission using an LED light source. | Same |
| Sterilization | Voyant Sheath and Voyant Imager:<br>• Supplied sterile by Ethylene Oxide<br>(EtO) to a sterility assurance level of<br>$10^{-6}$ , for single use<br>Image Control Box: Supplied non-sterile<br>and reusable. | Same |
| Performance<br>Characteristics | Camera resolution: 1080p<br>Sensor resolution: 3280 x 2464 pixels<br>Camera sensor: 1/4" CMOS<br>Tubular retractor compressive force: $\geq$ 100N<br>Image resolution: $\geq$ 11 lp/mm at 63mm<br>Working length: 6-13cm | Camera resolution: Same<br>Sensor resolution: Same<br>Camera sensor: Same<br>Tubular retractor compressive<br>force: Same<br>Image resolution: $\geq$ 11 lp/mm at<br>80mm<br>Working length: 4-13cm |
### Design Comparison Table for Substantial Equivalence:
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for VISEON INC. The logo consists of a stylized letter V in two shades of blue, followed by the word "VISEON" in a sans-serif font, also in blue. Below the word "VISEON" is the abbreviation "INC." in a smaller font size and the same shade of blue.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for VISEON Inc. The logo features a stylized letter "V" in two shades of blue, with the left side being a lighter teal color and the right side a darker blue. To the right of the "V" is the company name, "VISEON," in a bold, sans-serif font, also in the darker blue color. Below the "ON" portion of the name, "INC." is written in a smaller font size.
#### G. Performance Data
Non-clinical performance verification testing was performed to demonstrate that the subject Voyant System is substantially equivalent to the predicate device. Testing included design verification testing, software validation, electrical safety and electromagnetic emissions and immunity testing was performed in accordance with the following standards in order to establish equivalence to the predicate device:
| Test/Document Description | Applicable Test Standard |
|----------------------------------------|--------------------------|
| Electrical Safety | IEC 60601-1 |
| Electromagnetic Emissions and Immunity | IEC 60601-1-2 |
| Software | IEC 62304 |
| Risk Management | EN ISO 14971 |
| Design Verification Testing | N/A |
The results demonstrate that the subject Voyant System is substantially equivalent to the predicate.
#### H. Conclusions
Based on the indications for use, technological characteristics, non-clinical performance data, and a comparison to the predicate device, the subject Voyant System has been shown to be substantially equivalent to the legally marketed predicate device for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.