K972501 · Sulzer Orthopedics, Inc. · JWH · Oct 1, 1997 · Orthopedic
Device Facts
Record ID
K972501
Device Name
APOLLO REVISION/CONSTRAINED KNEE SYSTEM
Applicant
Sulzer Orthopedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Oct 1, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Apollo™ Revision/Constrained Knee System is intended use in the following diagnostic indications: - Patient conditions, including but not limited to, inflammatory degenerative 1. joint disease (e.g., rheumatoid arthritis) and noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis). - Correctable valgus-varus deformity and moderate flexion contracture. 2. - Those patients with failed previous surgery where pain, deformity, or 3. dysfunction persist. - Revision of previously failed knee arthroplasty. 4. The Apollo Revision/Constrained Knee System is intended only for use with bone cement in the United States. Uncemented use of the Apollo Revision/Constrained Knee System is considered to be investigational in the United States. This device is intended for single use only.
Device Story
Apollo Revision/Constrained Knee System is a semi-constrained total knee replacement for primary and revision surgeries. Device components include cobalt chromium/molybdenum alloy femoral component, Ti-6Al-4V tibial baseplate/spacers, and UHMWPe tibial inserts. System features high medial/lateral constraint via a vertical Ti-6Al-4V pin connecting the tibial insert to the baseplate. Intended for use with bone cement; requires intact collateral ligaments. Used by orthopedic surgeons in clinical/OR settings to resurface knee joints, restore stability, and alleviate pain/dysfunction in patients with failed prior arthroplasty or degenerative disease. Output is a mechanical prosthetic joint replacement; clinical decision-making relies on surgeon assessment of ligamentous integrity and bone quality.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and design comparison to legally marketed predicate devices.
Indicated for patients with inflammatory (e.g., rheumatoid arthritis) or noninflammatory (e.g., osteoarthritis, avascular necrosis) degenerative joint disease, correctable valgus-varus deformity, moderate flexion contracture, or failed previous knee surgery requiring revision arthroplasty. Requires intact medial and lateral collateral ligaments; posterior cruciate ligament may be absent, deficient, or removed.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Apollo PS Knee System (Sulzer Orthopedics)
Kinemax Plus Superstabilizer (Howmedica)
PFC TC3 (Johnson & Johnson)
Reference Devices
Apollo All-Poly Elliptical Patella
Apollo PS All-Poly Tibia
Apollo Posterior Stabilized (PS) Tibial Insert
Apollo Tibial Baseplate I
Apollo All-Poly Single-Lug Patella
Natural-Knee II Revision Femoral Stems
Apollo Congruent Tibial Insert
Submission Summary (Full Text)
{0}------------------------------------------------
OCT - I 1997
K972501
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this is to serve as a Summary of Safety and Effectiveness for the Apollo™ Revision/Constrained Knee System.
| Submitter: | Sulzer Orthopedics Inc.<br>9900 Spectrum Drive<br>Austin, Texas 78717<br>(512) 432-9900 |
|-------------------------|--------------------------------------------------------------------------------------------------------------|
| Date: | July 1, 1997 |
| Contact Person: | Jacquelyn Hughes<br>Manager, Regulatory Affairs |
| Classification Name: | Prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer/ metal/polymer 21 CFR 888.3560 |
| Common/Usual Name: | Knee Prosthesis, Partially Constrained |
| Trade/Proprietary Name: | Apollo™ Revision/Constrained Knee System |
#### PRODUCT DESCRIPTION
The Apollo™ Revision/Constrained Knee System is a non-porous total knee replacement intended for primary and revision knee surgeries. The components are intended for resurfacing of the knee joint where the posterior cruciate ligament is absent, deficient, or has been removed at the time of surgery. The system is a semiconstrained design and both medial and lateral collateral ligaments must be intact.
The Apollo Revision/Constrained Knee femoral component is manufactured from cobalt chromium/molybdenum alloy (CoCr) with seperate left and right components. Femoral spacers manufactured from Ti-6Al-4V are also available for use with the Apollo Revision/Constrained Knee femoral component.
The Apollo Revision/Constrained Knee tibial components include a baseplate, spacers, and polyethylene insert. The tibial components are also symmetrical in design, eliminating the need for left/right orientations. The tibial baseplate and tibial spacers is manufactured from Ti-6A1-4V. The tibial inserts are manufactured from Ultra High Molecular Weight Polyethylene (UHMWPe). A Ti-6Al-4V pin passing vertically from the constrained condylar (high medial/lateral constraint) eminence of the tibial insert to the tibial baseplate provides additional stability.
The Apollo Revision/Constrained Knee System is to be used with the Apollo all poly patellar components.
{1}------------------------------------------------
The Natural-Knee II Revision Stems may be used with either the Apollo Revision/Constrained Knee System femoral or tibial components.
This device is intended for use with the following previously cleared devices:
- 비 Apollo All-Poly Elliptical Patella
- I Apollo PS All-Poly Tibia
- Apollo Posterior Stabilized (PS) Tibial Insert and Apollo Tibial Baseplate I
- 별 Apollo All-Poly Single-Lug Patella
- 트 Natural-Knee II Revision Femoral Stems
The Apollo Revision/Constrained Knee Tibial Baseplate may also be used with the Apollo Congruent Tibial Insert. The Apollo Congruent Tibial Insert is not compatible with the Apollo Revision/Constrained Knee Femoral Component.
## SPECIFIC DIAGNOSTIC INDICATIONS
The Apollo™ Revision/Constrained Knee System is intended use in the following diagnostic indications:
- Patient conditions, including but not limited to, inflammatory degenerative 1. joint disease (e.g., rheumatoid arthritis) and noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis).
- 2. Correctable valgus-varus deformity and moderate flexion contracture.
- 3. Those patients with failed previous surgery where pain, deformity, or dysfunction persist.
- Revision of previously failed knee arthroplasty. 4.
The Apollo™ Revision/Constrained Knee System is intended only for use with bone cement in the United States. Uncemented use of the Apollo™ Revision/Constrained Knee System is considered to be investigational in the United States. This device is intended for single use only.
### SUBSTANTIAL EQUIVALENCE
The Apollo™ Revision/Constrained Knee System is substantially equivalent to the Apollo PS Knee System (Sulzer Orthopedics), the Kinemax Plus Superstabilizer (Howmedica), and the PFC TC3 ( Johnson & Johnson).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure, with three overlapping profiles facing to the right. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT - 1 1997
Mr. Mitchell Dhority Requlatory Affairs Specialist Sulzer Orthopedics, Inc. 9900 Spectrum Drive Austin, Texas 78717
K972501 Re : Apollo Revision/Constrained Knee System Regulatory Class: II Product Code: JWH Dated: July 2, 1997 Received: July 3, 1997
Dear Mr. Dhority:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). This decision is based on this device being equivalent only to similar devices labeled and intended to be fixed within bone with acrylic "bone cement." You may, therefore, market your device subject to the general controls provisions of the Act and the following limitations:
:
- The thinnest tibial insert available is the nominal "11 1. mm" sized insert, which has a minimum polyethylene thickness under the condyles of 8.00 mm.
- This device may not be labeled or promoted for non-2. cemented use.
- 3. All labeling for this device, including package label and labeling included within the package, must prominently state that the device is intended for cemented use only.
{3}------------------------------------------------
## Page 2 - Mr. Mitchell Dhority
- Any non-cemented fixation of this device is considered 4 . investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulation under 21 CFR, Part 812. All users of the device for noncemented fixation must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) to conduct the investigation.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General requlation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be
{4}------------------------------------------------
# Page 3 - Mr. Mitchell Dhority
obtained from the Division of Small Manufacturers Assistance obcaincd from cho (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
C. Colla M. Witton, Ph.D., M.
Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: Apollo™ Revision/Constrained Knee System
# Indications For Use:
The Apollo™ Revision/Constrained Knee System is intended use in the following diagnostic indications:
- Patient conditions, including but not limited to, inflammatory degenerative 1. joint disease (e.g., rheumatoid arthritis) and noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis).
- Correctable valgus-varus deformity and moderate flexion contracture. 2.
- Those patients with failed previous surgery where pain, deformity, or 3. dysfunction persist.
- Revision of previously failed knee arthroplasty. 4.
The Apollo Revision/Constrained Knee System is intended only for use with bone cement in the United States. Uncemented use of the Apollo Revision/Constrained Knee System is considered to be investigational in the United States.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------|---------|
| Division of General Restorative Devices | |
| 510(k) Number | K972501 |
| Prescription Use | <div style="text-align:center;">X</div> |
|------------------|-----------------------------------------|
|------------------|-----------------------------------------|
OR
| Over-The-Counter Use |
|----------------------|
|----------------------|
000077
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.