Storz Medical Duolith SD1 T-Top & Tower System with C-Actor Sepia Handpiece
K202112 · Storz Medical AG · PZL · Feb 21, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202112
Device Name
Storz Medical Duolith SD1 T-Top & Tower System with C-Actor Sepia Handpiece
Applicant
Storz Medical AG
Product Code
PZL · General, Plastic Surgery
Decision Date
Feb 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4685
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The DUOLITH SD1 C-ACTOR Handpiece is indicated to provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm, which extend through the exidermis, dermis, tendon, or capsule, but without bone exposure. The DUOLITH SD1 with C-ACTOR is indicated for adult (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care.
Device Story
Device uses electromagnetic shock wave generation for chronic wound treatment. Control unit discharges high-voltage capacitor into handpiece coil; transient magnetic field induces eddy currents in metallic membrane; membrane repulsion creates pressure wave in water; parabolic reflector focuses wave onto target tissue. Operated by clinicians in clinical settings. System includes microprocessor control of parameters (intensity, frequency, pulse count). Output is focused acoustic pressure shockwaves. Clinicians use device to promote wound healing in conjunction with standard care. Benefits include reduction of wound area and support for complete healing in chronic diabetic ulcers.
Clinical Evidence
Clinical effectiveness supported by literature review of 89 chronic soft tissue wounds (including diabetes mellitus). Five independent studies reported 54-70% complete wound healing at 13-20 weeks and 35-80% reduction in wound area at 7-20 weeks. Nonclinical testing confirmed design specifications, biocompatibility (ISO 10993-1), and electrical safety (IEC 60601 series).
Technological Characteristics
Electromagnetic shock wave generation; high-voltage capacitor/coil system; parabolic reflector. Dimensions: 454 x 187 x 460 mm. Energy flux density: 0.03 - 1.24 mJ/mm2. Pulse repeat rate: 1-8 Hz. Microprocessor-controlled. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601-1, IEC 60601-2-36. EMC per IEC 60601-1-2 and AIM 7351731.
Indications for Use
Indicated for adult patients (22+ years) with chronic, full-thickness diabetic foot ulcers (>30 days duration) measuring ≤16 cm2, extending through epidermis, dermis, tendon, or capsule, without bone exposure. Used with standard diabetic ulcer care.
Regulatory Classification
Identification
An extracorporeal shock wave device for treatment of chronic wounds is a prescription device that focuses acoustic shock waves onto the dermal tissue. The shock waves are generated inside the device and transferred to the body using an acoustic interface.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must be conducted to demonstrate that the system produces anticipated and reproducible acoustic pressure shock waves.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance data must demonstrate that the reusable components of the device can be reprocessed for subsequent use.
(4) Performance data must be provided to demonstrate the electromagnetic compatibility and electrical safety of the device.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Performance data must support the use life of the system by demonstrating continued system functionality over the labeled use life.
(7) Physician labeling must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) A detailed summary of the device's technical parameters;
(iii) Validated methods and instructions for reprocessing of any reusable components; and
(iv) Instructions for preventing hearing loss by use of hearing protection.
(8) Patient labeling must include:
(i) Relevant contraindications, warnings, precautions, adverse effects, and complications;
(ii) Information on how the device operates and the typical course of treatment;
(iii) The probable risks and benefits associated with the use of the device;
(iv) Post-procedure care instructions; and
(v) Alternative treatments.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
February 21, 2021
Storz Medical AG % Michael Dayton Principal Consultant Biomed Research. Inc. 3959 Van Dyke Road, Suite 245 Lutz, Florida 33558
Re: K202112
Trade/Device Name: Storz Medical DUOLITH SD1 T-Top & Tower System with C-ACTOR Sepia Handpiece Regulation Number: 21 CFR 878.4685 Regulation Name: Extracorporeal Shock Wave Device for Treatment of Chronic Wounds Regulatory Class: Class II Product Code: PZL Dated: January 20, 2021 Received: January 22, 2021
Dear Michael Dayton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K202112
Device Name
DUOLITH SD1 T-Top & Tower System with C-ACTOR Sepia Handpiece
Indications for Use (Describe)
The DUOLITH SD1 T-Top & Tower System with C-ACTOR Sepia Handpiece is indicated to provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm2, which extend through the epidermis, dermis, tendon, or capsule, but without bone exposure. The DUOLITH SD1 with C-ACTOR is indicated for adult (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 14px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 14px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### 1. SPONSOR
Storz Medical AG Lohstampfestr. 8 CH-8274 Tägerwilen Switzerland
| Contact Person: | Pavel Novak, Ph.D. |
|-----------------|---------------------|
| Telephone: | 41 (0) 71 677 45 13 |
| FAX: | 41 (0) 71 677 45 05 |
Date Prepared: February 19, 2021
#### 2. DEVICE NAME
| Proprietary Name: | DUOLITH SD1 T-Top & Tower System with C-ACTOR Sepia Handpiece |
|-----------------------|------------------------------------------------------------------|
| Regulation Name: | Extracorporeal Shock Wave Device for Treatment of Chronic Wounds |
| Classification Panel: | General and Plastic Surgery Devices |
| Regulatory Number: | 21 CFR 878.4685 |
| Product Code: | PZL |
| Device Class: | II |
#### 3. PREDICATE DEVICES
Equivalence is claimed to the predicate devices: DERMAPACE device (DEN160037; primary) and the ORTHOGOLD device (K191961).
#### 4. DEVICE DESCRIPTION
The DUOLITH SD1 with the C-ACTOR Sepia Handpiece uses an electromagnetically generated shock wave produced within the hand-held applicator. The shock wave is generated by discharging a high voltage capacitor located in the Control Unit into a cylindrically shaped coil system in the C-ACTOR Handpiece which is surrounded by a cylindrical metallic membrane. The transient magnetic field produced by the coil induces eddy currents in the metal membrane, causing it to repel from the coil, producing a pressure wave. The membrane is immersed in water and the pressure wave produced by the membrane propagates through the water to a concentric parabolic reflector, where it is reflected to a focal point outside of the Handpiece in front of the reflector. The DUOLITH SD1 with C-ACTOR incorporates micro-processor control of the operating parameters.
{4}------------------------------------------------
#### ડાં INTENDED USE
The DUOLITH SD1 C-ACTOR Handpiece is indicated to provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm, which extend through the exidermis, dermis, tendon, or capsule, but without bone exposure. The DUOLITH SD1 with C-ACTOR is indicated for adult (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care.
#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The technological characteristics of the DUOLITH SD1 and the predicate devices are the same with the exception of acoustic wave generation. The predicate devices utilize a high voltage spark plug to initiate a shock wave, while the Duolith utilizes a high voltage electrical coil. Both methods generate a short, high amplitude pulse (shock wave) of about 5 micro-seconds duration and results in similar waveforms and output energies.
#### 7. PERFORMANCE TESTING
## Nonclinical Performance
Nonclinical verification and validation testing was performed and demonstrated that the DUOLITH SD1 with C-ACTOR meets the design specifications and is safe and effective for its intended use. All tests required by the verification and validation plan were completed and passed. The testing demonstrates that the DUOLITH SD1 with C-ACTOR is substantially equivalent to the predicate devices.
The DUOLITH SD1 software was validated and demonstrated to be of a Moderate level of concern; while hazard analysis / risk management was performed and demonstrated that all risks are mitigated to an acceptable level. There are no direct body contacting components since the treatment handpiece is separated from direct patient contact by a sterile barrier during treatment. However the treatment head component was nevertheless tested for biocompatibility (contact duration <24 hours) and found to conform to elements of ISO 10993-1:2009. The DUOLITH SD1 was tested and demonstrated to conform to the general safety requirements of IEC 60601-1:2012; IEC 60601-6:2013; and electromagnetic compatibility requirements of IEC 60601-1-2:2014; and the RFID electromagnetic immunity requirements of AMI 7351731.
No performance standards applicable to this device have been adopted under Section 514 of the Act. The DUOLITH SD1 System with C-ACTOR complies with the applicable requirements of the following international consensus standards:
- · ISO 14971: 2007: Medical devices: Application of risk management to medical devices.
{5}------------------------------------------------
- · ISO 10993-1:2009 (4th Ed.): Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process.
- · ISO 61846:1998 Ultrasonics Pressure pulse lithotripters Characteristics of fields.
- · IEC 60601-1:2012, (3rd Ed.): Medical electrical equipment Part 1: General requirements for basic safety and essential performance.
- · IEC 60601-2-36: 2014 (2nd Ed.): Medical electrical equipment Part 2-36: Particular requirements for the basic safety and essential performance of equipment for extracorporeally induced lithotripsy.
- · IEC 60601-1-2: 2014 (4th Ed.): Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
- · IEC 60601-1-6: 2013 (3.1 Ed.): Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- · AIM 7351731:2017 (Rev. 2.0): Medical Electrical Equipment and System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers.
- · IEC 62304:2015 (1.1 Ed.) Medical Device Software Software Life Cycle Process.
- · ISO 13485:2016: Medical devices Quality management systems Requirements for regulatory purposes.
## Clinical Performance
A review of published literature with the subject device demonstrated successful clinical use with the Duolith SD1 technology in both men and women with 89 chronic soft tissue wounds having a range of etiologies including diabetes mellitus. The reports from five independent studies show significant results in complete wound healing (54-70% at 13-20 weeks) and reduction of wound area (35-80% at 7-20 weeks). The data from these studies establish the effectiveness of the Duolith SD1 to treat chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm, which extend through the epidermis, dermis, tendon, or capsule, but without bone exposure.
The conclusions drawn from the nonclinical and clinical tests demonstrate that the Duolith SD1 is as safe, as effective, and performs as well as the legally marketed predicate devices.
#### 8. SUBSTANTIAL EQUIVALENCE COMPARISON
| Product Comparison | Predicate Device<br>dermaPACE | Predicate Device<br>OrthoGold 100 | Subject Device<br>DUOLITH SD1 w/<br>C-ACTOR |
|------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| 510(k) Number | DEN160037 | K191961 | K202112 |
| Product Class | II | - Same - | - Same - |
| Classification Panel | General and Plastic Surgery Devices | - Same - | - Same - |
| Product Code and<br>Regulation | 21 CFR 878.4685<br>PZL - Extracorporeal Shock Wave Device for<br>Treatment of Chronic Wounds | Same | - Same - |
| Indications for Use | To provide acoustic pressure shockwaves in the<br>treatment of chronic, full-thickness diabetic foot<br>ulcers with wound areas measuring no larger<br>than 16 cm2, which extend through the<br> | - Same - | - Same - |
| Modes of action | Focused pressure pulses | Unfocused pressure<br>pulses | Focused pressure<br>pulses |
| Mechanism of action | Extracorporeally induced pressure pulses | - Same - | - Same - |
| Operating mode | Continuous | - Same - | - Same - |
| Electrical Protection | Class I, B | - Same - | - Same - |
| User Interface | LCD with function keys | Touch Screen | Touch Screen |
| Firmware Controlled | Yes | Yes | Yes |
| Type of Energy | Pressure pulses | - Same - | - Same - |
| Energy Source | 100 - 240 V AC, 50-60 Hz | - Same - | - Same - |
| Type of Acoustic<br>Wave Generation | Electrohydraulic, pressure wave under water<br>caused by discharge of high voltage<br>condensers | Electrohydraulic,<br>pressure wave under<br>water caused by<br>discharge of high<br>voltage condensers | Electromagnetic,<br>pressure wave under<br>water caused by<br>discharge of high<br>voltage condensers |
| Selection of parameters<br>(Intensity, Frequency,<br>Number of Pulses) | Yes | Yes | Yes |
| Size of treatment<br>applicator heads | ø70 x 160 mm | ø70 x 230 mm | ø62 x 179 mm |
| Intensity Settings<br>Range (Min-max:<br>Energy Flux Density) | 1 to 6<br>(0.23 - 0.33mJ/mm2) | 1 to 16<br>(0.01 - 0.19mJ/mm2) | 0.03 - 1.24mJ/mm2 |
| Focus depth | 1.5 - 10.5mm | 0 - 38mm<br>"unfocused" | 0 - 30mm |
| Pulse repeat rate (1/s) | 1 - 4 Hz | 1 - 8 Hz | 1 - 8 Hz |
| Number of pulses<br>(min and max) | 31.25 - 500 | 100 - 1000 | 100 - 500 |
| System Dimensions<br>(W x H x D) | 473 x 219 x 427 mm | 400 x 218 x 459 mm | 454 x 187 x 460 mm |
| Operating<br>Temperature | N / A | 10° - 30°C | 10° - 30°C |
| Operating Relative<br>Humidity | N / A | 3 - 85% | 5 - 55% |
| Treatment Duration<br>(Typical) | 5 — 10 min | 10 - 20 min | 10 - 20 min |
{6}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.