Flexi-Seal PROTECT PLUS Fecal Management System (with ENFit Connector)
Applicant
Convatec Limited
Product Code
KNT · Gastroenterology, Urology
Decision Date
Apr 30, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Flexi-Seal™ PROTECT PLUS Fecal Management System (with ENFit™ connector) is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. The device is intended for use in adult patients.
Device Story
Indwelling fecal management catheter; manages fecal incontinence in adult patients by collecting liquid/semi-liquid stool; provides access for medication administration. Features include low-profile retention balloon, finger pocket for insertion, 2-dome fill indicator (optimal fill/over-inflation), and cinch clamp. New features: self-closing ENFit port for irrigation/medication; ENFit syringe; self-closing quick connector at catheter end; collection bag with Diamonds™ gelling/odor control sachet. Used in clinical settings; operated by healthcare professionals. Device diverts stool to collection bag; gelling sachet solidifies/deodorizes contents. Benefits include improved fecal management, reduced skin contact, and standardized connection safety via ENFit.
Clinical Evidence
No clinical data required. Substantial equivalence supported by bench testing, including joint strength, water leakage, and drug efficacy testing for the new ENFit connector and self-closing mechanisms.
Technological Characteristics
Silicone with Zeolite catheter; retention balloon; 2-dome visual fill indicator; ENFit (ISO 80369-3) irrigation/medication port; self-closing quick connector; collection bag with gelling/odor control sachet. Non-sterile. 29-day usage duration.
Indications for Use
Indicated for adult patients with fecal incontinence requiring management via collection of liquid to semi-liquid stool and access for medication administration.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Flexi-Seal™ PROTECT Fecal Management System (K162906)
Reference Devices
Bard™ DigniShield™ Stool Management System ENFit™
Bard™ DigniShield™ Stool Management System (K133251)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font below.
April 30, 2019
ConvaTec Limited Jason Skramsted Senior Regulatory Affairs Specialist GDC, First Avenue, Deeside Industrial Park Deeside, Flintshire CH5 2NU United Kingdom
Re: K190114
Trade/Device Name: Flexi-Seal™ PROTECT PLUS Fecal Management System (with ENFit™ connector) Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT, PIF Dated: April 3, 2019 Received: April 4, 2019
Dear Jason Skramsted:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K190114
Device Name
Flexi-Seal™ PROTECT PLUS Fecal Management System (with ENFit™ connector)
#### Indications for Use (Describe)
The Flexi-Seal™ PROTECT PLUS Fecal Management System (with ENFit™ connector) is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid stool and to provide access to administer medications. The device is intended for use in adult patients.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for ConvaTec. The logo consists of a blue circle with a white building icon inside, followed by the word "ConvaTec" in blue. The building icon has three columns and a triangular roof.
## Section 5 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
- 1. Applicant: ConvaTec Limited GDC, First Avenue Deeside Industrial Park Deeside Flintshire, CH5 2NU UK
Establishment Registration number 1000317571
- Contact: Jason Skramsted ConvaTec, Inc. Senior Regulatory Affairs Specialist Tel: 336-542-4753 Email: Jason.Skramsted@ConvaTec.com
23 January 2019 Date Prepared:
#### 2. Device Identification:
| Trade Name: | Flexi-Seal™ PROTECT PLUS Fecal Management<br>System (with ENFit™ connector) |
|------------------------------|-----------------------------------------------------------------------------|
| Common Name: | Fecal Management System |
| Regulation Number: | 21 CFR §876.5980 |
| Classification Panel: | Gastroenterology and Urology |
| Classification Name: | Gastrointestinal tube and accessories |
| Regulatory Class: | II |
| Product Classification Code: | KNT, PIF |
3. Predicate Devices:
Predicate Manufacturer: Predicate Trade Name: Predicate 510 (k):
ConvaTec Limited Flexi-Seal™ PROTECT Fecal Management System K162906
{5}------------------------------------------------
#### 4. Device Description
The subject device, Flexi-Seal™ PROTECT PLUS Fecal Management System (with ENFit™ connector), has the same intended use as the predicate device (K162906). Both devices include a low-profile retention balloon, a finger pocket to allow for easy balloon insertion, a green dome to indicate the optimal fill status of the retention balloon, a red dome to indicate the over inflation status, and the ability to administer medication through the use of a cinch clamp. In addition, the subject device incorporates new features, including:
- add a self-closing ENFit port for irrigation/medication and provide an ENFit (1) syringe in the kit,
- (2) add a sub-assembled self-closing quick connector to the catheter end, and
- add a matching bag connector in the collection bag with an option of (3) incorporating Diamonds™ gelling and odor control sachet (4 sachets per kit) to solidify and deodorize the fecal contents in the collection bag.
The subject device is comprised of the following components (in a clamshell pack with label card, IFU and 5 date labels) see Figure 1.
Figure 1: Flexi-Seal™ PROTECT PLUS Fecal Management System (with ENFit™ connector)
Image /page/4/Figure/8 description: The image shows a medical device with various components labeled with numbers. The device includes a tube-like structure labeled 4, with additional parts such as a connector labeled 7, a drainage bag labeled 11, and syringes labeled 9 and 10. There is also a small packet labeled 13, which appears to be a Diamonds ICE pack.
- Catheter Irrigation and Medication Port
- Balloon Inflation Port with 2-Dome Fill Indicator and Cap
- Sampling Por
-
-
- tention Balloon with
-
-
-
- ENFit Syrir
- 11 Collection Bag
- 12 Cinch Clamp
- 13 Diamonds™ gelling and odor control sachet
{6}------------------------------------------------
Figure 2 represents an overview of the predicate device:
Image /page/5/Figure/2 description: The image shows a diagram of a catheter and its components. The components are labeled with numbers 1 through 11, and a corresponding list describes each component. Some of the components include a catheter irrigation and medication port, a balloon inflation port, a sampling port, and a Luer-lock syringe. Other components include a collection bag with an integrated cap and a cinch clamp.
#### Figure 2: Flexi-Seal™ PROTECT Fecal Management System
#### 5. Intended Use
The Flexi-Seal™ PROTECT PLUS Fecal Management System (with ENFit™ connector) is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. The device is intended for use in adult patients.
#### 6. Performance Testing - Bench
There are no differences in the intended use or performance of the subject device and the predicate device. The device specifications are identical to those of the predicate device K162906, except for the new features described above. In addition to biocompatibility and stability tests, the following performance tests of the subject device were performed (results are detailed in Section 18 of this Submission):
(1) Joint strength of new ENFit™ connector to the catheter
(2) Joint strength of ENFit™ connector to the irrigation port housing
(3) Drug efficacy in ENFit™ connector
(4) Water leakage test of the ENFit™ connector
(5) Water leakage test of the self-closing catheter connector (both during use or when the bag is removed)
Based on non-clinical testing of the subject device for functional and structural parameters, the modifications in the subject device do not affect the safety and effectiveness. Flexi-Seal™ PROTECT PLUS Fecal Management System (with ENFit™ connector) described in this submission is, in our opinion, substantially equivalent to the predicate device. Flexi-Seal™ PROTECT Fecal Management System (K162906).
{7}------------------------------------------------
#### 7. Comparison of technological characteristic with the predicate device
It has been demonstrated through comparison of design features and performance testing, that the proposed device and its predicate have been found to be substantially equivalent.
The intended use, technological characteristics and principles of operation of the subject device remain the same as those of the predicate device.
#### 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## Sterilization & Shelf-life Testing
See Section 14
# Biocompatibility Testing
See Section 15
Electrical safety and electromagnetic compatibility (EMC) Not Applicable
Software Verification and Validation Testing Not Applicable
# Mechanical and acoustic Testing
Not Applicable
#### Animal Study
Animal performance testing was not required to demonstrate the safety and effectiveness of the proposed device.
## Clinical Studies
Clinical testing was not required to demonstrate the safety and effectiveness of the proposed device. Instead, substantial equivalence is based upon benchtop performance testing.
{8}------------------------------------------------
#### Table 1: Comparison of Flexi-Seal™ PROTECT PLUS Fecal Management System (with ENFit™ connector) vs Flexi-Seal™ PROTECT Fecal Management System (K162906), Bard™ DigniShield™ Stool Management System ENFit™ and Bard™ DigniShield™ Stool Management System
| Name | Subject Device:<br>Flexi-Seal™<br>PROTECT PLUS Fecal<br>Management System (with<br>ENFit™ connector) | Predicate Device:<br>Flexi-Seal™<br>PROTECT Fecal<br>Management System<br>(K162906) | Bard™ DigniShield™ Stool<br>Management System<br>ENFit™ | Bard™ DigniShield™ Stool<br>Management System<br>(K133251) |
|---------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Fecal Management | Fecal Management | Fecal Management | Fecal Management |
| Indications | The Flexi-Seal™<br>PROTECT PLUS Fecal<br>Management System is an<br>indwelling fecal<br>management catheter<br>intended for use to manage<br>fecal incontinence through<br>the collection of liquid or<br>semi-liquid stool and to<br>provide access to administer<br>medications. | The Flexi-Seal™<br>PROTECT Fecal<br>Management System is an<br>indwelling fecal<br>management catheter<br>intended for use to manage<br>fecal incontinence through<br>the collection of liquid or<br>semi-liquid stool and to<br>provide access to administer<br>medications. | Bard DigniShield™ Stool<br>Management System is<br>intended for fecal<br>management by diverting<br>and collecting liquid and<br>semi-liquid stool to<br>minimize skin contact in<br>bedridden patients and to<br>provide access for the<br>administration of<br>medications. Adult Use<br>Only. | Bard DigniShield™ Stool<br>Management System is<br>intended for fecal<br>management by diverting<br>and collecting liquid and<br>semi-liquid stool to<br>minimize skin contact in<br>bedridden patients and to<br>provide access for the<br>administration of<br>medications. Adult Use<br>Only. |
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile | Non-Sterile |
| Max Usage of Device | 29 days | 29 days | 29 days | 29 days |
| Balloon fill volume<br>per IFU | 35 - 45 ml based on<br>fill indicator | 35 - 45 ml based on<br>fill indicator | 45 ml | 45 ml |
| Optimal balloon fill<br>volume to be<br>indicated by a visual<br>fill indicator | Yes | Yes | No | No |
| Over-inflation<br>indicator | Yes | Yes | No | No |
| Odor Barrier Catheter | Yes | Yes | Yes | Yes |
| Catheter Material | Silicone with Zeolite | Silicone with Zeolite | Thermoplastic Elastomer and Silicone | Thermoplastic Elastomer and Silicone |
| Medication Delivery | Yes | Yes | Yes | Yes |
| ENFitTM Port for irrigation and medication (ISO 80369-3) | Yes | No (Luer) | Yes | No (Luer) |
| Self-Closing ENFitTM Port | Yes | No | No | No |
| ENFitTM Syringe in the kit | Yes | No | No | No |
| Self-Closing Catheter Connector | Yes | No | Yes | Yes |
| Collection Bag | Close with a cap (not self-closing) | Close with a cap (not self-closing) | Close with a cap (not self-closing) | Close with a cap (not self-closing) |
| Odor Control Sachet or Spray | DiamondsTM Gelling and odor control sachets for collection bag | No | Odor eliminator spray | Odor eliminator spray |
{9}------------------------------------------------
Note: Bard DigniShield™ Stool Management System ENFit™ is used as a comparison device for the ENFit and self-closing catheter connector only
ENFit™ is a trademark of Global Enteral Device Supplier Association (GEDSA)
Flexi-Seal™, Signal™, Diamonds™ - Trademark of ConvaTec Inc., NJ
BardTM and DigniShield™- Trademark of CR Bard, Inc. Covington,
{10}------------------------------------------------
#### 9. Conclusion
The modifications made to the existing device do not alter its intended use, change fundamental performance, or introduce new questions of safety and efficacy when compared with the predicate device. This comparison demonstrates that Flexi-Seal™ PROTECT PLUS Fecal Management System (with ENFit™ connector) is substantially equivalent to ConvaTec's previously cleared Flexi-Seal™ PROTECT Fecal Management System (K162906).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.