K123459 · Zimmer, Inc. · MBH · Feb 22, 2013 · Orthopedic
Device Facts
Record ID
K123459
Device Name
ZIMMER PERSONA TM PERSONALIZED KNEE SYSTEM
Applicant
Zimmer, Inc.
Product Code
MBH · Orthopedic
Decision Date
Feb 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3565
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is indicated for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. Collagen disorders, and/or avascular necrosis of the femoral condyle. Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. Moderate valgus, varus, or flexion deformities. The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. Porous coated components may be used cemented or uncemented (biological fixation). All other femoral, tibial baseplate, and all-polyethylene (UHMWPE and VEXLPE) patella components are indicated for cemented use only.
Device Story
Zimmer Persona Personalized Knee System is a semiconstrained modular knee prosthesis for resurfacing femoral, tibial, and patellar bones. System includes Constrained Posterior Stabilized (CPS) Vivacit-E articular surface components providing moderate varus/valgus and internal/external rotation constraint. Device components are implanted by orthopedic surgeons in a clinical setting. CPS components are intended for use with cemented nonporous tibial and femoral components. Porous-coated components allow for cemented or uncemented biological fixation. Device provides mechanical stability and joint articulation to alleviate pain and restore function in patients with severe knee pathology. Clinical benefit is derived from improved joint mechanics and wear resistance compared to conventional polyethylene.
Clinical Evidence
No clinical data was required or presented for this device. Substantial equivalence was established through non-clinical bench testing, including anterior tibiofemoral constraint evaluation, anterior/posterior liftoff testing, spine/underspine fatigue testing, varus/valgus loading tests, contact area/pressure evaluation, wear testing, and EO residual testing.
Technological Characteristics
Modular semiconstrained knee prosthesis. Materials include Vivacit-E (highly crosslinked polyethylene) and metal components. Features include modular tibial plates and articular surfaces. Sterilization via ethylene oxide (EO). Designed for cemented or uncemented (porous-coated) fixation. MRI compatible.
Indications for Use
Indicated for patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle, post-traumatic joint configuration loss, or moderate valgus/varus/flexion deformities. Also indicated for salvage of failed surgical attempts or cases requiring improved stability in flexion.
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
{0}------------------------------------------------
# FEB 2 2 2013
P.O. Box 708 Warsaw. IN 46581-0708 574 267-6131
### Summary of Safety and Effectiveness
Sponsor:
Contact Person:
Date:
Trade Name:
Product Codes / Device:
Regulation Numbers / Description:
Predicate Device:
Zimmer, Inc. P.O. Box 708 Warsaw, IN 46581-0708
Daniel J. Williman Project Manager, Regulatory Affairs Telephone: (574) 371-8065 Fax: (574) 372-4605
January 25, 2013
Zimmer® Persona™ Personalized Knee System
MBH, OIY, JWH
21 CFR § 888.3565 - Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis
21 CFR § 888.3560 - Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis
PersonaTM Personalized Knee System (K113369, cleared March 27, 2012)
Zimmer® Persona™ The Personalized Knee System (K121771, cleared November 7, 2012)
NexGen Complete Knee Solution Legacy Posterior Stabilized (PS) and Constrained Condylar Knee (LCCK). manufactured by Zimmer, Inc. (K960279 cleared April 26, 1996)
NexGen Complete Knee Solution LPS-Flex Prolong Highly Crosslinked Polyethylene Articular Surfaces, manufactured by Zimmer, Inc. (K042271, cleared October 13, 2004)
{1}------------------------------------------------
#### Device Description:
The Zimmer Persona Personalized Knee System is a semiconstrained modular knee prosthesis designed to resurface the articulating surface of the femoral, tibial and patellar bones. The Persona Knee System utilizes a modular design between the tibial plates and articular surfaces. The addition of the Constrained Posterior Stabilized (CPS) Vivacit-E® articular surface components will provide surgeons with the ability to obtain moderate varus/valgus and/or internal/external rotation constraint compared to the existing Persona PS articular surfaces. CPS articular surfaces are for use with cemented nonporous tibial and femoral components only.
This device is indicated for patients with severe knee pain and disability due to:
- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polvarthritis.
- Collagen disorders, and/or avascular necrosis of the femoral condyle.
- Post-traumatic loss of ioint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- Moderate valgus, varus, or flexion deformities.
- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
Porous coated components may be used cemented or uncemented (biological fixation). All other femoral, tibial baseplate, and all-polyethylene (UHMWPE and VEXLPE) patella components are indicated for cemented use only.
The proposed Zimmer Persona Personalized Knee System components are similar or identical in intended use. materials, sterility, and performance characteristics to the predicate devices.
Clinical Performance and Conclusions:
Clinical data and conclusions were not needed for this device.
Non-Clinical Performance and Conclusions:
Intended Use:
Comparison to Predicate Device:
Performance Data (Nonclinical and/or Clinical):
{2}------------------------------------------------
Vivacit-E material characteristics for the CPS articular surfaces are identical to the material characteristics of the currently marketed Zimmer Vivacit-E Polyethylene Liners (K120370) and Persona Vivacit-E articular surfaces (K121771). In contrast to conventional polyethylene, the Vivacit-E material is delamination resistant and exhibits a reduction in wear according to knee simulator bench testing.
Bench testing outlined below was conducted in support of this device:
| Property or Characteristics | Test Results |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Anterior Tibiofemoral Constraint<br>Evaluation of the Persona CPS Vivacit-E<br>UHMWPE Articular Surfaces with the<br>Lowest Anterior Jump Height | Demonstrated that the Persona CPS Vivacit-E articular<br>surfaces provide adequate anterior subluxation constraint<br>through the necessary tibiofemoral flexion movements. |
| Anterior and Posterior Liftoff Testing of the<br>Persona CPS Vivacit-E UHMWPE Articular<br>Surfaces | Demonstrated sufficient locking mechanism strength to<br>survive potential worst case anterior and posterior liftoff<br>loading conditions, respectively. |
| Evaluation of Interactions of the Zimmer<br>Persona Primary and Persona Porous Knee<br>Implant Systems with the Magnetic Fields in<br>the Magnetic Resonance Imaging (MRI)<br>Environment | Demonstrated safety and compatibility of the Persona<br>knee system within the MRI environment. |
| Spine Fatigue and Underspine Fatigue<br>Evaluation of the Persona CPS Vivacit-E<br>UHMWPE Articular Surfaces | Demonstrated sufficient resistance to spine fracture,<br>underspine fracture, and locking mechanism dissociation<br>during worst case walking gait conditions. |
| Spine Fatigue Evaluation of the Persona<br>CPS Vivacit-E Articular Surfaces | Demonstrated that the spine of the Persona Vivacit-E CPS<br>articular surfaces has sufficient strength to survive<br>expected in-vivo loading conditions. |
| Spine Fracture and Locking Mechanism<br>Resistance of Persona CPS Vivacit-E<br>UHMWPE During Varus/Valgus Loading | Demonstrated sufficient locking mechanism and articular<br>surface spine strength to withstand worst case varus or<br>valgus loading conditions. |
| Tibiofemoral Constraint Evaluation of the<br>Persona CPS Vivacit-E UHMWPE | Determined the anterior-posterior, medial-lateral, internal-<br>external and varus-valgus constraint of the Persona CPS<br>Vivacit-E articular surfaces through the necessary<br>tibiofemoral flexion movements. |
{3}------------------------------------------------
.
.
.
| Property or Characteristics | Test Results |
|------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tibiofemoral Contact Area and Contact<br>Pressure Evaluation of the Persona CPS<br>Vivacit-E UHMWPE Articular Surfaces | Determined the contact area and contact pressure of the<br>Persona CPS Vivacit-E articular surfaces through the<br>necessary tibiofemoral flexion movements. |
| Wear Testing of Persona CPS Vivacit-E<br>UHMWPE Articular Surfaces | Demonstrated that the wear characteristics of the Persona<br>CPS Vivacit-E articular surfaces, when articulated against<br>the Persona PS femoral component, are sufficient to<br>survive expected in vivo loading conditions. |
| EO Residual Testing of Persona CPS<br>Vivacit-E UHMWPE Articular Surfaces | Demonstrated acceptable residual levels following EO<br>sterilization. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines extending from its head, resembling feathers or wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
February 22, 2013
Zimmer, Incorporated % Mr. Daniel Williman Project Manager, Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708
Re: K123459
Trade/Device Name: Zimmer® Persona™ Knee System Regulation Number: 21 CFR 888.3565 Regulation Name: Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: MBH, JWH, OIY Dated: January 25, 2013 Received: January 28, 2013
Dear Mr. Williman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
### Page 2 - Mr. Daniel J. Williman
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Image /page/5/Picture/7 description: The image shows the name "Erin Keith" in a stylized font. The letters "I", "L", and "K" are decorated with geometric patterns. The rest of the letters are in a plain, bold font. The name appears to be a logo or a design element.
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known): K123459
#### Device Name:
Zimmer® Persona™ Personalized Knee System
#### Indications for Use:
This device is indicated for patients with severe knee pain and disability due to:
- । Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- | Collagen disorders, and/or avascular necrosis of the femoral condyle.
- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- Moderate valgus, varus, or flexion deformities. ー
- The salvage of previously failed surgical attempts or for a knee in which satisfactory l stability in flexion cannot be obtained at the time of surgery.
Porous coated components may be used cemented or uncemented (biological fixation). All other femoral, tibial baseplate, and all-polyethylene (UHMWPE and VEXLPE) patella components are indicated for cemented use only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
3.02.21 14:58:17
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anton E. Dmitriev, PhD Division of Orthopedic Devices
13
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.