K173729 · Photonmd, Inc. · OAP · Feb 28, 2018 · Physical Medicine
Device Facts
Record ID
K173729
Device Name
REVIAN RED
Applicant
Photonmd, Inc.
Product Code
OAP · Physical Medicine
Decision Date
Feb 28, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
REVIAN RED is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.
Device Story
REVIAN RED is a wearable, hands-free, textile cap for home-based, OTC treatment of androgenetic alopecia. The device uses 119 red LEDs to deliver Modulated Light Therapy (MLT) to the scalp. It is powered by a rechargeable lithium polymer battery and controlled via a wireless Bluetooth connection to a mobile application. The system includes driver electronics and a flexible printed circuit board. The mobile app provides treatment reminders, messages, and controls the 10-minute daily treatment session. The device includes safety features that pause therapy if the user's head moves outside the radiation zone and automatically powers down after the 10-minute session. The healthcare provider does not operate the device; it is for patient self-use. The device aims to promote hair growth through light therapy.
Clinical Evidence
No clinical trial data was submitted. Substantial equivalence was demonstrated through non-clinical performance testing, including electrical safety (IEC 60601-1, IEC 60601-1-2, ANSI/AAMI HA60601-1-11:2015), photobiological safety (IEC 62471), and biocompatibility (ISO 10993-1). Human factors evaluation was conducted per ANSI/AAMI/IEC 62366-1 and 60601-1-6.
Technological Characteristics
Wearable textile cap with 119 red LEDs (620-660 nm). Powered by rechargeable lithium polymer battery. Connectivity via Bluetooth to mobile application. Safety features include automatic power-down and radiation zone monitoring. Biocompatible materials per ISO 10993-1. Conforms to IEC 60601-1, IEC 60601-1-2, ANSI/AAMI HA60601-1-11:2015, IEC 60601-2-57, and IEC 62471.
Indications for Use
Indicated for males (Norwood-Hamilton IIa-V) and females (Ludwig-Savin I-1 to I-4, II-1, II-2 or frontal) with Androgenetic Alopecia and Fitzpatrick Skin Types I-IV to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 28, 2018
PhotonMD. Inc Mr. William Knape Senior Director, Regulatory Quality and Clinical Affairs 627 Davis Drive, Suite 400 Morrisville, North Carolina 27560
Re: K173729
Trade/Device Name: Revian Red Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP Dated: December 4, 2017 Received: December 6, 2017
Dear Mr. Knape:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
K173729
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173729
Device Name REVIAN RED
#### Indications for Use (Describe)
REVIAN RED is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|----------------------------------------------|--------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div>☑</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
PhotonMD
#### 510(K) Summary 1
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 890.5500.
#### 1.1 Submitter Information:
| Sponsor: | PhotonMD, Inc.<br>627 Davis Drive, Suite 400<br>Morrisville, NC 27560 |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | William A. Knape<br>Senior Director of Regulatory, Quality and Clinical Affairs<br>Phone: (919) 757-2033<br>Email: bknape@photonmd.com |
| Submission Date: | December 4, 2017 |
# 1.2 Subject Device Information
| Trade/Proprietary Name: | REVIAN RED (Model 101) |
|-------------------------|-----------------------------------|
| Common/Usual Name: | Lamp for Promotion of Hair Growth |
| Regulation Name: | Infrared Lamp per 21 CFR 890.5500 |
| Review Panel: | General & Plastic Surgery |
| Product Code: | OAP (laser, comb, hair) |
| Regulation Number: | 21 CFR 890.5500 |
| Regulatory Class: | Class 2 |
#### 1.3 Predicate Device Information
| Predicate Devices: | Device | 510(k)<br>Number | Manufacturer |
|--------------------|----------------------------|------------------|-------------------------|
| | iHelmet Hair Growth System | K162782 | Slinph Technologies |
| | HairMD | K152019 | Trophy Skin |
| | HairMax LaserComb 82 | K142573 | Lexington International |
| | LH80 PRO | K122950 | Theradome |
# 1.4 Device Description
REVIAN RED is a non-invasive, hands-free device indicated to treat androgenetic alopecia in both men and women. It is operated via a mobile application to promote hair growth using Modulated Light Therapy (MLT) for a recommended 10-minute daily treatment. REVIAN RED is a system comprised of a wearable soft textile Cap using driver electronics, a rechargeable battery, and integrated light emitting diode (LED) flexible printed circuit board. REVIAN RED is designed as an over-the-counter (OTC) home-based device to treat
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for PhotonMD. The word "Photon" is in gray, with a colorful atom symbol in the middle of the word. The letters "MD" are in green.
#### February 26, 2018 Traditional 510(k) Premarket Notification
the entire scalp area of the user and is controlled via a wireless Bluetooth connection to a mobile application operating on a wireless device.
# 1.5 Indications for Use
The REVIAN RED device is indicated to treat Androgenetic Alopecia and to Promote Hair Growth in males who have Norwood-Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and Promote Hair Growth in females who have Ludwig-Savin Scale I-1 to I-4, II -1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.
# 1.6 Performance Test Summary
No clinical trial data for the REVIAN RED System was submitted for this 510(k). Nonclinical performance testing to demonstrate substantial equivalence to the predicate device included evaluation to IEC 60601-1 and 60601-1-2 to confirm the device's electrical safety and electromagnetic compatibility and conformance with ANSI/AAMI HA60601-1-11:2015, IEC 60601-2-57 (Edition 1.0), and IEC 62471 (Edition 1.0). Biocompatibility testing of the Cap per ISO 10993-1 confirmed no effects of the applied parts were noted for cytotoxicity, skin irritation or skin sensitization. Performance verification and validation testing demonstrates REVIAN RED meets user needs and design inputs and met requirements for its intended use.
#### Studies Related to Human Factors Evaluation
Human factors evaluation of REVIAN RED was tested following ANSVAAMI/IEC 62366-1 and 60601-1-6 standards. The study protocol was designed to evaluate representative users of the device under expected use conditions during simulated and actual use of the device. The Human Factors Engineering and Usability Engineering process involved several methodologies, including observational user research, formative and summative user testing.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for PhotonMD. The word "Photon" is in gray, with a colorful graphic in the middle of the word. The letters "MD" are in green. The logo is simple and modern.
#### February 26, 2018 Traditional 510(k) Premarket Notification
### 1.7 Comparison to Predicate Devices
REVIAN RED has similar technological characteristics as the iHelmet Hair Growth System (K162782) including wavelength, light energy delivery method, mobile app interface with interactive audible timers and intended use. REVIAN RED is also similar technologically to the other helmet designed devices; with HairMD (K152019) and LH80 PRO (K122959) having fewer diodes (51 laser diodes/LEDS and 80 laser diodes, respectively) and longer treatment regimens of 20 minutes.
| 510(k) # | K173729 | K162782 | K152019 | K142573 | K122950 |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | REVIAN RED | iHelmet Hair<br>Growth System<br>(LTD200S) | HairMD | HairMax<br>LaserComb 82<br>(HMLC 82) | LH80 PRO |
| Company | PhotonMD, Inc. | Slinph<br>Technologies | Trophy Skin | Lexington<br>International | Theradome |
| Product Code | OAP | OAP | OAP | OAP | OAP |
| Rx/OTC | OTC | OTC | OTC | OTC | OTC |
| Intended Use/<br>Indications for<br>Use | Treat<br>Androgenetic<br>Alopecia and<br>promote hair<br>growth in males<br>who have<br>Norwood-<br>Hamilton<br>Classification IIa<br>to V patterns of<br>hair loss and to<br>treat Androgenetic<br>Alopecia and<br>promote hair<br>growth in females<br>who have<br>Ludwig-Savin<br>scale I-1 to I-4,<br>II-1, II-2 or frontal<br>patterns of hair<br>loss; both with<br>Fitzpatrick Skin<br>Types I to IV | Promote hair<br>growth in females<br>with<br>Androgenetic<br>Alopecia who<br>have Ludwig-<br>Savin<br>Classification of<br>I-II, in males with<br>Androgenetic<br>Alopecia who<br>have Norwood<br>Hamilton<br>Classifications of<br>IIa-V and for<br>both, Fitzpatrick<br>Classification of<br>Skin Phototypes I<br>- IV. | Promote hair<br>growth in females<br>with<br>Androgenetic<br>Alopecia who<br>have Ludwig-<br>Savin<br>Classification of<br>I-II, in males with<br>Androgenetic<br>Alopecia who<br>have Norwood<br>Hamilton<br>Classifications of<br>IIa-V and for<br>both, Fitzpatrick<br>Classification of<br>Skin Phototypes I<br>- IV. | Treat<br>Androgenetic<br>Alopecia and<br>promote hair<br>growth in males<br>who have<br>Norwood<br>Hamilton<br>Classification<br>IIa to V patterns<br>of hair loss and<br>treat<br>Androgenetic<br>Alopecia and,<br>promote hair<br>growth in females<br>who have Ludwig<br>(Savin) scale I-4,<br>II-1, II-2 or<br>frontal, both with<br>Fitzpatrick skin<br>types I to IV. | Treat<br>Androgenetic<br>Alopecia, to<br>promote hair<br>growth in<br>females with<br>female pattern<br>hair loss (FPHL)<br>on the Ludwig<br>and Savin Hair<br>Loss Scale I-II,<br>Fitzpatrick Skin-<br>Types I to IV. |
| Wearable<br>Mounting | Textile Cap | Helmet | Helmet | Head Band | Helmet |
| Wave-Length(s) | 620 - 660 nm | 650 nm ±10nm | 650 nm | 655 nm ±10nm | 678 nm |
| Visible Light<br>source | 119 Red<br>LEDs | 200 Red Laser<br>Diodes | 21 Red Laser<br>Diodes and<br>30 Red LEDs | 82 Red Laser<br>Diodes | 80 Red Laser<br>Diodes |
| Battery | Lithium<br>Polymer | Lithium<br>Polymer | Powered by 120V<br>Wall Plug | Lithium<br>Polymer | Lithium<br>Polymer |
#### Table 1 – Summary of Predicate Devices Compared to REVIAN RED
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for PhotonMD. The word "Photon" is in gray, with a graphic of colorful people surrounding a white sphere in the middle of the word. The letters "MD" are in green. The logo is simple and modern.
#### February 26, 2018 Traditional 510(k) Premarket Notification
#### Table 1 - Summary of Predicate Devices Compared to REVIAN RED (cont'd)
| 510(k) # | K173729 | K162782 | K152019 | K142573 | K122950 |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Treatment Time | Every day<br>10 mins | Every other day<br>20-35 mins | Every other day<br>20 mins | 3 times / week<br>90 seconds | 2 times / week<br>20 mins |
| Digital Mobile<br>Controller | Yes | Yes | No | No | No |
| Restrictions on<br>Daily Irradiance | Yes, daily limit<br>for treatments | No | No | No | No |
| Other Design<br>Characteristics | Cap pauses<br>therapy if subject's<br>head moves<br>outside zone of<br>radiation; after<br>completion of<br>treatment, unit<br>powers down,<br>limits daily<br>treatment to 10<br>mins, and<br>provides treatment<br>reminders and<br>messages via<br>mobile app<br>controller | Helmet pauses<br>therapy if<br>subject's head<br>moves outside<br>zone of radiation;<br>after completion<br>of time, unit<br>powers down and<br>audible beep. | Helmet pauses<br>therapy if<br>subject's head<br>moves outside<br>zone of radiation;<br>after completion<br>of time, unit<br>powers down and<br>audible beep. | Vibrate so user<br>can move to the<br>next section, 3<br>sections @ 30<br>seconds each,<br>proximity sensor<br>to shut off lasers<br>if not in contact,<br>7.5 cm scalp<br>section, automatic<br>turn off | Helmet pauses<br>therapy if<br>subject's head<br>moves outside<br>zone of<br>radiation; after<br>completion of<br>time, unit<br>powers down<br>and audible<br>beep. |
# 1.8 Conclusion
This notification contains all information required by 21 CFR 807.87. Performance testing including electrical, electromagnetic compatibility, biocompatibility, software, and human factors demonstrates REVIAN RED meets the requirements for its intended use and does not raise any new types of safety or effectiveness questions when compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.