The Flexi-Seal PROTECT FMS is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. The device is not intended for use on pediatric patients.
Device Story
Indwelling fecal management catheter; manages fecal incontinence by collecting liquid to semi-liquid stool; provides access for medication administration. Device consists of collapsible catheter with distal soft annular silicone retention balloon; proximal end connects to disposable 1L collection bag with gas filter. Operated by clinicians in clinical settings. Features include 2-dome fill indicator (green for optimal 35-45ml inflation, red for over-inflation), inflation port cap, and color-coded sampling port. Input is fecal matter; output is collected waste. Benefits include containment of stool, protection of patient skin, and simplified medication delivery. Modifications from predicate include updated fill indicator, inflation port cap, and reduced number of collection bags.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included fill indicator functionality (optimal/over-inflation), joint strength (fill indicator, luer valve, cap), seal integrity (sampling port), and journey hazard testing to confirm structural integrity of modifications.
Technological Characteristics
Indwelling catheter; silicone rubber construction with 3% zeolite; soft annular retention balloon (45ml capacity). Dimensions: 1,550-1,670mm length, 56.82mm inflated balloon diameter. Connectivity: Luer-lock inflation/irrigation ports. Non-sterile. Packaging: PETG/APET clamshell. Features 2-dome visual fill indicator and external cinch clamp.
Indications for Use
Indicated for management of fecal incontinence in adult patients via collection of liquid to semi-liquid stool and administration of medications. Not for use on pediatric patients.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Flexi-Seal® SIGNAL™ Fecal Management System (K150350)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a series of interconnected profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 7, 2017
ConvaTec Karen Howes Regulatory Affairs Manager GDC, First Avenue, Deeside Industrial Park Deeside, Flintshire CH5 2NU United Kingdom
Re: K162906
> Trade/Device Name: Flexi-Seal® PROTECT™ Fecal Management System Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT Dated: February 2, 2017 Received: February 6, 2017
Dear Karen Howes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) Unknown K162906
Device Name
Flexi-Seal® PROTECT TM Fecal Management System
#### Indications for Use (Describe)
The Flexi-Seal® PROTECT ™ Fecal Management System is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid stool and to provide access to administer medications. The device is not intended for use on pediatric patients.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 243-1741 EF
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image features the logo of ConvaTec, a global medical products and technologies company. The logo consists of two distinct elements: a circular icon on the left and the company name "ConvaTec" on the right. The circular icon is a blue circle with a white graphic inside, resembling a building with columns. The text "ConvaTec" is written in a bold, sans-serif font, also in blue, matching the color of the circular icon.
# Section 5 510(k) Summary
Applicant: ConvaTec Limited.
Applicant Address: GDC, First Avenue Deeside Industrial Park Deeside Flintshire CH5 2NU UK The Establishment Registration number is 1000317571
Contact: Karen Howes Regulatory Affairs Manager ConvaTec Ltd. GDC, First Avenue, Deeside industrial Park Deeside Flintshire, CH5 2NU UK Tel: +44(0) 1244832195
Date Prepared: 1 December 2016
Device Identification:
Trade Name: Flexi-Seal® PROTECT™ Fecal Management System Common Name: Fecal Management System
Regulation Number: 21 CFR §876.5980
Classification Panel: Gastroenterology and Urology
Classification Name: Gastrointestinal tube and accessories
Regulatory Class: II
Product Code: KNT
Predicate Devices: 510(K) number: K150350 Flexi-Seal® SIGNAL™ Fecal Management System
{5}------------------------------------------------
### Device Description
The new device is called Flexi-Seal® PROTECT™ Fecal Management System which is an upqrade to the predicate device K150350. Flexi-Seal® Signal™ Fecal Management System (also known as Flexi-Seal® Signal™ Fecal Management System with Odor Barrier), launched in October 2015. The new device uses the same design and components except
- (1) the change from the 1-dome fill indicator to a new 2-dome fill indicator to provide visual cue by using
- (a) a green color dome to indicate an optimal fill status of retention balloon which is similar to the existing fill indicator except the color, and
- (b) a red color dome to indicate the over inflation status of retention balloon,
- (2) adding a white cap that can close off the inflation port,
- (3) adding a dark blue color in the sample port, and
- (4) Privacy bag (one bag instead of 3).
The device is comprised of the following components (in a clamshell pack with label card, IFU and 5 date labels): see diagram
Image /page/4/Figure/9 description: The image shows a medical catheter with different parts labeled with numbers 1-11. The parts include the catheter irrigation and medication port, balloon inflation port, sampling port, catheter, position indicator line, low-pressure retention balloon, connector, hanging strap, luer-lock syringe, collection bag, and cinch clamp. The catheter is a long, flexible tube that is inserted into the body to drain fluids or deliver medications.
# New device Flexi-Seal® PROTECT™ Fecal Management System
{6}------------------------------------------------
Predicate device: Flexi-Seal® SIGNAL™ Fecal Management System
Image /page/5/Figure/1 description: This image shows a diagram of a catheter system with different parts labeled with numbers 1-11. The numbers correspond to the following parts: Catheter Irrigation and Medication Port, Balloon Inflation Port with Fill Indicator, Sampling Port, Catheter, Position Indicator Line, Low-Pressure Retention Balloon with Finger Pocket, Connector, Hanging Strap, Luer-lock Syringe, Collection Bag with Integrated Cap, and Cinch Clamp. The diagram shows the different parts of the catheter system and how they are connected.
### Intended Use
The Flexi-Seal PROTECT FMS is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications. The device is not intended for use on pediatric patients.
# Performance Testing - Bench
There are no differences in the intended use or performance of the subject and the predicate device. Non clinical testing has been conducted on the predicate and subject devices. The list of non-clinical testing for which the results are detailed in Sec. 18 includes:
- 1. Fill indicator testing:
- a. Fill indication for optimal inflation volume of retention balloon
- b. Fill Indictor for overfilling status of retention balloon
- c. Fill Indicator joint strength to catheter
- d. Luer valve joint strength to Inflation Port
- e. Green ink adhesion
- Red ink adhesion f.
- 2. Inflation port cap testing:
- a. Cap joint strength to the fill indicator
- b. Cap removal force
- c. Cap repeated removal 5 times
- 3. Sampling port testing:
- a. Sampling port seal against air
- b. Sampling port seal against water
- 4. Privacy bag testing:
{7}------------------------------------------------
- a. Width of clear window at 10 25 mm
- b. Graduation at 500 ml is within ± 15%
- c. Graduation at 1000 ml is within ± 15%
- d. Pouch has 1 Liter capacity
- e. Ink Adhesion
- 5. Journey hazard testing:
- a. Journey hazard testing
- b. Post-JHT Product Release Testing Visual, Color, and Cap
- c. Post-JHT Product Release Testing Fill indication for optimal inflation volume of retention balloon
- d. Post-JHT Product Release Testing Fill Indictor for overfilling status of retention balloon
Based on non-clinical testing of the subject device for functional and structural parameters, the modifications in the subject device do not affect the safety and effectiveness of the predicate device. Flexi-Seal® PROTECT™ Fecal Management System described in this submission is, in our opinion, substantially equivalent to the predicate device, Flexi-Seal® SIGNAL™ Fecal Management System per K150350.
### Substantial Equivalence Conclusion
It has been demonstrated through comparison of design features and performance testing, that the proposed device and its predicate have been found to be substantially equivalent; see following pages.
{8}------------------------------------------------
### Substantial Equivalence Discussion
The following table compares the similarities and differences between the subject device: Flexi-Seal® PROTECT™ Fecal Management System and the predicate device: Flexi-Seal® SIGNAL™ Fecal Management System (ref; K150350) and outlines the product characteristics and specifications which form the basis of the substantial equivalence discussion.
The intended use, technological characteristics and principles of operation of the fecal management system remains the same as those of the predicate device.
| | Subject Device | Predicate | Comparison | |
|----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Parameter | Flexi-Seal® PROTECT™ Fecal Management<br>System | Flexi-Seal® SIGNAL™ Fecal Management System<br>(K150350) | Similarities | Differences |
| FDA Product Code | KNT | KNT | Both devices are the same | None |
| FDA Classification<br>Regulation | 21 CFR 876.5980 | 21 CFR 876.5980 | Both devices are the same | None |
| Regulatory Class | Class II | Class II | Both devices are the same | None |
| Intended Use | To manage fecal incontinence through the<br>collection of liquid to semi-liquid stool and to<br>provide access to administer medications | To manage fecal incontinence through the collection<br>of liquid to semi-liquid stool and to provide access to<br>administer medications | Both devices are the same | None |
| Functional<br>configuration | Collapsible catheter with distal end secured<br>in the rectum and proximal end connected to<br>a collection bag | Collapsible catheter with distal end secured in the<br>rectum and proximal end connected to a collection<br>bag | Both devices are the same | None |
| Retention feature | Soft annular balloon | Soft annular balloon | Both devices are the same | None |
| Balloon material | Silicone rubber | Silicone rubber | Both devices are the same | None |
| Inflation<br>management | Sealed water filled balloon | Sealed water filled balloon | Both devices are the same | None |
| Retention balloon | Water/saline (45ml) | Water/saline (45ml) | Both devices are the same | None |
| inflation material | | | same | |
| Retention balloon<br>inflated diameter | 56.82mm | 56.82mm | Both devices are the<br>same | None |
| Retention balloon<br>inflated height | 25.96mm | 25.96mm | Both devices are the<br>same | None |
| Catheter tube<br>material | Silicone rubber plus 3% zeolite, by weight | Silicone rubber plus 3% zeolite, by weight | Both devices are the<br>same | None |
| Catheter tube<br>surface roughness | <15µm | <15µm | Both devices are the<br>same | None |
| Fill indication for<br>optimal inflation<br>volume of<br>retention balloon | Green dome inflated at 35 - 45 ml | White dome inflated at 35 - 45 ml | Except color, both<br>devices provided the<br>same visual indication<br>for optimal balloon<br>inflation volume. The<br>dimension of the fill<br>indicator is similar. | Color |
| Fill Indictor for<br>overfilling status of<br>retention balloon | Red dome inflated beyond 45 ml and above | None | A new feature in the<br>subject device using a<br>similar fill indicator<br>design | A new feature in the<br>subject device |
| Cap for inflation<br>port | A removable silicone cap to close off the<br>inflation port with sufficient removal force<br>and joint strength | None | A new feature in the<br>subject device to<br>differentiate the<br>inflation port from the<br>irrigation port | A new feature in the<br>subject device |
| Catheter tube<br>hardness | Shore A 38.5 - 51 | Shore A 38.5 - 51 | Both devices are the<br>same | None |
| Catheter tube wall<br>thickness | 0.40 - 0.65mm | 0.40 - 0.65mm | Both devices are the<br>same | None |
| Catheter tube<br>tensile strength | 245N | 245N | Both devices are the<br>same | None |
| Balloon-to-tube<br>bonding strength | 25.4N | 25.4N | Both devices are the<br>same | None |
| Catheter tube<br>length | 1,550mm - 1,670mm | 1,550mm - 1,670mm | Both devices are the<br>same | None |
| Length of<br>sphincter section | >10cm | >10cm | Both devices are the<br>same | None |
| Drain channel | Large main drain tube | Large main drain tube | Both devices are the<br>same | None |
| Inflation and<br>irrigation port<br>connections | Luer | Luer | Both devices are the<br>same | None |
| Inflation lumen<br>diameter | Two lumens at 1.45 mm | Two lumens at 1.45 mm | Both devices are the<br>same | None |
| Inflation lines | 2 | 2 | Both devices are the<br>same | None |
| Irrigation lines | 1 | 1 | Both devices are the<br>same | None |
| Irrigation lumen<br>size | 8mm x 1.6mm (collapsible) | 8mm x 1.6mm (collapsible) | Both devices are the<br>same | None |
| Sampling port | Closable snap seal for catheter syringe<br>access - dark blue color | Closable snap seal for catheter syringe access -<br>natural color | Except color, both<br>devices provided the<br>same seal against<br>water and air. | Color |
| Waste lumen<br>diameter collapsed | 10mm | 10mm | Both devices are the<br>same | None |
| Collection bag<br>configuration | Disposable with gas filter | Disposable with gas filter | Both devices use the<br>same filter, flange,<br>and film panels in the<br>collection bag. | One bag in the subject<br>device vs. three bags in<br>the predicate device |
| Collection bag size | 1 litre disposable | 1 litre disposable | Both devices are the<br>same | None |
| Collection bag<br>material | Ethyl vinyl acetate | Ethyl vinyl acetate | Both devices are the<br>same | None |
| Collection bag -<br>front panel color | White with a clear window | Clear | Except color in the<br>front panel, both bags<br>perform the same | Color |
| Method of bed<br>connection | Hanging strap on bag connector | Hanging strap on bag connector | Both devices are the<br>same | None |
| Accessory<br>components | Syringe and cinch clamp | Syringe and cinch clamp | Both devices are the<br>same | None |
| Flow stop<br>mechanism | External cinch clamp | External cinch clamp | Both devices are the<br>same | None |
| Sterility status<br>(how provided) | Non-sterile | Non-sterile | Both devices are the<br>same | None |
| Packaging | Thermoformed plastic clamshell PETG<br>(POLYETHYLENE TEREPHTHALATE<br>GLYCOL-MODIFIED) or APET<br>(AMORPHOUS POLYETHYLENE<br>TEREPHTHALATE), both sides coated with<br>2% silicone. | Thermoformed plastic clamshell PETG<br>(POLYETHYLENE TEREPHTHALATE GLYCOL-<br>MODIFIED) or APET (AMORPHOUS<br>POLYETHYLENE TEREPHTHALATE), both sides<br>coated with 2% silicone. | Both devices use the<br>same packaging<br>material | None |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.