K211178 · Msi · GAW · Sep 2, 2022 · General, Plastic Surgery
Device Facts
Record ID
K211178
Device Name
DURAMESH Mesh Suture
Applicant
Msi
Product Code
GAW · General, Plastic Surgery
Decision Date
Sep 2, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DURAMESH™ is a permanent implant intended for single use by surgeons for the purpose of approximating soft tissues such as muscle, tendons, ligaments, and fascia with the exception of epithelial tissues such as skin, bowel, viscera and bladder mucosa. DURAMESH™ is not to be used in the closure of skin and other epithelial tissues, in the treatment of pelvic organ prolapse and stress urethral incontinence, as a dural substitute in intracranial or spinal surgery, in the closure of infected or dirty CDC Class IV wounds, or when an allergy to polypropylene is present. DURAMESH™ is not designed to be removed.
Device Story
DURAMESH™ is a polyfilament, non-absorbable, sterile surgical suture made of high molecular weight isotactic polypropylene. It consists of 12-18 braided filaments forming a hollow cylinder. The device is used by surgeons in clinical settings to approximate soft tissues (muscle, tendons, ligaments, fascia) during procedures like midline laparotomy. The surgeon uses an attached swaged needle to pass the suture through tissue and secures it with self-holding knots. The mesh structure allows for tissue ingrowth and incorporation. By providing a secure, permanent approximation of tissues, it facilitates healing in clean wounds. It is not intended for skin, epithelial, cardiovascular, ophthalmic, or neurological applications.
Clinical Evidence
Clinical trial of 53 patients (80 surgery sites). Primary endpoint: surgical site occurrence (SSO). Secondary endpoints: rehospitalization, return to OR, wound complications. Results: 0% early or late SSO; no device-related rehospitalizations or returns to OR. 15 subjects underwent 12-month follow-up, showing equivalent or improved outcomes compared to a cohort of 11 patients treated with standard sutures.
Technological Characteristics
Polyfilament, non-absorbable, synthetic suture; isotactic polypropylene polymer; Phthalocyaninato (2-) Copper dye. Hollow braided cylinder design (1.5-4mm diameter). Sterilized via Ethylene Oxide (AAMI TIR28:2016, ISO 11135:2014). Meets USP <871> (needle attachment) and USP <881> (tensile strength); does not meet USP <861> (diameter) due to oversized design.
Indications for Use
Indicated for general soft tissue approximation and/or ligation in clean (CDC Class I) wounds, including tendon repair and midline laparotomy where the bowel is not entered. Not for use in skin, epithelial surfaces, cardiovascular, ophthalmic, or neurological procedures.
Regulatory Classification
Identification
Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 2, 2022
MSi Greg Dumanian Medical Officer One Magnificent Mile 980 N. Lake Shore Drive, Suite 1400 Chicago, Illinois 60611
Re: K211178
Trade/Device Name: DURAMESH™M Mesh Suture Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable Polypropylene Surgical Suture Regulatory Class: Class II Product Code: GAW Dated: April 19, 2022 Received: April 20, 2022
Dear Greg Dumanian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Deborah Fellhauer, RN, BSN, CQIA Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211178
Device Name Duramesh™ Mesh Suture
#### Indications for Use (Describe)
Duramesh™ Mesh Suture is indicated for general soft tissue approximation and/or ligation in clean (CDC Class I) wounds, including tendon repair and midline laparotomy where the bowel is not entered. Duramesh™ is not for use in the skin and other epithelial surfaces, cardiovascular, ophthalmic, and/or neurological procedures.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b></b></span> | <span style="text-decoration: underline;"><b></b></span> |
| <div style="display:inline-block;"><div><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div></div> | <div style="display:inline-block;"><div><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div></div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### Contact Details
| Applicant Name | MSi |
|-----------------------------|-------------------------------------------------------------------------------------------------------------|
| Applicant Address | One Magnificent Mile<br>980 N. Michigan Ave.<br>Suite 1400<br>Chicago, IL 60611<br>United States of America |
| Applicant Contact Telephone | 202 737 9633 |
| Applicant Contact | Jeffrey Shapiro |
| Applicant Contact Email | JShapiro@hpm.com |
| Date Prepared | September 2, 2022 |
| Device Name | |
| Device Trade Name | Duramesh™ Mesh Suture |
| Common Name | Surgical Suture |
| Classification Name | Suture, Nonabsorbable, Synthetic, Polypropylene |
| Regulation Number | 878.5010 |
| Product Code | GAW |
## Legally Marketed Predicate Device
| Predicate # | Predicate Trade Name |
|-------------|----------------------------------|
| K131224 | SharpointTM polypropylene suture |
#### Indications for Use
DURAMESH™ Mesh Suture is indicated for general soft tissue approximation and/or ligation in clean (CDC Class I) wounds, including tendon repair and midline laparotomy where the bowel is not entered. DURAMESH™ is not for use in the skin and other epithelial surfaces, cardiovascular, ophthalmic, and/or neurological procedures.
{4}------------------------------------------------
### Device Description
DURAMESH™ is a polyfilament, synthetic, non-absorbable, sterile surgical suture composed of isotactic polypropylene polymer of high molecular weight. The suture is provided dyed (blue) and is uniform in appearance. The blue dye pigment is Phthalocyaninato (2-) Copper. The product meets all requirements established by the United States Pharmacopeia (USP) for nonabsorbable surgical suture, except for diameter.
DURAMESH™ is passed on either side of tissues using an attached swaged needle as an introducing agent and tied to itself with at least two knots composed of two throws each to maintain tension.
DURAMESH™ is supplied sterile and is available with two needle configurations (small and large) in four USP sizes (2-0 through 2), each with a length of approximately 36 inches (91 cm). Each box contains one DURAMESH™ housed in a Tyvek envelope within which the DURAMESH™ is packaged in a sterile protective medical grade tube fixed to a sterile card.
#### Intended Purpose
DURAMESH™ is a permanent implant intended for single use by surgeons for the purpose of approximating soft tissues such as muscle, tendons, ligaments, and fascia with the exception of epithelial tissues such as skin, bowel, viscera and bladder mucosa. DURAMESH™ is not to be used in the closure of skin and other epithelial tissues, in the treatment of pelvic organ prolapse and stress urethral incontinence, as a dural substitute in intracranial or spinal surgery, in the closure of infected or dirty CDC Class IV wounds, or when an allergy to polypropylene is present. DURAMESH™ is not designed to be removed.
| | Subject Device | Predicate Device (K131224) |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | DURAMESH™ | Sharpoint™ polypropylene suture |
| Indications for use | DURAMESH™ Mesh Suture is indicated<br>for general soft tissue approximation and/or<br>ligation in clean (CDC Class I) wounds,<br>including tendon repair and midline<br>laparotomy where the bowel is not entered.<br>DURAMESH™ is not for use in the skin<br>and other epithelial surfaces,<br>cardiovascular, ophthalmic, and/or<br>neurological procedures. | Sharpoint™ polypropylene suture is<br>indicated for general soft tissue<br>approximation and/or ligation, including use<br>in cardiovascular, ophthalmic and<br>neurological procedures. |
| Design, needle | Grade 470 surgical needle with standardized<br>shapes and sizes | Grade 470 surgical needle with standardized<br>shapes and sizes |
| Design, suture<br>body | Polyfilament suture comprised of 12 to 18<br>polypropylene filaments loosely braided to<br>form a hollow cylinder with outer wall 1.5<br>mm to 4 mm in diameter. Diameter changes<br>with lateral or axial tension. | Monofilament suture with filament diameter<br>ranging from 0.03 mm to 0.5 mm. Diameter<br>does not change with lateral or axial tension |
| Natural / Synthetic | Synthetic | Synthetic |
### Comparison Table
{5}------------------------------------------------
| Material | Isotactic polypropylene polymer of high<br>molecular weight, Phthalocyaninato (2-)<br>Copper | Isotactic polypropylene polymer of high<br>molecular weight, Phthalocyaninato (2-)<br>Copper |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide |
| Available Size<br>Range | USP 2-0 through 2 (oversized) | USP 10-0 through 1 |
| Specifications and<br>Properties | Each size exceeds USP minimums for knot-<br>pull tensometry (USP <881>) and needle<br>attachment tensometry (USP <871>). Does<br>not follow USP for diameter (USP <861>);<br>the suture is oversized in diameter. | Each size exceeds USP minimums for knot-<br>pull tensometry (USP <881>) and needle<br>attachment tensometry (USP <871>). Does<br>not follow USP for diameter (USP <861>) |
| Principle of<br>Operation | The implant is threaded through two or<br>more separated tissues with the aid of a<br>surgical needle driver. The surgical needle<br>pierces the tissue to allow the trailing suture<br>body to be threaded. The surgeon applies<br>axial tension on the implant thereby<br>approximating the separated tissues.<br>Tension is maintained with a self-holding<br>knot created by the surgeon. | The implant is threaded through two or<br>more separated tissues with the aid of a<br>surgical needle driver. The surgical needle<br>pierces the tissue to allow the trailing suture<br>body to be threaded. The surgeon applies<br>axial tension on the implant thereby<br>approximating the separated tissues.<br>Tension is maintained with a self-holding<br>knot created by the surgeon. |
| Absorption /<br>residual strength | Polypropylene suture is not absorbed, nor is<br>there any significant change in tensile | Polypropylene suture is not absorbed, nor is<br>there any significant change in tensile |
| profile | strength retention known to occur in vivo. | strength retention known to occur in vivo. |
DURAMESH™ has the same principles of operation as its predicate device and achieves soft tissue approximation and/or ligation with substantial equivalence to its predicate. Like its predicate. DURAMESH™ is introduced through tissues with an attached needle in order to achieve soft tissue approximation and/or ligation. Like its predicate, DURAMESH™ is tied with self-holding knots to maintain the required tension for tissue approximation. DURAMESH™ is made of polypropylene and is identical to the predicate in its synthetic material of fabrication and material chemical composition. Both DURAMESH™ and its predicate are sterilized with Ethylene Oxide, are not absorbed, and have no known significant changes in tensile strength occurring in vivo. The DURAMESH™ Indications for use specify use in clean wounds and exclude use in the skin and other epithelial surfaces, cardiovascular, ophthalmic, and/or neurological procedures while the predicate indications for use do not. DURAMESH™ differs from its predicate device in its design: it is a polyfilament mesh suture composed of multiple braided predicate filaments separated by air, while its predicate is a solid monofilament suture. The DURAMESH™ outer diameter is oversized compared to the predicate's, and changes with axial or lateral tension while the predicate diameter does not change. DURAMESH™ is available in USP sizes 2-0 through 2 with oversized outer diameters while the predicate is available in USP sizes 10-0 through 1 and has USP <861> - compliant outer diameters.
### Non-Clinical Data
### Biocompatibility
A biological risk assessment of DURAMESH™ was performed according to the requirements of the current ISO 10993-1, ISO 14971, and the 2020 FDA Biocompatibility Guidance. Based on cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, rabbit pyrogen,
{6}------------------------------------------------
genotoxicity, muscle implantation, and chemical characterization studies, DURAMESH™ was shown to be biocompatible per ISO 10993-1 for use as intended.
### Packaging
Transport testing for DURAMESH™ in the final packaging and shipping configuration was conducted according to ASTM D4169-16, standard practice for performance testing of shipping containers and systems. Protection of the primary container (sterility) was confirmed via ASTM F1929 - 15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration.
### Sterilization
DURAMESH™ is sterilized with Ethylene Oxide following AAMI TIR28:2016 guidelines [Product Adoption And Process Equivalence For Ethylene Oxide Sterilization]. The adoption standard augments ISO 11135:2014 |Sterilization of health-care products -- Ethylene oxide --Requirements for the development, validation and routine control of a sterilization process for medical devices]. DURAMESH™ meets the ISO 10993-7: 2008 limits for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) for implanted devices.
## Shelf-Life
Accelerated and real-time aging stability testing was completed to support a 5-year shelf for DURAMESH™. Data support no loss of tensile strength over the proposed shelf life.
## Bench Data
DURAMESH™ passes prospectively defined acceptance criteria for needle attachment strength and tensile strength per FDA-recognized consensus standards USP43-NF38: 2020 <871> Sutures - Needle Attachment and USP43-NF38: 2020 <881> Sutures - Tensile Strength, respectively. DURAMESH™ diameter decreases with increasing axial load per Diameter vs. Load attribute testing. In Knot Size and Knot Strength testing, DURAMESH™ knots were shown to be stronger than predicate knots for the same number of throws. Testing shows that DURAMESH™ knot security is achieved with 4 throws as recommended in DURAMESH™M labeling. DURAMESH™ passes prospectively defined acceptance criteria for tensile strength, per USP requirements, under three separate loading conditions: simulated surgical deployment, cycle loading, and dynamic loading. An internal company study demonstrated that "stripped" samples of DURAMESH™ (with a loss of bonds between filaments) maintain adequate tensile strength to satisfy USP requirements.
### Animal Data
Two large-animal porcine implantation studies for midline incision closure of 1- and 3-month durations were performed with endpoints of adverse tissue reaction (manifesting as excessive fibrosis, bowel adhesions, and infection) and dehiscence. In the 1- month and 3-month porcine studies, DURAMESH™ was able to close a midline incision with either a running or interrupted
{7}------------------------------------------------
suturing technique. DURAMESH™ exhibited tissue ingrowth and implant incorporation and was a slight irritant in comparison to standard polypropylene suture on semi-quantitative histology.
### Clinical Data
Implantation of DURAMESH™ was performed in a single-investigator clinical trial of 53 patients with 80 individual surgery sites treated with DURAMESH™. The primary outcome was surgical site occurrence (SSO). Secondary outcomes included subject device related rehospitalization and return to the operating room, and wounds for any reason.
Fifteen of the total 53 DURAMESH™ subjects were brought back for a long-term assessment (12-month) of healing, function, pain, and quality of life. The remaining 38 subjects did not have evaluation at the 12-month endpoint. The 15 subjects were compared to a cohort of 11 patients who underwent similar orthopedic procedures for cerebral palsy with standard sutures.
The primary outcome showed no early or late SSO. Secondary outcomes showed no DURAMESH™ related rehospitalization or return to the operating room, and no wounds for any reason. One-year cohort comparison of DURAMESH™ to non-DURAMESH™ patients demonstrated equivalent or slightly improved outcomes with DURAMESH™.
### Conclusion
All data and information lead to the conclusion that, despite the technological difference of polyfilament mesh suture versus monofilament suture design, DURAMESH™ is substantially equivalent to the predicate device for the proposed indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.