K181805 · Xuzhou Kernel Medical Equipment Co., Ltd. · FTC · Jan 11, 2019 · General, Plastic Surgery
Device Facts
Record ID
K181805
Device Name
UV Phototherapy
Applicant
Xuzhou Kernel Medical Equipment Co., Ltd.
Product Code
FTC · General, Plastic Surgery
Decision Date
Jan 11, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UV Phototherapy is intended for use, by or under the direction of a physician for the treatment of psoriasis, vitiligo, and atopic dermatitis (eczema). It is intended for use on all skin types (I-VI). It can be used in hospitals, clinics and households.
Device Story
Handheld ultraviolet (UV) phototherapy device for partial skin treatment (excluding eyes). Device emits UVA or UVB light to treat psoriasis, vitiligo, and atopic dermatitis. Operated by physician or patient in hospitals, clinics, or households. Features include adjustable irradiation time (0-30 minutes), button-based control, and optional light comb for scalp treatment. Select models (KN-4003AL2S, KN-4003BL2S, KN-4003AL2DS, KN-4003BL2DS, KN-4006AL1S, KN-4006BL1S, KN-4006AL1DS, KN-4006BL1DS) incorporate SD card slots allowing physicians to write recommended treatment parameters to the device. Output provides targeted UV exposure to affected skin areas to manage dermatologic conditions.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Handheld UV lamp; UVA or UVB source; 1 or 2 lamps. Materials comply with ISO 10993 (cytotoxicity, irritation, sensitization, systemic toxicity). Electrical safety per IEC 60601-1, IEC 60601-2-57, IEC 60601-1-11. EMC per IEC 60601-1-2. Battery safety per IEC 62133. Models with SD card support physician-programmed treatment parameters.
Indications for Use
Indicated for patients with psoriasis, vitiligo, and atopic dermatitis (eczema) across all skin types (I-VI).
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Xuzhou Kernel Medical Equipment Co., LTD. Ms. Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai. 200120 CN
January 11, 2019
Re: K181805
Trade/Device Name: UV Phototherapy Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet Lamp For Dermatologic Disorders Regulatory Class: Class II Product Code: FTC Dated: November 30, 2018 Received: December 10, 2018
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
| Neil R | Digitally signed by Neil R |
|----------|----------------------------|
| Ogden -S | Ogden -S |
| | Date: 2019.01.11 |
| | 09:28:12 -05'00' |
Ogden -S
Date: 2019.01.11
09:28:12-05'00'
MD MRA FAC
Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181805
Device Name UV Phototherapy
Indications for Use (Describe)
The UV Phototherapy is intended for use, by or under the direction for the treatment of psoriasis, vitiligo, and atopic dermatitis (eczema). It is intended for use on all skin types (I-VI). It can be used in hospitals, clinics and households.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Tab #7 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K181805
- 1. Date of Preparation: 01/07/2019
- 2. Sponsor Identification
Xuzhou Kernel Medical Equipment Co., LTD. Kernel Mansion, Economic Development District, Xuzhou, Jiangsu Province, China
Establishment Registration Number: 3008393409
Contact Person: Jing Wang Position: Management Representative Tel: +86-516-87732208 Fax: +86-516-87732277 Email: wjkernel@126.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: UV Phototherapy Common Name: Ultraviolet lamp for dermatologic disorders
### Regulatory Information
Classification Name: Light, Ultraviolet, Dermatological Classification: II Product Code: FTC Regulation Number: 21 CFR 878.4630 Review Panel: General& Plastic Surgery
### Intended Use Statement:
The UV Phototherapy is intended for use, by or under the direction of a physician for the treatment of psoriasis, vitiligo, and atopic dermatitis (eczema). It is intended for use on all skin types (I-VI). It can be used in hospitals, clinics and households.
### Device Description
The proposed device is a hand held ultraviolet phototherapy device, intended for partial treatment excluding eyes. Irradiation time can be adjusted from 0~30min and the treatment status can be controlled by the button on the device. The light comb equipped on the device is intended for easier treatment of target skin covered by hair, such as the scalp. The device is available in sixteen specifications. The device model KN-4003AL2S, KN-4003BL2S, KN-4003AL2DS, KN-4003BL2DS, KN-4006AL1S, KN-4006BL1S, KN-4006AL1DS and KN-4006BL1DS are designed with SD card. The recommended treatment can be written into the SD card by a physician. The device can be used in hospitals, clinics and households.
#### న్. Identification of Predicate Device
510(k) Number: K132643 Device Name: UV Phototherapy
#### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
{5}------------------------------------------------
- ア ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity
- > ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
- > ISO 10993-11:2017 Biological evaluation of medical devices- Part 11: Test for systemic toxicity
- > IEC60601-1:2005+CORR.1:2006+CORR.2:2007+AM1:2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- ▲ IEC 60601-2-57: 2011 Medical electrical equipment Part2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/ aesthetic use
- A IEC 60601-1-2:2014 Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility-Requirements and tests
- > IEC 60601-1-11 Medical electrical equipment-Part 1-11: General requirements for basic safety and essential performance-Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- > IEC 62133:2012 Secondary Cells And Batteries Containing Alkaline Or Other Non-Acid Electrolytes - Safety Requirements For Portable Sealed Secondary Cells, And For Batteries Made From Them, For Use In Portable Applications [Including: Corrigendum 1 (2013)]
- Clinical Test Conclusion 7.
No clinical study is included in this submission.
{6}------------------------------------------------
### Substantially Equivalent (SE) Comparison 8.
| Table 1 Comparison of Technology Characteristics | | |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | Proposed Device | Predicate Device<br>K132643 |
| Product Code | FTC | FTC |
| Regulation Number | 21 CFR 878.4630 | 21 CFR 878.4630 |
| Intended Use | The UV Phototherapy is intended for<br>use, by or under the direction of a<br>physician for the treatment of<br>psoriasis, vitiligo, and atopic<br>dermatitis (eczema). It is intended for<br>use on all skin types (I-VI). It can be<br>used in hospitals, clinics and<br>households. | The UV phototherapy, model<br>KN-4001/KN-4002/KN-4003/KN-4004/<br>KN4005/KN4006, deliver ultraviolet<br>(UV) light to targeted affected skin. It is<br>intended for use, by or under the<br>direction of a physician, for the treatment<br>of psoriasis, vitiligo, and atopic<br>dermatitis (eczema) on all skin types<br>(I-VI). |
| Single Use | No | No |
| Sterile | N/A | N/A |
| Mode of operation | Handheld | Handheld |
| Treatment area | Partial treatment excluding eyes | Partial treatment excluding eyes |
| Treatment Time | 0~30min | 0~30min |
| Treatment times | 3~5 times per week | 3~5 times per week |
| Lamp quantities | 1 or 2 | 1 or 2 |
| UV spectral | UVA or UVB | UVA or UVB |
| Biocompatibility | No cytotoxicity<br>No irritation<br>No sensitization<br>No evidence of acute toxicity<br>No pyrogen | Comply with ISO 10993 |
| Electrical safety | Comply with IEC60601-1 | Comply with IEC60601-1 |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
# Table 1 Comparison of Technology Characteristics
- 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.