K181947 · Zimmer, Inc. · JWH · Oct 25, 2018 · Orthopedic
Device Facts
Record ID
K181947
Device Name
Persona Revision Knee System
Applicant
Zimmer, Inc.
Product Code
JWH · Orthopedic
Decision Date
Oct 25, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. Collagen disorders, and/or avascular necrosis of the femoral condyle. Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. Moderate valgus, varus, or flexion deformities. The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. Porous components may be used cemented or uncemented (biological fixation). Augments may be attached via bone cement or screw to the tibial plates and/or femoral components. Splined stem extension components are intended to be used press-fit (uncemented). All other femoral component, tibial plate, stem extension, and femoral and tibial augment components are indicated for cemented use only.
Device Story
Persona Revision Knee System is a semi-constrained total knee prosthesis; provides modular components to resurface femoral and tibial articulating surfaces. Includes femoral components, articular surfaces, tibial components, stem extensions, augments, and cones. Used by orthopedic surgeons in primary or revision total knee arthroplasty (TKA) to address bone and joint conditions. Modular design allows various configurations to optimize joint stability and function. Components are either cemented or press-fit (uncemented) depending on type. Device restores joint configuration and stability, potentially reducing pain and disability for patients with severe knee pathology.
Clinical Evidence
No clinical data was deemed necessary. Evidence consists of non-clinical bench testing, including wear testing (ISO 14243-1), constraint testing (ASTM F2083, ASTM F1223), contact area/pressure analysis (ASTM F1672, ASTM F2083), fatigue testing (ASTM F1800, ASTM F2009), and finite element analysis for stress and impingement. Backwards compatibility with the Persona Primary Knee System was also verified via engineering analysis.
Technological Characteristics
Semi-constrained total knee prosthesis. Materials: Metal/Polymer. Modular components: femoral, tibial, articular, stem extensions, augments, cones. Fixation: Cemented or uncemented (porous/press-fit). Standards: ASTM F1672 (patellofemoral), ASTM F2083/F1223 (constraint), ASTM F1800 (fatigue), ASTM F2009 (taper connection), ISO 14243-1 (wear). Analysis: Finite element analysis used for stress, fatigue, and impingement. MR conditional labeling included for specific stem extensions.
Indications for Use
Indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle, post-traumatic joint loss, or moderate valgus/varus/flexion deformities. Also indicated for salvage of failed surgical attempts or knees lacking satisfactory stability in flexion.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
October 25, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Zimmer Inc. Nicole Meredith Regulatory Project Manager P.O. Box 708 Warsaw, Indiana 46581-0708
Re: K181947
Trade/Device Name: Persona Revision Knee System Regulation Number: 21 CFR 888.3565 Regulation Name: Knee Joint Patellofemorotibial Metal/Polymer Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: MBH, JWH, OIY Dated: September 26, 2018 Received: September 27, 2018
Dear Ms. Meredith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Peter G. Allen -S 2018.10.25 18:22:34 -04'00'
Mark N. Melkerson FOR Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K181947
Device Name Persona® Revision Knee System
#### Indications for Use (Describe)
This device is indicated for patients with severe knee pain and disability due to:
- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Collagen disorders, and/or avascular necrosis of the femoral condyle.
- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- Moderate valgus, varus, or flexion deformities.
- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
Porous components may be used cemented (biological fixation). Augments may be attached via bone cement or screw to the tibial plates and/or femoral components. Splined stem extension components are intended to be used press-fit (uncemented). All other femoral component, tibial plate, stem extension, and femoral and tibial augment components are indicated for cemented use only.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Persona® Revision Knee System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | Zimmer Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708<br>Establishment Registration Number: 1822565 |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Nicole J. Meredith<br>Regulatory Project Manager<br>Telephone: (574) 377-3718<br>Fax: (574) 372-4710 |
| Date: | July 18, 2018 |
| Subject Device: | Trade Name: Persona® Revision Knee System<br>Common Name: Knee Prosthesis |
| | Classification Name:<br>● MBH – Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer (21 CFR 888.3565)<br>● JWH – Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (21 CFR 888.3560)<br>● OIY – Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive (21 CFR 888.3560) |
| Predicate Devices: | Zimmer Persona K133737<br>The Personalized Knee System K130143<br>K123459<br>K122765<br>K121771<br>K113369<br>NexGen Complete Knee Solution K173057<br>K152494 |
{4}------------------------------------------------
| | | K960279 |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| | | K946150 |
| | | K933785 |
| | NexGen TM Augments and Cones | K103517 |
| | | K040630 |
| | | K031962 |
| | | K024161 |
| | Vanguard SSK 360 Revision Knee<br>System | K171054 |
| | | K121149 |
| | | K093293 |
| | | K042757 |
| | Natural-Knee II | K173057 |
| | | K023528 |
| | | K013031 |
| | | K982903 |
| | | K973412 |
| | | K972501 |
| Purpose and Device Description: | The purpose of this submission is for introduction of the<br>Persona Revision Knee System. | |
| | The Persona Revision Knee System is a semi-constrained<br>total knee prosthesis consisting of anatomically shaped<br>components designed to resurface the articulating surface<br>of the femoral and tibial bones including: | |
| | • Femoral components | |
| | • Articular surfaces | |
| | • Tibial components | |
| | • Stem extensions | |
| | • Femoral and tibial augments | |
| | • Femoral and tibial cones | |
| | The large modularity of the componentry of the Persona<br>Revision knee system including articular surfaces with<br>different levels of constraint, augments, cones and stem<br>extensions provides numerous possible configurations to<br>optimally address the bone and joint condition of the<br>patient in a primary or revision TKA surgery. | |
In addition, this submission includes the addition of MR conditional labeling for the Persona 14 x 30mm Tapered
{5}------------------------------------------------
Stem Extension cleared under the K133737 on February 2, 2014.
Intended Use and Indications for Use: This device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, । polyarthritis. -Collagen disorders, and/or avascular necrosis of the femoral condyle. Post-traumatic loss of joint configuration, particularly | when there is patellofemoral erosion, dysfunction or prior patellectomy. | Moderate valgus, varus, or flexion deformities. The salvage of previously failed surgical attempts or | for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. Porous components may be used cemented or uncemented (biological fixation). Augments may be attached via bone cement or screw to the tibial plates and/or femoral components. Splined stem extension components are intended to be used press-fit (uncemented). All other femoral component, tibial plate, stem extension, and femoral and tibial augment components are indicated for cemented use only. Summary of Technological The rationale for substantial equivalence is based on Characteristics: consideration of the following characteristics: ● Intended Use: Identical to the predicates Indications for Use: Identical to the predicates . ● Materials: Identical to the predicates
- Design Features: Similar to the predicates ●
- Sterilization: Identical to the predicates ●
## Summarv of Performance Data (Nonclinical and/or Clinical)
#### Non-Clinical Tests: ● Femoral Components
- Patellofemoral durability per internal test method O
- Patellofemoral lateral constraint per ASTM F1672 O
- Patellofemoral contact area and contact pressure O per ASTM F1672
- Posterior condyle fatigue per engineering analysis O
{6}------------------------------------------------
- O Intercondylar notch and anterior chamfer strength per engineering analysis
- Anterior flange strength via finite element O analysis
- Stem housing cantilever fatigue per internal test O method
## Articular Surfaces
- Wear per ISO 14243-1 o
- Anterior spine impingement via finite element o analysis
- Constraint per ASTM F2083 and ASTM F1223 O
- Contact area and contact pressure per ASTM O F2083
- Anterior lift-off per engineering analysis O
- Posterior lift-off per engineering analysis O
- Combined internal/external (I/E) and varus/valgus O (V/V) spine loading via finite element analysis
- Static varus/valgus constraint per internal test O method
- o Assembly/insertion force per internal test method
- Secondary lockdown screw assembly strength per O internal test method
- Secondary lockdown screw torque after gait-type o loading per internal test method
### Tibial Components
- Cantilever fatigue per ASTM F1800 o
- Tibial stem housing cantilever fatigue per internal O test method
- Locking mechanism static shear strength per o engineering analysis
- Fixation stability per internal test method O
### Augments
- O Shear attachment strength per internal test method
- Attachment screw assembly strength per internal O test method
### Cones
- Tibial cone stress under physiologic loading per O finite element analysis
- Femoral cone stress under physiologic loading per o finite element analysis
{7}------------------------------------------------
## Stem Extensions
- Tibial taper connection axial pull-off and torsional O strength per ASTM F2009
- Femoral taper connection axial pull-off and O torsional strength per ASTM F2009
- Splined stem tip fatigue strength per internal test o method
- Cantilever fatigue via finite element analysis per O ASTM F1800
## System
- Fretting/corrosion of modular junctions per o engineering analysis
- MR compatibility per engineering analysis O
- Testing to establish product non-pyrogenicity O
## Backwards Compatibility with Persona Primary Knee System
- Persona Revision femoral components wear on o Persona Primary PS and CPS articular surfaces per engineering analysis
- Persona Revision femoral components posterior O crush loading of the Persona Primary PS and CPS articular surfaces per finite element analysis
- Persona Revision femoral components anterior O lift-off of the Persona Primary PS and CPS articular surfaces per engineering analysis
- Persona Revision femoral components posterior O lift-off of the Persona Primary PS and CPS articular surfaces per engineering analysis
- Contact area and contact pressure of the Persona o Revision femoral components with the Persona Primary PS and CPS articular surfaces per engineering analysis
- Tibiofemoral constraint of the Persona Revision O femoral components with the Persona Primary PS and CPS articular surfaces per engineering analysis
- Persona Revision femoral components anterior о spine impingement with the Persona Primary PS and CPS articular surfaces per engineering analysis
- Cantilever fatigue of the Persona Revision tibial O components with the Persona Primary PS and CPS articular surfaces and Persona Primary tibial
{8}------------------------------------------------
components with Persona Revision articular surfaces per engineering analysis
#### . Clinical Tests:
- Clinical data was not deemed necessary for the subject device.
## Substantial Equivalence Conclusion
The subject device has the same intended use and indications for use as the predicate devices. The subject device has similar technological characteristics to the predicates, and the performance data and analyses demonstrate that:
- any differences do not raise new questions of safety and effectiveness; and
- . the proposed device is at least as safe and effective as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.