K981322 · Genzyme Surgical Product (Genzyme Biosurgery) · GAR · Jul 7, 1998 · General, Plastic Surgery
Device Facts
Record ID
K981322
Device Name
CUTALON NYLON POLYAMIDE SURGICAL SUTURE
Applicant
Genzyme Surgical Product (Genzyme Biosurgery)
Product Code
GAR · General, Plastic Surgery
Decision Date
Jul 7, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cutalon Nonabsorbable Polyamide Surgical Sutures are provided sterile and are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
Device Story
Cutalon is a sterile, wet-packed, extruded nylon monofilament (Nylon 6) surgical suture; designed for general soft tissue approximation and ligation. Suture is packaged in a solution containing isopropyl alcohol, propylene glycol, water, glycerin, lactic acid, diethanolamine, sodium nitrite, and ammonia to maintain suppleness and ease of handling. Device is provided on a plastic winder within a sealed polyethylene blister and Tyvek pouch; sterilized via Gamma Irradiation. Used by surgeons in clinical settings (OR) for cardiovascular, ophthalmic, and neurological procedures. Wet packaging facilitates easier twisting and tying compared to dry-packed nylon sutures. Clinical benefit derived from reliable tissue approximation and improved handling characteristics.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing, including USP muscle implant studies (1 week and 90 days), cytotoxicity, sensitization, intracutaneous testing, acute systemic toxicity, hemocompatibility, and genotoxicity. All results met USP and ISO 10993-1 requirements for nonabsorbable nylon sutures.
Technological Characteristics
Extruded Nylon 6 monofilament; wet-packed in solution (isopropyl alcohol, propylene glycol, water, glycerin, lactic acid, diethanolamine, sodium nitrite, ammonia). USP sizes 6-0 to 0. Sterilized via Gamma Irradiation (Cobalt 60) per AAMI/ANSI/ISO 11137 to SAL 10^-6. Meets USP requirements for nonabsorbable sutures.
Indications for Use
Indicated for general soft tissue approximation and/or ligation in patients requiring surgical closure, including cardiovascular, ophthalmic, and neurological procedures.
Regulatory Classification
Identification
Nonabsorbable polyamide surgical suture is a nonabsorbable, sterile, flexible thread prepared from long-chain aliphatic polymers Nylon 6 and Nylon 6,6 and is indicated for use in soft tissue approximation. The polyamide surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. monograph for nonabsorbable surgical sutures; it may be monofilament or multifilament in form; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
JUL 7 1998
#### Section 2 510(k) Summary
# 510(k) Summary Genzyme "Cutalon" Nonabsorbable Polyamide Surgical Suture
#### 1. DATE PREPARED
April 7, 1998
### 2. SPONSOR INFORMATION
| Address: | Genzyme Surgical Products<br>600 Airport Road<br>Fall River, Massachusetts 02720 |
|------------|----------------------------------------------------------------------------------|
| Telephone: | 508-677-6679 |
| Contact: | Jim Kenney<br>Director, Quality Assurance/Regulatory Affairs |
#### 3. DEVICE NAME
Proprietary Name: "Cutalon" Nonabsorbable Polyamide Surgical Suture Common/Usual Name: Nylon Nonabsorbable Polyamide Suture Classification Name: Nylon Nonabsorbable Polyamide Suture
### 4. DEVICE DESCRIPTION AND INTENDED USE
"Cutalon" Nonabsorbable Polyamide Surgical Sutures are composed of the long-chain aliphatic polymers extruded Nylon (Nylon 6) sutures which conform to the United States Code of Federal Regulations, 21 CFR Part, 878.5020.
"Cutalon" Nonabsorbable Polyamide Surgical Sutures are provided sterile and are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
{2}------------------------------------------------
# 5. COMPARISON TO PREDICATE DEVICE(S)
"Cutalon" Nonabsorbable Polyamide Surgical Sutures are identical in design, material, manufacturing process, function, and intended use to Nylon (Nylon 6) Nonabsorbable Surgical Sutures which are have been marketed by the Sponsor Company since 1976.
Section 3 Proposed Labeling
Draft proposed labeling for the "Cutalon" Nonabsorbable Polyamide Surgical Suture includes the actual device labeling and the package insert. Sample draft labeling for the device is provided in Appendix A.
{3}------------------------------------------------
### Section 4 Device Description
"Cutalon" is a wet-packed, extruded, nylon monofilament suture. The base polymer material is Nylon 6. (As required in CFR 878.5020) The nylon suture material is foilpackaged "wet" in a solution containing isopropyl alcohol, water and other components (see below description of formulation).
The packaging fluid contains: Isopropyl Alcohol 73.65% 1.2-Propylene Glycol 15.00% Water 8.25% Glycerin 2.30% Lactic Acid 0.45% Diethanolamine 0.20% Sodium Nitrite 0.10% Ammonia 0.05%
The purpose of the "wet" packaging is to provide a nylon suture that is supple (easier to twist and tie) straight from the package due to the water absorption into the nylon.
Genzyme "Cutalon" Surgical Suture are available in USP sizes 6-0 through 0 (metric sizes 0.7 through 3.5). Genzyme Cutalon Sutures meet all requirements established by the United States Pharmacopeia (USP) for nonabsorbable sutures except for diameter.
The "Cutalon" Suture is identical in design, materials (Nylon 6), manufacturing process and intended use to the Nylon Suture already marketed by Genzyme Surgical Products. This "Cutalon" Suture is simply a change in the packaging.
Genzyme's Nylon Suture has been on the market since December 1976. In Appendix D, please see a copy of NDA 85-060 FDA letter from December 1976 and also see a copy of the FDA's 510(k) letter dated July 26, 1994 which confirms that Genzyme has followed the instructions for "Suture Labeling Guidance" dated October 9, 1992.
{4}------------------------------------------------
## Design, Materials, Process, Packaging
The "Cutalon" Suture is designed to maintain sterility of the product until the time of use. Each individual suture/needle assembly is wrapped around a plastic winder and the combination is placed inside an open-ended pouch. The suture assemblies in their pouches are then loaded into the packaging machine. This machine forms a polyethylene "blister" and seals a medical grade paper lid stock onto it after it is filled with packaging fluid and product. The sealed blister is placed into a purchased film/Tyvek pouch and sealed on a bar sealer. (The bar sealers were validated for outer pouch sealing for the "wet" products and a packaging machine was validated for wet packed inner blister packaging. The pouch is designated to allow sterilization with the pouch and maintain a sterile environment until the pouch is opened. The Cutalon Nylon suture is foil-packaged "wet" in a solution containing isopropyl alcohol, water and other components. The purpose once again of the "wet" packaging is to provide a Nylon suture that is supple (easier to twist and tie) straight from the package due to the water absorption into the nylon.
The "wet" pouches are placed into solid bleached sulfite board boxes, with instructions for use. Polyamide suture packages are opened by peeling open the pouch. The loaded carrier is then aseptically opened in the sterile field. There is a clear film overwrap applied to the shelf box as a dust cover. Outer corrugate shipping cartons of various sizes are used to ship the suture boxes.
# STERILIZATION:
| a. Sterilization Method: | Gamma Irradiation (Cobalt 60) |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| b. Validation Method: | AAMI/ANSI/ISO 11137 Sterilization of health care<br>products-Requirements for validation and routine<br>control-Radiation sterilization. |
| c. Sterility Assurance<br>Level | $10^{-6}$ |
{5}------------------------------------------------
#### Substantial Equivalence and Comparison Information Section 5
The "Cutalon" Nonabsorbable Polyamide Surgical Suture is identical in design, materials, manufacturing process, function and intended use to the predicate device, Genzyme's Nylon Suture.
Genzyme's Nylon Suture has been on the market since December 1976. In Appendix C, please see a copy of NDA 85-060 FDA letter from December 1976 and also see a copy of the FDA's 510(k) #K930738 letter dated July 26, 1994 which confirms that Genzyme has followed the instructions for "Suture Labeling Guidance" dated October 9, 1992.
| Table 1<br>"Cutalon" Nylon Sutures Compared to Predicate Suture Product | | |
|-------------------------------------------------------------------------|---------------------------------------------------|---------------------------------------------------|
| | "Cutalon" Suture | Nylon Suture |
| Intended Use | General soft tissue approximation and/or ligation | General soft tissue approximation and/or ligation |
| Suture Material | Nylon 6 or 6.6 | Nylon 6 or 6.6 |
| Sterilization Method | Gamma Irradiation | Gamma Irradiation or ETO |
| How provided | Sterile for Single Use Only | Sterile for Single Use Only |
| USP Sizes | 6-0 thru 0 | 10-0 thru 5 |
| Suture Tensile Strength | Meets USP Requirement | Meets USP Requirement |
| Packaging | Wet Packaged (sterile) | Dry Packaged (sterile) |
{6}------------------------------------------------
#### Section 6 Biocompatibility Information
Biocompatibility testing was performed on Genzyme "Cutalon" Nylon Surgical Sutures which satisfied the requirements of ISO 10993-1 Biological Evaluation of Medical Devices. The biocompatibility tests conducted were:
< USP Muscle Implant (1 week)
USP Muscle Implant (90 days)
USP Physico-Chemical
<Sensitization (saline and CSO)
USP Intracutaneous Testing (saline and CSO)
/ Acute Systemic Toxicity
Hemocompatibility
- Cytotoxicity (USP Elution Method)
Genotoxicity (DMSO extract)
- Genotoxicity (Saline extract)
The results of the biocompatibility testing of Genzyme "Cutalon" Nylon Surgical Sutures show the sutures to be non-cytotoxic, non-hemolytic, non-allergenic, non-mutagenic, nonirritant, meeting requirements for systemic toxicity, meeting implant requirements, and meeting USP requirements for long-chain aliphatic polymer Nylon 6. Reference Appendix D for copies of the actual test reports.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Genzyme Surgical Products Mr. Stephen M. Page Director of Regulatory Affairs 600 Airport Road Fall River, Massachusetts 02720
7 1998 JUL
Re: K981322
Trade Name: "Cutalon" Nylon Polyamide Surgical Suture Regulatory Class: II Product Code: GAR Dated: April 8, 1998 Received: April 10, 1998
Dear Mr. Page:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices that were regulated as transitional devices and that have been reclassified into class II. Notice of this reclassification was published in the Federal Register on Friday, May 31, 1991 (Vol. 56, No. 105, Pages 24684 and 24685). A copy of this Federal Register can be obtained by calling the Division of Small Manufacturers Assistance (DSMA) at (800) 638-2041 or (301) 443-6597. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act) and the following limitations:
- 1. The "Cutalon" Nylon Polyamide Surgical Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
- 2. This device may not be manufactured from any long chain aliphatic polymers other than nylon 6 and/or nylon 6,6. In addition, you must maintain documentation at your premises regarding vendor certification for raw or semiprocessed source material, all manufacturing and quality control release procedures, and validation of sterilization procedures used in the manufacture of the "Cutalon" Nylon Polyamide surgical suture. Any deviation of the source material or processing as described in this 510(k) notification requires submission of a new premarket notification and Food and Drug Administration (FDA) clearance prior to commercial distribution of the modified device.
{8}------------------------------------------------
Page 2 - Mr. Stephen Page
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
[signature]
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a series of numbers and a letter written in a stylized, calligraphic font. The characters appear to be 'K981322'. The characters are bold and black, contrasting with the white background. The style of writing gives the impression of a handwritten or artistic representation of the alphanumeric sequence.
Device Name: "Cutalon" Nonabsorbable Polyamide Surgical Suture
Indications for Use:
(Per 21 CFR 801.109)
"Cutalon" Nonabsorbable Polyamide Surgical Sutures are provided sterile and are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
These indications for use are exactly the same as for the device we have on the market, Nylon Nonabsorbable Polyamide Surgical Sutures. The Nylon suture was approved for marketing by way of PreMarket Approval Application (PMA) # N85-060 "Approval Letter" for the original PMA dated July 26, 1976.
The only change between "Cutalon" and Nylon is the method of packaging. Nylon is dry packaged. "Cutalon" is wet packaged. [The difference is explained in detail in the body of this 510(k) ].
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE) (201.108 AFF 801.101 Prescription Use Genzyme Surgical Products of General Resto 600 Airport Road (O(k) Number Fall River, Massachusetts 02720-4740 Prescription Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.