K163441 · United Orthopedic Corporation · LZO · Aug 25, 2017 · Orthopedic
Device Facts
Record ID
K163441
Device Name
Locking Cage, Full XPE Cup
Applicant
United Orthopedic Corporation
Product Code
LZO · Orthopedic
Decision Date
Aug 25, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For Locking Cage 1. Revision of previous unsuccessful acetabular replacement. 2. Class III segmental and/or cavitary acetabular defects which make it difficult to achieve satisfactory results while using standard total hip replacement acetabular components and procedures. For Full XPE Cup 1. Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, and painful hip dysplasia; 2. Inflammatory degenerative joint disease such as rheumatoid arthritis; 3. Correction of function deformity; 4. Revision procedures where other treatments or devices have failed; 5. Treatment of nonunion and femoral neck fractures of the proximal femur with head involvement that is unmanageable using other techniques. This device is a single use implant and intended for cemented use only.
Device Story
The United Locking Cage and Full XPE Cup are orthopedic implants for hip arthroplasty. The Locking Cage provides stable fixation for severely deficient acetabula; it includes an ischial flange, hook, cancellous locking screw, and auto break-off locking nut. The Full XPE Cup is a cemented acetabular component featuring a PMMA spacer and X-ray marking wire. The system is used by orthopedic surgeons in clinical settings to restore hip function. The Locking Cage is used for revision or complex defects; the Cup is used for degenerative joint disease or fractures. The components are assembled with bone cement. The device benefits patients by providing structural support and articulation in cases where standard components are insufficient.
Clinical Evidence
No clinical data provided. Safety and effectiveness were established through non-clinical bench testing, including structural compression stiffness, range of motion (ROM), bending fatigue, locking strength, endurance testing, and bacterial endotoxin testing per USP <161>.
Technological Characteristics
Locking Cage: Ti6Al4V (ASTM F136-13). Full XPE Cup: Highly cross-linked UHMWPE (ASTM F2565-06, ASTM F648-14, ISO 5834-1:2005), PMMA spacer, and Co-20Cr-15W-10Ni alloy (ASTM F90-14) X-ray wire. Cemented fixation. Single-use.
Indications for Use
Indicated for patients requiring acetabular revision or treatment of severe acetabular defects (Locking Cage), and patients with degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, hip dysplasia, rheumatoid arthritis), functional deformity, failed prior treatments, or proximal femur fractures (Full XPE Cup).
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 25, 2017
United Orthopedic Corporation Ms. Gimpel Chien Regulation and Document Management No 57, Park Ave 2, Science Park Hsinchu 300 Taiwan
Re: K163441 Trade/Device Name: Locking Cage, Full XPE Cup Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, JDI Dated: July 31, 2017 Received: August 2, 2017
Dear Ms. Gimpel Chien:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163441
Device Name Locking Cage, Full XPE Cup
#### Indications for Use (Describe)
For Locking Cage
1. Revision of previous unsuccessful acetabular replacement.
2. Class III segmental and/or cavitary acetabular defects which make it difficult to achieve satisfactory results while using standard total hip replacement acetabular components and procedures.
For Full XPE Cup
1. Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, and painful hip dysplasia;
2. Inflammatory degenerative joint disease such as rheumatoid arthritis;
3. Correction of function deformity;
4. Revision procedures where other treatments or devices have failed;
5. Treatment of nonunion and femoral neck fractures of the proximal femur with head
involvement that is unmanageable using other techniques.
This device is a single use implant and intended for cemented use only.
Type of Use (*Select one or both, as applicable*)
| <span> <b> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> |
|-------------------------------------------------------------------------------------------|
| <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
# Submitter Information
| Name | United Orthopedic Corporation |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | No 57, Park Ave 2, Science Park, Hsinchu 300, Taiwan |
| Phone Number | +886-3-5773351 ext. 2217 |
| Fax Number | +886-3-577156 |
| Name of Contact Person | Gimpel Chien<br>Regulation and Document Management |
| Date prepared | August 18, 2017 |
| Name of Device | |
| Trade Name | Locking Cage, Full XPE Cup |
| Common Name | Acetabular component |
| Regulation Name and<br>Number | The device classification for Locking Cage, Full XPE Cup<br>are “prosthesis, hip, semi-constrained, metal/ceramic/polymer,<br>cemented or non-porous, uncemented” and “prosthesis, hip,<br>semi-constrained, metal/polymer, cemented” which are<br>contained in the Code of Federal Regulation, under 21CFR<br>888.3353 and 21 CFR 888.3350, respectively. This falls<br>under the Orthopedic Panel. |
| Device Class | Class II |
| Classification Panel | Orthopaedics |
| Product Code | LZO, JDI |
| Predicate Device | 1. “Stryker Howmedica Osteonics” GAP-II Restoration<br>Acetabular Shells (K980774)<br>2. “Smith & Nephew” Cemented all polyethylene Acetabular<br>component (K991026)<br>3. “United” U-Motion II Acetabular Component (K122185) |
{5}------------------------------------------------
# Locking Cage, Full XPE Cup
- 4. "Howmedica" Trident® Crossfire® Polyethylene Liners (K021911)
### Device Description:
### "United" Locking Cage
The "United" Locking Cage is designed to achieve stable and lasting fixation of the severely deficient acetabulum. The "United" Locking Cage includes locking cage and several accessory components including: ischial flange, hook, cancellous locking screw and auto break-off locking nut. They are all manufactured from Ti6A14V which confirms to ASTM F136-13.
### "United" Full XPE Cup
The "United" Full XPE Cup is designed for cemented use and assembled with a PMMA Spacer and an X-ray marking wire. The Full XPE Cup is manufactured from highly cross-linked UHMWPE which conforms to ASTM F2565-06. The UHMWPE raw material is in accordance with ASTM F648-14 and ISO 5834-1:2005. The PMMA Spacer and X-ray marking wire are made of PMMA (Medical Grade) and Co-20Cr-15W-10Ni alloy (ASTM F90-14), respectively. The X-ray marking wire is designed for X-ray image identification purpose. The "United" Full XPE Cup is available in a range of sizes to fit varying anatomical requirements.
The "United" Locking Cage can be used with correspondent sizes of "United" Full XPE Cup with bone cement. The "United" Full XPE Cup is also compatible with "United" Metal Femoral Head (K994078, K022520, K111546 and K122504) and "United" Ceramic Femoral Head (K103497) in correspondent sizes. The "United" Femoral Heads can be used with various types of "United" hip stems (K003237, K062978, K111546, K123550, K132207, K151316 and K152530).
{6}------------------------------------------------
# Locking Cage, Full XPE Cup
### Indications for Use:
For Locking Cage
Revision of previous unsuccessful acetabular replacement. 1.
Class III segmental and/or cavitary acetabular defects which make it difficult to 2. achieve satisfactory results while using standard total hip replacement acetabular components and procedures.
### For Full XPE Cup
- Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, 1. ankylosis, and painful hip dysplasia;
- 2. Inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. Correction of function deformity;
- 4. Revision procedures where other treatments or devices have failed;
- 5. Treatment of nonunion and femoral neck fractures of the proximal femur with head involvement that is unmanageable using other techniques.
This device is a single use implant and intended for cemented use only.
### Comparison to Predicate Device:
### "United" Locking Cage
The "United" Locking Cage is substantially equivalent to "Stryker Howmedica Osteonics" GAP-II Restoration Acetabular Shells (K980774) in indications, design rationale, dimension characteristic and sterilization method.
## "United" Full XPE Cup
The Full XPE Cup is substantially equivalent to "Smith & Nephew" REFLECTION cross-linked UHMWPE Acetabular Components (K991026) in indications, material, design rationale, dimension characteristic and sterilization method.
{7}------------------------------------------------
### Performance Data:
### ● Non-clinical Performance
Tests as follows were conducted to evaluate the safety and effectiveness of the subjected device:
- a. Structural compression stiffness of locking cage
- b. ROM for Full XPE Cup
- c. Bending fatigue testing of flanges
- d. Locking strength of locking cage and cemented cup
- e. Endurance testing
- f. Bacterial endotoxin testing was conducted and met the endotoxin limit as specified in USP <161>.
Performance data demonstrate the device is as safe and effective and is substantially equivalent to the legally marketed predicate devices.
## ● Clinical Performance Data/Information
None provided as a basis for substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.