BARE PLATINUM FRAMING COIL, MODEL(S) 100408BPFC-V,100412BPFC-V,100515BPFC-V100518BPFC-V,100522BPFC-V,100618BPFC-V100622B

K082461 · MicroVention, Inc. · HCG · Oct 2, 2008 · Neurology

Device Facts

Record IDK082461
Device NameBARE PLATINUM FRAMING COIL, MODEL(S) 100408BPFC-V,100412BPFC-V,100515BPFC-V100518BPFC-V,100522BPFC-V,100618BPFC-V100622B
ApplicantMicroVention, Inc.
Product CodeHCG · Neurology
Decision DateOct 2, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.5950
Device ClassClass 2
AttributesTherapeutic

Intended Use

The MicroPlex Coil System (MCS) is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.

Device Story

Bare Platinum Framing Coils (BPFC) are platinum alloy implant coils designed in 3D spherical structures for neurovascular embolization. Coils attach to a V-Trake™ delivery pusher via a polymer filament. The pusher features radiopaque markers for positioning. The system is used with a hand-held, battery-powered V-Grip™ Detachment Controller. During endovascular procedures, a physician navigates the coil to the target site; activation of the Detachment Controller triggers the release of the implant. The device permanently obstructs blood flow to aneurysms or vascular malformations, providing a therapeutic benefit by reducing the risk of rupture or hemorrhage. Used in clinical settings by neurovascular specialists.

Clinical Evidence

Bench testing only.

Technological Characteristics

Implant: Platinum alloy coils in 3D spherical structures. Delivery: V-Trake™ MCS pusher with radiopaque markers. Detachment: Hand-held, battery-powered V-Grip™ Detachment Controller. Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for endovascular embolization of intracranial aneurysms, arteriovenous malformations, and arteriovenous fistulae; vascular occlusion of neurovascular blood vessels to obstruct flow to aneurysms or malformations; and arterial/venous embolizations in peripheral vasculature.

Regulatory Classification

Identification

A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.

Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary | Trade Name: | Bare Platinum Framing Coils, BPFC | | |-----------------|--------------------------------------------------------------------|--------------| | Generic Name: | Neurovascular Embolization Device | OCT - 2 2008 | | Classification: | Class II, 21 CFR 882.5950 | | | Submitted By: | MicroVention, Inc<br>75 Columbia<br>Aliso Viejo, California U.S.A. | | | Contact: | Florin Truuvert | | ## Predicate Device: | Number | Description | Predicate For | Clearance Date | |---------|------------------------------------------------------|--------------------------------|----------------| | K050954 | MicroVention Inc.,<br>MicroPlex Coil<br>System (MCS) | Bare Platinum Framing<br>Coils | June 28, 2005 | ### Device Description The Bare Platinum Framing Coils, BPFC, consist of implant coil made of platinum alloy. The coils are designed in 3D spherical structure in various loop sizes and lengths. The coil is attached to a V-Trake™ MCS delivery pusher via a polymer filament. The delivery pusher contains radiopaque positioning markers at the distal end. The proximal end is inserted into a hand held battery powered V-Grip™ Detachment Controller. The implant segment detaches upon activation of the Detachment Controller. #### Indication For Use The BPFC as a member of the MicroPlex Coil System (MCS) is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird-like figure with three curved lines representing wings or ribbons. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # OCT - 2 2008 Ms. Florin Truuvert Director, Regulatory Affairs Microvention, Incorporated 75 Columbia, Suite A Aliso Vicjo, California 92656-1408 Re: K082461 Trade/Device Name: Bare Platinum Framing Coils (BPFC) Regulation Number: 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: II Product Code: HCG Dated: August 25, 2008 Received: September 4, 2008 Dear Ms. Truuvert: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2- Ms. Truuvert Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N Millerson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known): K082461 Device Name: Bare Platinum Framing Coils (BPFC) Indications For Use: The MicroPlex Coil System (MCS) is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The MCS is also intended for wascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. Prescription Use AND/OR Over-The-Counter Use Page 1 of (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|--| |--------------------------------------------------------|--| (Division Sign-Off) Reception Control, Dental Devices Yhil Ko Ogden for axn 510 (k) Number: (Division Sign-Off) Division of General, Restorative, and Neurological Devices Page 14 of 113 **510(k) Number** K082461
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...