The Osteonics® GAP-IIS Restoration Acetabular Shells are total hip replacement components are intended to resurface the acetabulum, and are intended to be used in conjunction with commercially available Osteonics femoral stems and compatible Osteonics modular femoral bearing heads. The Osteonics® GAP-II and GAP-IIS Restoration Acetabular Shells are single-use devices. The metal shells are intended for screw fixation within the prepared acetabulum, while the polyethylene inserts are intended for cemented fixation within the metal shells. The Osteonics® GAP-II and GAP-IIS Restoration Acetabular Shells have been specifically designed to address cases involving major bony defects of the acetabulum which may require extensive reconstruction and bone grafting. The indications for the Osteonics® GAP-IIS Restoration Acetabular Shells include the following: - Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis. - Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure. - Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results. - Bone stock which is of poor quality or which is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum. - Class III cavitary or segmental defects.
Device Story
Acetabular shell component for total hip replacement; used in conjunction with femoral stems and modular bearing heads. Designed for patients with major acetabular bony defects requiring reconstruction. Shells feature spherical geometry, dome screw holes, superior plates, superior/posterior hood, and inferior notch hook for fixation. Metal shell secured to acetabulum via bone screws; polyethylene liner affixed to shell interior using PMMA bone cement. Used in clinical/surgical settings by orthopedic surgeons. Provides articulating surface for hip joint; aids in stable fixation in cases of poor bone quality; facilitates bone grafting and reconstruction.
Clinical Evidence
Bench testing only. Mechanical testing performed to demonstrate fatigue strength and shell/insert assembly integrity compared to predicate designs.
Technological Characteristics
Material: ASTM F-67 Commercially Pure Titanium (CP Ti). Geometry: Spherical shell with dome screw holes, superior plates, 20° superior/posterior hood, and inferior acetabular notch hook. Surface: Satin-finished interior, grit-blasted exterior. Fixation: Screw fixation to bone; PMMA bone cement for polyethylene insert. Single-use.
Indications for Use
Indicated for patients with painful, disabling hip joint disease (degenerative, rheumatoid, post-traumatic arthritis, or late-stage avascular necrosis), revision of failed previous procedures, or clinical scenarios where alternative reconstruction is less likely to succeed. Specifically indicated for patients with poor bone stock or acetabular deficiencies, including Class III cavitary and segmental defects.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
Osteonics® Restoration Acetabular Cup Series
Submission Summary (Full Text)
{0}------------------------------------------------
APR 2 9 1998
K980774
## 510(k) Premarket Notification Summary of Safety and Effectiveness for the Osteonics® GAP-II Restoration Acetabular Shells
#### Submission Information
Name and Address of the Sponsor of the 510(k) Submission:
Contact Person:
Date of Summary Preparation:
Device Identification
Proprietary Name:
Common Name:
Classification Name and Reference:
Osteonics Corporation 59 Route 17 Allendale, NJ 07401-1677
Kate Sutton Regulatory Affairs Specialist
February 26, 1998
Osteonics® GAP-II Restoration Acetabular Shells
Acetabular Shell
Hip Joint, Metal/Ceramic/Polymer, Semi-Constrained. Cemented or Non-Porous Uncemented Prosthesis 21 CFR 8888.3353 3
#### Predicate Device Identification
The Osteonics® GAP-II Restoration Acetabular Shells employ features which are substantially equivalent to features of the following Osteonics predicate device, which has been cleared for marketing via the 510(k) process:
> . Osteonics® Restoration Acetabular Cup Series
{1}------------------------------------------------
### Device Description
The Osteonics® GAP-II and GAP-IIS Restoration Acetabular Shells are single use components which are intended for placement within the acetabulum, and which are intended to provide an articulating surface for corresponding Osteonics' femoral head/stem components. The two iliac plates on the GAP-IIS shell are shorter than those of the GAP-II shell. Each Osteonics® GAP-II or GAP-IIS Restoration Acetabular Cup is assembled from two separate components: an Osteonics® GAP-II or GAP-IIS Restoration Acetabular Shell and a predicate Osteonics® Omnifit® Cup Insert, ABC Cementable Polyethylene Insert, Osteonics® Concentric Polyethylene Acetabular Cup, or Osteonics® Flanged Polyethylene Acetabular Cup. The polyethylene cups and inserts are intended to be affixed to the shell via polymethylmethacrylate (PMMA) bone cement.
The Osteonics® GAP-II and GAP-IIS Restoration Acetabular Shells are manufactured from ASTM F-67 Commercially Pure Titanium (CP Ti). The shells are available in a range of outer diameter shell sizes. The Osteonics® GAP-II and GAP-IIS Restoration Acetabular Shells feature a basic spherical geometry. 18 to 26 acetabular dome screw holes (depending on the size of the shell), two superior plates, a 20° superior/posterior hood, an inferiorly located acetabular notch hook, a satin-finished interior surface, and a grit-blasted exterior surface.
Bone screws placed through the dome and/or lip of the acetabular shell are used to secure the Osteonics® GAP-II or GAP-IIS Restoration Acetabular Shell within the prepared acetabulum. The shells are designed to allow some of the bone cement (used to fix the polyethylene liner to the shell) to be extruded through any unoccupied acetabular dome screw holes. In additional acetabular dome screws, and in order to further enhance the potential for initial and long term stable shell fixation despite large bony defects, the Osteonics® GAP-IIS Restoration Acetabular Shells employ an inferior hook, superior plates, and a superior posterior hood. The inferior hook is crimped around the acetabular notch. The shell hood covers the superior rim of the acetabular cavity, and the superior plates and can be secured to the acetabulum and/or ilium with Osteonics® 6.5mm Restoration GAP Plate Screws.
#### Intended Use:
The Osteonics® GAP-IIS Restoration Acetabular Shells are total hip replacement components are intended to resurface the acetabulum, and are intended to be used in conjunction with commercially available Osteonics femoral stems and compatible Osteonics modular femoral bearing heads. The Osteonics® GAP-II and GAP-IIS Restoration Acetabular Shells are single-use devices. The metal shells are intended for screw fixation within the prepared acetabulum, while the polyethylene inserts are intended for cemented fixation within the metal shells. The Osteonics® GAP-II and GAP-IIS Restoration Acetabular Shells have been specifically designed to address cases involving major bony defects of the acetabulum which may require extensive reconstruction and bone grafting. The indications for the Osteonics® GAP-IIS Restoration Acetabular Shells include the following:
{2}------------------------------------------------
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid . arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other . procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are . less likely to achieve satisfactory results.
- Bone stock which is of poor quality or which is inadequate for other reconstructive techniques ● as indicated by deficiencies of the acetabulum.
- Class III cavitary or segmental defects. .
#### Performance Data:
Mechanical testing has been performed to demonstrate the substantial equivalence of this Osteonics acetabular shell design to predicate acetabular shell designs in terms of its fatigue strength and shell/insert assembly method.
#### Statement of Technological Comparison:
The features of the Osteonics® GAP-II or GAP-IIS Restoration Acetabular Shell, either alone or in combination, are substantially equivalent to corresponding features of Osteonics predicate device as follows.
The Osteonics® GAP-II and GAP-IIS Restoration Acetabular Shells are substantially equivalent, in terms of substrate material, indications for use, and availability of screw holes for the employment of acetabular screw fixation, to the legally marketed plated shell in the predicate Osteonics® Restoration Acetabular Cup Series.
The cup geometry of the Osteonics® GAP-II and GAP-IIS Restoration Acetabular Shells is spherical, and as such, is substantially equivalent in terms of its basic geometry to the predicate Osteonics® Restoration Acetabular Cup Series.
The inferior hook and superior plates of the Osteonics® GAP-IIS Restoration Acetabular Shells are substantially equivalent in terms of design and function to similar characteristics featured on the iliac plated cup version in the Osteonics® Restoration Acetabular Cup Series.
The inserts used with the Osteonics® GAP-II or GAP-IIS Restoration Acetabular Shells are the predicate Osteonics® Omnifit® Cup Inserts, Osteonics® Concentric Polyethylene Acetabular Cup, or Osteonics® Flanged Polyethylene Acetabular Cup. These inserts are assembled to the Osteonics® GAP-II or GAP-IIS Restoration Acetabular Cup Shells through the use of bone cement. This assembly method is predicated by the Osteonics® Restoration Acetabular Cups.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads facing to the right. The eagle is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 9 1998
Ms. Elizabeth A. Staub Director, Quality Assurance and Regulatory Affairs Osteonics Corporation 59 Route 17 Allendale, New Jersey 07401-1677
Re : K980774 Trade Name: GAP-II Restoration Acetabular Shells Regulatory Class: LZO and JDI Product Code: II February 26, 1998 Dated: March 2, 1998 Received:
Dear Ms. Staub:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and __________________________________________________________________________________________________________________________ prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
#### Page 2 - Ms. Elizabeth A. Staub
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D.
Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 510(k) Number (if known): K980774
Device Name: Osteonics® GAP-II Restoration Acetabular Shells
Indications For Use:
The indications for the use of the Osteonics® GAP-II Restoration Acetabular Shells, in keeping with those of other legally marketed Osteonics acetabular components, are as follows:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results. ------------
- . Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
- Class III cavitary and segmental defects. .
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) (Divisor Sign-Off) Division is General Restorative 510(k) Number Prescription Use y la OR Over-The-Counter Use (Per 21 CFR 801.109) (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.