ERBE ESU Model VIO dV with Accessories

K150364 · Erbe Elektromedizin GmbH · GEI · May 5, 2016 · General, Plastic Surgery

Device Facts

Record IDK150364
Device NameERBE ESU Model VIO dV with Accessories
ApplicantErbe Elektromedizin GmbH
Product CodeGEI · General, Plastic Surgery
Decision DateMay 5, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ERBE ESU Model VIO dV with Accessories is intended to deliver High Frequency (HF) current for the cutting and/or coagulation of tissue.

Device Story

Electrosurgical unit (ESU) generates high-frequency electrical current for tissue cutting and coagulation; features multiple modes (Auto Cut, Dry Cut, Swift Coag, Forced Coag, Bipolar Soft Coag, BiClamp). Device includes touchscreen interface and physical controls; connects to monopolar/bipolar cables and footswitches. Integrates with da Vinci Xi Surgical System via Remote Communication Bus (RCB) for remote HF-activation and setting adjustments. Used in clinical surgical settings by physicians. Output affects tissue via thermal effects; enables precise surgical intervention. Benefits include surgical flexibility and remote control integration for robotic-assisted procedures.

Clinical Evidence

Bench testing only. No clinical or animal data required. Verification and validation of software version 3.1 for remote control functions and integrated system testing (VIO dV and da Vinci system) were performed. BiClamp mode output parameters were measured to confirm performance against predicate specifications.

Technological Characteristics

Electrosurgical generator; non-sterile, reusable. Energy source: High Frequency (HF) electrical current. Connectivity: Remote Communication Bus (RCB) for integration with da Vinci Xi Surgical System. User interface: Touchscreen and physical controls. Software: Version 3.1. Modes: Auto Cut, Dry Cut, Swift Coag, Forced Coag, Bipolar Soft Coag, BiClamp.

Indications for Use

Indicated for cutting and/or coagulation of tissue using High Frequency (HF) electrical current in surgical applications.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized design of an eagle or bird-like figure with three human profiles incorporated into its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the emblem. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 5, 2016 ERBE Elektromedizin GmbH Mr. Axel Retzlaff Director of Regulatory Affairs Waldhoernlestr. 17 Tuebingen, Germany 72072 Re: K150364 Trade/Device Name: ERBE ESU Model VIO dV with Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device Regulatory Class: Class II Product Code: GEI Dated: May 2, 2016 Received: May 4, 2016 Dear Mr. Retzlaff: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Jennifer R. Stevenson -A For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K150364 Device Name ERBE ESU Model VIO dV with Accessories Type of Use (Select one or both, as applicable) Indications for Use (Describe) The ERBE ESU Model VIO dV with Accessories is intended to deliver High Frequency (HF) current for the cutting and/ or coagulation of tissue. | | X Prescription Use (Part 21 CFR 801 Subpart D)<br>Over=The=Counter Use (21 CFR 801 Subpart C) | | | |----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | | | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*<br>The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | | | | | | | | Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff(@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY # As Required by 21 CFR 807.92(c) | Owner/Operator: | ERBE Elektromedizin GmbH<br>Waldhoernlestr. 17<br>72072 Tuebingen<br>Germany<br>Phone +49-7071-755-0<br>Fax +49-7071-755-179<br>Contact: Axel Retzlaff, Director RA | |------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date Prepared: | 20/01/2015 | | Trade (Proprietary) Name: | ERBE ESU Model VIO dV with Accessories | | Common Name: | Electrosurgical Unit (ESU/Generator) System | | Classification Name: | Electrosurgical Cutting and Coagulation Device<br>and Accessories (21 CFR 878.4400) | | Product Code: | GEI | | Legally Marketed Predicate Device: | ERBE ESU Model VIO dV with Accessories,<br>510(k) Number K133180<br><br>ERBE ESU Model 300 D with Accessories, 510(k)<br>Number K083452 | Description of the Device: The ERBE ESU Model VIO dV with Accessories is an ElectroSurgical Unit that generates High Frequency (HF) electrical current to cut and/or coagulate tissue. The ESU has clearly defined Cutting and Coagulation Modes with different electrical waveforms and electrical parameters, which are programmed with defined Effect levels. The available Modes are Auto Cut, Dry Cut, Swift Coag, Forced Coag, and Bipolar Soft Coag with and without Auto Stop and BiClamp. Each Effect level corresponds to a specific voltage. The Modes provide the {4}------------------------------------------------ physician flexibility in interventional applications. Thus the Unit may be used for a broad array of surgical applications. The ESU user interface consists primarily of a touchscreen surrounded by a small number of physical controls, such as a power switch and connection points for the instruments and accessories with which the generator is compatible. Various hand instruments and neutral electrodes from ERBE and different manufacturers may be attached to and operated by the generator. The standard Accessories for the ESU consist of reusable Footswitches as well as Monopolar and Bipolar Cables The VIO dV has a RCB (Remote Communication Bus) for remote control operation by the da da Vinci Xi Surgical System by Intuitive Surgical Inc. To have the unit installed as part of the da Vinci Xi Surgical System, contact Intuitive Surgical ## Intended Use: The ERBE ESU Model VIO dV with Accessories is intended to deliver High Frequency (HF) current for the cutting and/or coagulation of tissue. #### Summary of the Technological Characteristics: #### Similarities between modified and predicate devices: #### Unit: The proposed device ERBE ESU Model VIO dV (remote) compares to the predicate device ERBE ESU Model VIO dV (standalone) as follows: Both devices have the same intended use, the same basic technology, protective circuits, and use the same basic accessories. Both devices are identical in hardware design including all materials and are using the same energy source. Both generators have the same user interface displays to select modes, power settings, etc. and both have the same Modes Auto Cut, Dry Cut, Swift Coag, Forced Coag, Bipolar Soft Coag with and without AutoStop function. Also, both units have audio and visual error monitoring. {5}------------------------------------------------ The subject ESU and the predicate devices are manufactured by ERBE Elektromedizin GmbH in Germany and are supplied as non-sterile and are reusable. The packaging is also the same for both devices with similar labeling (e.g. Outer Package Label, User Manual, etc.). The proposed device ERBE ESU Model VIO dV (remote) compares to the predicate device ERBE ESU Model VIO 300 D as follows: Both devices have the same BiClamp Mode. #### Accessories: The one pedal and two pedal footswitches are the same for both the subject and the predicate devices. The packaging is also the same with equal labeling. #### Differences between modified and predicate devices: The proposed device ERBE ESU Model VIO dV (remote) differs from the predicate device ERBE ESU Model VIO dV (standalone) as follows: The proposed device has an updated Software Version 3.1 (predicate: 3.0) as the device is intended for remote control including HF-activation and change of settings via Intuitive Surgical, Inc.'s Da Vinci Surgical System. The BiClamp Mode which was disabled in the predicate ERBE ESU Model VIO dV is enabled in the proposed device. The User Manual was modified to appropriately address both changes. ### Performance Data to determine Substantial Equivalence The release of the software version 3.1 for the RCB remote control of the ERBE ESU Model VIO dV has been verified and validated as part of the design control. Software validation for remote control functions has been carried out for the VIO dV Software 3.1 as well as for the integrated system (VIO dV and da Vinci system). Testing/measurement of the output parameters of the BiClamp-Mode was carried out to show that the ESU's BiClamp-Mode performs as specified and in accordance with the specifications of the predicate device ERBE ESU Model VIO 300 D. ERBE Elektromedizin GmbH, in accordance with established procedures, directs and controls design activities. These procedures involve design and development planning, {6}------------------------------------------------ design input, design review, design verification/design validation, design transfer, as well as design change control. The activities provide the oversight and forum for project approval, formal management/design review, evaluation, and final project approval. This work is documented within design review and the approvals are documented as part of the current change control system. The ERBE ESU Model VIO dV, software version 3.1 for the RCB remote control complies to FDA's "Recognized Consensus Standards". Animal or clinical performance testing was not considered necessary. #### Conclusion: The software release version 3.1 and modification of the ERBE ESU Model VIO dV has the same intended use, principles of operation and technological characteristics as the predicate devices in the previously cleared 510(k)s. The subject device has been verified and validated in design control. There are no issues with the subject device that would raise additional safety or efficacy issues, when compared to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...