Hair Growth Device is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II, in males with androgenetic alopecia who have Norwood-Hamilton Classifications of IIa-V and for both, Fitzpatrick Classification of Skin Phototypes I to IV.
Device Story
Hair Growth Device is a portable, non-invasive, over-the-counter low-level laser therapy (LLLT) system for androgenetic alopecia. Device consists of a helmet containing 56 to 200 laser diodes (650 nm, 5 mW), a USB cable, and a silicone pad. Operation involves placing the helmet on the head; device emits laser light penetrating 3-5 mm into the scalp to stimulate hair follicles. Treatment is self-administered by the patient for 30 minutes every two days for 16 weeks. Device provides audible feedback upon activation. No communication or connectivity functions. Intended to promote hair growth by stimulating follicles via LLLT.
Clinical Evidence
No clinical data was used to support the determination of substantial equivalence. Evidence is based on bench testing, including electrical safety, electromagnetic compatibility, laser safety, and biocompatibility.
Technological Characteristics
LLLT device using 650 nm (±20 nm) diode lasers; 5 mW per lamp. 56-200 laser diodes. Class 3R laser product. Materials include silicone. Powered via USB. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60825-1, ISO 10993-5, and ISO 10993-10.
Indications for Use
Indicated for females with androgenetic alopecia (Ludwig-Savin I-II) and males with androgenetic alopecia (Norwood-Hamilton IIa-V), Fitzpatrick Skin Phototypes I-IV, to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 29, 2025
Dongguan Boyuan Intelligent Technology Co., Ltd.
% Reanny Wang
General Manager
Shenzhen Reanny Medical Devices Management Consulting Co.Ltd
Room 1509, Jingting Building, Dongzhou Community,
Guangming Street, Guangming District
Shenzhen, Guangdong 518107
China
Re: K250308
Trade/Device Name: Hair Growth Device (KCA441-200, KCA441-120, KCA441-88, KCA441-56, KCA441-200S, KCA441-120S, KCA441-88S, KCA441-56S, KCA451-200, KCA451-120, KCA451-88, KCA451-56, KCA451-200S, KCA451-120S, KCA451-88S, KCA451-56S)
Regulation Number: 21 CFR 890.5500
Regulation Name: Infrared Lamp
Regulatory Class: Class II
Product Code: OAP
Dated: January 27, 2025
Received: February 3, 2025
Dear Reanny Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250308 - Reanny Wang
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K250308 - Reanny Wang
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.04.29
15:17:28 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250308 | |
| Device Name Hair Growth Device (KCA441-200, KCA441-120, KCA441-88, KCA441-56, KCA441-200S, KCA441-120S, KCA441-88S, KCA441-56S, KCA451-200, KCA451-120, KCA451-88, KCA451-56, KCA451-200S, KCA451-120S, KCA451-88S, KCA451-56S) | |
| Indications for Use (Describe) Hair Growth Device is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II, in males with and rogenetic alopecia who have Norwood-Hamilton Classifications of IIa-V and for both, Fitzpatrick Classification of Skin Phototypes I to IV. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) number: K250308
1. **Information of Submitter and Correspondent**
**Submitter's information:**
Company Name: Dongguan Boyuan Intelligent Technology Co., Ltd.
Street Address: Room 801, Building 3, No.3 Kuiqiao Road, Puxin Lake, Tangxia Town, Dongguan, Guangdong, China.
City: Dongguan
State/ Province: Guangdong
Country: China
Telephone: +86-13422889315
Contact Person: Li Jinlong
Contact Title: General Manager
Contact Email: 978953150@qq.com, reanny@reanny.com
Date Prepared: Apr. 28, 2025
**Submission correspondent's information:**
Shenzhen Reanny Medical Devices Management Consulting Co., Ltd
Address: Room 1509, Jingting Building, Dongzhou Community, Guangming Street, Guangming District, Shenzhen 518107, China
Contact Person: Reanny Wang
E-mail: reanny@reanny.com
Phone: +86(755) 27391220
2. **Device Information**
Trade Name: Hair Growth Device
Model: KCA441-200, KCA441-120, KCA441-88, KCA441-56, KCA441-200S, KCA441-120S, KCA441-88S, KCA441-56S, KCA451-200, KCA451-120, KCA451-88, KCA451-56, KCA451-200S, KCA451-120S, KCA451-88S, KCA451-56S
Common Name: Laser, Comb, Hair
1 / 7
{5}
Classification Name: Infrared lamp
Regulation: 21 CFR 890.5500
Device Class: Class 2
Product Code: OAP
## 3. Identification of Predicate Device(s)
| Substantial equivalent devices | Sponsor | Trade Name | 510(k) number | Product Code | Regulation Number | Approval Date |
| --- | --- | --- | --- | --- | --- | --- |
| Predicate device | Slinph Technologies Co., Ltd. | iHelmet Hair Growth System | K162782 | OAP | 21 CFR 890.5500 | April 4, 2017 |
| Reference device | DermaScalp LLC | DermaScalp Laser Caps | K173846 | OAP | 21 CFR 890.5500 | March 21, 2018 |
| Reference device | PhotonMD, Inc | REVIAN RED | K173729 | OAP | 21 CFR 890.5500 | February 28, 2018 |
## 4. Description of Device
The product is composed of main unit, USB cable, and silicone pad. The main unit consists of laser diodes that are spread throughout the helmet.
Low-energy laser therapy (LLLT) is a safe and effective treatment for androgenic alopecia (AGA). The Hair Growth Device employs intelligent LLLT technology, utilizing a 650 nm, 5-milliwatt low-energy laser that penetrates the scalp surface to a depth of 3-5 mm, reaching the hair follicles.
Hair Growth Device is a small over-the-counter, portable, noninvasive, low-level laser device intended to treat Androgenetic Alopecia (Hair Loss) and to promote hair growth in Males and Females. The product uses diode lasers to cover the entire area of the head that is normally covered with hair, and this unique design allows the treatment of the entire scalp without manual movement. Hair Growth Device consists of main unit, USB cable and silicone strap and no communication function. The device will then turn on and you'll hear an audible beep.
## 5. Indications for Use
Hair Growth Device is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II, in males with androgenetic alopecia who have Norwood-Hamilton Classifications of IIa-V and for both, Fitzpatrick Classification of Skin Phototypes I to IV.
## 6. Discussion of Comparison to Predicate Devices.
The Hair Growth Device submitted in this 510(k) submission is compared in intended use, technological characteristics and performance to the cleared iHelmet Hair Growth System K162782, DermaScalp Laser Caps K173846 and REVIAN RED K173729.
2 / 7
{6}
| Device | Subject device | Primary Predicate device | Primary reference device | Primary reference device | Comparison |
| --- | --- | --- | --- | --- | --- |
| Manufacturer | Dongguan Boyuan Intelligent Technology Co., Ltd. | Slinph Technologies Co., Ltd. | DermaScalp LLC | PhotonMD, Inc | NA |
| 510(K) number | Pending | K162782 | K173846 | K173729 | NA |
| Product name | Hair Growth Device | iHelmet Hair Growth System | DermaScalp Laser Caps | REVIAN RED | NA |
| Classification | Class II Device, OAP (21 CFR890.5500) | Class II Device, OAP (21 CFR890.5500) | Class II Device, OAP (21 CFR890.5500) | Class II Device, OAP (21 CFR890.5500) | Same |
| Classification Name | Infrared Lamp | Infrared Lamp | Infrared Lamp | Infrared lamp | Same |
| Indications for Use (IFU) | Hair Growth Device is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II, in males with androgenetic alopecia who have Norwood-Hamilton Classifications of IIa-V and for both, Fitzpatrick Classification of Skin Phototypes I to IV. | iHelmet Hair Growth System (Model: LTD200S) is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications I - II, in males with androgenetic alopecia who have Norwood Hamilton Classifications IIa - V and for both, Fitzpatrick Classification of Skin Phototypes of I - IV. | The family of DermaScalp Laser Cap devices are indicated to treat Androgenetic Alopecia and promote hair growth in Males who have Norwood Hamilton Classifications of IIa to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in Females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV. | REVIAN RED is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV. | Same |
| Model | KCA441-200, KCA441-120, KCA441-88, KCA441-56, KCA441-200S, KCA441-120S,KCA441-88S, KCA441-56S, KCA451-200, KCA451-120, KCA451-88, KCA451-56, KCA451-200S, KCA451-120S, KCA451-88S, KCA451- | LTD200S | -- | --- | NA |
{7}
| Device | Subject device | | Primary Predicate device | Primary reference device | Primary reference device | Comparison |
| --- | --- | --- | --- | --- | --- | --- |
| | 56S | | | | | |
| OTC or Rx | OTC | | OTC | OTC | OTC | Same |
| Classification Product Code | OAP | | OAP | OAP | OAP | Same |
| Waveform | Visible red laser | | Visible red laser | 119 Red LEDs | 119 Red LEDs | Same |
| Wavelength | 650nm±20nm | | 650nm±10nm | 650 nm | 620 - 660 nm | Same |
| Amounts of Laser Lamp | 56, 88, 120, 200 | | 200 | 50, 80, 120, 148, 180, 202, 224, 272 | 51, 80 | Minor different Note 1 |
| Energy of per Laser Lamp | 5mW | | 4~5mW | < 5mW | -- | Same |
| Classification according to IEC60825-1 | Class 3R | | Class 3R | Class 3R | Class 3R | Same |
| Treatment Time | Each Treatment: 30 minutes. Total Treatment: once every 2 days, for 16 weeks | | Each Treatment: 20-35 min Total Treatment: every other day, for 16 weeks | Treatment - 17weeks, every other day (indefinite) | -- | Minor different Note 1 |
| Treatment Area | 390cm² Mathematically Max. derived | | 424.93 cm² Mathematically Max. derived | -- | -- | Minor different Note 1 |
| Irradiance (power per area) | 2.56mW/cm² Mathematically Max. derived | | 2.3533 mW/cm² Mathematically Max. derived | -- | 1.67 mW/cm² | Minor different Note 1 |
| Fluence | KCA441-200, KCA451-200, KCA441-200S, KCA451-200S | 4.62J/cm²; | 4.9420 J/ cm² Mathematically Max. derived | -- | -- | Minor different Note 1 |
| | KCA441-88, KCA451-88, KCA441-120S, KCA451-120S | 2.03J/cm² | | | | |
{8}
| Device | Subject device | | Primary Predicate device | Primary reference device | Primary reference device | Comparison |
| --- | --- | --- | --- | --- | --- | --- |
| | KCA441-120, KCA451-120: KCA441-88, SKCA451-88S | 2.77J/cm²; | | | | |
| | KCA441-56, KCA451-56, KCA441-56S, KCA451-56S | 1.29J/cm² | | | | |
| Dimension | KCA441size: 26cm x 21cm x 12.5cm
KCA451size: 26cm x 21cm x 13cm | | 266mm x 196mm x 135mm (L x W x H) | -- | Height: 15.87 cm (6.25 in) x Width: 20 cm (7.87 in) x Length: 22.86 cm (9 in) | Minor different Note 2 |
| Weight | KCA441 size:572g
KCA451 size: weight:567g | | 600g | -- | 155 g | Minor different Note 2 |
| Environment for Operation | Temperature: 5°C~40°C
Humidity: ≤90%RH
Atmosphere range: 80~110kPa | | Temperature: 15~30°C
Humidity: 30~80% RH
Atmosphere range: 90~110kPa | -- | 5-35°C (41-95°F) Humidity: Up to 90%, non-condensing
Atmospheric Pressure: 700 hPa to 1060 hPa | Minor different Note 3 |
| Environment for Storage | Temperature: -20°C~55°C
Humidity: ≤93%RH
Atmosphere range: 50~110kPa | | Temperature: -20~65°C
Humidity: 0~ 80% RH
Atmosphere range: 50~110kPa | -- | Temperature: 15-35°C (59-95°F)
Humidity: Up to 90%, non-condensing
Atmospheric Pressure: 700 hPa to 1060 hPa | Minor different Note 3 |
| Standards | IEC 60601-1
IEC 60601-1-2
IEC 60825-1
IEC 60601-1-11
ISO 10993-5
ISO 10993-10 | | IEC 60601-1
IEC 60601-1-2
IEC 60825-1
IEC 60601-1-11
ISO 10993-5
ISO 10993-10 | IEC 60601-1
IEC 60601-1-2
IEC 60825-1
IEC 60601-1-11
ISO 10993-5 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-57
ISO 10993-1
IEC 62366-1
IEC 60601-1-6 | Minor different Note 4 |
5 / 7
{9}
Note 1:
Although the "Amounts of Laser Lamp", "Treatment Time", "Treatment Area", "Irradiance", and "Fluence" of subject device and predicate device are a little difference, the energy and power parameters' range of subject device can be covered by predicate device's several models' range; they are very similar. So these parameters' differences will not raise any safety or effectiveness issue.
Note 2:
The Dimensions and Weigh of subject device, predicate devices and reference devices are difference, but they belong to portable device, only the size, weight and service life are difference, they are both complied with IEC 60601-1, so the differences do not affect the safety and effectiveness.
Note 3:
The "Operating environment" and "Transportation & Storage environment" of subject devices may be differ from the predicate devices and reference devices, but they are both compliance with the IEC 60601-1 and IEC 60601-1-11 standards, so the difference will not raise any safety or effectiveness issue.
Note 4:
They are both complied with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11 and ISO 10993-1. IEC 60601-2-57 is for non-laser light source equipment, so the difference will not raise any safety or effectiveness issue.
Conclusions:
The subject device Hair Growth Device (Model: KCA441-200, KCA441-120, KCA441-88, KCA441-56, KCA441-200S, KCA441-120S, KCA441-88S, KCA441-56S, KCA451-200, KCA451-120, KCA451-88, KCA451-56, KCA451-200S, KCA451-120S, KCA451-88S, KCA451-56S) has been compared with iHelmet Hair Growth System (K162782), DermaScalp Laser Caps (K173846) and REVIAN RED (K173729). The subject device has same intended use, similar technological characteristics as that of predicate devices and reference devices. Although there are several specifications that are different between the subject device, predicate devices and reference devices, the comparison analysis has been completed to demonstrate that the differences between these parameters would not adversely impact the safety and effectiveness of the subject device. The subject device has undergone safety and performance tests, and the results complied with the test requests. Therefore, the differences between the subject device, the predicate devices and reference devices do not raise any concerns with respect to substantial equivalence. The subject device is substantially equivalent to the predicate devices.
6 / 7
{10}
7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
7.1 Non-clinical testing
A series of safety and performance tests were conducted on the subject device.
- Product service life
- Software validation
- Biocompatibility evaluation
- Electromagnetic compatibility and electrical safety
- Safety of laser products
- Performance test
All the test results demonstrate Hair Growth Device meets the requirements of its predefined acceptance criteria and intended use, and it is substantially equivalent to the predicate devices.
7.2 Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
8. Conclusions
The results of the tests described above demonstrate that Hair Growth Device is as safe and effective as the predicate devices and supports a determination of substantial equivalence.
7 / 7
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.