HYDROSOFT AND HYDROSOFT PLUS EMBOLIZATION COIL SYSTEMS

K080666 · MicroVention, Inc. · HCG · Jul 11, 2008 · Neurology

Device Facts

Record IDK080666
Device NameHYDROSOFT AND HYDROSOFT PLUS EMBOLIZATION COIL SYSTEMS
ApplicantMicroVention, Inc.
Product CodeHCG · Neurology
Decision DateJul 11, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5950
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HydroSoft and HydroSoft Plus Embolization Coil Systems with HES-HC-HS (10) Coils are intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and artiovenous fistula. The HydroSoft and HydroSoft Plus Embolization Coils Systems also are intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature.

Device Story

Platinum embolization coils with inner hydrogel core; delivered via V-Trak delivery pusher. Used for endovascular embolization of intracranial aneurysms, neurovascular abnormalities, and peripheral vascular occlusion. Operated by physicians in clinical settings. Device permanently obstructs blood flow to aneurysms or vascular malformations. Benefits include effective vascular occlusion.

Clinical Evidence

No clinical data provided; substantial equivalence based on technical and performance attributes.

Technological Characteristics

Platinum embolization coils with inner hydrogel core; V-Trak delivery pusher. Dimensions include larger secondary diameters and longer lengths compared to predicate. No software or electronic components.

Indications for Use

Indicated for endovascular embolization of intracranial aneurysms, arteriovenous malformations, and arteriovenous fistulae; vascular occlusion of neurovascular blood vessels to obstruct blood flow to aneurysms or malformations; and arterial and venous embolization in the peripheral vasculature.

Regulatory Classification

Identification

A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.

Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K080666 ### 510(k) Summary JUL 11 2008 | Trade Name: | HydroSoft® and HydroSoft <i>Plus</i> Embolization Coil Systems with HES-<br>HC-HS (10) Coils | |-----------------------|----------------------------------------------------------------------------------------------| | Generic Name: | Artificial Embolization Device | | Classification: | Class II, 882.5950<br>Class II, 870.3300 | | Submitted By: | MicroVention, Inc.<br>75 Columbia, Suite A<br>Aliso Viejo, CA 92656 | | Contact: | Kevin E. Daly | | Predicate<br>Devices: | HydroCoil® Embolic System with (HES) with the HES-HC-HS (10)<br>Coils (K070656) | Device Description: The HydroSoft® and HydroSoft Plus Embolization Coil Systems with HES-HC-HS (10) Coils are platinum embolization coils with an inner hydrogel core, and a V-TrakTM Delivery Pusher. ## Indications for Use: The HydroSoft and HydroSoft Plus Embolization Coil Systems with HES-HC-HS (10) Coils are intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and artiovenous fistula. The HydroSoft and HydroSoft Plus Embolization Coils Systems also are intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. ## Comparison to Predicate Device: The HydroSoft and HydroSoft Plus Embolization Coils Systems are similar in materials of construction and intended use to the predicate HydroCoil device, but are available in larger secondary diameters and longer lengths to suite the needs of the physician. ## Summary: Based upon the technical and performance attributes of the HydroSoft and HydroSoft Plus Embolization Coils Systems, these devices are substantially equivalent to the cited predicate device. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MicroVention, Inc. % Mr. Kevin E. Daly Vice President, Regulatory Affairs and Quality Assurancc 75 Columbia, Suite A Aliso Viejo, California 92656 JUL 11 2008 Re: K080666 Trade/Device Name: HydroSoft® Plus Embolization Coil Systems with HES-HE-HS (10) Coil Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular embolization device Regulatory Class: II Product Code: HCG Dated: June 6, 2008 Received: June 9, 2008 Dear Mr. Daly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {2}------------------------------------------------ Page 2 - Mr. Kevin E. Daly forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark M. Millman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # 1080666 ## Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: HydroSoft® and HydroSoft Plus Embolization Coil Systems with HES-HC-HS (10) Coils Indications for Use: The HydroSoft and HydroSoft Plus Embolization Coil Systems with HES-HC-HS (10) Coils are intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and artiovenous fistula. The HydroSoft and HydroSoft Plus Embolization Coils Systems also are intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. | Prescription Use | X | |-----------------------------|---| | (Part 21 CFR 801 Subpart D) | | AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | |------------------------------------------------------------| | Division of General, Restorative, and Neurological Devices | | 510(k) Number | 1080666 | |---------------|---------| |---------------|---------| 004
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...