MICROPLEX COIL SYSTEM (MCS), HYDROCOIL EMBOLIC SYSTEM (HES)
K132952 · MicroVention, Inc. · HCG · Dec 19, 2013 · Neurology
Device Facts
Record ID
K132952
Device Name
MICROPLEX COIL SYSTEM (MCS), HYDROCOIL EMBOLIC SYSTEM (HES)
Applicant
MicroVention, Inc.
Product Code
HCG · Neurology
Decision Date
Dec 19, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. Also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
Device Story
MicroPlex Coil System (MCS) and HydroCoil Embolic System (HES) are neurovascular embolization devices. MCS consists of bare platinum/tungsten alloy coils; HES includes the same alloy with a hydrogel inner core. Both attach to a V-Trak Advanced delivery pusher via a polyolefin elastomer filament. Used in neurointerventional procedures, the device is navigated to the target site. A hand-held, battery-powered V-Grip Detachment Controller applies electrical current to the pusher, heating the polyolefin filament to detach the coil. The device permanently obstructs blood flow to aneurysms or vascular malformations. Operated by physicians in clinical settings, the system provides mechanical occlusion to treat neurovascular abnormalities and peripheral vascular conditions.
Clinical Evidence
Bench testing only. Testing included visual inspection, simulated use (deployment, repositioning, detachment), pusher resistance (36.7-53.0 Ω), and detachment zone tensile strength (≥0.08 lbf). Biocompatibility testing for MCS, HES, and delivery pusher followed ISO 10993 standards (MEM elution, sensitization, hemolysis, systemic toxicity, pyrogenicity, and implantation studies). All tests met acceptance criteria.
Technological Characteristics
Implantable coils made of platinum/tungsten (92/8%) alloy. HES includes hydrogel core. Delivery via V-Trak Advanced pusher with polyolefin elastomer detachment filament. Coupler: platinum/iridium (90/10%). Adhesive: DYMAX 1128-AM-VT UV. Detachment via electrical heating from external V-Grip controller. Sterilization method not specified.
Indications for Use
Indicated for patients requiring endovascular embolization of intracranial aneurysms, arteriovenous malformations, arteriovenous fistulae, and vascular occlusion of neurovascular or peripheral arterial and venous vessels.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}------------------------------------------------
K132952
DEC 1 9 2013
### 510(K) SUMMARY
(Prepared November 12, 2013)
MicroPlex Coil System (MCS), HydroCoil Embolic System (HES)
Laraine Pangelina, Sr. Regulatory Affairs Project Manager
Trade Name:
Generic Name: Neurovascular Embolization Device
Class II, 21 CFR 882.5950 Classification:
Submitted By: MicroVention, Inc 1311 Valencia Avenue Tustin, California 92780 U.S.A.
Contact:
Predicate Devices:
MicroPlex Coil System (MCS) (K103758, K091641, K111451, K103758, K08241, K093919, K090891, K093358, K102365) HydroCoil Embolic System (HES) (K091641, K103758, K113457, K112226, K120908, K090357, K100454, K091641, K080666)
Intended for the endovascular embolization of Indications for Use: intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. Also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to
an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
Treatment of cerebral aneurysms can be performed Device Description: either by surgical clipping or endovascular techniques (e.g., embolization coils). Endovascular methods to
embolize intracranial aneurysms with implantable platinum embolization coils were developed in the 1970s. Embolization coils including the MicroVention MCS and HES products are now widely used to treat all aneurysms.
The MCS consists of an implantable coil made of bare platinum alloy (Platinum/Tungsten), and the HES consists of an implantable coil made of the same platinum alloy with a hydrogel inner core. The coil is attached to a delivery pusher via a polyolefin elastomer material. The delivery pusher contains radiopaque positioning markers at the distal end. The proximal end is inserted into a hand held battery powered V-Grip™ Detachment Controller. When the Detachment Controller is activated, the flow of electrical current heats the polyolefin elastomer filament, resulting in detachment of the implant segment. The V-Grip is packaged and sold separately.
The modified MCS and HES devices are substantially equivalent to the cleared predicate devices with regard to intended use, principal of operation, materials, manufacturing processes, packaging configuration, and sterilization method.
{1}------------------------------------------------
#### MicroVention : ﺗ
#### Pre-Clinical Testing:
| Design Verification and Validation Bench Test Summary | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|-------------------------------------------------------------------|-------------------------------------|
| Test / Test Description | Acceptance Criteria | Test & Acceptance<br>Criteria same as<br>predicate device?<br>Y/N | Result |
| Visual Inspection<br>Perform visual inspection<br>and measurements using<br>device drawing | Per product drawing | Y | Pass<br>Met established<br>criteria |
| Simulated Use<br>The test simulates a<br>neurointerventional<br>embolization procedure to<br>assess deployment,<br>repositioning, and<br>detachment in an<br>aneurysm. | All performance ratings<br>shall be ≥ 3 | Y | Pass<br>Met established<br>criteria |
| Pusher Resistance<br>Using a digital multimeter,<br>measure the resistance<br>through a microcatheter. | 36.7-53.0 $Ω$ | Y | Pass<br>Met established<br>criteria |
| Detachment Zone Tensile<br>Using an Instron tester,<br>measure the breakforce at<br>the detachment zone. | ≥0.08 lbf | Y | Pass<br>Met established<br>criteria |
CONCLUSION: The results of the bench testing demonstrate that the subject device is safe and
effective when used according to the instructions for use and performs equivalen device. The testing was used in support of the risk analysis documentation for the subject device.
| Biocompatibility Test Summary - HES Coil Implant | | | |
|--------------------------------------------------|--------------|--------|--|
| Test | Requirement | Result | |
| MEM Elution | 10993-5 | Passed | |
| ISO Cell Culture Agar Overlay | 100 10993-2 | Passed | |
| Sensitization-Guinea Pig Maximization | ISO 10993-10 | Passed | |
| ISO Intracutaneous Reactivity Evaluation | ISO 10993-10 | Passed | |
| Hemolysis | ISO 10993-4 | Passed | |
| Prothrombin Time Assay - ISO | ISO 10993-4 | Passed | |
| Systemic toxicity (IV injection) | ISO 10993-11 | Passed | |
| Rabbit Pyrogen Test (material mediated) | ISO 10993-11 | Passed | |
| Bacteria Reverse Mutation Assay (Ames) | 180 10993-3 | Passed | |
| 7-day Muscle Implantation | 180 10993-6 | Passed | |
| 13-week Intramuscular Implantation Test | ISO 10993-6 | Passed | |
| 26-week Intramuscular Implantation Test | ISO 10993-6 | Passed | |
{2}------------------------------------------------
## AlleroVention
| Biocompatibility Test Summary - MCS Coil Implant | | | | |
|--------------------------------------------------|--------------|--------|--|--|
| Test | Requirement | Result | | |
| MEM Elution | ISO 10993-5 | Passed | | |
| ISO Cell Culture Agar Overlay | ISO 10993-5 | Passed | | |
| Sensitization-Guinea Pig Maximization | ISO 10993-10 | Passed | | |
| ISO Intracutaneous Reactivity Evaluation | ISO 10993-10 | Passed | | |
| Hemolysis | ISO 10993-4 | Passed | | |
| Prothrombin Time Assay - ISO | ISO 10993-4 | Passed | | |
| Systemic toxicity (IV injection) | ISO 10993-11 | Passed | | |
| Rabbit Pyrogen Test (material mediated) | ISO 10993-11 | Passed | | |
| Bacteria Reverse Mutation Assay (Ames) | SO 10993-3 | Passed | | |
| 7-day Muscle Implantation | ISO 10993-6 | Passed | | |
| 90-day Intramuscular Implantation Test | ISO 10993-6 | Passed | | |
| 100-day Intramuscular Implantation Test | ISO 10993-6 | Passed | | |
| Biocompatibility Test Summary - Delivery Pusher | | | | |
| Test | Requirement | Result | | |
| MEM Elution | ISO 10993-5 | Passed | | |
| ISO Cell Culture Agar Overlay | ISO 10993-5 | Passed | | |
| Sensitization-Guinea Pig Maximization | ISO 10993-10 | Passed | | |
| ISO Intracutaneous Reactivity Evaluation | ISO 10993-10 | Passed | | |
| Hemolysis | ISO 10993-4 | Passed | | |
| Prothrombin Time Assay - ISO | ISO 10993-4 | Passed | | |
| Systemic toxicity (IV injection) | ISO 10993-11 | Passed | | |
| Rabbit Pyrogen Test (material mediated) | ISO 10993-11 | Passed | | |
Predicate / Subject Technological Comparison:
| Feature | MCS Predicate Devices | HES Predicate Devices | Subject Devices |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-----------------|
| Indications for Use | Intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. | Same | Same |
| Device Overview | An implantable coil attached to a delivery pusher via a polyolefin elastomer material. The delivery pusher contains radiopaque | Same | Same |
{3}------------------------------------------------
#### Special 510(k) MicroPlex Embolic System (MCS) & HydroCoil Embolic System (HES) with V-Trak Advanced Delivery Pusher
#### MicroVention
| | | TERUMO | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-------------------------------------------|
| | positioning markers at the distal end. The<br>proximal end is inserted into a hand held<br>battery powered Detachment Controller.<br>When the Detachment Controller is<br>activated, the flow of electrical current<br>heats the polyolefin elastomer filament,<br>resulting in detachment of the implant<br>segment. The Detachment Controller is<br>packaged and sold separately. | | |
| Coil shape | Helical, 3D | Same | Same |
| Coil OD (mm) | 1-24 | 1-24 | Same |
| Restrained coil<br>length (cm) | 1-68 | 1-50 | Same |
| Main coil wire<br>material | Platinum/Tungsten (92/8%) alloy | Same | Same |
| Coupler Material | Platinum (90%)/ iridium (10%) | Same | Same |
| Adhesive Material | DYMAX 1128-AM-VT UV Adhesive | Same | Same |
| Stretch resistance<br>filar material | Polyolefin Elastomer or PET | PET | Same |
| Implant-to-pusher<br>material | Polyolefin elastomer | Same | Same |
| Delivery method | V-Trak delivery pusher | V-Trak delivery<br>pusher | V-Trak<br>Advanced<br>delivery<br>pusher1 |
| 1 The subject device is the same as the predicate devices with the exception of the modified V-<br>Trak Advanced deliver pusher. Other than the colorant in the PET connector tube, the V-Trak<br>Advanced delivery pusher is essentially the same as the cleared V-Trak delivery pusher. The MCS<br>and HES devices with the V-Trak Advanced delivery pusher are the same with regard to intended<br>use and principal of operation. Any differences in technological characteristics do not introduce<br>any new issues of safety or effectiveness. Therefore, it is our conclusion that the MCS and HES<br>devices with the V-Trak Advanced delivery pusher is substantially equivalent to the predicate<br>devices. | | | |
#### Summary of Substantial Equivalence:
The subject devices are the same as the predicate devices with regard to intended use and principal of operation. Any differences in technological characteristics do not introduce any new issues of safety or effectiveness. Therefore, it is our conclusion that the MCS and HES devices with the V-Trak Advanced delivery pusher is substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized eagle symbol, which is a common emblem associated with the U.S. government.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### December 19, 2013
MicroVention Terumo c/o Ms. Laraine Pangelina Sr. Regulatory Affairs Project Manager MicroVention, Incorporated 1311 Valencia Avenue Tustin, CA 92780
Re: K132952
Trade/Device Name: Microplex Coil System (MCS), Hydrocoil Embolic System (HES) Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: November 18, 2013 Received: November 21, 2013
Dear Ms. Pangelina:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
Page 2 - Ms. Laraine Pangelina
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Joyce M. Whang -S
for Carlos Peña, Ph.D.
Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K132952
Device Name: _ MicroPlex Coil System (MCS), HydroCoil Embolic System (HES)
Indications For Use:
ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ
Intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. Also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
Prescription Use _ كا (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Joyce M. Whang -S
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.