K191118 · Avanti Orthopaedics, Inc. · HRS · Dec 13, 2019 · Orthopedic
Device Facts
Record ID
K191118
Device Name
Avanti Distal Radius and Forearm System
Applicant
Avanti Orthopaedics, Inc.
Product Code
HRS · Orthopedic
Decision Date
Dec 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The system intended for fixation of fractures, osteotomies, malunions, and non-unions, involving the radius, ulna, and carpals. Additionally, the wrist fusion plates may be used for wrist arthrodesis.
Device Story
System comprises metallic bone fixation plates (volar, wrist fusion, spanning, forearm, ulnar) and fasteners (screws, pegs); used for fracture fixation, osteotomies, malunions, non-unions, and wrist arthrodesis. Plates feature pre-contoured anatomy, low-profile design, and bi-planar angulated screw holes. Volar plates include pre-assembled PEEK inserts allowing variable angle (up to 11 degrees) for locking/non-locking fasteners. Surgeons operate device in clinical/OR settings to stabilize bone segments. Output is physical fixation of bone; aids in bone healing and structural restoration.
Clinical Evidence
Bench testing only. Performed static and dynamic plate loading (ASTM F382), screw pullout strength (ASTM F543), and locking mechanism push-out testing to demonstrate substantial equivalence.
Technological Characteristics
Plates/screws manufactured from medical grade stainless steel (ASTM F138/F139), electropolished (ASTM F86). Volar plates include PEEK inserts (Invibio, LTD). Features low-profile geometry, bi-planar angulated screw holes, and compression or locking cortical screw options. Mechanical fixation device; no software or energy source.
Indications for Use
Indicated for fixation of fractures, osteotomies, malunions, and non-unions of the radius, ulna, and carpals; wrist arthrodesis using wrist fusion plates. Patient population includes individuals requiring orthopedic bone fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Biomet Distal Radius Volar Rim (DVR) Plating System (K132704)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 13, 2019
Avanti Orthopaedics, Inc. % Linda Braddon President/CEO Secure BioMed Evaluations 7828 Hickory Flat Highway Suite 120 Woodstock, Georgia 30188
Re: K191118
Trade/Device Name: Avanti Distal Radius and Forearm System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: November 13, 2019 Received: November 14, 2019
Dear Linda Braddon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801: medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page.
510(k) Number (if known)
K191118
Device Name
Avanti Distal Radius and Forearm System
Indications for Use (Describe)
The system intended for fixation of fractures, osteotomies, and non-unions, involving the radius, ulna, and carpals. Additionally, the wrist fusion plates may be used for wrist arthrodesis.
Type of Use (Select one or both, as applicable) □ Over-The-Counter Use (21 CFR 801 Subpart C) 区 Prescription Use (Part 21 CFR 801 Subpart D)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
K191118 Page 1/2
Image /page/3/Picture/1 description: The image shows the logo for Avanti Orthopaedics. The logo is in blue and features a stylized letter "A" with a swooping line extending from the bottom of the "A" to underline the rest of the word "vanti". Below the word "Avanti" is the word "ORTHOPAEDICS" in smaller, sans-serif font.
In accordance with 21 CFR 807.87 (h) and 21 CRF 807.92, the 510(k) summary for the Avanti Distal Radius and Forearm System is provided below.
| Date Summary<br>Prepared | November 13, 2019 |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Avanti Orthopaedics, Inc<br>4606 Simon Road<br>Wilmington, DE 19803<br>Phone 302-530-6369<br>Fax 302-351-4896 |
| 510(k) Contact | Secure BioMed Evaluations<br>Linda Braddon, Ph.D.<br>7828 Hickory Flat Highway<br>Suite 120<br>Woodstock, GA 30188<br>770-837-2681<br>Regulatory@SecureBME.com |
| Trade Name | Avanti Distal Radius and Forearm System |
| Common Name | Single/Multiple component metallic bone fixation appliances and accessories |
| Code -Classification | HRS, 21 CFR 888.3030: Class II |
| Primary Predicate | K132704 Biomet Distal Radius Volar Rim (DVR) Plating System |
| Additional<br>Predicates | K042355 Synthes LCP Wrist Fusion Plates<br>K951304 Synthes 2.7 MM Cannulated Screw and Threaded Washer<br>K000684 Synthes Small Fragment Dynamic Compression Locking (DCL) System<br>K160995 In2Bones SAS Neoview Plate |
| Device Description | The Avanti Distal Radius and Forearm System includes volar plates, wrist fusion<br>plates, spanning plates, forearm plates, ulnar plates, and cannulated screws,<br>with screw and peg fasteners, for the fixation of fractures, malunions, and<br>osteotomies. Distal radius and forearm fixation plates may be placed on the<br>volar or dorsal surface of the radius and along the shaft of the radius and ulna<br>on the carpal bones. Features of the system include pre-contoured plates to fit<br>the anatomy of the distal radius, a low-profile design to minimize soft tissue<br>irritation, low contact design to optimize blood supply, and bi-planar angulated<br>screw holes for ease of drilling and to optimize fixation and subchrondral<br>support.<br>All plates are manufactured from medical grade stainless steel per ASTM F138<br>or ASTM F139 and are electropolished per ASTM F-86. The Volar PEEK Plate has<br>a polyetheretherkeytone that is preassembled to the plate and is sourced from<br>Invibio, LTD. The PEEK insert allows insertion of both locking and non-locking<br>fasteners with an infinite variable angle up to 11 degrees using the drill guide |
| | provided in the system. All plates include low-profile geometry to minimize soft<br>tissue irritation and the option of compression or locking cortical screws.<br><br>The Avanti Distal Radius and Forearm System includes self-tapping screws, listed<br>below, that are manufactured from medical grade stainless steel per ASTM<br>F138, electropolished per ASTM F-86, and are compatible with all Avanti plate<br>configurations. All fasteners but the Cannulated Screw are intended to be used<br>with the plates. The cannulated screws can be used for fracture fixation that is<br>appropriately sized for the device which is determined through surgeon<br>discretion. |
| Intended Use | The system intended for fixation of fractures, osteotomies, malunions, and non-<br>unions, involving the radius, ulna, and carpals. Additionally, the wrist fusion<br>plates may be used for wrist arthrodesis. |
| Technological<br>Characteristics | The subject Avanti Distal Radius and Forearm System is technological similar to<br>other similarly intended devices in that it has the same technological<br>characteristics to its predicate devices through comparison in areas including<br>design, intended use, material composition, function, and range of sizes. The<br>subject system differences technologically in that the volar plate comes with a<br>pre-assembled PEEK insert. The PEEK insert allows insertion of both locking and<br>non-locking fasteners with variable angle up to 11 degrees using the drill guide<br>provided in the system. |
| Non-Clinical<br>Performance<br>Testing Conclusion | Non-clinical testing was performed to demonstrate the Avanti Distal Radius and<br>Forearm System is substantially equivalent to other predicate devices. The<br>following tests were performed:<br>Static and dynamic plate loading per ASTM F382 Screw pullout strength via ASTM F543 Locking Mechanism Push-Out Testing |
| Substantial<br>Equivalence<br>Summary | The Avanti Distal Radius and Forearm System has been shown to be substantially<br>equivalent to legally marketed predicate devices. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Avanti Orthopaedics. The logo is blue and features the word "Avanti" in a stylized font. Below "Avanti" is the word "ORTHOPAEDICS" in a smaller, sans-serif font. The logo is simple and professional.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.