K200971 · Xuzhou Kernel Medical Equipment Co., Ltd. · FTC · Dec 15, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200971
Device Name
308nm Excimer System
Applicant
Xuzhou Kernel Medical Equipment Co., Ltd.
Product Code
FTC · General, Plastic Surgery
Decision Date
Dec 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 308nm Excimer System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.
Device Story
308nm Excimer System is a handheld ultraviolet light therapy instrument; utilizes Xenon-Chlorine (XeCl) excimer lamp to emit monochromatic 308nm UVB light. Device features 8-inch touch LCD screen; multi-spec treatment handle filters; one-click light source output; user management; and fingerprint entry (KN-5000D). Operated by clinicians in clinical settings for targeted small-area treatment. Device performs manual Minimal Erythema Dose (MED) tests to determine patient-specific skin characteristics. Output is controlled via software to limit maximum dosage to 5 J/cm². Accurate calibration system (KN-5000D) ensures reliable dose output. Healthcare providers use device to deliver controlled UV radiation to skin lesions; intended to improve patient outcomes for psoriasis and vitiligo.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified with IEC 60601-1 (safety), IEC 60601-2-57 (non-laser light source), IEC 60601-1-2 (EMC), IEC 62471 (photobiological safety), and ISO 10993-5/10 (biocompatibility/cytotoxicity/irritation).
Technological Characteristics
Xenon-Chlorine (XeCl) excimer lamp; 308nm ± 2nm wavelength; handheld irradiation; air cooling; 8-inch touch LCD; AC100-240V power; software-controlled dose limits (max 5 J/cm²); automatic calibration (KN-5000D). Complies with IEC 60601-1, IEC 60601-2-57, IEC 60601-1-2, IEC 62471, ISO 10993-5, and ISO 10993-10.
Indications for Use
Indicated for treatment of psoriasis and vitiligo on intact skin.
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 15, 2020
Xuzhou Kernel Medical Equipment Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China
Re: K200971
Trade/Device Name: 308nm Excimer System Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet Lamp For Dermatologic Disorders Regulatory Class: Class II Product Code: FTC Dated: October 15, 2020 Received: October 19, 2020
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200971
Device Name 308nm Excimer System
Indications for Use (Describe)
The 308nm Excimer System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
#### 1.0 Submitter's Information
Name: Xuzhou Kernel Medical Equipment Co., Ltd. Address: Kernel Mansion, Economic Development District, Xuzhou City, Jiangsu Province, 221004 China Tel: +86-516-87732209 Fax: +86-516-87732210 Contact: Jing Wang Date of Preparation: Dec. 10th, 2020
#### Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Address: Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn
#### 2.0 Device Information
Trade name: 308nm Excimer System Ultraviolet Lamp for Dermatologic Disorders Common name: Classification name: Light, Ultraviolet, Dermatological Model(s): KN-5000C; KN-5000D
#### 3.0 Classification
Production code: FTC Regulation number: 21 CFR 878.4630 Classification: Class II Review Panel: General& Plastic Surgery
#### 4.0 Predicate Device Information
Predicate Device Information 1: Manufacturer: Chongqing Peninsula Medical Technology Co.,Ltd. 308nm Excimer System Device:
{4}------------------------------------------------
510(k) number: K172273
Predicate Device Information 2:
Manufacturer: Chongqing Peninsula Medical Technology Co.,Ltd. Device: 308nm Excimer System (Model: XECL-308E) 510(k) number: K192642
#### 5.0 Indication for Use Statement
The 308nm Excimer System is intended to be used for the treatment of psoriasis and vitiligo. It is to be used on intact skin only.
#### 6.0 Device Description
The proposed device, 308nm Excimer System is a new type of ultraviolet light therapy instrument, which could emit high levels of UV light. It has the following features:
- High intensity UV light. -
- The instrument is suitable for small area treatment with good targeting. -
- -8 inches Super Large color touch LCD screen which make the operation display more convenient;
- Manual MED tests can also be performed according to the skin characteristics of the patient.
- One-click Control light source Output. -
- Powerful user management capabilities. -
- Can realize the export and view of user information; -
- -Multi-spec treatment handle light filter. The user can choose filter according to the size of the treatment area. Avoid excess radiation output.
- -Fingerprint entry makes it easy for doctors to quickly access patient information (for KN-5000D):
- Accurate calibration system. Enhanced reliability of instrument dose output -(KN-5000D applicable).
#### Non-Clinical Test Conclusion
Non clinical tests were conducted by lab bench testing to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1:2012, Medical electrical equipment-Part 1: General requirements for basic safety, and essential performance.
IEC 60601-2-57: 2011 Medical electrical equipment Part2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and
{5}------------------------------------------------
cosmetic/ aesthetic use.
IEC 60601-1-2:2014, Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
IEC 62471:2006, Photobiological safety of lamps and lamp systems
ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
### 7.0 Clinical Test Conclusion
No clinical study is included in this submission.
#### 8.0 Substantial Equivalence Comparison
| Table1-Technological Characteristic Comparison | | | | | |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|---------------|
| Item | Proposed device<br>K200971 | Predicated device 1<br>K172273 | Predicated device 2<br>K192642 | Remark | |
| Product<br>Code | FTC | FTC | FTC | SE | |
| Regulation<br>No. | 21 CFR 878.4630 | 21 CFR 878.4630 | 21 CFR 878.4630 | SE | |
| Class | II | II | II | SE | |
| Intended<br>Use | The 308nm Excimer<br>System is intended to be<br>used for the treatment of<br>psoriasis and vitiligo. It is<br>to be used on intact skin<br>only. | The 308nm Excimer<br>System is intended to<br>be used for the<br>treatment of psoriasis<br>and vitiligo. It is to be<br>used on intact skin<br>only. | The 308nm Excimer<br>System(model:<br>XECL-308E) is an<br>Ultraviolet Light<br>Emitting<br>Medical<br>Device.It is intended<br>to be used for the<br>treatment of<br>psoriasis, vitiligo,<br>atopic dermatitis, and<br>leukoderma. It is to<br>be used on intact skin<br>only. | SE | |
| Prescription<br>Required | Yes | Yes | Yes | SE | |
| Weight | 5kg (Treatment<br>handle:1.4kg ) | < 2.6 kg | Not publicly available | -- | |
| Dimensions<br>(H x W x D) | 34.1cm×29.7cm×<br>22.1cm(For Host) | 26 cm x 24 cm x 27<br>cm(For lamp)<br>5 cm x 9.5 cm x 18 cm | 548.0mm*253.6<br>mm*181.4mm | -- | |
| | 28.0cm(After placing the<br>Treatment handle) | cm<br>(For Adapter) | | | |
| Mode of<br>Operation | Continuous operation | Continuous operation | Not publicly available | SE | |
| Wavelength | 308 nm ± 2nm | 308 nanometers (nm)<br>±3nm | 308±2 nm | SE | |
| Light Source | Xenon-Chlorine (XeCI)<br>excimer lamp produces<br>monochromatic UVB light | Xenon-Chlorine<br>(XeCI)<br>excimer lamp<br>produces<br>monochromatic UVB<br>light | Xenon-Chlorine<br>(XeCl) excimer<br>lamp produces<br>monochromatic<br>UVB light | SE | |
| Irradiation<br>mode | Handheld irradiation | Handheld irradiation | Handheld irradiation | SE | |
| Cooling of<br>light source | Air cooling | Air cooling | Cooling fan | SE | |
| Treatment<br>Area | $20cm^2±10%$ | 16 cm² (4 x 4 cm) | $6cm^2~28.8 cm^2$ | Analysis<br>1 | |
| Max.UV<br>Irradiation<br>Intensity | 50 mW/cm² | 50 mW/cm² | 50 mW/cm² | SE | |
| Max.<br>Irradiation<br>Power | 1000 mW | 800 mW | 1440 mW | Analysis<br>1 | |
| Treatment<br>Time | 0~120s | 1s to 40 s | 1-140s | Analysis<br>1 | |
| Maximum<br>dose | 5 J/cm² | 2 J/cm² | 7 J/cm² | Analysis<br>2 | |
| Power<br>Source | AC100-240V, 50/60Hz<br>Input Power:<br>KN-5000C: 160VA<br>KN-5000D: 200VA | Adapter Input:<br>100~240Vac, 50/60<br>Hz<br>Main unit input: 48<br>Vdc,2.94 A,135 VA<br>max | Not publicly available | Analysis<br>3 | |
| Power<br>Calibration<br>Method | KN-5000D:<br>Automatic calibration<br>KN-5000C:NA | Internal, automatic | Not publicly available | SE | |
| MED Dose<br>Determination | Manual MED test | Menu driven | Not publicly available | Analysis<br>4 | |
| Display | 8 inches touch LCD<br>screen | Touch Screen Control<br>Panel | Not publicly available | SE | |
| Security<br>Type | Classification by type of<br>anti-electric shock: Class | Classification by type<br>of anti-electric shock: | Not publicly available | SE | |
| | | I; Classification according<br>to the degree of<br>anti-electric shock: Type B. | Class I; Classification<br>according to the<br>degree of anti-electric<br>shock: Type B. | | |
| | Patient<br>Leakage<br>Current | Complied with IEC<br>60601-1 and IEC<br>60601-2-57 | Complied with IEC<br>60601-1 and IEC<br>60601-2-57 | Complied with IEC<br>60601-1 and IEC<br>60601-2-57 | SE |
| | Operating<br>Environment | Temperature: 5~35℃<br>Relative humidity: ≤85%<br>Atmospheric pressure::<br>700hPa~1060hPa | Temperature:<br>15~35°C,<br>Humidity: ≤ 80%RH,<br>Atmospheric<br>Pressure:<br>86 ~ 106 kPa | Temperature:<br>5℃~30℃<br>Humidity: ≤80%RH<br>Atmospheric<br>Pressure: 700 ~<br>1060 hPa | Analysis<br>4 |
| | Storage<br>Environment | Ambient temperature:<br>-40 ~ 55 °C<br>Relative humidity: ≤<br>90%<br>Atmospheric pressure:<br>500 ~ 1060hpa | Temperature:<br>-20~45°C,<br>Humidity: 10-85%RH<br>Atmospheric<br>Pressure:<br>50 ~ 106 kPa | Temperature:<br>0℃~40℃<br>Humidity:≤85%RH<br>Atmospheric<br>Pressure: 700 ~<br>1060 hPa | Analysis<br>4 |
| | Electrical<br>Safety/Performance | Comply with IEC60601-1<br>and IEC 60601-2-57 | Comply with<br>IEC60601-1 and IEC<br>60601-2-57 | Comply with<br>IEC60601-1 and IEC<br>60601-2-57 | SE |
| | Sterile | N/A | N/A | N/A | SE |
| | Single Use | No | No | No | SE |
| | EMC | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | SE |
| | Biocompatibility | Cytotoxicity (ISO<br>10993-5:2009)<br>Irritation (ISO<br>10993-10:2010<br>Sensitization (ISO<br>10993-10:2010) | Under the conditions<br>of the study, Comply<br>with the requirements | Under the conditions<br>of the study, Comply<br>with the requirements | SE |
| | Label and<br>Labeling | Conforms to FDA<br>Regulatory Requirements | Conforms to FDA<br>Regulatory<br>Requirements | Conforms to FDA<br>Regulatory<br>Requirements | SE |
# Table1 Tochnological Charactoristic Comparis
{6}------------------------------------------------
{7}------------------------------------------------
Comparison Discussion:
#### Analysis1:
Although some output specifications as "Max Power Output", "Treatment area", and "Treatment time" of the proposed device are a little different from the predicate devices, but they are considered substantially equivalent, and they all comply with the standards: IEC 60601-1, IEC 60601-1-2 and IEC 60601-2-57, and range between the predicate device K172273 and predicate
{8}------------------------------------------------
device K192642. So the we can think that the slight differences in specification will not raise any safety or effectiveness issue.
#### Analysis 2:
The software limits the maximum dosage of the proposed device to no more than 5 J/cm²,it is different from the predicate devices, but the value range between the predicate device K172273 and predicate device K192642. So the we can think that the slight differences in specification will not raise any safety or effectiveness issue.
#### Analvsis 3:
The " Power Source" is a little different from the predicate devices, but they all comply with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57 requirements. So the differences will not raise any safety or effectiveness issue.
#### Analvsis 4:
The " Power Source", "MED Dose Determination" and "Operating and Storage Environment" are a little different from the predicate devices, but they all comply with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57 requirements. So the differences of function specification will not raise any safety or effectiveness issue.
#### Final Conclusion:
Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device. The proposed device "308nm Excimer System" is Substantial Equivalent to the predicate devices.
#### 9.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the leqally marketed predicated device and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.