K243279 · E.M.S Electro Medical Systems S.A · PZL · Aug 13, 2025 · General, Plastic Surgery
Device Facts
Record ID
K243279
Device Name
DOLORCLAST Focused Shock Waves
Applicant
E.M.S Electro Medical Systems S.A
Product Code
PZL · General, Plastic Surgery
Decision Date
Aug 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4685
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DOLORCLAST Focused Shock Waves System is intended to provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm2, which extend through the epidermis, dermis, tendon, or capsule, but without bone exposure. The DOLORCLAST Focused Shock Waves system is indicated for adult (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care.
Device Story
System generates focused acoustic pressure shockwaves for treating chronic diabetic foot ulcers. Input: high-voltage electrical impulses from console; Transformation: piezoelectric elements on concave handpiece surface expand to generate focused shockwaves; Output: acoustic pressure pulses delivered to patient skin via interchangeable gel pads. Used in clinical settings by healthcare providers. Provider selects intensity, frequency, and pulse count via touch screen. Output affects wound healing; benefits patient by providing non-invasive treatment for chronic ulcers. System includes console, handpiece, gel pads, and coupling gel.
Clinical Evidence
Bench testing only. Acoustic pulse characterization performed per IEC 61846 and IEC 63045. System functional performance, lifetime testing (7 years console, 5M pulses handpiece), software validation (IEC 62304), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993-5, -10, -11, -23) were conducted to demonstrate safety and effectiveness.
Technological Characteristics
Piezoelectric shock wave generation; high-voltage impulse control. Console-based system with handpiece and interchangeable gel pads. Electrical: 100–127/220–240 VAC, 50-60Hz. Biocompatible materials per ISO 10993. Connectivity: Touch screen interface. Software compliant with IEC 62304. Standards: IEC 60601-1, IEC 60601-1-2, IEC 61846, IEC 63045.
Indications for Use
Indicated for adult (22+ years) diabetic patients with chronic, full-thickness diabetic foot ulcers >30 days duration, wound area ≤16 cm2, extending through epidermis, dermis, tendon, or capsule, without bone exposure. Used with standard diabetic ulcer care.
Regulatory Classification
Identification
An extracorporeal shock wave device for treatment of chronic wounds is a prescription device that focuses acoustic shock waves onto the dermal tissue. The shock waves are generated inside the device and transferred to the body using an acoustic interface.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must be conducted to demonstrate that the system produces anticipated and reproducible acoustic pressure shock waves.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance data must demonstrate that the reusable components of the device can be reprocessed for subsequent use.
(4) Performance data must be provided to demonstrate the electromagnetic compatibility and electrical safety of the device.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Performance data must support the use life of the system by demonstrating continued system functionality over the labeled use life.
(7) Physician labeling must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) A detailed summary of the device's technical parameters;
(iii) Validated methods and instructions for reprocessing of any reusable components; and
(iv) Instructions for preventing hearing loss by use of hearing protection.
(8) Patient labeling must include:
(i) Relevant contraindications, warnings, precautions, adverse effects, and complications;
(ii) Information on how the device operates and the typical course of treatment;
(iii) The probable risks and benefits associated with the use of the device;
(iv) Post-procedure care instructions; and
(v) Alternative treatments.
Predicate Devices
DUOLITH SD1 T-Top & Tower System with C-ACTOR Sepia Handpiece (K202112)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 13, 2025
E.M.S. Electro Medical Systems S.A.
% Sheila Hemeon-Heyer
President
Heyer Regulatory Solutions
125 Cherry Lane
Amherst, Massachusetts 01002
Re: K243279
Trade/Device Name: DOLORCLAST Focused Shock Waves
Regulation Number: 21 CFR 878.4685
Regulation Name: Extracorporeal Shock Wave Device For Treatment Of Chronic Wounds
Regulatory Class: Class II
Product Code: PZL
Dated: July 22, 2025
Received: July 22, 2025
Dear Sheila Hemeon-Heyer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243279 - Sheila Hemeon-Heyer
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243279 - Sheila Hemeon-Heyer
Page 3
Sincerely,
JAMES H. JANG -S
Digitally signed by
JAMES H. JANG -S
Date: 2025.08.13
09:29:05 -04'00'
James Jang, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243279 | |
| Device Name | |
| DOLORCLAST Focused Shock Waves | |
| Indications for Use (Describe) | |
| The DOLORCLAST Focused Shock Waves System is intended to provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm2, which extend through the epidermis, dermis, tendon, or capsule, but without bone exposure. The DOLORCLAST Focused Shock Waves system is indicated for adult (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
K243279
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
A. Submitter: E.M.S. Electro Medical Systems S.A.
Ch. de la Vuarpillière 31
1260 Nyon, Switzerland
Contact: Vivian Yee
Phone: +41 22 99 44 700
Email: vyee@ems-ch.com
B. Date Prepared: August 12, 2025
C. Device Name and Classification Information:
| Trade name: | DOLORCLAST Focused Shock Waves |
| --- | --- |
| Regulation Name | Extracorporeal shock wave device for treatment of chronic wounds. |
| Review Panel | General & Plastic Surgery |
| Classification regulation: | 21 CFR 878.4685 |
| Product code: | PZL |
| Class: | Class II |
D. Predicate Device(s):
Predicate Device: K202112, DUOLITH SD1 T-Top & Tower System with C-ACTOR Sepia Handpiece by Storz Medical AG
Reference: K191961, OrthoGold 100™ by Tissue Regeneration Technologies, LLC
E. Device Description:
The DOLORCLAST Focused Shock Waves system is composed of the DolorClast console, a handpiece, interchangeable gel pads (3 different sizes) and optional cart. EMS DolorClast Gel, previously approved under P050004 and K220538, is also provided for transmitting shock waves between the piezoceramic elements to the gel pad and coupling the treatment gel pad to the patient's skin.
The DOLORCLAST Focused Shock Waves system generates acoustic pressure waves generated using piezoelectric technology. A high-voltage impulse, controlled by the console, is used to stimulate the piezoceramic elements arranged across a concave surface in the emission source (the handpiece), causing the elements to expand simultaneously and briefly by a few micrometers, thus generating a focused shock wave. The shock waves are delivered to the patient's body via the handpiece, through a gel pad in a focused manner, and the treatment depth varies in line with the gel pad used.
Page 1 of 5
{5}
510(k) Summary
K243279
# F. Indications for Use Statement
The DOLORCLAST Focused Shock Waves System is intended to provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm2, which extend through the epidermis, dermis, tendon, or capsule, but without bone exposure. The DOLORCLAST Focused Shock Waves system is indicated for adult (22 years and older) diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care.
# G. Comparison with Predicate Devices
| Parameter | Proposed Device | Primary Predicate Device (K202112) |
| --- | --- | --- |
| Device Trade Name | DOLORCLAST Focused Shock Waves | DUOLITH SD1 w/ C-ACTOR |
| Device Manufacturer | E.M.S. Electro Medical Systems S.A. | Storz Medical AG |
| Regulation | 21 CFR 878.4685 | 21 CFR 878.4685 |
| Product Code | PZL | PZL |
| Device Class | II | II |
| Indications for Use Statement | To provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm2, which extend through the epidermis, dermis, tendon, or capsule, but without bone exposure. Indicated for adult (22 years and older) diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care. | To provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm2, which extend through the epidermis, dermis, tendon, or capsule, but without bone exposure. Indicated for adult (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care. |
| Modes of action | Focused pressure pulses | Focused pressure pulses |
| Mechanism of action | Extracorporeally induced pressure pulses | Extracorporeally induced pressure pulses |
| Operating mode | Continuous | Continuous |
| Electrical protection | Class I, B | Class I, B |
| User Interface | Touch Screen | Touch Screen |
| Firmware controlled | Yes | Yes |
| Type of Energy | Pressure pulses | Pressure pulses |
| Energy Source | 100–127/220–240 VAC, 50-60Hz | 100 – 240 V AC, 50–60 Hz |
| Type of Pressure/Shock Wave Generation | Piezoelectric, pressure wave caused by discharge of high voltage condensers | Electromagnetic, pressure wave under water caused by discharge of high voltage condensers |
| Operator selection of parameters (Intensity, Frequency,# of Pulses) | Yes | Yes |
{6}
510(k) Summary
K243279
| Parameter | Proposed Device | Primary Predicate Device (K202112) |
| --- | --- | --- |
| Intensity Settings and corresponding ranges of Energy Flux Density (Total EFD) | 0.1-20 (0.03–0.97 mJ/mm²) | No discrete settings or step increments stated (0.03 - 1.24 mJ/mm²) |
| Diameter of treatment applicator head | 99 mm | 62 mm |
| Number of applicator head offsets | Two | Two |
| Depth of applicator head offsets* | 15mm: Gel Pad M (A)
30mm: Gel Pad S (B) | 15 mm: Short Standoff (A)
30 mm: Long Standoff (B) |
| Therapeutically effective penetration depth (5 MPa) at maximum intensity setting:
- No offset
- Offset A
- Offset B | - 65 mm
- 50 mm
- 35 mm | - 65 mm
- 45 mm
- 30 mm |
| Depth of focus:
- No offset
- Offset A
- Offset B | - 40 mm
- 25 mm
- 10 mm | - 30 mm
- 15 mm
- 0 mm |
| Depth of focal zone
- No offset
- Offset A
- Offset B | - 30 - 50 mm
- 15 - 35 mm
- 0 - 20 mm | - 20 - 40 mm
- 0 - 20 mm
- 0 - 10 mm |
| Pulse repeat rate (l/s) (min – max) | 1 - 8 Hz | 1 - 8 Hz |
| Number of pulses per treatment (min - max) | 100-4,000 | 100-4,000** |
| System Dimensions (W x H x D) | 340x 180 x 450 (console) | 454 x 187 x 460 mm |
| Operating temperature | 10 °C - 35 °C | 10°C - 30°C |
| Operating Relative Humidity | Maximum 85% | 5-55% |
| Treatment Duration (Typical) | 5-15 min | 10-20 min |
*The offsets are called Gel Pads in the DOLORCLAST Focused Shock Waves system and are called Standoffs in the Duolith SD1 with C-Actor Handpiece system. For the DOLORCLAST Focused, the no offset (i.e., flat) treatment scenario is achieved using the D (i.e., deep penetration) gel pad, offset A is achieved using the M gel pad (i.e., medium penetration), and offset B is achieved using the S (i.e., shallow penetration) gel pad. For the Duolith SD1 C-Actor handpiece, the no offset treatment scenario (deep penetration) is achieved with the bare treatment head (no standoff), offset A is achieved using the Short (S) standoff (for medium penetration) and offset B is achieved using the Long (L) standoff (for shallow penetration).
**Based on Duolith US instruction manual the maximum impulses per treatment session available with the device is up to 4000. As with any ESWT system, the operator can always treat longer during a session by restarting the treatment.
## H. Performance Testing
The following testing was included in this 510(k) to demonstrate the safety and effectiveness of the device and is substantially equivalent to the predicate devices:
Page 3 of 5
{7}
510(k) Summary
K243279
# Acoustic Pulse Characterization
The purpose of this study was to characterize the pressure field output of the DOLORCLAST Focused Shock Waves device. The method used was based on IEC 61846 Ultrasonics – Pressure pulse lithotripters – Characteristics of fields – First edition – 1998 and IEC 63045 - Ultrasonics – Non-focusing short pressure pulse sources including ballistic pressure pulse sources – Characteristics of fields – Edition 1.0 – 2020-05, The measurement method used a hydrophone to record pressure data generated by the device at different positions of the sensor relative to the focal point when set to different intensity levels from minimum to maximum. Both temporal and spatial measurements were recorded. The results were compared to publicly available information on the pressure pulse output of the Duolith SD1 with C-Actor Handpiece to demonstrate the substantial equivalence of the proposed and primary predicate devices.
# System functional performance testing
- The operation of the DOLORCLAST Focused Shock Waves console was tested under an internal EMS protocol to confirm correct operation of the console at all treatment settings. All tests were passed.
- The operation of the DOLORCLAST Focus handpiece was tested under an internal EMS protocol to confirm correct operation of the handpiece including adjustment of the intensity, starting and stopping treatment. All tests were passed.
# Lifetime Testing
Testing was conducted under internal EMS protocols to confirm that the DOLORCLAST Focused Shock waves console and handpiece continued to function and meet their performance specifications for the claimed useful lifetimes of:
- Console: 7 years
- Handpiece: 2 years or 5,000,000 pulses
# Software Validation
The DOLORCLAST Focused Shock Waves system software was validated in both software testing and system level testing with the software integrated into the final device. Software development and testing was compliant with IEC 62304 Edition 1.1:2015-06 Medical device software – Software life-cycle processes.
# Electrical Safety and EMC
The device was tested and demonstrated to comply with the following standards:
- IEC 60601-1:2005 + A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
Page 4 of 5
{8}
510(k) Summary
K243279
## Biocompatibility Testing
The device labeling requires that the diabetic foot ulcer be covered with one of the recommended sterile wound dressings and contact gel throughout the treatment duration. Testing confirmed that the OpSite™ Wound Dressing (K852211 and K895408) and 3M Tegaderm™ +Pad (K811291 and K812678) are able to withstand worst case treatment conditions using the DOLORCLAST Focused Shock Waves without compromising their integrity. In addition, testing confirmed that the patient contacting materials of the DOLORCLAST Focused Shock Waves are biocompatible per the following ISO 10993 test standards for a surface device in contact with breached or compromised skin for prolonged duration (>24 h to 30 d, due to repeat treatment sessions):
- ISO 10993-1:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization
- ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity (Acute Systemic Toxicity and Material Mediated Pyrogenicity)
- ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests for irritation
## I. Conclusion
The information presented in this 510(k) demonstrates that DOLORCLAST Focused Shock Waves system is substantially equivalent to the predicate device under 21 CFR 878.4685, Product Code PZL for the treatment of diabetic foot ulcers.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.