K101961 · Laax, Inc. · GDW · Oct 29, 2010 · General, Plastic Surgery
Device Facts
Record ID
K101961
Device Name
TIGERPAW SYSTEM
Applicant
Laax, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Oct 29, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TigerPaw® System is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, without assistance from a camera, endoscope, or any other viewing technology. This includes procedures performed by sternotomy (full or partial) as well as thoracotomy (single or multiple).
Device Story
TigerPaw System is an implantable fastener and delivery tool for left atrial appendage (LAA) occlusion during open cardiac surgery. The implant consists of linearly spaced connectors overmolded with soft silicone to conform to anatomy. The disposable delivery tool features a pistol-grip handle with triggers that actuate jaws to mate the fastener sides together. Used in OR by surgeons under direct visualization (no cameras/endoscopes). The device provides mechanical occlusion of the LAA, intended to reduce risk of thromboembolic events. Clinical benefit includes complete LAA closure confirmed by TEE, preventing leakage or communication between LAA and left atrium.
Clinical Evidence
Prospective clinical study of 60 patients undergoing open cardiac surgery. Primary safety endpoint: device-related adverse events (AEs) at 30 days. Primary effectiveness endpoint: complete LAA occlusion confirmed visually and via TEE at 90 days. Results: 100% (60/60) visual confirmation of occlusion; 100% (54/54) TEE confirmation at 90 days. Zero incidences of bleeding or leakage. One minor device-related tissue tear repaired with suture. No unanticipated AEs or late device-related AEs reported.
Technological Characteristics
Implantable fastener composed of linearly spaced connectors overmolded with soft silicone. Delivery tool is a hand-held, disposable, pistol-grip mechanical device. Sterile, single-patient use. No electronic components or software.
Indications for Use
Indicated for occlusion of the left atrial appendage (LAA) in patients undergoing open cardiac surgical procedures (e.g., valve repair/replacement, CABG, cardiac ablation). Requires direct visualization without endoscopic assistance.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Synovis Surgical Innovations, Peri-Strips Staple Line Reinforcement (K040415)
Related Devices
K111064 — TIGERPAW SYSTEM II · Laax, Inc. · May 18, 2011
K190884 — TigerPaw Pro Left Atrial Appendage Occlusion Fastener And Delivery Tool · Datascope Corp. · Dec 26, 2019
K093679 — ATRICLIP LAA EXCLUSION SYSTEM WITH PRELOADED GILLINOV-COSGROVE CLIP, GILLINOV-COSGROVE SECTION GUIDE MODEL VARIOUS LAA0 · AtriCure, Inc. · Jun 10, 2010
K160454 — AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip · AtriCure, Inc. · Mar 18, 2016
K180010 — AtriClip FLEX-V · AtriCure, Inc. · Jan 31, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## 510(k) Number: K101961
#### Applicant Information:
Date Prepared: July 13, 2010
| Name: | LAAx Inc. |
|----------|--------------------------------------------------|
| Address: | 151 Lindbergh Ave, Suite I<br>Livermore CA 94551 |
| Phone: | 925-449-4108 |
| Contact Person: | Michael A. Daniel |
|-------------------|------------------------------------------|
| Phone Number: | Office: 925-254-5228 / Cell 415-407-0223 |
| Facsimile Number: | (925) 254-5187 |
#### Device Information:
| Classification: | Class II |
|----------------------|--------------------------------------------------------------------|
| Trade Name: | TigerPaw® System |
| Common Name: | Implantable Fastener and Accessories |
| Classification Name: | Staple, Implantable, GDW / Staple Line Reinforcement Material, FTL |
| Regulation Number: | 21 CFR 878.4750<br>21 CFR 870.3470 |
#### Predicate Devices:
The TigerPaw System is substantially equivalent in intended use and method of operation to a combination of the following predicate devices:
| USGI Auto Suture TA & GIA Staplers | K032696 |
|---------------------------------------------------------------------|---------|
| Synovis Surgical Innovations, Peri-Strips Staple Line Reinforcement | K040415 |
#### Device Description:
The TigerPaw System consists of an implantable Fastener and Delivery Tool designed for occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures. The Fastener is composed of linearly spaced connectors overmolded with a soft silicone. The spacing between the connectors along with the silicone gives the Fastener an ability to conform to the anatomy and completely exclude the LAA. The TigerPaw® System also includes a hand-held disposable Delivery Tool. The Delivery Tool
{1}------------------------------------------------
consists of a pistol grip handle with triggers that when actuated close jaws that in turn mate the two sides of the permanently implanted Fastener together.
### Intended Use:
The TigerPaw® System is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, without assistance from a camera, endoscope, or any other viewing technology. This includes procedures performed by sternotomy (full or partial) as well as thoracotomy (single or multiple).
#### Comparison to Predicate Device(s):
The design of the TigerPaw® System is similar to the USGI device in that they are devices designed to place titanium or stainless steel staples in various types of tissue. The TigerPaw® System is similar to the Synovis predicate in they both use a non-absorbable material to reinforce suture-lines and staple-lines. All devices are packaged sterile and are for single patient use. Further, the TigerPaw® System and the predicate devices have the same or similar intended use.
#### Device Evaluation:
Extensive bench, animal, and cadaver evaluations were performed on the TigerPaw System. This testing consisted of in vitro bench testing, including connector tensile testing and pressure leak testing of fasteners; in vivo animal evaluation composed of a total of six acute canine and six chronic canine survived for six months with corresponding histopathology; and in situ cadaver evaluation comprising 14 specimens with varying disease history including gross pathology. Finally, a 60 patient clinical study was conducted.
Clinical study: This study was conducted to assess the safety and effectiveness of surgical Left Atrial Appendage (LAA) occlusion, using the LAAx, Inc. TigerPaw System (delivery system and implant/Fastener), as a concomitant procedure to other open chest cardiac surgery procedures including valve repair / replacement, coronary artery bypass grafting (CABG), and cardiac ablation procedures. The primary safety outcome was the rate of device related adverse (AE) and serious adverse events (SAE) assessed peri-operatively, and at 30 days post procedure. The primary effectiveness outcome was the percentage of patients with complete occlusion of the LAA assessed peri-operatively, both visually (implant/Fastener applied fully across the LAA Os) and via Transesophageal Echocardiography (TEE), and at 90 days post procedure also using TEE. Secondary safety endpoints included Major Adverse Cardiac Event (MACE) rates. Secondary effectiveness endpoints included the assessment of the size of residual LAA cavity remaining after application of the TigerPaw® System compared with a pre-defined target (≤ 6 mm).
There were no unanticipated adverse events (AE) with one minor device-related tissue tear repaired with suture. Complete Left Atrial Appendage (LAA) occlusion was visually confirmed in each of the 60 treated patients and confirmed by Transesophageal Echocardiography (TEE) in 56 patients with one patient requiring adjunctive use of suture to achieve LAA closure. There were zero incidences of bleeding or leakage from the area of the device footprint. Fifty-four out
{2}------------------------------------------------
of 54 evaluable patients followed up at the final 90 day period were confirmed to have complete LAA occlusion via TEE with zero leaks or communication between the LAA and LA and no late device-related AEs.
## Summary:
Based upon the intended use, descriptive information, and the in vitro, in vivo, in situ, and clinical performance evaluation provided in this pre-market notification, the TigerPaw System has been shown to be substantially equivalent to currently marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
LAAx, INC c/o Mr. Michael A. Daniel Regulatory and Clinical Affairs 151 Lindbergh Avenue, Suite I Livermore CA 94551
OCT 2 9 2010
Re: K101961 LAAx Inc. TigerPaw® System Regulation Number: 21 CFR 870.4750 Regulation Name: Staple, Implantable Regulatory Class: Class II (two) Product Code: GOW Dated: September 13, 2010 Received: September 14, 2010
Dear Mr. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Mr. Michael A. Daniel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Dmma D. V. hmer
Image /page/4/Picture/8 description: The image shows a signature on the left side and the words "Branch" and "Director" on the right side. The signature appears to be a series of curved lines and loops. The text is aligned vertically, with "Branch" above "Director". The image is in black and white.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K101961
Device Name: TigerPaw® System
## Indications For Use:
The TigerPaw® System is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, without assistance from a camera, endoscope, or any other viewing technology. This includes procedures performed by stemotomy (full or partial) as well as thoracotomy (single or multiple).
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
sumer R. v. hines
(Division Sign-Off) (Division Sign-Olf)
Division of Cardiovascular Devices
510(k) Number_K10 | Jle
Page 1 of 1
.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.