K250779 · Curative Sound Therapeutics · PZL · Jul 2, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250779
Device Name
CS-Pro MED
Applicant
Curative Sound Therapeutics
Product Code
PZL · General, Plastic Surgery
Decision Date
Jul 2, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4685
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CS-Pro MED is intended to provide acoustic shock waves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm2, which extend through the epidermis, dermis, tendon, or capsule but without bone exposure. The CS-Pro MED is indicated for adults (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care.
Device Story
Handheld, battery-operated extracorporeal shock wave therapy (ESWT) device; utilizes piezocomposite transducer array to generate focused acoustic pressure pulses. Operator (healthcare professional) uses touch-sensitive display to adjust energy level, pulse count, and repetition frequency (2-12 Hz). Device includes five interchangeable elastomeric standoffs to control focal depth. Pressure pulses propagate through standoffs into patient tissue to stimulate biological healing response. Used in clinical settings for chronic diabetic foot ulcers. Output is focused acoustic energy; device provides non-invasive treatment to promote wound healing. Battery-powered design improves portability compared to AC-mains predicate.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via electrical safety (IEC 60601-1, IEC 60601-2-36), electromagnetic compatibility (IEC 60601-1-2), and acoustic characterization (IEC 61846). Biocompatibility evaluated per ISO 10993-1.
Technological Characteristics
Handheld, battery-operated (3.7V Li-ion). Piezoelectric transducer array generates focused acoustic shock waves. Five elastomeric standoffs for depth control (0-30mm). Energy flux density: 0.02-0.22 mJ/mm². Pulse repetition rate: 2-12 Hz. Touch-screen interface. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601-1 and IEC 60601-2-36.
Indications for Use
Indicated for adults (22+) with chronic, full-thickness diabetic foot ulcers (>30 days duration) measuring ≤16 cm2, extending through epidermis, dermis, tendon, or capsule, without bone exposure. Used with standard ulcer care.
Regulatory Classification
Identification
An extracorporeal shock wave device for treatment of chronic wounds is a prescription device that focuses acoustic shock waves onto the dermal tissue. The shock waves are generated inside the device and transferred to the body using an acoustic interface.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must be conducted to demonstrate that the system produces anticipated and reproducible acoustic pressure shock waves.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance data must demonstrate that the reusable components of the device can be reprocessed for subsequent use.
(4) Performance data must be provided to demonstrate the electromagnetic compatibility and electrical safety of the device.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Performance data must support the use life of the system by demonstrating continued system functionality over the labeled use life.
(7) Physician labeling must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) A detailed summary of the device's technical parameters;
(iii) Validated methods and instructions for reprocessing of any reusable components; and
(iv) Instructions for preventing hearing loss by use of hearing protection.
(8) Patient labeling must include:
(i) Relevant contraindications, warnings, precautions, adverse effects, and complications;
(ii) Information on how the device operates and the typical course of treatment;
(iii) The probable risks and benefits associated with the use of the device;
(iv) Post-procedure care instructions; and
(v) Alternative treatments.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 2, 2025
Curative Sound Therapeutics
Les Bogdanowicz
Quality Manager
11611 N. Meridian Street
Suite 425
Carmel, Indiana 46032
Re: K250779
Trade/Device Name: CS-Pro MED
Regulation Number: 21 CFR 878.4685
Regulation Name: Extracorporeal Shock Wave Device For Treatment Of Chronic Wounds
Regulatory Class: Class II
Product Code: PZL
Dated: March 13, 2025
Received: March 14, 2025
Dear Les Bogdanowicz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
2K250779 - Les Bogdanowicz
Page
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
3K250779 - Les Bogdanowicz
Page
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James H. Jang -S
Digitally signed by
James H. Jang -S
Date: 2025.07.02
23:17:12 -04'00'
James Jang, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
K250779
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Page 1 of 1
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
| Submission Number (if known) |
| --- |
| K250779 |
| Device Name |
| CS-Pro MED |
| Indications for Use (Describe) |
| The CS-Pro MED is intended to provide acoustic shock waves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm2, which extend through the epidermis, dermis, tendon, or capsule but without bone exposure. The CS-Pro MED is indicated for adults (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care. |
| Type of Use (Select one or both, as applicable) |
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K250779
Page 1 of 7
Curative Sound
CS-Pro MED
510(k) Summary
Page 1 of 7
# 510(k) Summary
## Sponsor
Curative Sound, LLC
11611 N. Meridian
Street Suite 425
Carmel, IN 46032
## Contact
Les Bogdanowicz PhD, MBA
VP of Quality and Regulatory Affairs
Ph: 224.585.3870
Les.Bogdanowicz@curative-sound.com
Date of Submission: February 28, 2025
Proposed Class: Class II
Proprietary Name: CS-Pro MED
Regulation Name: Extracorporeal Shock Wave Device for Treatment of Chronic Wounds
Classification Panel: General and Plastic Surgery Devices
Regulatory Number: 21 CFR 878.4685
Product Code: PZL
## Predicate Devices
Equivalence is claimed to the predicate device: DuoLith SD1 with C-ACTOR (K202112).
## Device Description
The CS-Pro MED is a handheld, battery-operated device that produces high-pressure, focused, acoustic shock waves. It is designed to provide healthcare professionals with an effective and easy-to-use device for administering focused Extracorporeal Shock Wave Therapy (ESWT) treatment. The CS-Pro MED is battery-powered and utilizes an array of piezocomposite transducers to create acoustic shock waves. The device delivers a drive signal to the piezocomposite transducers creating pressure pulses that propagate through the elastomeric standoffs and into the patient's tissue. Each pressure pulse is calibrated to arrive simultaneously at the focal point. The device includes a touch-sensitive display that indicates useful information and allows the operator to adjust the energy level, set the number of shock wave pulses, and adjust the pulse repetition frequency. The CS-Pro MED includes five standoffs of varying sizes to control the depth at which the therapy is focused, allowing the physician to target a specific tissue depth based on the treatment protocol.
{5}
K250779
Page 2 of 7
Curative Sound
CS-Pro MED
510(k) Summary
Page 2 of 7
# Indications for Use
The CS-Pro MED is intended to provide acoustic shock waves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm², which extend through the epidermis, dermis, tendon, or capsule but without bone exposure. The CS-Pro MED is indicated for adults (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care.
# Performance Data
The Curative Sound CS-Pro MED device complies with voluntary standards for electrical safety, electromagnetic compatibility, biocompatibility and characterization of pressure pulses. The following performance data are provided in support of the substantial equivalence determination:
## Electrical Safety
- Electrical Safety testing in accordance with IEC 60601-1:2005/A1:2012 + A2:2020 (Edition 3.2), Medical Electrical Equipment – Part 1: General requirements for Basic Safety and Essential Performance
- Electrical Safety testing in accordance with ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012, Medical Electrical Equipment – Part 1: General requirements for Basic Safety and Essential Performance Electromagnetic Compatibility
- Electrical Safety testing in accordance with IEC 60601-2-36 Edition 2.0: 2014-04, Medical electrical equipment - Part 2-36: Particular requirements for the safety of equipment for extracorporeally induced lithotripsy
## Electromagnetic Compatibility
- Electromagnetic compatibility testing in accordance with IEC 60601-1-2:2014/A1:2020 (Edition 4.1), Medical electrical equipment – Part 1-2: General requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
## Risk Management
- Risk analysis in accordance with ISO 14971:2019 (Third Edition), Medical Device – Application of Risk Management to Medical Devices
- Hazard Analysis / Risk Management was performed and demonstrated that all risks are mitigated to an acceptable level
{6}
K250779
Page 3 of 7
Curative Sound
CS-Pro MED
510(k) Summary
Page 3 of 7
# Biocompatibility
- There are no direct body patient contacting components since the CS-Pro MED standoff is separated from direct patient contact by a sterile barrier during treatment. Nevertheless, the treatment applicator head (standoff) component was tested to and meets applicable biocompatibility requirements.
- Biocompatibility evaluation was completed according to the FDA guidance document “Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process,” dated September 04, 2020
# Bench Testing
- Characterization of acoustic was performed per Per IEC 61846:1998, “Therapeutic focused short pressure pulse sources – Characteristics of fields”
- Nonclinical verification and validation testing was performed and demonstrated that the CS-Pro MED meets the design specifications and is safe and effective for its intended use. All tests required by the verification and validation plan were completed and passed. The testing demonstrates that the CS-Pro MED is substantially equivalent to the predicate devices.
# Clinical Information
Not applicable. Bench and performance testing support the conclusion of substantial equivalence in this submission.
# Technological Characteristics and Substantial Equivalence
The mechanism of action and technological characteristics of the CS-Pro MED and the predicate device are described in the comparison table below. Any differences between subject and predicate device do not raise any new questions of safety or effectiveness.
The CS-Pro MED generates extracorporeally induced acoustic shock waves to stimulate a biological healing response in treated tissue. The CS-Pro MED utilizes an array of piezoelectric transducers driven by an electrical signal to generate the acoustic shock waves.
{7}
K250779
Page 4 of 7
Curative Sound
CS-Pro MED
510(k) Summary
Page 4 of 7
# Substantial Equivalence Comparison Table
| Product Characteristic | Subject Device
CS-Pro MED | DuoLith SD1
with C-ACTOR | Comparison |
| --- | --- | --- | --- |
| 510(k) Number | To be assigned | K202112 | |
| Indication of Use | Indicated to provide acoustic pressure shock waves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm², which extend through the epidermis, dermis, tendon, or capsule, but without bone exposure. Indicated for adult (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care. | Indicated to provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm², which extend through the epidermis, dermis, tendon, or capsule, but without bone exposure. Indicated for adult (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care. | Same |
| Product Class | II | II | Same |
| Classification Panel | General and Plastic Surgery Devices | General and Plastic Surgery Devices | Same |
| Product Code and Regulation | 21 CFR 878.4685 | 21 CFR 878.4685 | Same |
| Modes of Action | Focused Pressure Pulses | Focused Pressure Pulses | Same |
| Mechanism of Action | Extracorporeally induced pressure pulses | Extracorporeally induced pressure pulses | Same |
| Operating Mode | Continuous | Continuous | Same |
{8}
K250779
Page 5 of 7
510(k) Summary
Page 5 of 7
Curative Sound
CS-Pro MED
| Product Characteristic | Subject Device CS-Pro MED | DuoLith SD1 with C-ACTOR | Comparison |
| --- | --- | --- | --- |
| Focus Depth Range | 0 - 30mm | 0 - 30 mm | Same |
| Energy Flux Density (EFD) Setting Range | 0.02 - 0.22mJ/mm² | 0.03 - 1.24mJ/mm² | Not Significantly Different |
| Number and Size of Treatment Applicator Heads | 5 standoffs | ø62 x 179 mm | Not Significantly Different |
| | 58 x 38mm | | |
| Device / Transducer Lifetime | 5M shocks | 5M shocks | Same |
| Pulse Repeat Rate | 2 - 12 Hz | 1 - 8 Hz | Not Significantly Different |
| Adustable Number of Pulses (min and max) | Variable, 100 - 5000 | Variable, 100 - 3000 | Not Significantly Different |
| Operating Temperature | 5° – 37°C | 10° – 30°C | Not Significantly Different |
| Weight and System Dimensions (WxHxD) | 3.7 kg | Unknown | Different
Subject Device is significantly smaller in size and lighter in weight. |
| | 203 x 76 x 44.5 mm | 454 x 187 x 460 mm | |
| Type of Acoustic Wave Generation | Piezoelectric, high voltage pulse applied to an array of transducers | Electromagnetic, pressure wave under water caused by discharge of high voltage condensers | Not Significantly Different
Differences are caused by alternative technology used to generate acoustic shock waves. |
| User Interface | Touch Screen | Touch Screen | Same |
{9}
K250779
Page 6 of 7
510(k) Summary
Page 6 of 7
| Product Characteristic | Subject Device CS-Pro MED | DuoLith SD1 with C-ACTOR | Comparison |
| --- | --- | --- | --- |
| Power Supply | 3.7 VDC, 2300 mAh, 8.5 Wh Lithium-ion battery | 100 – 240 VAC (AC-mains) | Different Subject Device is battery powered whereas predicate is AC-mains powered. |
| Treatment Duration | 5 - 10 min | 10 - 20 min | Not Significantly Different |
**Energy Flux Density (EFD) Setting Range** – The subject device uses alternative technology that is more suitable for a handheld, battery-operated device. The EFD range of the subject device is lower or within the setting range of the predicate device.
**Number and Size of Treatment Standoff Heads** – The number of standoff heads is different from the predicate device, but the area of the subject device standoff heads in contact with the patient is comparable and within the range of contact area of application pads / heads of the predicate device. All materials in contact with patient have been assessed for biocompatibility.
**Pulse Repeat Rate** – The subject device can be configured to administer 2 to 12 shock wave pulses per second (2 to 12 Hz), whereas the predicate can be configured to administer between 1 and 8 shock wave pulses per second. (1 to 8 Hz). Shock wave pulses from subject device are on the order of 2 microseconds in total duration. At the maximum pulse repeat rate setting for subject device (12 Hz), the device administers energy for only 24 microseconds per second, or put another way, the device is **not** administering energy 99.9976% of the time. There is no significant difference between 8 Hz and 12 Hz maximum pulse repeat rate.
**Adjustable Number of Pulses** – The subject device allows the clinician to set the number of pulses per treatment as needed between 100 and 5,000 pulses. This parameter is similar to that of the predicate device. This difference is not significant and does not raise new questions of safety or effectiveness.
**Operating Temperature** – The subject device has a slightly broader operating temperature range than the predicate.
**Weight and System Dimensions** – The subject device is smaller and lighter in weight than the predicate, making it far more convenient for handheld use.
**Type of Acoustic Wave Generation** – The subject device provides therapy by means of extracorporeally induced acoustic pressure waves that are generated by means of a focused array of piezoelectric elements. The acoustic pressure waves are coupled to the patient tissue by means of an elastomeric standoff with
{10}
K250779
Page 7 of 7
Curative Sound
CS-Pro MED
510(k) Summary
Page 7 of 7
specific acoustic properties. This technology creates the same type of acoustic pressure wave as the predicate device, with comparable acoustic parameters thereby achieving the same mechanism of action.
**Power Supply** – The power supply of the subject device differs from that of the predicate. The subject device is powered by a 3.8 V, 2300 mAh, 8.5 Wh rechargeable lithium-ion battery, whereas the predicate device is powered from AC-mains power.
**Treatment Duration** – The subject device is able to deliver shock wave pulses at a higher pulse repeat rate, providing for shorter treatment durations compared to the predicate device. This difference is not significant and does not raise new questions of safety or effectiveness.
## Substantial Equivalence
The CS-Pro MED is substantially equivalent to the predicate device DuoLith SD1 with C-ACTOR (K202112). The subject device is safe and effective for its intended use.
Any differences between the predicate and CS-Pro MED have no significant influence on the safety or effectiveness of the subject device. Therefore, the CS-Pro MED is substantially equivalent to the predicate device.
## Conclusion
Based upon the intended use and known technical information provided in this pre-market notification, the CS-Pro MED has been shown to be substantially equivalent to the currently marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.