TigerPaw Pro Left Atrial Appendage Occlusion Fastener And Delivery Tool
K190884 · Datascope Corp. · GDW · Dec 26, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190884
Device Name
TigerPaw Pro Left Atrial Appendage Occlusion Fastener And Delivery Tool
Applicant
Datascope Corp.
Product Code
GDW · General, Plastic Surgery
Decision Date
Dec 26, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TigerPaw Pro is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures. Direct visualization, in this context, requires that the surgeon is able to see the heart directly, without assistance from a camera, endoscope, etc., or any other viewing technology. This includes performed by sternotomy (full or partial) as well as thoracotomy (single or multiple).
Device Story
TigerPaw Pro is a flexible, implantable occlusion fastener and delivery tool for the left atrial appendage (LAA) ostium. Used during open cardiac surgery by surgeons; provides alternative to sutures/staplers. Device consists of a pre-loaded fastener (35mm or 45mm) and a delivery tool. Surgeon positions tool at LAA ostium; single-trigger mechanism with lock button and shaft rotator knob deploys fastener. Blunted needle barbs reduce tissue injury risk. Output is a secured LAA closure. Benefits include effective LAA occlusion during open heart procedures.
Clinical Evidence
No human clinical data. Evidence includes bench testing (tensile strength, closure integrity, corrosion resistance per ISO 10555-1, jaw flexure, rotational force testing) and a GLP acute animal study (n=7 canines, 14 fasteners). Animal study evaluated access, deployment, repositioning, and safety via epicardial/endocardial assessment and necropsy. Biocompatibility testing performed per ISO 10993.
Technological Characteristics
Implantable fastener and delivery tool. Materials: biocompatible per ISO 10993. Sterilization: Ethylene Oxide (ISO 11135-1:2014) to SAL 10^-6. Mechanical actuation: single-trigger delivery system with lock mechanism and shaft rotation. Fastener lengths: 35mm and 45mm. Corrosion testing per ISO 10555-1 Annex A.
Indications for Use
Indicated for occlusion of the left atrial appendage in patients undergoing open cardiac surgical procedures requiring direct visualization (sternotomy or thoracotomy).
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
December 26, 2019
Datascope Corp. Mark Dinger Sr. Regulatory Affairs Specialist 15 Law Drive Fairfield, New Jersey 07004
Re: K190884
Trade/Device Name: TigerPaw Pro Left Atrial Appendage Occlusion Fastener and Delivery Tool Regulation Number: 21 CFR 878.4750, 21 CFR 870.3470 Regulation Name: Implantable staple, Staple Line Reinforcement Material Regulatory Class: Class II Product Code: GDW, FTL Dated: April 2, 2019 Received: April 4, 2019
Dear Mr. Dinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190884
Device Name
TigerPaw Pro Flexible LAA Occlusion Fastener and Delivery Tool
Indications for Use (Describe)
The TigerPaw Pro is indicated for the occlusion of the left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.
Direct visualization, in this context, requires that the surgeon is able to see the heart directly, without assistance from a camera, endoscope, etc., or any other viewing technology. This includes performed by sternotomy (full or partial) as well as thoracotomy (single or multiple).
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## TigerPaw Pro Left Atrial Appendage Occlusion Fastener and Delivery Tool 510(k) Summary
Prepared in accordance with 21 CFR Part 807.92
| 510(k) Number: | K190884 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | 23 December 2019 |
| Submitter: | Datascope Corp.<br>15 Law Drive<br>Fairfield NJ 07004<br>United States of America |
| Contact Person: | Mark Dinger |
| Title: | Sr. Regulatory Affairs Specialist |
| Email: | mark.dinger@getinge.com |
| Phone: | 973-709-7691 |
| Fax: | 973-909-9954 |
| Trade Name: | TigerPaw Pro Left Atrial Appendage Occlusion Fastener and<br>Delivery Tool |
| Common Name: | Implantable Fastener and Accessories |
| Classification: | Class II |
| Classification Name: | Staple, Implantable, GDW<br>Staple Line Reinforcement Material, FTL |
| Regulation Numbers: | 21 CFR 878.4750<br>21 CFR 870.3470 |
| Predicate Device: | (K111064) TigerPaw System II (SE: 11 May 2011) |
| Device Description: | The TigerPaw Pro is a flexible, implantable occlusion fastener to<br>be placed on the ostium of the left atrial appendage (LAA) during<br>open cardiac surgical procedures. TigerPaw Pro consists of a<br>Fastener implant that is pre-loaded onto a Delivery Tool and is for<br>single use only. It is designed to provide an alternative to<br>conventional suture and surgical staplers for occlusion of the |
{4}------------------------------------------------
| | ostium of the LAA. The TigerPaw Pro implantable fastener comes<br>in two lengths; 35mm and 45mm. Surgeon judgment based on<br>patient anatomy should determine what length fastener to apply. |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use: | The TigerPaw Pro is indicated for the occlusion of the left atrial<br>appendage, under direct visualization, in conjunction with other<br>open cardiac surgical procedures. |
| | Direct visualization, in this context, requires that the surgeon is<br>able to see the heart directly, without assistance from a camera,<br>endoscope, etc., or any other viewing technology. This includes<br>procedures performed by sternotomy (full or partial) as well as<br>thoracotomy (single or multiple). |
| Technological<br>Characteristics | The TigerPaw Pro and the predicate device have<br>the following similarities: |
| | the same intended use, the same operating principles, incorporates the same basic design and materials, sterilized using the same materials and processes has same packaging. The TigerPaw Pro and the predicate device have the following<br>differences: |
| | Delivery Tool The Delivery Tool handle was redesigned. The predicate device has two triggers which are actuated in sequence to deploy the fastener. The TigerPaw Pro Delivery tool contains one trigger with a trigger lock button to deploy the fastener. The internal components, jaw, clamp box, roller pins and trigger, were modified to prevent incomplete closure of the TigerPaw Fastener on the LAA. New patient-contacting materials were introduced into the delivery tool. The design of the TigerPaw delivery tool was modified to include a trigger lock mechanism and shaft rotator knob to make the device more user |
{5}------------------------------------------------
| | friendly and to ensure the user performs the deployment steps in the correct sequence. |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • | Fastener |
| o | The tips of the fastener needle barbs were blunted to reduce the risk of tissue injury during the procedure |
| | The differences were made to improve device safety and are not considered technological differences that raise new questions of safety and effectiveness. |
| <b>Safety and Performance:</b> | Datascope Corp., development process completed the following activities to ensure the TigerPaw Pro met its clinically relevant design input requirements and specifications: |
| o | Performance testing, which included the following: |
| | o Ability to completely close the fastener after deployment and prevent leakage from a mock LAA |
| | o Tensile strength of the individual fastener barbs after closure |
| | o Assessment of TigerPaw Pro Delivery Tool jaw component characteristics, including angle measurements |
| | o Force required to actuate the Delivery Tool trigger, force required to unlock the Delivery Tool lock button, torque required to rotate the Delivery Tool shaft |
| | o Corrosion testing in accordance with ISO 10555-1 Annex A was performed on the delivery tool and fastener |
| | o A comparison of TigerPaw II and TigerPaw Pro's Tool jaw to resist flexure while deploying a fastener on a worst-case thickness mock LAA ostium |
| | o A comparison of TigerPaw II and TigerPaw Pro's ability to deploy and secure the fastener connectors under different worst-case rotational forces on a mock ostiumUsability testing on the TigerPaw Pro to obtain clinical user feedback and identify potential usability hazards |
| | o GLP animal study |
| | o An acute animal study was performed to evaluate the usability of the device and the risk of tissue injury on the LAA during deployment and securement of the fastener. This was demonstrated via an acute Good Laboratory Practices (GLP) animal study on seven canines. Atrial |
{6}------------------------------------------------
appendage closure was performed on the right and left atrial appendages on each animal for a total of 14 deployed fasteners. Operative procedures that were evaluated included the ability to access the LAA ostium with the Delivery Tool, assessment of the Delivery Tool and Fastener safety- pre-deployment, ability to reposition the Delivery Tool, deploy the Fastener and remove the Delivery Tool, epicardial assessment of Delivery Tool and Fastener safety and performance-post-deployment and endocardial assessment of Delivery Tool and Fastener safety and performance, post-deployment. A full necropsy examination of the local tissues post device implantation was also performed by a board-certified pathologist. All acceptance criteria were met for this animal study performed using the market version of this device. Biocompatibility Testing in accordance with ISO 10993 which O included cytotoxicity, sensitization, irritation, acute systemic toxicity, and material mediated pyrogenicity testing Sterilization - Sterilization validation was completed for the TigerPaw Pro to the requirements of ISO 11135-1:2014. Sterilization is performed with Ethylene Oxide to a Sterility Assurance Level (SAL) of 10-6. Residual limits are in accordance with ISO 10993-7:2008. The results of the in-vitro tests conducted demonstrate that the
functionality and performance characteristics of the device are comparable to the previously cleared TigerPaw System II.
Based upon the information submitted in this Traditional 510(k) Conclusion: premarket notification, Datascope's TigerPaw Pro is substantially equivalent to the predicate device. The TigerPaw Pro is similar to the predicate device in the intended use and the fundamental scientific technology of the device. The performance and other testing established that the TigerPaw Pro is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.