K981211 · Sulzer Orthopedics, Inc. · LPH · May 22, 1998 · Orthopedic
Device Facts
Record ID
K981211
Device Name
APR -T REVISION HIP STEM
Applicant
Sulzer Orthopedics, Inc.
Product Code
LPH · Orthopedic
Decision Date
May 22, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The APR-T Revision Hip Stem is intended for prosthetic replacement of the proximal portion of the femur during hip arthroplasty. This hip stem can be used for total hip replacements (i.e., replacements where both the femur and acetabulum are replaced) or for hemi-arthroplasties where the hip stem-head assembly articulates directly with wellpreserved articular cartilage. For this indication, the stem must be paired with one of the following Sulzer Orthopedics endoprosthesis: - the Unipolar with 12/14 Taper, cleared for use via 510(k)s K833403 and K934159, or - the Bipolar component which was cleared for use via 510(k)s K833404 and K873815 and is indicated for use with a CoCr modular femoral head. Specific diagnostic indications include: - 1. Patient conditions of inflammatory degenerative joint disease (e.g., rheumatoid arthritis) and noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis); - 2. Those patients with failed previous surgery where pain, deformity, or dysfunction persist; and, - 3. Revision of previously failed hip arthroplasty. The APR-T Revision Hip Stem is intended for use with or without bone cement.
Device Story
APR-T Revision Hip Stem is a femoral replacement component for primary and revision hip arthroplasty. Features asymmetric proximal wedge, circumferential CSTi porous coating, and porous-coated medial collar for proximal fixation. Anatomic design follows diaphyseal canal curve; distal coronal slot reduces stiffness in larger sizes. Used by orthopedic surgeons in clinical settings. Device provides structural support for hip joint; can be used with or without bone cement. Paired with specific Sulzer Orthopedics unipolar or bipolar components for hemi-arthroplasty.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design and material characteristics compared to legally marketed predicate devices.
Technological Characteristics
Femoral stem prosthesis featuring asymmetric proximal wedge, circumferential CSTi porous coating, and distal coronal slot. Designed for cementless or cemented use. Materials and design are consistent with established hip arthroplasty standards.
Indications for Use
Indicated for patients requiring proximal femoral replacement during hip arthroplasty (total or hemi-arthroplasty). Applicable for inflammatory (e.g., rheumatoid arthritis) and noninflammatory (e.g., osteoarthritis, avascular necrosis) degenerative joint disease, failed previous surgery, or revision of failed hip arthroplasty.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
K981211
## 510(k) SUMMARY
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this is to serve as a Summary of Safety and Effectiveness for the use of zirconia ceramic heads with the femoral stems listed below.
| Submitter: | Sulzer Orthopedics Inc.<br>9900 Spectrum Drive<br>Austin, Texas 78717<br>(512) 432-9900 |
|-------------------------|--------------------------------------------------------------------------------------------------------|
| Date: | April 1, 1998 |
| Contact Person: | Mitchell A. Dhority<br>Manager, Regulatory Affairs |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous coated<br>uncemented prosthesis, 21 CFR 888.3358 |
| Common/Usual Name: | Total Hip Prosthesis, Semi-constrained |
| Trade/Proprietary Name: | APR-T® Revision Hip Stem |
#### SPECIFIC DIAGNOSTIC INDICATIONS
The APR-T Revision Hip Stem is intended for prosthetic replacement of the proximal portion of the femur during hip arthroplasty. This hip stem can be used for total hip replacements (i.e., replacements where both the femur and acetabulum are replaced) or for hemi-arthroplasties where the hip stem-head assembly articulates directly with wellpreserved articular cartilage. Specific diagnostic indications include:
- Patient conditions of inflammatory degenerative joint disease (e.g., rheumatoid arthritis) . and noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis) :
- . Those patients with failed previous surgery where pain, deformity, or dysfunction persist; and,
- Revision of previously failed hip arthroplasty. .
The APR-T Revision Hip Stem is intended for use with or without bone cement.
#### PRODUCT DESCRIPTION
The APR-T Revision Hip Stem is part of a family of femoral replacement components which are used during primary and revision hip arthroplasties. Its proximal design features an asymmetric wedge and circumferential CSTi ™ porous coating. In addition, a porouscoated medial collar helps achieve improved proximal fixation.
The stem's anatomic design follows the natural curve of the diaphyseal canal for better fit, and features a coronal slot at the distal end to reduce stiffness in the larger sizes.
{1}------------------------------------------------
### SUBSTANTIAL EQUIVALENCE
This 510(k) is being submitted in response to an update of the design such that certain features are uniform across the entire APR stem line. The characteristics of the APR-T Revision Stem, either alone or in combination, are substantially equivalent to the following legally marketed predicate Sulzer Orthopedics and/or competitive devices in terms of intended use, materials and design characteristics: the APR® Collared Revision Stem (Sulzer Orthopedics Inc.), the APR® Porous Hip Stem (Sulzer Orthopedics Inc.), the BLAS™ Total Hip Prosthesis (Zimmer, Inc.), the PCA ™ Long Stem (Howmedica, a Division of Pfizer Hospital Products Group, Inc.), and the Natural-Hip™ Porous Stem (Sulzer Orthopedics Inc.).
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure, with three curved lines representing the body and head. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is written in all capital letters.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 2 1998
Ms. Shavawn Evans Parduhn Requlatory Affairs Specialist Sulzer Medica Sulzer Orthopedics Inc. 9900 Spectrum Drive Austin, Texas 78717
K981211 Re : APR-T® Revision Stem Trade Name: Requlatory Class: II Product Codes: LPH and JDI Dated: April 1, 1998 Received: April 2, 1998
Dear Ms. Parduhn:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
#### Page 2 - Ms. Shavawn Evans Parduhn
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
MAY-14-98 THU 15:25
Page 1 of 1
510(k) Number (if known): K981211
Device Name: APR-T Revision Hip Stem
# Indications For Use:
The APR-T Revision Hip Stem is intended for prosthetic replacement of the proximal portion of the femur during hip arthroplasty. This hip stem can be used for total hip replacements (i.e., replacements where both the femur and acetabulum are replaced) or for hemi-arthroplasties where the hip stem-head assembly articulates directly with wellpreserved articular cartilage. For this indication, the stem must be paired with one of the following Sulzer Orthopedics endoprosthesis:
" in "w
- the Unipolar with 12/14 Taper, cleared for use via 510(k)s K833403 and K934159, or ●
- the Bipolar component which was cleared for use via 510(k)s K833404 and K873815 and is . indicated for use with a CoCr modular femoral head.
Specific diagnostic indications include:
- 1. Patient conditions of inflammatory degenerative joint disease (e.g., theumatoid arthritis) and noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis);
- 2. Those patients with failed previous surgery where pain, deformity, or dysfunction persist; and,
- 3. Revision of previously failed hip arthroplasty.
OR
The APR-T Revision Hip Stern is intended for use with or without bone cement.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
biollez
Prescription Use
Over-The Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.