The Renovis A400 Surgical Hip Joint Replacement Prosthesis is indicated for patients suffering from the following: 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 2. Rheumatoid arthritis; 3. Correction of functional deformity; 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and 5. Revision procedures where other treatment or devices have failed. The Renovis A400 Surgical Hip Joint Replacement Prosthesis is intended for cementless applications.
Device Story
Renovis A400 is a cementless total hip replacement system; components include titanium alloy femoral stem and acetabular cup, CoCr or Biolox delta ceramic femoral head, and mechanically annealed HXL polyethylene acetabular liners (with/without alpha-tocopherol). System replaces both femur and acetabulum in patients with degenerative joint disease, arthritis, or fractures. Used in surgical settings by orthopedic surgeons. Components provided sterile for single use. Design relies on standard mechanical fixation; no software or active components. Clinical benefit derived from restoring joint function and mobility in patients with failed prior treatments or severe joint pathology.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by comprehensive bench performance testing, including distal/proximal stem fatigue, snap-lock testing, range of motion analysis, hip simulator wear testing (normal and abrasive), wear debris characterization, porous coating integrity testing, and extensive polyethylene material characterization. Biocompatibility testing performed per ISO 10993.
Technological Characteristics
Cementless hip prosthesis. Materials: Titanium alloy (stem/cup), CoCr/Biolox delta ceramic (head), HXL polyethylene (liner). Testing standards: ISO 7206 (fatigue), ASTM F1820 (snap-lock), ISO 14242 (wear), ASTM F1877 (debris), ASTM F1978/F1160/F1044/F1147/F1854 (porous coating), ISO 10993 (biocompatibility). Mechanical fixation. No software or electronics.
Indications for Use
Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or proximal femur fractures (non-union, femoral neck, trochanteric) requiring hip replacement, including revision procedures. Intended for cementless applications.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
# RENOVIS
#### RENOVIS Surgical Technologies, Inc.
#### 510(k)# K112897
### 510(k) Summary
- Submitter's Name: Renovis Surgical Technologies, Inc.
- Address: 1901 W. Lugonia Ave, Ste 340 Redlands, CA 92374
- Contact Person: Anthony DeBenedictis Vice President of Quality Assurance Phone: 909.557.2360
Date Prepared: May 31, 2012
| Classification Name: | Prosthesis, Hip Revision System |
|----------------------|-------------------------------------------------|
| Common Name: | Surgical Hip Joint Replacement System |
| Model Number: | A400 |
| Product Code: | OQG, OQI, LPH, LZO |
| C.F.R. Section: | 888.3353-Hip joint |
| | metal/ceramic/polymer semi-constrained |
| | cemented or nonporous uncemented<br>prosthesis. |
| | 888.3358-Hip joint |
| | metal/polymer/ |
metal semi-constrained porouscoated uncemented prosthesis.
| Recommended Classification: ...................................... | Class II |
|--------------------------------------------------------------------|--------------------|
| Classification Panel: ...................................... | Orthopedics |
| Reason for Submission: ...................................... | Traditional 510(k) |
### Device Description:
The Renovis Surgical Technologies, Inc. A400 Surgical Hip Replacement System (Renovis A400 Hip System) is a family of components designed in a variety of standard clinically usable sizes for cementless total hip replacements (i.e. replacements where both the femur and acetabulum are replaced).
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text 'K112897' at the top. Below that is the word 'RENOVIS' in large, bold, sans-serif font. To the left of the word 'RENOVIS' is a circular graphic with three overlapping shapes.
The system consists of four main components available in a variety of clinically usable sizes.
A) Femoral Stem (titanium alloy)
B) Femoral Head (CoCr and Biolox delta ceramic)
C) Acetabular Cup (titanium alloy)
D) Acetabular Liners (Mechanically annealed HXL polyethylene with and without alphatocopherol)
The Renovis A400 Hip System implant components are provided sterile and are for single use.
### Indications for Use
The Renovis A400 Surgical Hip Replacement System is indicated for patients suffering from:
1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
2. Rheumatoid arthritis;
3. Correction of functional deformity;
4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and
5. Revision procedures where other treatment or devices have failed.
The Renovis A400 Surgical Hip Joint Replacement Prosthesis is intended for cementless applications.
### Predicate Devices
Predicate List
| 510(k) Number | Manufacturer |
|---------------|----------------------|
| K030055 | Biomet Manufacturing |
| K032396 | Biomet Manufacturing |
| K051411 | Biomet Manufacturing |
| K061312 | Biomet Manufacturing |
| K070364 | Biomet Manufacturing |
| K073102 | Biomet Manufacturing |
2 Page
{2}------------------------------------------------
### ×112897
## RENOVIS
| K091956 | DJO Surgical |
|---------|----------------------|
| K094035 | StelKast, Inc. |
| K043537 | Biomet Manufacturing |
| K093235 | Biomet Manufacturing |
| K981211 | Sulzer Orthopedics |
### Substantial Equivalence
The Renovis A400 Surgical Hip Replacement System claims substantial equivalence to commercially available hip prosthesis systems and related hip system components manufactured by Biomet Manufacturing Corporation (Biomet).
The Renovis A400 Hip System's intended use and indications for use are identical to the Biomet hip prosthesis system and associated components.
The Renovis A400 Hip System implanted components: femoral head; acetabular cup; acetabular liners, with vitamin E and without vitamin E are equivalent and raise no new question of safety or technology to those devices manufactured by Biomet. The materials used for all Renovis A400 Hip System components are similar to those used in the components of the Biomet hip prosthesis system predicate devices. All materials used in the Renovis A400 Surgical Hip Replacement System are biocompatible and are well known and commonly used in the orthopedic implant industry.
### Performance Testing
Comprehensive bench performance testing conducted on the hip system components: femoral stems, femoral heads, and acetabular liners demonstrated that the A400 Hip System design is safe for its intended use and is substantially equivalent to the predicate devices. The following bench/ performance testing was conducted on the A400 Hip System:
A) A400 Hip System – The following device tests were performed on the complete A400 Hip system and demonstrate the system is safe for its intended uses and is substantially equivalent to the predicate devices:
- Distal Stem Fatigue Testing per ISO 7206-4:2002/ Guidance for Industry and FDA 1. Staff - Non-clinical Information for Femoral Stem Prostheses, September 1997
- Proximal Stem Fatigue Testing per ISO 7206-6:1992/ Guidance for Industry and FDA 2. Staff – Non-clinical Information for Femoral Stem Prostheses, September 1997
- Push-out Snap-Lock Testing of Acetabular Liners and Shells per ASTM F1820-97 3.
{3}------------------------------------------------
K"2897
RENOVIS
- Lever-out and Torque-out Snap-Lock Testing of Acetabular Liners and Shells 4.
- 5. CoCr Head Attachment Testing per ISO 7206-10:2003
- BIOLOX delta Ceramic Heads Attachment Testing per FDA Guidance Document for 6. the Preparation of Premarket Notifications for Ceramic Ball Hip Systems
- Range of Motion Analysis 7..
- Hip Simulator Wear Testing under Normal Conditions per ISO 14242-1 8.
- Hip Simulator Wear Testing under Abrasive Wear Conditions per ISO 14242-1 ക്
- Wear Debris Characterization per ASTM F1877-05 10.
B) Femoral Stem and Acetabular Cup Porous Surface/Coating- the following tests were performed to evaluate the integrity of the porous surface/coating:
- 1. Abrasion resistance per ASTM F1978
- 2. Shear fatigue strength per ASTM F1160
- 3. Static shear strength per ASTM F1044
- 4. Static tensile strength per ASTM F1147
- 5. Characterization per ASTM F1854
C) Acetabular Liner – The following poly characterization testing, both chemical and mechanical, was performed on the HXL polyethylene materials:
- 1. Heat of fusion , melting point, and percent crystallinity per ASTM F2625/D3418
- 2. Small punch test properties (peak load, ultimate load, ultimate displacement, work to failure) per ASTM F2183
- 3. Oxidation measurement per ASTM F2102
- 4. Trans-vinylene vield per ASTM F2381
- 5. Crosslink density (Swell Ratio Testing) per ASTM F2214
- 6. Tensile properties (yield strength, ultimate strength, elongation at break, elastic modulus) per ASTM D638
- 7. Izod impact strength per ASTM D256 /F648
- 8. GCMS and LCMS extract analyses
- 9. Vitamin E content
- 10. Fatigue crack propagation per ASTM E647
- 11. Compressive modulus per ASTM D695
- 12. Uniformity of radiation dose
- 13. Free radical content (Electron Spin Resonance)
- 14. Consolidation verification
- 15. Density per ASTM D1505/D792
{4}------------------------------------------------
K112897
## SRENOVIS
All poly characterization test results demonstrate that the A400 Hip System design is safe for its intended use and is substantially equivalent to the predicate devices.
D) Polyethylene Biocompatibility - The following tests in accordance with ISO 10993 (permanent implant panel) were performed on the polyethylene material to demonstrate biocompatibility:
- 1. Systemic Toxicity: Materials Mediated Rabbit Pyrogen Test per ISO 10993-11 and -12
- 2. Systemic Toxicity: Acute Systemic Toxicity per ISO 10993-11 and -12
- 3. In Vitro Cytotoxicity Study Using the ISO Elution Method per ISO 10993-5
- 4. Genotoxicity: Chromosomal Aberration per ISO 10993-1, -3,-5, and -12.
- 5. Genotoxicity: In-Vivo Mouse Micronucleus Assay per ISO 10993-3 and -12
- 6. Genotoxicity: Bacterial Reverse Mutation per ISO 10993-3.
- 7. Irritation and Skin Sensitization: Intracutaneous Study per ISO 10993-2 and -10.
- 8. ISO Muscle Implantation Study in Rabbits- 2 Weeks per ISO 10993-2 and -6
- 9. ISO Muscle Implantation Study in Rabbits- 13 Weeks per ISO 10993-2 and -6
- 10. Irritation and Skin Sensitization: Guinea Pig Maximization Sensitization Test per ISO 10993-2 and -10
- 11. Systemic Toxicity: Sub-Chronic (14 Day) Intravenous Toxicity Study per ISO 10993-11 and -12
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle faces right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 1 3 2012
Renovis Surgical Technologies, LLC c/o David Teicher, DuVal & Associates 825 Nicollet Mall Minneapolis, Minnesota 55402
Re: K112897
Trade/Device Name: Renovis A400 Hip Joint Replacement Prosthesis Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: OQG, OQI, LPH, LZO Dated: May 4, 2012 Received: May 7, 2012
Dear Mr. Teicher:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### X112897
### RENOVIS Surgical Technologies, Inc.
### Indications for Use
510(k) Number (if known): K112897
Device Name: Renovis A400 Surgical Hip Joint Replacement Prosthesis
Indications for Use:
The Renovis A400 Surgical Hip Joint Replacement Prosthesis is indicated for patients suffering from the following:
1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
2. Rheumatoid arthritis;
3. Correction of functional deformity;
4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and 5. Revision procedures where other treatment or devices have failed.
The Renovis A400 Surgical Hip Joint Replacement Prosthesis is intended for cementless applications.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K112.897
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.