HYDROCOIL EMBOLIC SYSTEM (HES)

K070656 · MicroVention, Inc. · HCG · Jun 15, 2007 · Neurology

Device Facts

Record IDK070656
Device NameHYDROCOIL EMBOLIC SYSTEM (HES)
ApplicantMicroVention, Inc.
Product CodeHCG · Neurology
Decision DateJun 15, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5950
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HydroCoil® Embolic System (HES) is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.

Device Story

HydroCoil Embolic System (HES) consists of an implantable embolic coil attached to a V-TRAK Delivery Pusher. The pusher features a variable stiffness stainless steel mandrel with silver electrical leads and a platinum/stainless steel heater coil at the distal end. The system is used by neuro-interventionalists in clinical settings to treat intracranial aneurysms and vascular malformations. The coil is delivered via standard micro-catheters. Detachment is achieved using a hand-held, battery-powered V-GRIP Detachment Controller; when activated, electrical current heats and melts a polyolefin elastomer filament at the coupler junction, releasing the coil. This permanent occlusion obstructs blood flow to the target site, providing therapeutic benefit for aneurysms and vascular abnormalities.

Clinical Evidence

Bench testing only. Performance testing demonstrated that the HES with HES-HC-HS (10) coils is equivalent in performance to the predicate devices.

Technological Characteristics

Implantable embolic coil with polyolefin elastomer filament detachment mechanism. Delivery system: stainless steel mandrel, silver electrical leads, platinum/stainless steel heater, PET and polyimide tubing. Energy source: battery-powered electrical current for thermal detachment. Sterilization method not specified.

Indications for Use

Indicated for patients requiring endovascular embolization of intracranial aneurysms, arteriovenous malformations, arteriovenous fistulae, and vascular occlusion of neurovascular or peripheral arterial and venous vessels.

Regulatory Classification

Identification

A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.

Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K07D656 page # Section 2 - Summary of Safety and Effectiveness # Company Information (1) MicroVention, Inc. 75 Columbia Aliso Viejo, CA 92656 Telephone: (949) 461-3314 Fax: (949) 349-1360 www.microvention.com JUN 1 5 2007 # Contact Information (2) Vincent Cutarelli Telephone: (949) 951-0505 Fax: (949) 349-1360 E-mail: vinc@microvention.com #### (3) Device Name | Classification Name: | Device, Artificial Embolization | |-------------------------|---------------------------------| | Trade/Proprietary Name: | HydroCoil® Embolic System (HES) | | Common/Usual Name: | Embolization Coil | | Classification | Class II | # (4) Device Description The HydroCoil® Embolic System (HES) is based on electrical coil detachment. The HES consist of an implantable coil attached to a delivery system called a V-TRAK™ Delivery Pusher. The Delivery Pusher is a variable stiffness, stainless steel and tapered mandrel. Two silver electrical leads run along the outside of the mandrel from the proximal to the distal end. Platinum and stainless steel wires are wound around the distal end of the mandrel to form the electrical heater and provide kink-resistance. Two outer layers of PET tubing cover the distal end of the pusher assembly. A layer of polyimide tubing covers the proximal end. The proximal end of the HES coils incorporates a platinum coupler for attachment to the Delivery Pusher. A polyolefin elastomer filament is attached to the proximal end of the coil. This filament runs through the inner lumen of the coil coupler and is attached to the distal end of the Delivery Pusher. {1}------------------------------------------------ K070656 page 2/2 The Delivery Pusher is powered by a hand-held, battery-powered V-GRIP™ Detachment Controller designed specifically for the HES. A gold-plated stainless steel connector at the proximal end of the Delivery Pusher is used to connect it to the Detachment Controller, The Detachment Controller is provided separately. The coil is delivered to treatment site on the Delivery Pusher through standard neuro-interventional micro-catheters. A removable introducer sheath on the outside of the Delivery Pusher assists in the placement of the HES into the micro-catheter. Once the coil is deployed, the proximal end of the Delivery Pusher is connected to the Detachment Controller. When the Detachment Controller is turned on, the flow of current melts the polyolefin elastomer filament at the coupler/pusher junction resulting in coil detachment. # (5) Indications for Use The HydroCoil® Embolic System (HES) is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. # (6) Name of Predicate or Legally Marketed Device The HydroCoil® Embolic System (HES) with the HES-HC-HS (10) coils is substantially equivalent to the HydroCoil® Embolic System (HES) with the HES-HC-S (10) coils that was determined to be substantially equivalent on December 30, 2003 (reference K033836) and the MicroPlex® Coil System (MCS) and HydroCoil® Embolic System (HES) with a Modified Detachment System that was determined to be substantially equivalent on June 28, 2005 (reference K050954). # (7) Technological Characteristics and Substantial Equivalence The HydroCoil® Embolic System (HES) with the HES-HC-HS (10) coils is substantially equivalent in operating principle, method of applications for use, design, materials, packaging and sterilization to the predicate devices. # Performance Data Summary (8) Performance testing has demonstrated that the HydroCoil® Embolic System (HES) with the HES-HC-HS (10) coils is equivalent in performance to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of a human figure, represented by three overlapping profiles, suggesting a sense of community and support. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MicroVention, Inc % Mr. Vincent Cutarelli Vice President, Regulatory Affairs and Quality Assurance 75 Columbia, Suite A Aliso Viejo, California 92677 JUN 1 5 2007 Re: K070656 Trade/Device Name: HydroCoil® Embolic System with (HES) with the HES-HC-HS (10) coils Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular embolization device Regulatory Class: II Product Code: HCG, KRD Dated: May 30, 2007 Received: May 31, 2007 Dear Mr. Cutarelli: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Vincent Cutarelli forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index20tm1 Sincerely yours, Fox Po Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and ROMA DIRECT 6/14. Radiological Health Enclosure {4}------------------------------------------------ # Indications For Use 510(k) Number: Device Name: HydroCoil® Embolic System (HES) with the HES-HC-HS (10) coils Indications for Use: The HydroCoil® Embolic System (HES) is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. Concurrence of CDRH, Office of Device Evaluation (ODE): N (Division Sign-Off) Division of General, Restorative, and Neurological Devices **510(k) Number** L67065 Prescription Use: X (Per 21 CFR 801.109)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...