MD TECH BIOPSY SET FOR BONE AND BONE MARROW
Device Facts
| Record ID | K961959 |
|---|---|
| Device Name | MD TECH BIOPSY SET FOR BONE AND BONE MARROW |
| Applicant | Medical Device Technologies, Inc. |
| Product Code | GAA · General, Plastic Surgery |
| Decision Date | Jul 19, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4800 |
| Device Class | Class 1 |
Indications for Use
The MD Tech Biopsy-Set is intended for the purpose of harvesting bone and/or bone marrow specimens.
Device Story
Device: MD Tech Biopsy-Set; percutaneous bone/bone marrow biopsy needle. Operation: mechanical insertion adjacent to biopsy site; mechanical cutting to harvest specimen; negative pressure aspiration via attached syringe to outer cannula. Specimen containment: within outer cannula during withdrawal. Usage: clinical setting; physician-operated. Benefit: facilitates diagnostic tissue sampling.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Mechanical biopsy needle; percutaneous insertion; manual operation; utilizes cutting surfaces and negative pressure aspiration via syringe attachment; stainless steel construction (implied by biopsy needle class).
Indications for Use
Indicated for patients requiring bone or bone marrow biopsy/aspiration for diagnostic purposes.
Regulatory Classification
Identification
A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.
Predicate Devices
- Bone Marrow Biopsy Needle (K890925)
- Super Core Biopsy Needle (K930732)
- Bone Marrow Harvest Needle (K940025)
- Monoject Bone Marrow Needle (K945109)