K133017 · Estech, Inc. · GEI · Oct 7, 2013 · General, Plastic Surgery
Device Facts
Record ID
K133017
Device Name
ESTECH COBRA(R) GENESIS BIPOLAR CLAMP
Applicant
Estech, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 7, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Estech COBRA® Genesis Bipolar Clamp is intended for the coagulation of soft tissue during general surgery. The system may also be used to coagulate blood and soft tissue to produce hemostasis.
Device Story
Estech COBRA® Genesis Bipolar Clamp is a sterile, single-use electrosurgical device; guides RF energy to target tissue via electrodes in clamp jaws. Operates in bipolar mode; one jaw serves as ablative electrode, other as indifferent electrode. Used by physicians in surgical settings to coagulate soft tissue and achieve hemostasis. Design features non-rotatable jaws and single active electrode for improved stability and ease of use. Device connects to an external RF generator (not described) to deliver energy. Clinical benefit includes effective tissue coagulation and hemostasis during general surgical procedures.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including mechanical design validation and ablation performance comparison against the predicate. Biocompatibility testing performed per ISO 10993. Electrical safety testing performed per IEC 60601-2-2.
Technological Characteristics
Bipolar electrosurgical clamp; RF energy delivery. Materials are biocompatible per ISO 10993. Electrical safety conforms to IEC 60601-2-2. Features non-rotatable jaws and single active electrode. Sterile, single-use device.
Indications for Use
Indicated for coagulation of soft tissue and hemostasis during general surgery in patients requiring such procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
K100224 — ESTECH COBRA BIPOLSR II · Endoscopic Technologies, Inc. D/B/A Estech · Apr 7, 2010
K023288 — COBRA BIPOLAR CLAMP, MODEL 15902 · Boston Scientific EP Technologies · Apr 2, 2003
K053100 — ESTECH COBRA BIPOLAR SYSTEM · Endoscopic Technologies, Inc. · Dec 19, 2005
K021046 — COBRA COOLED SURGICAL PROBE, MODEL 1596X · Boston Scientific Corp · Apr 16, 2002
K230432 — Single Use Bipolar Forceps · Hg Innovations, Ltd. · May 18, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
#### K133017
# SECTION 5: Special 510(k) Summary of Safety and Effectiveness
| A. Device Information: | |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Category | Comments |
| Sponsor: | Endoscopic Technologies, Inc., (dba<br>Estech)<br>2603 Camino Ramon<br>Suite 100<br>San Ramon, CA 94583<br>Tel: 925-543-2110 |
| Correspondent: | Tamer Ibrahim<br>Vice President<br>ESTECH, Inc |
| Contact Information: | Tel: 925-543-2110<br>Fax: 925-866-7117 |
| Device Common Name: | Electrosurgical cutting and coagulation<br>device and accessories |
| Device Proprietary Name: | Estech COBRA® Genesis Bipolar Clamp |
| Device Classification: | Class II, GEI (21 CFR 878.4400) |
### Predicate Device Information:
이
| Predicate Devices: | Estech COBRA Bipolar System II<br>(K100224) |
|-----------------------------------------|-------------------------------------------------------------------|
| Predicate Device Manufacturers: | Endoscopic Technologies, Inc (dba Estech) |
| Predicate Device Common Name: | Electrosurgical cutting and coagulation<br>device and accessories |
| Predicate Device Classification: | 21 CFR 878.4400 |
| Predicate Device Classification Number: | Class II, GEI |
# B. Date Summary Prepared
25 September 2013
# C. Description of Device
The Estech COBRA® Genesis Bipolar Clamp is a sterile, single-use device designed to guide RF energy to the target tissue via electrodes in the jaws of a surgical clamp style device. RF energy is delivered in a bipolar mode, with one jaw of the device as the ablative electrode and the other jaw as the indifferent electrode.
# D. Intended Use
The Estech COBRA® Genesis Bipolar Clamp is intended for the coagulation of soft tissue during general surgery. The system may also be used to coagulate blood und soft tissue to produce hemostasis.
{1}------------------------------------------------
# E. Comparison to Predicate Device
1
The Estech COBRA® Genesis Bipolar Clamp is identical in indications for use, technology, manufacturing, packaging and sterilization to the predicate ESTECH Cobra Bipolar II System (K100224).
There are minor design differences between the application and predicate devices. These design differences (non-rotatable jaws, single active electrode) provide improved stability and ease of use for the physician, but do not alter the performance profile seen in the predicate device.
# F. Summary of Supporting Data
The testing supplied in Section 18 demonstrates that the joints and other mechanical aspects of the design of the Estech COBRA® Genesis Bipolar Clamp are adequate for use. Additionally, the data demonstrates that the ablation performance of the Estech COBRA® Genesis Bipolar Clamp is equivalent to that of the predicate.
The patient contacting materials of the Estech COBRA® Genesis Bipolar Clamp have been demonstrated to be biocompatible, in accordance with ISO 10993: Biological Evaluation of Medical Devices, and the safety of the electrical design is in conformance with the pertinent sections of IEC 60601-2-2: Medical electrical equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment (see reference test reports in Sections 15 and 17).
#### G. Summary of Supporting Data
Estech concludes that the Estech COBRA® Genesis Bipolar Clamp is substantially equivalent to the predicate ESTECH Cobra Bipolar II System (K100224). The indications for use, technology, manufacturing, materials, packaging, and sterilization methods are identical to that of the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
October 7. 2013
Food and Drug Administration 10903 New Hampshife Avenue Document Control Center - WO06-G609 Silver Spring, MD 20993-0002
Endoscopic Technologies, Incorporated, dba Estech Mr. Tamer Ibrahim Vice President, R&D/RA/QA 2603 Camino Ramon, Suite 100 San Ramon, California 94583
Re: K133017
Trade/Device Name: Estech COBRA* Genesis Bipolar Clamp Regulation Number: 21 CFR 878,4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 25, 2013 Received: September 26, 2013
Dear Mr. Ibrahim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that I'DA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21) CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Tamer Ibrahim
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetw/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
- Mark N. Meikerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K133017
Device Name: Estech COBRA® Genesis Bipolar Clamp
Indications for Use:
The Estech COBRA® Genesis Bipolar Clamp is intended for the coagulation of soft tissue during general surgery. The system may also be used to coagulate blood and soft tissue to produce hemostasis.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
:
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|------------------------------|--------------------------------------------------------|
| Lona H. :<br>Chen -A | Page 1 of<br>for MXM |
| (Division Sign-Off) | |
| Division of Surgical Devices | |
| 510 (k) Number: K133017 | |
Section 4, Page 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.