IM C-MENT ACETABULAR CAP
Device Facts
| Record ID | K955137 |
|---|---|
| Device Name | IM C-MENT ACETABULAR CAP |
| Applicant | Implant Ideas, LLC |
| Product Code | JDI · Orthopedic |
| Decision Date | Oct 10, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3350 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The IM C-Ment Acetabular Cup will be used in indications which are the same as that for other acetabular components of cemented total hip replacement systems. Specifically, the IM C-Ment Acetabular Cup is intended to resurface the acetabular socket in cases of femoral head and neck fractures, nonunion of femoral head and neck fractures, femoral head aseptic necrosis, osteoarthritis, rheumatoid arthritis and revision of failed prostheses. The IM C-Ment Acetabular Cup is intended to be used in cemented applications only.
Device Story
Acetabular component for cemented total hip replacement; hemispherical all-polyethylene design. Features three concentric outer grooves and three circular indentations for cement fixation. Used by orthopedic surgeons in clinical settings to reconstruct hip joint articulation. Implanted in conjunction with femoral stem and head to restore joint spacing. Benefits patients by providing stable articulation in cases of degenerative joint disease or fracture.
Clinical Evidence
Bench testing only; no clinical data provided. Equivalence established through design comparison and material history.
Technological Characteristics
Material: UHMWPE. Geometry: Hemispherical, 28mm inner diameter, 40-60mm outer diameter. Features: Three concentric outer grooves, three circular indentations, x-ray markers. Application: Cemented fixation only.
Indications for Use
Indicated for patients requiring acetabular socket resurfacing due to femoral head/neck fractures, nonunion of fractures, aseptic necrosis, osteoarthritis, rheumatoid arthritis, or revision of failed prostheses.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
- M.E. Muller Type Acetabular Cups (DePuy)
- Charnley Type Acetabular Cup (DePuy)
- New England Baptist Hip Cups (DePuy)
Related Devices
- K964003 — METAGEN ALL POLY ACETABULAR CUP SYSTEM · Metagen, LLC · Mar 21, 1997
- K980774 — OSTEONICS GAP-II RESTORATION ACETABULAR SHELLS · Osteonics Corp. · Apr 29, 1998
- K963671 — OSTEONICS FLANGED POLYETHYLENE ACETABULAR CUP · Osteonics Corp. · Feb 3, 1997
- K051407 — LEGEND ALL-POLY CUP · Ortho Development Corp. · Sep 8, 2005
- K983837 — HEDROCEL ACETABULAR CAGE SYSTEM · Implex Corp. · Jul 1, 1999