HF Surgery Generator 400KHZ

K230194 · Richard Wolf Medical Instruments Corporation · GEI · Mar 22, 2023 · General, Plastic Surgery

Device Facts

Record IDK230194
Device NameHF Surgery Generator 400KHZ
ApplicantRichard Wolf Medical Instruments Corporation
Product CodeGEI · General, Plastic Surgery
Decision DateMar 22, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The product is used for generating high frequency electrical current for cutting and coagulating tissue electrosurgically.

Device Story

HF Surgery Generator 400KHZ converts electrical grid energy into high-frequency current for tissue cutting and coagulation; supports monopolar and bipolar modes. Operated by specialized medical personnel in clinical settings via touchscreen interface and footswitch. Device provides multiple current waveforms optimized for surgical requirements. Output affects tissue via thermal effects; enables precise surgical intervention. Benefits include controlled tissue management during micro- and macrosurgical procedures. System includes generator, footswitch, and connection cables; features LAN/USB connectivity for data/maintenance.

Clinical Evidence

Bench testing only. Evidence includes validation of electromagnetic compatibility and electrical safety per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2. Performance testing included waveform comparison, tissue performance studies, and verification of software functions, power consumption, and cooling methods.

Technological Characteristics

Metal and plastic housing; AC powered (100-240V, 50/60Hz). Protection Class I, CF applied part, defibrillator-proof. Output frequency 338-500 kHz. Connectivity: USB, LAN, potential equalization connector. Touchscreen interface. Software version 1.4.3. Standards: IEC 60601-1:2005, IEC 60601-1-2:2014, IEC 60601-2-2:2017.

Indications for Use

Indicated for adult patients requiring electrosurgical cutting and coagulation of tissue in micro- and macrosurgical operations. Exclusively for use by specialized, medically qualified, and adequately instructed personnel.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. March 22, 2023 Richard Wolf Medical Instruments Corporation Mike Mcandrew US Head of Regulatory - OA/OC 353 Corporate Woods Parkway Vernon Hills, Illinois 60061 Re: K230194 Trade/Device Name: HF Surgery Generator 400KHZ Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: January 23, 2023 Received: January 24, 2023 Dear Mike Mcandrew: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark Digitally signed by Mark Trumbore -S Trumbore -S Date: 2023.03.22 15:31:45 -04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) To be determined Device Name HF SURGERY GENERATOR 400KHZ Indications for Use (Describe) The product is used for generating high frequency electrical current for cutting and coagulating tissue electrosurgically. Type of Use (Select one or both, as applicable) | [X] Prescription Use (Part 21 CFR 801 Subpart D) | |--------------------------------------------------| | [ ] Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in black, while the word "WOLF" is in a dark red color. To the right of the text is a logo featuring a stylized dog standing next to a vertical line with a small object at the top and a paw print at the bottom, all enclosed in a rounded square. # 510(k) Summary 5 ### Submitter l Richard Wolf Medical Instruments Corporation 353 Corporate Woods Parkway Vernon Hills, IL 60061 Phone: 847-913-1113, ext. 380 Mobile: 847-636-0887 Fax: 847-913-1489 E-Mail: MMcAndrew@richardwolfusa.com Contact Person, Mike McAndrew, US Head of Regulatory – QA/QC Date Prepared: March 21, 2023 # Legal Manufacturer Richard Wolf GmbH Pforzheimer Straße 32 75438 Knittlingen ### ll Device | Product<br>Specification | Regulation<br>Number and | Product<br>Code | Device<br>Class | |--------------------------|--------------------------|-----------------|-----------------| |--------------------------|--------------------------|-----------------|-----------------| Table 5-1: Device classification | Brand name | Trade name | Model Number | Product<br>classification<br>name | Regulation<br>Number and<br>Name | Product<br>Code | Device<br>Class | Review<br>Panel | |------------|-----------------------------------|--------------|----------------------------------------------------------------|----------------------------------------------------------------------------------------|-----------------|-----------------|-----------------------------------| | 2260 | HF SURGERY<br>GENERATOR<br>400KHZ | 2260003 | Electrosurgical<br>, Cutting &<br>Coagulation &<br>Accessories | 878.4400<br>Electrosurgical<br>cutting and<br>coagulation<br>device and<br>accessories | GEI | Class II | General<br>and Plastic<br>Surgery | | N/A | FOOTSWITCH 2<br>PEDALS L 5M | 2260021 | | | | | | Document 001 510(k) Summary {4}------------------------------------------------ Image /page/4/Picture/1 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in black, while the word "WOLF" is in red. To the right of the words is a logo of a dog with a rod through its mouth, inside of a rounded square. | N/A<br>HF CONNECTION<br>CABLE BIPO L 5M | N/A<br>HF CONNECTION<br>CABLE BIPO L 3M | |-----------------------------------------|-----------------------------------------| | 8108253 | 8108233 | ### ��� Predicate Device Trade name of Predicate Device: ARC 400 510(k) Number: K193591 Requlatory Class: Class II Product Code: GEI Manufacturer: BOWA-Electronics GmbH & Co. KG The predicate has not been subject to a design-related recall. No reference devices were used in this submission. ### Subject Device Description 5.1 #### 5.1.1 Device Identification Table 5-2: Device identification | Brand<br>name | Model<br>Number | Trade name | Package unit | |---------------|-----------------|-------------------------------|--------------| | 2260 | 2260003 | HF SURGERY GENERATOR 400KHZ | 1 | | N/A | 2260021 | FOOTSWITCH 2 PEDALS L 5M | 1 | | N/A | 8108233 | HF CONNECTION CABLE BIPO L 3M | 1 | | N/A | 8108253 | HF CONNECTION CABLE BIPO L 5M | 1 | ### 5.1.2 Device characteristics # Delivered sterile / non-sterile The HF SURGERY GENERATOR 400KHZ is delivered non-sterilization is not required. # Single use / reusable The devices are reusable and do require cleaning and reprocessing during their use-life, i.e., users are required to clean and disinfect the device before every application and before returning for repairs. Methods of cleaning and reprocessing are detailed in the Instruction for Use. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in black, while the word "WOLF" is in a dark red color. To the right of the text is a logo featuring a stylized image of a wolf's head with a nail through it, enclosed in a rounded square. # Software The HF SURGERY GENERATOR 400KHZ contains software. The software is classified as a Major Level of Concern. # Materials with patient contact The HF SURGERY GENERATOR 400KHZ is not intended to contact the patient directly or indirectly. ### 5.1.3 Brief written description of the Device The HF SURGERY GENERATOR 400KHZ is used for electrosurgical cutting and coagulation of living human tissue. For this purpose, electrical energy from the power supply network is transformed into high frequency current which allows this surgical property. The device offers a multitude of different profiled current forms which are optimized for the different surgical requirements. It is equipped for monopolar and bipolar cutting and coagulation in micro- and macrosurgical operations. Image /page/5/Figure/8 description: The image shows a Richard Wolf HF-GENERATOR 2260. The device has a screen that displays the settings for different modes, including "M" and "B". The screen also shows that the device is currently "ON" and set to "Laparoscopy - Standard" mode, with options for "Mode" and "Menu". Figure 05-1: Richard Wolf HF SURGERY GENERATOR 400KHZ #### 5.1.4 Materials of Use The HF SURGERY GENERATOR 400KHZ consists of a metal housing which does not have any patient contact. ### 5.2 Indications for Use # Statement The product is used for generating high frequency electrical current for cutting and coagulating tissue electrosurqically. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in a smaller, black font above the word "WOLF", which is in a larger, red font. To the right of the text is a logo featuring a dog's head next to a vertical line with a small hand at the bottom, all enclosed in a rounded square. # User This product is exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately instructed persons. # Patient population The product is intended for adult patients. The patient group intended for the application of the medical product herein described is not limited with regard to ethnicity, gender, body height, Before use, the attending physician must make sure that the product in view of its dimensions or settings can be used safely in the patient. # Explanation on differences to the predicate device In comparison to the predicate device, the usage of the subject device is further restricted to adults. However, the full indication of use for the subject device is covered by the predicate device and reducing the scope does not lead to new questions regarding safety and effectiveness. # Comparison of Technological Characteristics with the Predicate Device 5.3 ### 5.3.1 Overview Table Table 5-3: Comparison of technical characteristics | ltem | Subject device | Predicate devices | Comparison | | | | |-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--|--|--| | General information | | | | | | | | Product name | HF SURGERY<br>GENERATOR 400KHZ | ARC 400 Electrosurgical<br>Unit | N/A | | | | | Brand name | 2260 | N/A | N/A | | | | | Type No. | 2260003 | 900400 | N/A | | | | | 510(k) number | TBD / subject device | K193591 | N/A | | | | | Manufacturer | Richard Wolf GmbH | BOWA-electronics GmbH<br>& Co. KG | N/A | | | | | Product Code | GEI | GEI | Same | | | | | Regulation<br>Number | 878.4400 | 878.4400 | Same | | | | | Indications for<br>Use | The product is used for<br>generating high frequency<br>electrical current for<br>cutting and coagulating<br>tissue electrosurgically. | The HF device is intended<br>exclusively for the<br>generation of electrical<br>power for monopolar and<br>bipolar cutting and<br>coagulation on tissue<br>structures in surgical<br>operations. | Equivalent | | | | | Contraindications | Contraindications directly<br>related to the product are<br>presently unknown. If<br>medical contraindications<br>for the mentioned<br>indications are known, the<br>use of the products is not<br>permitted.<br>On the basis of the<br>patient's general condition<br>the doctor in charge must<br>decide whether or not the<br>planned use is possible. | Do not use the HF device<br>if, in the opinion of an<br>experienced physician or<br>according to current<br>professional literature,<br>such use would endanger<br>the patient, due for<br>example to the general<br>condition of the patient, or<br>if other contraindications<br>are present. | Equivalent | | | | | Prescription/<br>over-the-counter-<br>use | RX only | RX only | Same | | | | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in a smaller, sans-serif font above the word "WOLF", which is in a larger, blocky, red font. To the right of the text is a logo featuring a stylized drawing of a wolf's head with a vertical line through it, enclosed in a rounded square. | Item | Subject device | Predicate devices | Comparison | |--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Sterile/Non-sterile<br>device | Non-sterile device | Non-sterile device | Same | | Single use/<br>Reusable | Reusable device | Reusable device | Same | | Sterilization<br>method | Non-sterile device | Non-sterile device | Same | | Reprocessing | Cleaning and disinfection<br>solution/wipes for low<br>level should be based on<br>alcohol. Examples for<br>suitable substances for<br>cleaning and disinfection<br>of surfaces:<br>• Propanol<br>• Ethanol<br>• Didecyldimethyla<br>mmoniumchloride<br>• Dimethylbenzyla<br>mmoniumchloride | Apply the cleaning agent<br>and disinfectant.<br>Wipe down the HF device<br>only with cleaning agents<br>and disinfectants that are<br>approved in the country of<br>use for surface cleaning. | Equivalent | | Maintenance<br>intervals | 12 months | 12 months | Same | | Operating<br>Condition | Ambient temperature +10<br>°C to +40 °C<br>Relative humidity 30% to<br>75% (non-condensing)<br>Atmospheric pressure 700<br>hPa to 1060 hPa | Temperature: +10 °C to<br>+40 °C<br>Relative humidity: 30 to<br>75%, non-condensing<br>Atmospheric pressure:<br>700 to 1600 hPa<br>Operating altitude (max.)<br>4000 m above sea level | Equivalent | | Storage and<br>Transport<br>Conditions | Ambient temperature -20<br>°C to +60 °C<br>Relative humidity 10% to<br>90% (non-condensing)<br>Atmospheric pressure 700<br>hPa to 1060 hPa | Temperature: -20 °C to<br>+50 °C<br>Relative humidity 0 to 75<br>%, non-condensing<br>Atmospheric pressure:<br>500 to 1600 hPa | Equivalent | | Operational modes / settings | | | | | Mode: Monopolar cutting | | | | | Modes | 12 | 12 | Same | | Max. Power | 400 W (at 500 Ω) | 400 W (at 200 Ω) | Same | | Output Frequency | 403 - 500 kHz | 350 kHz | Different | | Max. Voltage<br>Output | 3150 Vp | 1600 Vp | Different | | Crest Factor | 1.4 - 4.4 | 1.5, 3.5 | Different | | Wave Forms | Sinusoidal Constant<br>Sinusoidal Modulated<br>Sinusoidal Alternating | Sinusoidal Constant<br>Sinusoidal Modulated<br>Sinusoidal Alternating<br>Cut/Coag/Pause Phases | Equivalent | | Mode: Monopolar coagulation | | | | | Modes | 15 | 9 | Different | | Max. Power | 200 W (at 350 Ω) | 250 W (at 500 Ω) | Different | | Output Frequency | 375 - 500 kHz | 350 kHz | Different | | Max. Voltage<br>Output | 5000 Vp | 5000 Vp | Equivalent | | Crest Factor | 1.4 - 7.6 | 1.6 - 7.4 | Different | | Wave Forms | Sinusoidal Constant<br>Sinusoidal Modulated<br>Pulse Modulated | Sinusoidal Constant<br>Sinusoidal Modulated<br>Pulse Modulated | Same | | Item | Subject device | Predicate devices | Comparison | | Modes | 8 | 4 | Different | | Max. Power | 300 W (at 100 Ω) | 400W (at 75 Ω) | Different | | Output Frequency | 355 - 381 kHz | 350 kHz | Different | | Max. Voltage | 750 Vp | 500 Vp | Different | | Output | | | | | Crest Factor | 1.4 - 2.3 | 1.5 - 1.6 | Different | | Wave Forms | Sinusoidal Constant | Sinusoidal Constant | Same | | Mode: Bipolar coagulation | | | | | Modes | 9 | 9 | Equivalent | | Max. Power | 250 W (at 75 Ω) | 350 W (at 25 Ω) | Different | | Output Frequency | 338 - 381 kHz | 350 kHz | Different | | Max. Voltage | 425 Vp | 550 Vp | Different | | Output | | | | | Crest Factor | 1.4 - 1.7 | 1.5 - 3.8 | Different | | Wave Forms | Sinusoidal Constant | Sinusoidal Constant | Same | | | Pulse Modulated | Pulse Modulated | | | Technological characteristics | | | | | Standard | IEC 60601-1:2005 (19-4) | IEC 60601-1: 2005 (19-4) | Same | | compliance | IEC 60601-1-2:2014 (19-8) | IEC 60601-1-2: 2014 (19-8) | | | | IEC 60601-2-2:2017 (6-389) | IEC 60601-2-2:2017 (6-389) | | | Voltage [V] | 100-240 V | 230 V | Equivalent | | Frequency [Hz] | 50/60 Hz | 50/60 Hz | Same | | Power | 45 without HF output | 85 power consumption in | Equivalent | | consumption [VA] | 650 at max. output power | standby mode | | | / Current rating | | 975 maximum power | | | [A] | | consumption with 300 W<br>HF output power | | | Protection class | I | I | Same | | to EN/IEC 60601-<br>1;<br>(UL 2601-1/CSA<br>C22.2 No.601.1 -<br>for USA) | | | | | Protection against | CF applied part,<br>defibrillator-proof | CF | Same | | electric shock | | | | | Degree of | IPX0 | IP 21 | Equivalent | | protection against | | | | | liquids | | | | | Weight [kg] | 10.2 kg (22.4 lbs) | approx. 12.5 kg | Different | | Dimensions | 300 x 160 x 402 mm | 430 x 180 x 475 mm | Different | | Material | Metal & plastics | Metal & plastics | Equivalent | | Single-Use | No | No | Same | | Light Source | No | No | Same | | Battery Operated | No | No | Same | | AC Powered | Yes | Yes | Same | | Software version | 1.4.3 | 1.0.0.6.17.9 | Different | | USB Port | Yes | Yes | Same | | Potential | Yes | Yes | Same | | equalization | | | | | connector | | | | | LAN (Ethernet)<br>Network | Yes | Yes | Same | | Item | Subject device | Predicate devices | Comparison | | Foot switch<br>socket connector | Yes | Yes | Same | | Control elements<br>/ user interface | Touchscreen | Touchscreen | Same | | Materials with patient contact | | | | | Materials with<br>direct patient<br>contact | N/A - no materials with<br>direct patient<br>contact | N/A - no materials with<br>direct patient<br>contact | Same | | Materials with<br>indirect patient<br>contact | N/A - no materials with<br>indirect patient<br>contact | N/A - no materials with<br>indirect patient<br>contact | Same | | Instruction for use | | | | | Version | GA-A339 / en / US / V0.6 | 900-<br>400_IFU_V1.0.0.6.17.9_2<br>0140-S2-20130604-EN | Different | {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for Richard Wolf. The words "RICHARD" and "WOLF" are stacked on top of each other, with "RICHARD" in black and "WOLF" in red. To the right of the text is a square with rounded corners containing a stylized image of a wolf's head and a medical instrument. The logo is simple and modern, with a clear focus on the company name. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image contains the words "RICHARD WOLF" in a stylized font. The word "RICHARD" is in a smaller, sans-serif font above the word "WOLF", which is in a larger, bolder, red font. To the right of the text is a logo featuring a stylized wolf head with a nail through it, enclosed in a rounded square. # 5.3.2 Discussion The subject device and the predicate device share the same fundamental technology. They both convert electrical energy from the grid into a high-frequency current which provides the surgical property. For that purpose, monopolar and bipolar high frequency current modes were provided to the corresponding socket module. The fundamental frequencies of about 400 kHz provided by the subject and the predicate device are comparable and substantial equivalent. Both generators feature a maximum power output of 400 W. The general and technical characteristics that differ between the predicate and the subject device are differences in PCB layouts, software including the graphical user interface, number of output sockets and other technical characteristics such as power consumption and weight, dimensions. These differences do not question safety and effectiveness of the subject devices, several performance tests were performed, such as: - a. Validation of the electromagnetic compatibility and electrical safety (Section 17 - Electromagnetic Compatibility and Electrical Safety) - b. Validation of the performance on tissue (Section 19 Performing Testing Animal) - Validation of the performance based on waveform comparison and the tissue study (Section C. 18 - Performance Testing Bench) - ರ. Performance testing due to the different software functions, the change in power consumption, cooling method, and the "Dialog" function (Section 18 - Performance Testing Bench) - e. Transportation validation (Section 18 Performance Testing Bench) ### Performance Testing 5.4 - 5.4.1 Sterilization N/A ### 5.4.2 Biocompatibility N/A {10}------------------------------------------------ Image /page/10/Picture/1 description: The image contains the words "RICHARD WOLF" in a logo format. The word "RICHARD" is in a smaller, sans-serif font above the word "WOLF", which is in a larger, bolder, red font. To the right of the words is a logo of a wolf with a medical device. The logo is contained within a rounded square. ### 5.4.3 Electromaqnetic Compatibility and Electrical Safety The predicate device and the device covered by this registration are classified in protection class I according to 60601-1. They both fulfil the electrical safety requirements of IEC 60601-2-2 and IEC 60601-1-2. Based on that, the devices are considered to be comparable in terms of electromagnetic compatibility and electrical safety. #### 5.4.4 Performance and Operational testing The efficacy and safety of the HF SURGERY GENERATOR 400KHZ is documented by the verification and validation testing which confirms that the product meets all the requirements and specifications for overall design, basic safety, essential performance, and that the design inputs and specifications are met. #### રે. રે Conclusions The HF SURGERY GENERATOR 400KHZ has the same intended use and field of application as the legally marketed predicate device named in this submission. The same performance tests were concluded for the subject device as for the predicate device. Therefore, no comparison testing was performed. Non-clinical performance testing, including Electromagnetic Compatibility. Electrical Safety, and Software Verification and Validation. demonstrated that the safety and performance of the device is equivalent to the predicate device. Differences in characteristics described in Section 12.2 do not raise any new questions regarding the safety and effectiveness of the HF SURGERY GENERATOR 400KHZ compared to the predicate device. The HF SURGERY GENERATOR 400KHZ is substantially equivalent to the legally marketed predicate device ARC 400 (K193591).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...