INDWELLING FECAL MANAGEMENT SYSTEM

K012113 · Bowel Management Systems, LLC · KNT · May 3, 2002 · Gastroenterology, Urology

Device Facts

Record IDK012113
Device NameINDWELLING FECAL MANAGEMENT SYSTEM
ApplicantBowel Management Systems, LLC
Product CodeKNT · Gastroenterology, Urology
Decision DateMay 3, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5980
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Indwelling Fecal Management System (IFMS) is indicated for diversion of fecal matter to minimize external contact with the patient, to facilitate the collection of fecal matter for patients requiring stool management, and to provide access for colonic irrigation to trigger a defecatory response, and administration of enemas/medications.

Device Story

Indwelling Fecal Management System (IFMS) is a rectal irrigation and drainage catheter; used for fecal diversion to minimize external contact and facilitate collection. Device features include an external retention balloon, silicone lumen with flared capped port, intraluminal (ARV) balloon for irrigation, and external silicone faceplate with anchor tabs for drainage flow suspension. Operates by insertion into the rectum; allows for colonic irrigation to trigger defecatory response and administration of enemas/medications via port access. Compatible with catheter-tip syringes for flushing and sampling. Designed for clinical use to manage patient bowel function; provides a closed system for waste management. Benefits include reduced skin contact with fecal matter and controlled access for therapeutic bowel interventions.

Clinical Evidence

No clinical data provided; substantial equivalence is based on bench testing and design comparison.

Technological Characteristics

Silicone catheter with external retention balloon, intraluminal (ARV) irrigation balloon, and external silicone faceplate for drainage suspension. Features a flared, capped port for sampling and fluid administration. Sterile, single-use device. No electronic components or software.

Indications for Use

Indicated for patients requiring stool management for fecal diversion, colonic irrigation, or administration of enemas/medications.

Regulatory Classification

Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ 510(k) Summary Indwelling Fecal Management System (IFMS) Bowel Management Systems, LLC # 510(k) Summary The following safety and effectiveness summary has been prepared pursuant to requirement for 510(k) summaries specified in 21CFR 807.92(a). ## 807.92(a)(1) ## Submitter Information | Chad Coberly<br>Bowel Management Systems, LLC | | | | |-----------------------------------------------|------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------| | 1886 South 14th Street, Suite 6 | | | | | Fernandina Beach, FL 32034 | | | | | Phone: | (904) 261-2169 | | | | Facsimile: | (904) 261-2172 | | | | Contact Person: | | Colleen Hittle, Official Correspondent - (317) 849-1916 ext. 102<br>Carri Graham - (317) 849-1916 ext. 103 | | | Date: | February 1, 2002 | | | | 807.92(a)(2) | | | | | Trade Name: | | Indwelling Fecal Management System (IFMS) | | | Common Name: | | Rectal Irrigation tube | | | Classification Name(s): | | Tubes, Gastrointestinal | | | Classification Number: | | 78KNT | | | 807.92(a)(3) | | | | | | | Predicate Device(s) | | | HDC | | Colo-Vage System | K841289 | | Leon's Fecal Sanitary<br>Tube Manufacture | | Leon's Fecal Sanitary Tube | K813526 | | Bard | | Virden Rectal Catheter | Exempt per 21CFR 876.5980<br>(barium enema retention<br>catheter) | Additional Substantial Equivalence Information is provided in the following substantial Equivalence Comparison Table. K012113 {2}------------------------------------------------ ### 807.92(a)(5) #### Intended Use(s) The Indwelling Fecal Management System (IFMS) is indicated for diversion of fecal matter to minimize external contact with the patient, to facilitate the collection of fecal matter for patients requiring stool management, and to provide access for colonic irrigation to trigger a defecatory response, and administration of enemas/medications. {3}------------------------------------------------ 510(k) Summary Indwelling Fecal Management System (IFMS) Bowel Management Systems, LLC on Chart for Substantial Equivalence | Function | | | | | |--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|---------------------------------------------------------------------|--------------------------------------------------------| | | | Device Feature | | | | | IFMS Catheter<br>(Zassi Medical Evolutions) | Colo-Vage System<br>(HDC Corporation) | Virden Rectal<br>Catheter<br>(Bard) | Fecal Sanitary Tube<br>(Leon's Fecal Sanitary<br>Tube) | | | This submission | K841289 | Exempt per<br>21CFR876.5980<br>(barium enema<br>retention catheter) | K813526 | | Indications for Use | Diversion of fecal matter to<br>minimize external contact<br>with the patient, to facilitate<br>the collection of fecal matter<br>for patients requiring stool<br>management, and to provide<br>access for colonic irrigation<br>to trigger a defecatory<br>response, and administration<br>of enema/medications. | Colonic irrigation and<br>drainage of fecal<br>matter. | Administration of<br>barium during<br>radiopaque evaluation. | Collection of fecal<br>excretion | | Bowel Retention | External Balloon<br>Silicone lumen with flared,<br>capped port termination | External Balloon<br>Silicone lumen with<br>flared termination | External Balloon<br>No irrigation | External Ring<br>No irrigation | | Bowel Irrigation | Intraluminal (ARV) balloon | | | | | Drainage Flow Suspension | External silicone retention<br>faceplate w/ anchor tabs | | No drainage suspension | No drainage suspension | | Anti-Internal Migration | Mid-line silicone access port | | No anti-internal<br>migration | No anti-internal<br>migration | | Flush / Stool Sampling | compatible with catheter tip<br>syringe | | No flush / stool<br>sampling | No flush / stool sampling | | | IFMS Catheter<br>(Zassi Medical Evolutions) | Colo-Vage System<br>(HDC Corporation) | Virden Rectal<br>Catheter<br>(Bard) | Fecal Sanitary Tube<br>(Leon's Fecal Sanitary<br>Tube) | | | This submission | K841289 | | K813526 | | Enema / Medication<br>Administration | Silicone lumen with flared,<br>capped port termination | Silicone lumen with<br>flared termination | Natural rubber latex<br>lumen with flared<br>termination | No enema / medication<br>administration | | Drainage / Collection | Waste collection bag not<br>provided in kit | Sterile drainage unit<br>not provided | No drainage /<br>collection | Waste collection bag<br>integral to the device | | Port Access | Sampling / fluid<br>administration | Sampling / fluid<br>administration | Not specified | Not specified | | Sterile | Yes | Yes | No | Yes | Page ್ರಿ of 4 {4}------------------------------------------------ Footactes: Allows for temporary closure of catheter drains of bowel irrigation or for retention of enemalmedications after administration. : ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﺘﻲ ﺗﻘﺮﻳﺒﺎ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﺘﻲ ﺗﻘﺮﻳﺒﺎ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘ 1 す Page 4 of : ・ {5}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol consisting of three stylized human profiles facing to the right, with flowing lines representing hair or movement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## MAY 0-3-2002 Bowel Management Systems, LLC c/o Ms. Colleen J. Densmore Official Correspondent The Anson Group, LLC 7992 Castleway Drive Indianapolis, Indiana 46250 Re: K012113 Trade/Device Name: Indwelling Fecal Management System (IFMS) Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: 78 KNT Dated: February 1, 2002 Received: February 4, 2002 Dear Ms. Densmore: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ #### Page 2 - Ms. Colleen Densmore Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loast of activitied that I Drivistion that your device complies with other requirements of the Act that I Drinas made a actuations administered by other federal agencies. You must or any it coloral statutes and regainents, including, but not limited to: registration (21 CFR Part eomply with an alle 110 - 100 - 104 - 10 - 10 - 10 - 2011 - 10 - 10 - 10 - 2011 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 200 007), noung (21 CF Reason the quality systems (QS) regulation (21 CFR Part 820); and if requirements as electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) r ms letter will and w you to organ maing of substantial equivalence of your device to a legally prematic notification: "I results in a classification for your device and thus, permits your marketed proceed to the market. If you desire specific advice for your device on the labeling ucerio to proced to the manew additionally Part 809.10 for in vitro diagnostic devices), please rogulation office of Compliance at (301) 594-4616. Additionally, for questions on the contact the Office of Compilance and evice, please contact the Office of Compliance at promotion and acrossible se note the regulation entitled, "Misbranding by reference to (2011) 371 1057. 1180, pread 807.97). Other general information on your responsibilities promation act may be obtained from the Division of Small Manufacturers, International and ander the Ace may be obtained from and ber (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely vours. Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ Page 510(k) Number (if known): _ K012113 Device Name:_Indwelling Fecal Management System (IFMS) Indications For Use: Diversion of fecal matter to minimize external contact with the patient, to facilitate the collection of fecal matter for patients requiring stool management, and to provide access for colonic irrigation to trigger a defecatory response, and administration of enemas/medications. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Maurice Broadon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices K012113 510(k) Number(Optional Format 3-10-98) Prescription Use (Per 21 CFR 801.109)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...