U2 HIP STEM, TI PLASMA SPRAY

K062978 · United Orthopedic Corp. · LWJ · Jan 16, 2007 · Orthopedic

Device Facts

Record IDK062978
Device NameU2 HIP STEM, TI PLASMA SPRAY
ApplicantUnited Orthopedic Corp.
Product CodeLWJ · Orthopedic
Decision DateJan 16, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3360
Device ClassClass 2
AttributesTherapeutic

Intended Use

This device is indicated in hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: - 1. Non-inflammatory degenerative joint disease such as osteoathritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; - 2. Inflammatory degenerative joint disease such as rheumatoid arthritis; - 3. Correction of function deformity; - 4. Revision procedures where other treatments or devices have failed; and - 5. Treatment of nonunion, femoral neck, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. This device is a single use implant and intended for cementless use only.

Device Story

U2 Hip Stem is a cementless femoral component for revision hip arthroplasty. Device features titanium alloy (ASTM F620) femoral stem with CP Ti (ASTM F1580) plasma spray coating for press-fit fixation. Available in collar/collarless types, 180mm/230mm lengths, and 7 diameters. Includes 130° neck angle and Morse taper for modular femoral head attachment. Distal end is cylindrical with polished bullet-shaped tip; 230mm version includes anterior bow. Implanted by orthopedic surgeons in clinical settings to replace damaged femoral bone, restore hip function, and relieve pain. Single-use device.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Materials: Titanium alloy (ASTM F620) stem, CP Ti (ASTM F1580) plasma spray coating (500 um +/- 100 um thickness). Design: 130° neck angle, Morse taper, cylindrical distal stem with polished tip. Fixation: Cementless press-fit. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients requiring hip arthroplasty for pain relief or improved function due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity, revision of failed treatments, or nonunion/fractures of the proximal femur.

Regulatory Classification

Identification

A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K062978 Summary # 510(k) Summary of Safety and Effectiveness JAN 1 6 2007 | Company: | United Orthopedic Corporation | |----------------|--------------------------------------------------------| | Address: | No 57, Park Ave. 2, Science Park, Hsinchu, 300, Taiwan | | Phone Number: | 886-3-5773351 | | Fax Number: | 886-3-5777156 | | Date Prepared: | September 28, 2006. | | Device Name: | U2 Hip stem, Ti plasma spray | |----------------------|---------------------------------------------------------------------------------------| | Common Name: | Cementless hip stem | | Classification Name: | Hip joint femoral (hemi-hip) metallic cemented or uncemented<br>prosthesis (888.3360) | | Predicate Device: | Reach® Hip Stem, Ti plasma spray (K000760)<br>United U2 HA/Ti Plasma Spray (K003237) | #### Device Description: The U2 Hip Stem, Ti plasma spray is composed of a metallic femoral stem, which is designed to articulate with commercially available U1 and U2 acetabular components and They are manufactured from titanium alloy (ASTM F620). The stems femoral heads. are designed to provide secure fit and fixation in revision hip arthroplasty for anteroposterior bone loss, calcar area defects, metaphyseal bone loss The U2 Hip Stem have a 130° neck angle and Morse taper to receive modular situation. The U2 Hip Stem are available in with collar and collarless types. Each femoral heads. type is available in two lengths, 180mm and 230mm, and 7 diameters: 11mm, 12mm, 13mm, 14mm, 15mm, 16.5mm, 18mm. Distally, the stem is cylindrical with a polished bullet shape tip. The 180mm U2 Hip Stem is available as a straight stem, the 230mm femoral stem has an anterior bow for left and right specific applications. The U2 Hip Stem are fully coated with CP Ti (ASTM F1580). The coating is sprayed with CP Ti powder in thickness 500 um +/-100 um using 200~350 um powder to establish a rough surface for press-fit fixation. This device is a single use implant and intended for cementless use only. ### Intended Use: This device is indicated in hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: - 1. Non-inflammatory degenerative joint disease such as osteoathritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; - 2. Inflammatory degenerative joint disease such as rheumatoid arthritis; - 3. Correction of function deformity; - 4. Revision procedures where other treatments or devices have failed; and - 5. Treatment of nonunion, femoral neck, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. This device is a single use implant and intended for cementless use only. {1}------------------------------------------------ #### છ U2 HIP STEM, Ti Plasma Spray ## Basis for Substantial Equivalence: Features comparable to predicate devices, United U2 HA/Ti Plasma Spray (K003237), Reach® Hip Stem, Ti Plasma Spray (K000760), include same materials, design and indications. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an abstract design of an eagle with outstretched wings, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 United Orthopedic Corporation % Mr. Gene Huang Regulatory Affairs Coordinator No. 57. Park Avenue 2, Science Park Hsinchu, Taiwan 300 JAN 16 2007 Re: K062978 Trade/Device Name: U2 Hip Stem, Ti Plasma Spray Regulation Number: 21 CFR 888.3360 Regulation Name: Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis Regulatory Class: Class II Product Code: LWJ Dated: December 28, 2006 Received: January 3, 2007 Dear Mr. Huang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Gene Huang forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Soubey Brichum Mark N. Melk erson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510 (k) Number (if known):____________________________________________________________________________________________________________________________________________________ K062978 Device Name: _ U2 Hip stem, Ti plasma spray Indications for Use: This device is indicated in hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: - · Non-inflammatory degenerative joint disease such as osteoathritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; - Inflammatory degenerative joint disease such as rheumatoid arthritis; - Correction of function deformity; - Revision procedures where other treatments or devices have failed; and . - Treatment of nonunion, femoral neck, and trochanteric fractures of the proximal -. femur with head involvement that are unmanageable using other techniques. This device is a single use implant and intended for cementless use only. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Barbara huchins vision Sion Division of General, Restorative, and Neurological Devices Page 1 of 1 **510(k) Number** Ka 2978