FLEXI-SEAL SIGNAL FECAL MANAGEMENT SYSTEM

K112342 · Convatec, Inc. · KNT · Apr 26, 2012 · Gastroenterology, Urology

Device Facts

Record IDK112342
Device NameFLEXI-SEAL SIGNAL FECAL MANAGEMENT SYSTEM
ApplicantConvatec, Inc.
Product CodeKNT · Gastroenterology, Urology
Decision DateApr 26, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5980
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Flexi-Seal® SIGNAL™ Fecal Management System is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications.

Device Story

Flexi-Seal® SIGNAL™ Fecal Management System is an indwelling catheter for managing fecal incontinence. Device consists of a collapsible silicone rubber drain tube, a retention balloon (polyurethane or silicone), a collection bag, a Luer syringe, and a cinch clamp. Clinician inserts the device into the patient's rectum; retention balloon is inflated via a dedicated port. The inflation port provides visual and tactile feedback when the balloon reaches optimal volume. A second port allows for device flushing and medication administration. A cinch clamp is used to temporarily stop flow for medication retention. Device is used in clinical settings by healthcare professionals. It benefits patients by providing a hygienic method for collecting liquid/semi-liquid stool and facilitating targeted medication delivery.

Clinical Evidence

Bench testing only. Performance evaluated for functional and structural parameters including material strength, catheter size, and balloon size. Biocompatibility testing conducted in accordance with ISO-10993. No clinical data presented.

Technological Characteristics

Materials: silicone rubber, polyurethane. Sensing: tactile/visual inflation indicator for balloon volume. Energy: manual (syringe-operated). Connectivity: none. Sterilization: non-sterile. Form factor: indwelling catheter with retention balloon, collection bag, and access ports.

Indications for Use

Indicated for patients requiring management of fecal incontinence via collection of liquid to semi-liquid stool and access for medication administration.

Regulatory Classification

Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ APR 2 6 2012 # 510(K) SUMMARY . . · · · | Subject Device: | Flexi-Seal® SIGNAL™ Fecal Management System | |----------------------|-------------------------------------------------------------------------------------------------------------------------| | Date Prepared: | 4/24/2012 | | Applicant: | ConvaTec Inc.<br>200 Headquarters Park Drive<br>Skillman, New Jersey 08558 | | Contact: | Katrina Fiedler<br>Associate Director, US Regulatory Affairs<br>ConvaTec Inc.<br>Tel: 908-904-2541<br>Fax: 908-904-2235 | | Device Trade Name: | Flexi-Seal® SIGNAL™ Fecal Management System | | Classification Name: | Gastrointestinal Tube and Accessories (ref. 21 CFR 876.5980; Product Code KNT) | | Device Class: | Class II | Predicate Devices: | Trade Name: | Flexi-Seal® Fecal Management System - ConvaTec Inc. | |---------------------------------|--------------------------------------------------------------------------------| | Classification Name: | Gastrointestinal Tube and Accessories (ref. 21 CFR 876.5980; Product Code KNT) | | Device Class: | Class II | | 510(k) Substantial Equivalence: | K032734 – determined substantially equivalent on April 8, 2004 | | Trade Name: | Indwelling Fecal Management System - Bowel Management Systems, LLC | | Classification Name: | Gastrointestinal Tube and Accessories (ref. 21 CFR 876.5980; Product Code KNT) | | Device Class: | Class II | | 510(k) Substantial Equivalence: | K012113 – determined substantially equivalent on May 3, 2002 | {2}------------------------------------------------ #### Device Description: The Flexi-Seal® SIGNAL™ Fecal Management System is comprised of a soft catheter tube assembly, a Luer Syringe, a collection bag with filter and a cinch clamp to pinch off flow in the catheter when required for medication retention. The components are contained in a snap closed tray. The catheter main drain tube is fabricated from collapsible silicone rubber. The drain tube has a low-pressure polyurethane or silicone retention balloon at the distal end and a connector for attaching a collection bag (provided with the device and separately) at the proximal end. There is a recess under the balloon for the clinician's finger allowing the device to be positioned digitally. Two ports are attached to the side of the catheter. One port is used to inflate the retention balloon with water or saline after the device has been inserted into the patient's rectum. This port also provides a visual and tactile signal of when the low pressure retention balloon is filled to its optimal volume. The other port is used to flush the device if needed and administer medication, if prescribed. A syringe (provided with the device) is used to fill and evacuate the retention balloon for insertion and removal. The device, collection bay and syringe are intended for single us, are provided nonsterile, and have no components made with animal products, natural rubber latex or DEHP. This 510(k) concerns modifications to the indications for use, materials and design featured with the predecessor device, the ConvaTec Fecal Management System (ref. K032734). #### Intended use: The Flexi-Seal® SIGNAL™ Fecal Management System is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. ### Summary of Technological Characteristics: The Flexi-Seal® SIGNAL™ Fecal Management System has the same intended use and indications for use but with the addition of the access to administer medication. The approximate geometry of the new device is the same as the predicate devices. The materials and construction are the same as the predicate devices with the exception of a polyurethane balloon version additional to the original silicone retention balloon. This just provides an alternate substitute material for the retention balloon. The only differences in operation between the Flexi-Seal® SIGNAL™ Fecal Management System and the predicate devices are the use of a port to gather samples rather than the bay coupling and the use of an external separate cinch clamp to retard waste flow when desired. One of the predicate devices Indwelling Fecal Management System (ref. K012113) uses an internal inflatable balloon to block flow and has a sampling port. ## Summary of Performance (Non-Clinical Testing) Data: {3}------------------------------------------------ K 112342 page 3 of 3 Non-clinical testing of the subject device for functional and structural parameters has been performed. In this testing, the device's performance has been found to be substantially equivalent to the aforcmentioned predicate devices both functionally and structurally (material strength, catheter size, balloon size, etc.). The device has also been evaluated for biocompatibility in accordance with the US Food and Drug Administration's guidance entitled Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing', issued May 1, 1995, and has been found safe in such respect for its intended use. In conclusion, the subject device has been demonstrated as safe and effective and substantially equivalent to the predicate devices. {4}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ms. Katrina Fiedler Associate Director. US Regulatory Affairs ConvaTec Inc. 200 Headquarters Park Drive SKILLMAN NJ 08558 APR 2 6 2012 Re: K112342 Trade/Device Name: Flexi-Seal® SIGNALTM Fecal Management System (FMS) Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: March 27, 2012 Received: March 29, 2012 Dear Ms. Fiedler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vathar sthates that stequirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {5}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bariano K. torkin Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications For Use 510(K) Number (if known): Not yet assigned Device Name: Flexi-Seal® SIGNALTM Fecal Management System (FMS) Indication for Use: For use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. Prescription Use XX (21 CFR 801 Subpart D) AND/OR Over the Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Hubert Reum Division Sign-Off vision of Reproductive, Gastro-Renal, Urological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...