K103517 · Zimmer Trabecular Metal Technology, Inc. · MBH · Feb 15, 2011 · Orthopedic
Device Facts
Record ID
K103517
Device Name
TRABECULAR METAL FEMORAL CONE AUGMENTS
Applicant
Zimmer Trabecular Metal Technology, Inc.
Product Code
MBH · Orthopedic
Decision Date
Feb 15, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3565
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Trabecular Metal Femoral Cone Augments are intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates total knee arthroplasty. When used with the NexGen Complete Knee Solution - Rotating Hinge Knee (RHK) System, the Trabecular Metal Femoral Cone Augments are for cemented use only. When used with the NexGen Complete Knee Solution - Legacy Constrained Condylar Knee System, the Trabecular Metal Femoral Cone Augments are for cementless or cemented use.
Device Story
Trabecular Metal Femoral Cone Augments are porous tantalum implants used as adjuncts in total knee arthroplasty. Designed for use with Zimmer's NexGen Complete Knee Solution (Legacy Constrained Condylar Knee or Rotating Hinge Knee systems). Implants are fixed to the superior (non-articulating) side of the femoral component via cement; apposition to bone in the medullary canal is achieved with or without cement depending on the system. Used by orthopedic surgeons in clinical settings to address bone loss or structural requirements during revision or complex primary knee procedures. Benefits include structural augmentation of the femoral component. Device line extension adds new sizes and height options to existing portfolio.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing (static compression, static shear, axial compression fatigue, rotational beam fatigue, wear testing, assembly torque) and comparative Finite Element Analysis (FEA) simulating gait activities (walking, stair ascent/descent, deep flexion).
Technological Characteristics
Material: Trabecular Metal (porous tantalum). Form factor: Tapered elliptical cross-section cone augments. Fixation: Cemented or cementless depending on host system. Mechanical properties defined by Young's modulus in tension/compression. No software or electronic components.
Indications for Use
Indicated for patients requiring total knee arthroplasty due to severe degeneration, trauma, or other knee joint pathology.
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
Zimmer's (formerly Implex) Continuum Knee System Trabecular Metal (formerly Hedrocel) Tibial Spacers (K982302)
Zimmer's (formerly Implex) Continuum Knee System Trabecular Metal (formerly Hedrocel) Revision Femoral Spacers (K0980781)
Submission Summary (Full Text)
{0}------------------------------------------------
K103517 * 1/3 FEB 15 2011
## 510(k) Summary
| Submitter: | Zimmer Trabecular Metal Technology, Inc.<br>10 Pomeroy Road<br>Parsippany, New Jersey 07054 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Judith Rosen<br>Senior Regulatory Affairs Specialist<br>Telephone: (973) 576-0032 ext 28138<br>Fax: (973) 884-8792 |
| Date: | November 30, 2010 |
| Trade Name: | Trabecular Metal™ Femoral Cone Augments |
| Common Name: | Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented and uncemented |
| Classification Name: | "Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis";<br>and<br>"Knee joint patellofemorotibial metal/polymer porous-coated<br>uncemented prosthesis" |
| Reference: | 21 CFR § 888.3560, JWH and 21 CFR § 888.3565, MBH |
## DEVICE DESCRIPTION
The NexGen® Trabecular Metal™ Femoral Cone Augments are manufactured wholly of Trabecular Metal™, porous tantalum. The existing, commercially available NexGen® Trabecular Metal™ Femoral Cone Augments have a tapered elliptical cross-section. The periphery of the distal end is larger than that of the proximal end. The cone is not fully enclosed at its top most proximal section (giving the impression of U-Shape through slot). A similar pattern is also present in the modified femoral cones, with the exception that these cones are fully enclosed at its top most proximal section.
Both the modified femoral cones augments and the existing, commercially available femoral cone augments are to be used in conjunction with Zimmer's NexGen® Complete Knee Systems -both the Legacy® Constrained Condylar Knee (LCCK) Stemmed Femoral Implant and the femoral component of Zimmer's Rotating Hinge Knee (RHK) System. When used with the RHK System, the Trabecular Metal™ Femoral Cone Augments are for cemented use only. Fixation of all of the cone augments to the femoral implant is accomplished by cementing the internal portion of the augment to the superior side (non-articulating side) of the femoral component. Apposition of the cone to bone inside the medullary canal is with or without cement when used with the LCCK femoral implant and with cement for the RHK femoral implant.
{1}------------------------------------------------
K103517A²/₃
The existing, commercially available augments come in three sizes; small, medium and large and three height options. The proposed additional femoral cones will come in the same three sizes: small, medium and large and two height options.
## INDICATIONS FOR USE
Trabecular Metal Femoral Cone Augments are intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates total knee arthroplasty. When used with the NexGen Complete Knee Solution - Rotating Hinge Knee (RHK) System, the Trabecular Metal Femoral Cone Augments are for cemented use only. When used with the NexGen Complete Knee Solution - Legacy Constrained Condylar Knee System, the Trabecular Metal Femoral Cone Augments are for cementless or cemented use.
# DEVICE TECHNOLOGICAL CHARACTERISTICS AND COMPARISON TO PREDICATE DEVICES
The NexGen® Trabecular Metal™ Femoral Cone Augments were shown to be substantially equivalent to legally marketed predicate devices. The predicate devices are Zimmer's Trabecular Metal™ Femoral Cone Augments (K040630, K051756, K053340), Zimmer's NexGen® Trabecular Metal™ Tibial Cone Augments (K031962, K053340), Zimmer's NexGen® Complete Knee Solution - Trabecular Metal™ Augments (K024161, K040487), Zimmer's (formerly Implex) Continuum Knee System Trabecular Metal™ (formerly Hedrocel®) Tibial Spacers (K982302), and Zimmer's (formerly Implex) Continuum Knee System Trabecular Metal™ (formerly Hedrocel®) Revision Femoral Spacers (K0980781).
The NexGen® Trabecular Metal™ Femoral Cone Augments have the same material as previously cleared predicate devices. The intended use and indications for use of the subject devices are identical to that of the predicate. This submission is for an extension of the product line; sizes, cross sectional dimensions, design features and overall geometry of the device in the current submission are similar to the predicate device.
There are no significant differences between the proposed NexGen® Trabecular Metal™ Femoral Cone Augments and the predicate currently being marketed that would adversely affect the use of the product. Any differences in technological characteristics do not raise new issues of safety and efficacy.
## PERFORMANCE DATA
Performance testing of Trabecular Metal™ material is included in a master file, on file at the FDA, which includes static compression, static shear, axial compression fatigue and rotational beam fatigue. Young's modulus of elasticity of Trabecular Metal has been measured in tension and compression. The results of testing and analysis define the properties of Trabecular Metal™, porous tantalum.
Mechanical testing was performed on Trabecular Metal predicate devices which include fatigue testing, wear testing, and assembly torque testing. The results of testing and analyses conducted indicate that predicates of the Trabecular Metal™ Femoral Cone Augments should maintain assembly integrity when assembled using the surgical technique instructions.
{2}------------------------------------------------
KI03517 #3/3
A comparative Finite Element Analysis (FEA) study was performed to evaluate the strength of the proposed femoral cone designs across different normal gait activities (walking, stair ascent/decent and deep flexion). The results of testing and analyses conducted demonstrate that the proposed implant adequately meets the predetermined requirements established for its mechanical performance, supporting substantial equivalence to the predicate.
### CONCLUSION
The NexGen® Trabecular Metal™ Femoral Cone Augments product line extension is the same as the predicate device with respect to intended use/indications for use, technological characteristics and basic principles of operation. This product line extension does not present any new issues of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Zimmer Trabecular Metal Technology, Inc. % Ms. Judith Rosen 10 Pomeroy Road Parsippany, New Jersey 07054
FEB 15 201
Re: K103517
Trade/Device Name: Trabecular Metal Femoral Cone Augments Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH, JWH Dated: November 30, 2010 Received: November 30, 2010
Dear Ms. Rosen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
..
{4}------------------------------------------------
Page 2 – Ms. Judith Rosen
comply with all the Act's requirements, including, but not limited to: registration and lising (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Aliz B. nh
for
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
#### Indications for Use
510(k) Number (if known): K103517
Device Name: Trabecular Metal Femoral Cone Augment
Indications for Use:
ﺳﻴﺪ ﺳﻴﺔ
Trabecular Metal Femoral Cone Augments are intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates total knee arthroplasty. When used with the NexGen Complete Knee Solution - Rotating Hinge Knee (RHK) System, the Trabecular Metal Femoral Cone Augments are for cemented use only. When used with the NexGen Complete Knee Solution - Legacy Constrained Condylar Knee System, the Trabecular Metal Femoral Cone Augments are for cementless or cemented use.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Page*1*of*1*
for M.Mclkrøm
(Division Sign-Oft) (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103517
Zimmer TMT, Inc. Traditional 510(k) - Trabecular Metal Femoral Cone Augments
PROPRIETARY AND CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.