K160995 · In2bones Sas · HRS · Jun 10, 2016 · Orthopedic
Device Facts
Record ID
K160995
Device Name
NEOVIEW Plating System
Applicant
In2bones Sas
Product Code
HRS · Orthopedic
Decision Date
Jun 10, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The NEOVIEW® Plating System is intended for fixation of intra-articular fractures of the distal radius and reconstruction of the distal radius.
Device Story
The NEOVIEW® Plating System is a bone fixation device consisting of PEEK plates and titanium alloy screws. It is used in hospitals or outpatient surgery centers by surgeons to stabilize distal radius fractures. The plates are radiolucent and include tantalum markers for visualization. The system is supplied sterile for single use. Surgeons use the plates and screws to fix bone fragments, facilitating reconstruction and healing of the distal radius. The device provides mechanical stability to the fracture site.
Clinical Evidence
Bench testing only. Mechanical testing included static and dynamic bending tests performed by an independent laboratory. The device met all acceptance criteria. Animal and clinical testing were deemed not applicable.
Technological Characteristics
Plates: PEEK (ASTM F2026) with tantalum markers (ASTM F560). Screws: Titanium alloy Ti-6Al-4V (ISO 5832-3, ASTM F136). Principle: Mechanical bone fixation via locking or non-locking screws. Form factor: 4 plate designs, various screw lengths. Sterilization: Gamma irradiation.
Indications for Use
Indicated for fixation of intra-articular and extra-articular fractures of the distal radius and reconstruction of the distal radius in patients requiring bone fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Piccolo Composite™ Distal Volar Radius Plate (K102597)
Related Devices
K160002 — Piccolo Composite(R) Distal Volar Radius Plate System · CarboFix Orthopedics , Ltd. · Mar 9, 2016
K251132 — The Acumed Wrist Fixation System - Plates; The Acumed Wrist Fixation System - Screws · Acumed, LLC · Jul 14, 2025
K090374 — STERILE DVR ANATOMIC PLATES VARIOUS CONFIGURATIONS) MODEL (S) 131212001,2, 5-10 2.5MM THREADED AND SMOOTH PEGS, MODEL(S · DePuy Orthopaedics, Inc. · Mar 11, 2009
K050932 — DISTAL VOLAR RADIUS ANATOMICAL PLATE SYSTEM · Hand Innovations, Inc. · Apr 26, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling an abstract symbol. The symbol is a stylized representation of a human figure, possibly suggesting care and well-being. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 10, 2016
In2bones SAS % Norman Estrin, Ph.D. Managing Partner Estrin Consulting Group LLC 9109 Copenhaver Drive Potomac, Maryland 20854
Re: K160995
Trade/Device Name: NEOVIEW® Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: April 7, 2016 Received: April 11, 2016
Dear Dr. Estrin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K160995
Device Name NEOVIEW® Plating System
Indications for Use (Describe)
The NEOVIEW® Plating System is intended for fixation of intra-articular fractures of the distal radius and reconstruction of the distal radius.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;"> <span style="vertical-align: middle;"></span> <span style="vertical-align: middle;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> <span> <span style="vertical-align: middle;"></span> <span style="vertical-align: middle;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </span> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for IN2BONES. The logo is set on a gray oval shape. The text "IN2BONES" is in white, with the "2" in orange.
## 510(k) SUMMARY For In2Bones NEOVIEW® Plating System
| Sponsor identification | In2Bones SAS<br>28 chemin du Petit Bois<br>69130 Ecully – France<br>Phone: +33.4.72.29.26.26<br>Fax: +33.4.72.29.26.29 |
|------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>registration number | 3010470577 |
| Date of preparation | March 21st, 2016 |
| Contact person | Norman F. Estrin, Ph.D.<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Cell: 240-994-9999<br>Email: estrin@yourFDAconsultant.com |
| Authorized Agent in<br>the United States<br>NEOVIEW® Plating<br>System | Norman F. Estrin, Ph.D.<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Cell: 240-994-9999<br>Email: estrin@yourFDAconsultant.com |
| Proprietary Name | NEOVIEW® Plating System |
| Common name | NEOVIEW® Distal Radius Plate |
| Device classification<br>regulation | 21 CFR 888.3030: Single/multiple component metallic bone<br>fixation appliances and accessories<br>Class II |
| Device Product<br>Code and Panel | HRS: plate, fixation, bone<br>87 orthopedics |
{4}------------------------------------------------
| Device Description | The NEOVIEW® Plating System is composed of the<br>NEOVIEW® plates and the associated NEO screws.<br>The NEOVIEW® plate is design to adequately maintain the<br>bone fragments of distal radius and is made of PEEK, a<br>material recognized for its mechanical and radiolucent<br>properties.<br>The fixation of NEOVIEW® plate to the bone is performed<br>by the associated NEO screws that are available on locking<br>or not locking versions. |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Sizes: The NEOVIEW® plates include 4 different designs.<br>The associated NEO screws are available in one diameter<br>and a large range of lengths packaged individually of in a<br>kit.<br>Material: The NEOVIEW® plates are made of PEEK<br>according to standard ASTM F2026 and include a marker<br>made of tantalum according to ASTM F560. The NEO<br>screws are made of Titanium alloy Ti-6Al-4V according to<br>ISO 5832-3 and ASTM F136. |
| | Single use: The NEOVIEW® Plating System is designed for<br>single use only.<br>Sterilization: The NEOVIEW® Plating System is supplied<br>sterile, using gamma irradiation.<br>Place of use: The NEOVIEW® Plating System is indicated<br>for use in a hospital, or outpatient surgery center where |
| Predicate Devices | sterile field may be created and maintained.<br>Piccolo Composite™ Distal Volar Radius Plate (K102597) |
| Indications for use: | The NEOVIEW® Plating System is intended for fixation of<br>intra-articular and extra-articular fractures of the distal<br>radius and reconstruction of the distal radius. |
{5}------------------------------------------------
| Comparison of the<br>indications for use<br>with the predicate<br>devices: | The indications for use for the NEOVIEW® Plating System are<br>similar to the predicate device Piccolo Composite™ Distal Volar<br>Radius Plate (K102597) in intended use, design, material,<br>technological characteristics and principles of operation. |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison of<br>Technological<br>characteristics and<br>Substantial<br>Equivalence<br>Summary: | The NEOVIEW® Plating System is similar to the predicate device<br>Piccolo Composite™ Distal Volar Radius Plate (K102597) in<br>intended use, design, size ranges, principle of operation and<br>materials. |
| Summary<br>Performance Data | Performance testing of the NEOVIEW® Plating System was<br>assessed through mechanical bench testing performed by an<br>independent test laboratory, animal and clinical testing being<br>considered not applicable.<br>Testing performed included static and dynamic bending tests on<br>NEOVIEW® Plating System. The results of the testing performed<br>by the independent test laboratory indicate that the NEOVIEW®<br>Plating System met the acceptance criteria. |
| CONCLUSION | Based on the comparison of indications for use and<br>technological characteristics and the results of the testing<br>performed, the NEOVIEW® Plating System is substantially<br>equivalent to the predicate device identified in the 510(k)<br>submission. |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.