The ECHELON FLEX™, ECHELON™ 3000, and ECHELON ENDOPATH™ families of staplers and reloads are intended for transection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials. The instruments may also be used for transection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
Device Story
ECHELON 3000 is a sterile, single-patient-use, battery-powered surgical stapler; performs simultaneous tissue cutting and stapling. Input: manual user operation via handle/controls; device uses embedded software to manage motor firing speed and articulation. Output: six staggered rows of staples (three per side of cut line). Used in OR by surgeons for open or minimally invasive procedures. Features include rotating shaft, articulation mechanism for lateral access, haptic alerts, and multifunctional home button. Benefits: facilitates tissue management in complex surgeries; provides consistent staple line integrity. Healthcare providers use visual and haptic feedback to monitor firing; output directly affects surgical outcomes by ensuring hemostasis and tissue approximation.
Clinical Evidence
No clinical data provided. Substantial equivalence demonstrated via bench testing (staple height, line integrity, strength), animal studies (acute hemostasis, tissue healing), human factors/usability testing, and electrical/EMC safety testing.
Technological Characteristics
Powered, single-use, handheld surgical stapler. Materials: biocompatible. Energy: primary lithium battery. Features: rotating/articulating shaft, motorized firing, haptic feedback. Software: embedded control system for firing and articulation. Sterilization: not specified. Connectivity: none.
Indications for Use
Indicated for patients undergoing open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures requiring tissue transection, resection, or anastomosis creation, including liver parenchyma, pancreas, kidney, and spleen.
Regulatory Classification
Identification
Surgical stapler for external use: A specialized prescription device used to deliver compatible staples to skin during surgery. Surgical stapler for internal use: A specialized prescription device used to deliver compatible staples to internal tissues during surgery for resection, transection, and creating anastomoses.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(i) Performance testing must demonstrate that the stapler, when used with compatible staples, performs as intended under anticipated conditions of use. Performance testing must include the following:
(A) Evaluation of staple formation characteristics in the maximum and minimum tissue thicknesses for each staple type;
(B) For manual staplers only, measurement of the worst-case deployment pressures on stapler firing force;
(C) Measurement of staple line strength;
(D) Confirmation of staple line integrity; and
(E) In vivo confirmation of staple line hemostasis.
(ii) For powered staplers only, appropriate analysis/testing must demonstrate the electromagnetic compatibility and electrical, thermal, and mechanical safety of the device.
(iii) For powered staplers only, appropriate software verification, validation, and hazard analysis must be performed.
(iv) Human factors testing must demonstrate that the clinician can correctly select and safely use the device, as identified in the labeling, based on reading the directions for use.
(v) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
(vi) Performance data must demonstrate the sterility of the device.
(vii) Validation of cleaning and sterilization instructions must demonstrate that any reusable device components can be safely and effectively reprocessed per the recommended cleaning and sterilization protocol in the labeling.
(viii) Performance data must support the shelf life of the device by demonstrating continued device functionality, sterility, and package integrity over the identified shelf life.
(ix) Labeling of the device must include the following:
(A) Unless data demonstrates the safety of doing so, contraindications must be identified regarding use of the device on tissues for which the risk of stapling outweighs any reasonably foreseeable benefit due to known complications, including the stapling of tissues that are necrotic, friable, or have altered integrity.
(B) Unless available information demonstrates that the specific warnings do not apply, the labeling must provide appropriate warnings regarding how to avoid known hazards associated with device use including:
(
*1* ) Avoidance of use of the stapler to staple tissue outside of the labeled limits for maximum and minimum tissue thickness;(
*2* ) Avoidance of obstructions to the creation of the staple line and the unintended stapling of other anatomic structures;(
*3* ) Avoidance of clamping and unclamping of delicate tissue structures to prevent tissue damage;(
*4* ) Avoidance of use of the stapler on the aorta;(
*5* ) Establishing proximal control of blood vessels prior to stapling where practical and methods of blood vessel control in the event of stapler failure;(
*6* ) Ensuring stapler compatibility with staples; and(
*7* ) Risks specifically associated with the crossing of staple lines.(C) Specific user instructions for proper device use including measures associated with the prevention of device malfunction, and evaluation of the appropriateness of the target tissue for stapling.
(D) List of staples with which the stapler has been demonstrated to be compatible.
(E) Identification of key performance parameters and technical characteristics of the stapler and the compatible staples needed for safe use of the device.
(F) Information regarding tissues on which the stapler is intended to be used.
(G) Identification of safety mechanisms of the stapler.
(H) Validated methods and instructions for reprocessing of any reusable device components.
(I) An expiration date/shelf life.
(x) Package labels must include critical information and technical characteristics necessary for proper device selection.
Predicate Devices
ECHELON FLEXTM 45mm Powered Plus Articulating Endoscopic Linear Cutters (K163454)
ECHELON FLEXTM 60mm Powered Plus Compact Articulating Endoscopic Linear Cutters (K160521)
ECHELON FLEXTM 60mm Powered Plus Articulating Endoscopic Linear Cutters (K140560)
Reference Devices
ECHELON FLEX™ Powered Plus Articulating Endoscopic Linear Cutters (K202665)
ECHELON ENDOPATH™ Endoscopic Linear Cutter Reloads, 45mm (K163454)
ECHELON ENDOPATH™ Endoscopic Linear Cutter Reloads, 60mm (K183435)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 8, 2022 Ethicon Endo-Surgery, LLC Ekta Patel Senior Regulatory Affairs Specialist 4545 Creek Road Blue Ash, Ohio 45242
### Re: K213633
Trade/Device Name: ECHELON 3000 45mm Stapler, ECHELON 3000 60mm Stapler Regulation Number: 21 CFR 878.4740 Regulation Name: Surgical Stapler Regulatory Class: Class II Product Code: GAG Dated: March 8, 2022 Received: March 9, 2022
### Dear Ekta Patel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Trumbore Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213633
Device Name ECHELON™ 3000
### Indications for Use (Describe)
The ECHELON FLEX™, ECHELON™ 3000, and ECHELON ENDOPATH™ families of staplers and reloads are intended for transection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials. The instruments may also be used for transection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
Type of Use (Select one or both, as applicable)
| | <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--|-----------------------------------------------------------------------------------------------------------|
| | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
#### I. SUBMITTER
- Company: Ethicon Endo-Surgery, LLC 475 Calle C Guaynabo, PR 00969
- Contact: Ekta Patel Senior Regulatory Affairs Specialist Ethicon Endo-Surgery, Inc. Phone: 513-337-1124 Email: epatelho@its.jnj.com
Date Prepared: November 15, 2021
#### II. SUBJECT DEVICES
### Trade Names:
- ECHELON™ 3000 45mm Stapler ●
- ECHELON™ 3000 60mm Stapler ●
| Common or Usual Name: | Surgical Stapler |
|-----------------------|------------------------------------|
| Classification Name: | Surgical Stapler (21 CFR 878.4740) |
| Regulatory Class: | II |
| Product Code: | GAG |
#### PREDICATE DEVICES III.
| Predicate Device<br>510(k) Number | Predicate Device Name | Predicate Device<br>Product Codes |
|-----------------------------------|------------------------------------------------------------------------------------|-----------------------------------|
| K163454 | ECHELON FLEXTM 45mm Powered Plus Articulating<br>Endoscopic Linear Cutters | PSEE45A, PLEE45A,<br>PCEE45A |
| K160521 | ECHELON FLEXTM 60mm Powered Plus Compact<br>Articulating Endoscopic Linear Cutters | PCEE60A |
| K140560 | ECHELON FLEXTM 60mm Powered Plus Articulating<br>Endoscopic Linear Cutters | PSEE60A and<br>PLEE60A |
These predicates have not been subjected to a recall related to these design modifications.
{4}------------------------------------------------
Reference device(s):
- . K-number K202665: references 45mm and 60mm ECHELON FLEX™ Powered Plus Articulating Endoscopic Linear Cutters, cleared under a Special 510(k) Premarket Notification.
*Both K202665 and current subject device ECHELON™ 3000 has the same predicate device
- K-numbers K163454 and K183435: references 45mm and 60mm ECHELON . ENDOPATH™ (reloads) which are used in conjugation with ECHELON™ 3000 45 mm and 60mm Stapler respectively. There are no modifications to the reloads; this 510(k) submission is associated with the ECHELONTM 3000 stapler only.
| Reference Device K-Number | Reference Device Name | Reference Device Product Codes |
|---------------------------|------------------------------------------------------------------------------------------|----------------------------------------|
| K163454 | ECHELON ENDOPATH™ Endoscopic Linear Cutter Reloads, 45mm (+ Gripping Surface Technology) | GST45B, GST45D, GST45G, GST45T, GST45W |
| K183435 | ECHELON ENDOPATH™ Endoscopic Linear Cutter Reloads, 60mm (+ Gripping Surface Technology) | GST60B, GST60D, GST60G, GST60T, GST60W |
Table 5-1: Reference devices* (Reloads) used with the Subject Devices (Stapler)
*No change to stapler reloads
| | | | Table 5-2: Reload codes and corresponding reload color | |
|--|--|--|--------------------------------------------------------|--|
| | | | | |
| Reloads Codes for<br><b>45mm</b> | Reload Codes for 60mm | Corresponding Reload Color |
|----------------------------------|-----------------------|----------------------------|
| GST45B | GST60B | Blue |
| GST45D | GST60D | Gold |
| GST45G | GST60G | Green |
| GST45T | GST60T | Black |
| GST45W | GST60W | White |
*No change to stapler reloads
{5}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The ECHELONTM 3000 45 mm and 60 mm Staplers are sterile, single-patient-use instruments that simultaneously cut and staple tissue. There are six staggered rows of staples, three on either side of the cut line.
The ECHELON™ 3000 45 mm Staplers have a staple line that is approximately 45 mm long and a cut line that is approximately 42 mm long.
The ECHELON™ 3000 60 mm Staplers have a staple line that is approximately 60 mm long and a cut line that is approximately 57 mm long.
The shaft can rotate freely in both directions and an articulation mechanism enables the distal portion of the shaft to pivot to facilitate lateral access to the operative site.
The instruments are packaged with a primary lithium battery pack that must be installed prior to use. There are specific requirements for disposing of the battery pack. Refer to the Battery Pack Disposal section.
The instruments are packaged without a reload and must be loaded prior to use. A staple retaining cap on the reload protects the staple leg points during shipping and transportation. The instruments' lock-out feature is designed to prevent a used or improperly installed reload from being refired or an instrument from being fired without a reload.
#### V. INDICATIONS FOR USE
ECHELON FLEX™, ECHELON™ 3000, and ECHELON ENDOPATH™ families of staplers and reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue reinforcement materials. The instruments may also be used for transection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
Note: only the ECHELON™ 3000 staplers are the Subject of this 510(k) submission.
{6}------------------------------------------------
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
ECHELONTM 3000 Stapler is substantially equivalent to ECHELON FLEX™ Powered Plus Articulating Endoscopic Linear Cutters with respect to operating principle of device and Intended use of the device. Both the devices are sterile, single use device which are powered using lithium battery. The key technological difference between the subject and the predicate device is the embedded software in Subject device to fire the device.
The subject and predicate staplers have the following identical features:
- Indication for Use
- Intended use ●
- Contraindications
- Compatible Reloads (Subject device will be using the previously 510k cleared reloads used with the predicate device)
- Materials
- Operational principles
The following differences exist between the subject and predicate staplers:
- Control Mechanism Embedded software ●
- Sterilization Method
- Powered Articulation ●
- Increased Articulation and Jaw aperture
- Multifunctional "Home button"
- Haptic alerts
- Anvil Component Dimension Change ●
- Motor Firing Speed Change
#### VII. PERFORMANCE DATA:
The following performance data demonstrate that the subject device is substantially equivalent to the predicate device and the differences between the devices were found not to affect safety or performance.
Bench Testing: Formed Staple Height (FSH), Staple Line Integrity (SLI), Force to Close, Staple Line Visual Analysis, Staple Line Strength.
Animal Testing: In-vivo testing evaluations included
- Acute Hemostasis evaluation study
- Tissue Healing response, Survival Study ●
Clinical studies: The premarket submission did not rely on the assessment of clinical performance data to demonstrate device performance and equivalence.
Biocompatibility: studies was performed and confirmed that the Subject device is biocompatible for the intended patient contact profile.
{7}------------------------------------------------
Electrical Safety and Electromagnetic Compatibility: The Electrical Safety and Electromagnetic Compatibility of the subject device conforms with the requirements of the FDA recognized standards for Medical Electrical Equipment.
Human Factors Testing/Usability Study: was conducted to evaluate the performance of the subject device and Instruction for Use (IFU) and establish objective evidence that the device can be used safely as intended by representative users in the intended environment.
## VII. CONCLUSIONS
The conclusions of the testing criteria demonstrate that the subject device, ECHELON™ 3000 performs substantially equivalent to the predicate device and does not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.