K171717 · KARL STORZ Endoscopy-America, Inc. · GEI · Aug 2, 2017 · General, Plastic Surgery
Device Facts
Record ID
K171717
Device Name
AUTOCON III 400
Applicant
KARL STORZ Endoscopy-America, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The AUTOCON III 400 is intended for use by qualified surgeons to provide a high frequency electrical current for monopolar and bipolar cutting and coagulation.
Device Story
Electrosurgical unit (ESU) generating high-frequency (HF) electrical current for tissue cutting and coagulation; used in surgical operations by qualified surgeons. System comprises generator, foot switch, and power cord. Features touchscreen interface for mode selection (37 modes across monopolar/bipolar cut/coagulation), power settings, and effect adjustments. Includes bipolar, monopolar, and neutral electrode sockets with contact quality monitoring. Device transforms electrical input into specific waveforms (sinusoidal constant, modulated, pulse modulated) defined by power, crest factor, and duty cycle. Output affects tissue via thermal effects. Provides visual status of accessories and peripherals. Enhances surgical precision through controlled tissue effects.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance testing (IEC 60601-2-2). Bench testing verified output waveforms, power output across loads, accessory functionality, thermal effects on tissue (liver, kidney, muscle), and contact quality monitoring.
Technological Characteristics
Electrosurgical generator; 350kHz output frequency; 100-127V system voltage. Monopolar and bipolar energy delivery. Features contact quality monitoring system. Touchscreen interface. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2. Connectivity includes integration with KARL STORZ OR1.
Indications for Use
Indicated for qualified surgeons to provide high frequency electrical current for monopolar and bipolar cutting and coagulation of tissue structures during surgical operations.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 2, 2017
KARL STORZ Endoscopy-America, Inc. % Dawn Tibodeau Third Party 510(k) Project Coordinator TÜV SÜD America Inc. 1775 Old Highway 8 NW New Brighton, Minnesota 55112-1891
Re: K171717
Trade/Device Name: Autocon III 400 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: July 26, 2017 Received: July 31, 2017
Dear Dawn Tibodeau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv.
Image /page/1/Picture/9 description: The image shows the name "Jennifer R. Stevenson -S3" in a large, sans-serif font. The text is arranged vertically, with "Jennifer R." on the top line and "Stevenson -S3" on the second line. The background is a light color, and the text is dark, creating a high contrast.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) Not yet assigned K171717
Device Name AUTOCON III 400
Indications for Use (Describe)
The AUTOCON III 400 is intended for use by qualified surgeons to provide a high frequency electrical current for monopolar and bipolar cutting and coagulation of tissue structures during surgical operations.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
KARL STORZ Premarket Notification AUTOCON III 400 007_510(k) Summary
## 7. 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Submitter: | KARL STORZ Endoscopy-America, Inc. |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 2151 E. Grand Avenue<br>EI Segundo, CA 90245 |
| Contact: | AnnaLisa Smullin<br>Regulatory Engineer<br>Phone: (424) 218-8376<br>Fax: (424) 218-8519 |
| Date of<br>Preparation: | March 5, 2017 |
| Type of 510(k)<br>Submission: | Traditional |
| Device<br>Identification: | Trade Name: AUTOCON III 400<br>Classification Name: Electrosurgical Generator |
| Regulatory<br>Class: | II |
| Product Code: | GEI |
| Regulation: | 21 CFR part 878.4400 (Electrosurgical cutting and coagulation device and<br>accessories) |
| Predicate<br>Device(s): | Primary Predicate Device: AUTOCON II 400 (K062464)<br>Second Predicate Device: ERBE VIO 300D (K060484)<br>These predicate devices have not been subject to a design-related recall. |
| Device<br>Description: | The AUTOCON III 400 is an electrosurgical unit (ESU) that generates<br>High Frequency (HF) electrical current to cut and/or coagulate tissue.<br>AUTOCON III 400 offers a total of 37 different modes that can be divided<br>into 4 main categories: Monopolar cutting (13 modes), monopolar<br>coagulation (11 modes), bipolar cutting (4 modes), and bipolar<br>coagulation (9 modes). The various modes in each category represent<br>different settings that provide controlled cutting and coagulation. Each<br>mode can be defined by the waveform type, power, crest factor, and duty<br>cycle which correspond to different tissue effects. |
{4}------------------------------------------------
K171717
Image /page/4/Picture/2 description: The image shows the logo for Karl Storz Endoskope. The logo is in black and white and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. The letter "O" in the word "STORZ" is stylized as a circle with a smaller circle inside.
KARL STORZ Premarket Notification AUTOCON III 400 007_510(k) Summary
| | AUTOCON III 400 consists of a generator, a foot switch, and a power cord. A touchscreen user interface displays the connection status of accessories and peripherals and allows the user to select modes, power settings, effect, and various other settings. AUTOCON III 400 has a bipolar socket, a monopolar socket, and neutral electrode socket for the instruments and accessories with which the generator is compatible. Contact quality indicators are displayed to indicate if the neutral electrodes are correctly connected or not. | | | | |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|--|
| Intended Use: | The AUTOCON III 400 is intended for use by qualified surgeons to provide a high frequency electrical current for monopolar and bipolar cutting and coagulation. | | | | |
| Indications For Use: | The AUTOCON III 400 is intended for use by qualified surgeons to provide a high frequency electrical current for monopolar and bipolar cutting and coagulation of tissue structures during surgical operations. | | | | |
| Technological Characteristics: | The subject and predicate devices are electrosurgical generators used to cut and coagulate tissue. The technological differences between the systems are described in the table below. | | | | |
| | | AUTOCON III 400<br>(Proposed) | AUTOCON II 400<br>(K062464) | ERBE VIO 300D<br>(K060484) | |
| | Manufacturer | KARL STORZ<br>Endoscopy | KARL STORZ<br>Endoscopy | ERBE USA, Inc. | |
| | Product Code | GEI | GEI | GEI | |
| | Energy | Monopolar, Bipolar | Monopolar, Bipolar | Monopolar, Bipolar | |
| | Neutral<br>Electrode | Compatible FDA<br>cleared Single or<br>Split Electrodes | Compatible FDA<br>cleared Single or<br>Split Electrodes | Compatible FDA<br>cleared Single or<br>Split Electrodes | |
| | Electrode<br>Monitoring<br>System | Yes | Yes | Yes | |
| | System Voltage | 100-127 V | 100-127 V | 100-120 V | |
| | Connection to<br>KARL STORZ<br>OR 1TM | Yes | Yes | No | |
| | Monopolar Cut Modes | | | | |
| | # of Modes | 13 | 6 | 7 | |
| | Max Power | 400W (at 200Ω) | 300W (at 500Ω) | 300W (at 500Ω) | |
| | Output<br>Frequency | 350kHz | 350kHz | 350kHz | |
| | Max Voltage<br>Output | 1600Vp | 1450Vp | 1550Vp | |
| | Crest Factor | 1.5, 3.5 | 1.4-3.8 | 1.4 | |
| | Wave Forms | -Sinusoidal Constant | -Sinusoidal Constant | -Sinusoidal Constant | |
| | | -Sinusoidal<br>Modulated | -Sinusoidal<br>Modulated | -Sinusoidal<br>Modulated | |
| | | -Sinusoidal<br>Alternating | -Sinusoidal<br>Alternating | -Sinusoidal<br>Alternating | |
| | | Cut/Coag/Pause<br>Phases | Cut/Coag/Pause<br>Phases | Cut/Coag/Pause<br>Phases | |
| | | <b>Monopolar Coagulation Modes</b> | | | |
| | # of Modes | 11 | 3 | 8 | |
| | Max Power | 250W (at 500Ω) | 200W (at 500Ω) | 200W (at 500Ω) | |
| | Output<br>Frequency | 350kHz | 350kHz | 350kHz | |
| | Max Voltage<br>Output | 5000Vp | 4300Vp | 4300Vp | |
| | Crest Factor | 1.6-7.4 | 1.4-7.4 | 3.7-5.3 | |
| | Wave Forms | -Sinusoidal Constant<br>-Sinusoidal<br>Modulated<br>-Pulse Modulated | -Sinusoidal Constant<br>-Sinusoidal<br>Modulated<br>-Pulse Modulated | -Sinusoidal Constant<br>-Sinusoidal<br>Modulated<br>-Pulse Modulated | |
| | | <b>Bipolar Cut Modes</b> | | | |
| | # of Modes | 4 | 4 | 3 | |
| | Max Power | 400W (at 75Ω) | 370W (at 500Ω) | 370W (at 500Ω) | |
| | Output<br>Frequency | 350kHz | 350kHz | 350kHz | |
| | Max Voltage<br>Output | 500Vp | 740Vp | 770Vp | |
| | Crest Factor | 1.5-1.6 | 1.4 | 1.4 | |
| | Wave Forms | -Sinusoidal Constant | -Sinusoidal Constant | -Sinusoidal Constant | |
| | | <b>Bipolar Coagulation Modes</b> | | | |
| | # of Modes | 9 | 5 | 5 | |
| | Max Power | 350W (at 25Ω) | 200W (at 50Ω) | 200W (at 50Ω) | |
| | Output<br>Frequency | 350kHz | 350kHz | 350kHz | |
| | Max Voltage<br>Output | 550Vp | 190Vp | 560Vp | |
| | Crest Factor | 1.5-3.8 | 1.4 | 1.4-3.8 | |
| | Wave Forms | -Sinusoidal Constant<br>-Pulse Modulated | -Sinusoidal Constant<br>-Pulse Modulated | -Sinusoidal Constant<br>-Pulse Modulated | |
| Non-Clinical<br>Performance<br>Data: | <b>Electrical Safety and Electromagnetic Compatibility (EMC) Summary</b> | | | | |
| | The electrical safety and EMC data submitted for AUTOCON III 400 is in<br>compliance with the following FDA recognized standards: | | | | |
| | • IEC 60601-1 | | | | |
| | • IEC 60601-1-2 | | | | |
| | • IEC 60601-2-2 | | | | |
| | <b>Performance Testing – Bench Top Summary</b> | | | | |
| | Comparative verification testing has been performed per FDA Guidance<br>Premarket Notification (510(k)) Submissions for Electrosurgical Devices | | | | |
| | for General Surgery issued on August 15, 2016 for each major component<br>and for the system. | | | | |
| | • Electrosurgical Unit (ESU) | | | | |
| | o Technical Specifications | | | | |
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for Karl Storz Endoskope. The logo is in black and white and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.
KARL STORZ Premarket Notification AUTOCON III 400 007_510(k) Summary
{6}------------------------------------------------
K171717
Image /page/6/Picture/2 description: The image shows the logo for STORZ. The logo consists of the word "STORZ" in large, bold letters, with a circle in place of the letter "O". Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. The logo is black and white.
KARL STORZ Premarket Notification AUTOCON III 400 007_510(k) Summary
| | o Output Waveform at the rated load<br>o Power Output at maximum and half-of-maximum intensity<br>over the range of expected loads |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • | Miscellaneous Components/Accessories<br>o Functional testing of accessories |
| • | System Testing<br>o Thermal Effects on Tissue (liver, kidney, muscle)<br>o Contact Quality Monitoring (CQM)<br>Testing confirmed that comparable tissue effects and electrical waveforms<br>could be achieved for all modes of operation. |
| Clinical<br>Performance<br>Data: | Clinical testing was not required to demonstrate the substantial<br>equivalence to the predicate devices. Non-clinical bench testing was<br>deemed sufficient to assess safety and effectiveness and establish<br>substantial equivalence. |
| Conclusion: | The conclusions drawn from the non-clinical tests such as the bench top<br>performance data, the software data, the electrical safety data, and the<br>electromagnetic compatibility data demonstrated that the subject device is<br>as safe as and as effective as the predicate devices. As such, we concluded<br>that the substantial equivalence of the subject and the predicate devices<br>has been met and the differences between the subject and the predicate<br>devices do not raise new questions of safety and effectiveness. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.