Unistik® 3 Single-Use Safety Lancets:- Extra (21G), Normal (23G), Comfort (28G), Gentle (30G)

K231124 · Owen Mumford, Ltd. · FMK · Jun 27, 2023 · General, Plastic Surgery

Device Facts

Record IDK231124
Device NameUnistik® 3 Single-Use Safety Lancets:- Extra (21G), Normal (23G), Comfort (28G), Gentle (30G)
ApplicantOwen Mumford, Ltd.
Product CodeFMK · General, Plastic Surgery
Decision DateJun 27, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4850
Device ClassClass 2

Indications for Use

The Unistik® 3 single-use safety lancets are hand-held, disposable devices intended to be used to achieve a controlled skin puncture, typically on the fingertip, to obtain a capillary blood specimen.

Device Story

Single-use, hand-held safety lancet; obtains capillary blood samples via controlled fingertip puncture. Device consists of stainless steel needle, plastic housing, and internal spring mechanism. User twists off protective cap, presses device against skin, and activates via side button. Internal spring propels needle for puncture and automatically retracts it into housing post-activation; needle locks in shielded position to prevent reuse and needlestick injuries. Used by patients, caregivers, and clinicians in home or clinical environments. Benefits include standardized sampling, reduced risk of blood-borne pathogen transmission, and prevention of accidental reuse.

Clinical Evidence

Bench testing only. Verification testing included visual inspection, premature activation, torque, actuation force, needle retention, needle safety, depth of penetration, and drop testing. Simulated clinical use testing performed per FDA guidance for sharps injury prevention. Biocompatibility per ISO 10993-1. Sterilization validated per ISO 11137-1 to 10⁻⁶ SAL.

Technological Characteristics

Materials: plastic housing/holder, stainless steel needle, steel springs. Principle: spring-powered automatic lancing and retraction. Dimensions: 21G, 23G, 28G, 30G gauges; 1.5mm–2.0mm lancing depth. Connectivity: none. Sterilization: Gamma radiation (25-50 kGy). Features: integral sharps injury prevention, single-use lockout mechanism.

Indications for Use

Indicated for patients requiring capillary blood specimens for in-vitro diagnostic assays, such as blood glucose monitoring. Suitable for use by self-testing patients, caregivers, and healthcare professionals in home and clinical settings.

Regulatory Classification

Identification

The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.

Special Controls

*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (b) *Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) Handwashing instructions for the user before and after use of the device. (C) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device. (E) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Discard the entire device after use.” (B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.” (c) *Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that: (A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and (B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection. (ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use. (iii) The device must be demonstrated to be biocompatible. (iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin ( *e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected. (vi) Labeling must include: (A) Detailed descriptions, with illustrations, of the proper use of the device. (B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use. (C) Handwashing instructions for the user before and after use of the device. (D) Instructions on preparation ( *e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device. (F) Instructions for the safe disposal of the device. (G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades. (H) Labeling must be appropriate for the intended use environment. ( *1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.( *2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed: (A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.” (B) “Used lancet blades must be safely discarded after a single use.” (C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.” (d) *Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2) *Classification.* Class III (premarket approval).(3) *Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 26, 2023 Owen Mumford Ltd Darren Mansell Regulatory Affairs Manager Brook Hill Woodstock, Oxfordshire OX20 1TU United Kingdom Re: K231124 Trade/Device Name: Unistik® 3 Single-Use Safety Lancets:- Extra (21G), Normal (23G), Comfort (28G), Gentle (30G) Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: FMK Dated: April 18, 2023 Received: April 20, 2023 Dear Darren Mansell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K231124 - Darren Mansell Page 2 statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark Trumbore - S Digitally signed by Mark Trumbore - S Date: 2023.06.26 08:26:55 -04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2} # SECTION 4.0 ## INDICATIONS FOR USE STATEMENT {3} | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration **Indications for Use** | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K231124 | | | Device Name Unistik® 3 | | | Indications for Use (Describe) | | The Unistik® 3 single-use safety lancets are hand-held, disposable devices intended to be used to achieve a controlled skin puncture, typically on the fingertip, to obtain a capillary blood specimen. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. **DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) Page 1 of 1 FSC Publishing Services (703) 443-6341 {4} K231124 # SECTION 5.0 # 510(k) SUMMARY {5} K231124 # SECTION 5.0 # 510(k) SUMMARY ## 1. Submitter Prepared by: Owen Mumford Ltd Tel: +44(0)1993 812021 Fax: +44(0)1993 813466 Prepared for: Owner/ Operator Owen Mumford Ltd Brook Hill Woodstock Oxfordshire OX20 1TU United Kingdom Establishment Registration Number: 3003348846 Contact Person: Darren Mansell Regulatory Affairs Manager Owen Mumford Ltd, Tel: +44(0)1993 812021 Fax: +44(0)1993 813466 Email: darren.mansell@owenmumford.com Date Prepared: 18 April 2023 ## 2. Device Name of Device: Unistik® 3 Single-Use Safety Lancets Common Name: Blood lancets Classification Name: Single use only blood lancet with an integral sharps injury prevention feature Regulatory Class: II Product Code: FMK Section 5-1 {6} K231124 # 3. Predicate Devices Predicate Device Name: Unistik® 3 Single-Use Safety Lancets under 510k number K221126 (Cleared for: Prescription Use and Over-The-Counter Use) # 4. Description of The Device This submission covers the following devices: - Unistik® 3 Single-Use Safety Lancets 21G, 23G, 28G & 30G (new design variants) The Unistik® 3 sterile single-use safety lancets are hand-held disposable devices intended to be used to achieve a controlled skin puncture, typically on the fingertip, in order to obtain a capillary blood specimen. The Unistik® 3 sterile single-use safety lancets are indicated for use where a capillary blood specimen is required for the purposes of performing in-vitro diagnostic (IVD) assays, e.g. for blood glucose monitoring in patients with diabetes. The Unistik® 3 new design variant safety lancets are available in four different gauges; 21G, 23G, 28G and 30G, to facilitate appropriate blood flow rates from the skin puncture. The Unistik® new design variant safety lancets are designed for prescription and over-the-counter use and to be used by self-testing patients, care-givers and healthcare professionals. The devices are designed to perform a controlled skin puncture on the fingertip, in order for care-givers and healthcare professionals to obtain capillary blood specimens from patients for IVD assays, and also for lay (home) users to be able to perform a skin puncture on themselves where an IVD self-testing regime is required. The intended user population includes male and female, right or left-handed self-administering patients, care givers and healthcare professionals. The frequency of use and intended patient population is dependent on the given diagnostic regime. The Unistik® 3 new design variant safety lancets are sterile single-use devices with integral sharps protection whereby the lancet needle is shielded before and after use to prevent needlestick injuries, thereby mitigating the hazard of transmission of blood-borne infectious agents. Furthermore, the device automatically self-disables after a single use, thus preventing any hazards of re-use. The Unistik® 3 new design variant safety lancets are used by first twisting off the end cap, the needle tip is then exposed but remains safely shielded within the device housing. The user then presses the end face of the device against the sampling site, then activates the device by pressing the release button on the side of the device. The lancet needle is then automatically propelled forward by the internal pre-loaded spring to lance the skin and also automatically retracted by the spring back inside the device housing, where it is then automatically locked to prevent re-use. After firing, the locked position of the needle tip inside the device ensures that it remains safely shielded, and the device can be safely disposed of into an appropriate sharps receptacle. The Unistik® 3 new design variant safety lancets consist of a stainless steel lancet needle moulded into a plastic lancet holder component, which in turn is assembled into a moulded plastic outer housing with a pre-loaded steel firing spring for propelling the lancet holder forward when the device is activated and a retraction spring for subsequently automatically retracting it. After retraction, the lancet holder is automatically locked into the device such that the device cannot be re-used and the needle tip is safely shielded such that it is inaccessible within the body of the device. Section 5-2 {7} K231124 The lancet needles are moulded into the lancet holder component such that the needle tip is sealed by complete encapsulation in overmoulded plastic. The complete lancet holder component is then sterilised by gamma irradiation, so after irradiation the sterility of the needle tip is maintained by encapsulation within the plastic. The sterile seal is only broken when the user twists off and removes the lancet cap immediately before use. The needle tip is the exposed needle length after the cap is removed, and this is the only part of the needle that will penetrate the patient's skin during use. Therefore, the encapsulation of the needle tip by plastic overmoulding performs the function of primary packaging, whereby a sterile seal is maintained until the point of use. The purpose of this 510(k) application is to obtain both prescription-only clearance and over-counter clearance for the Unistik® 3 new design variant safety lancets. The intended use for the Unistik® 3 new design variant safety lancets remains the same as the predicate device. Table 5.1 on the following page illustrates the Unistik® 3 21G, 23G, 28G & 30G new design variants in comparison to the existing Unistik® 3 21G, 23G, 28G & 30G variants cleared under .K221126. Section 5-3 {8} K231124 Table 5.1 - Illustrations of Unistik® 3 21G, 23G, 28G & 30G new design variants in comparison to the existing Unistik® 3 variants (not to scale) | Unistik®3 safety lancet range – predicate design cleared under K221126 | Unistik®3 safety lancet range – new design variants in this submission | | --- | --- | | 30G Device - Predicate | 30G Device – New design variant | | 28G Device – Predicate | 28G Device – New design variant | | 23G Device – Predicate | 23G Device – New design variant | | 21G Device – Predicate | 21G Device – New design variant | Section 5-4 {9} K231124 # 5. Indications for Use The Unistik® 3 new design variant single-use safety lancets are hand-held disposable devices intended to be used to achieve a controlled skin puncture, typically on the fingertip, to obtain a capillary blood specimen. # 6. Technological Characteristics The Unistik® 3 new design variant safety lancets are substantially equivalent to the predicate device, the Unistik® 3 safety lancet. A comparison of the intended uses and technological characteristics of the Unistik® 3 new design variant safety lancets to the predicate Unistik® 3 devices is summarised in Table 5.2 below. Table 5.2 - Characteristics of the Unistik® 3 new design variant safety lancets (submission device) in comparison to the predicate device | Device Characteristic | Predicate Device: Unistik® 3 Safety Lancets - K221126 | Submission Device - Unistik® 3 Safety Lancets – 21G, 23G, 28G & 30G (new design variants) | | --- | --- | --- | | Indications For Use | The Unistik® 3 safety lancets are hand-held, disposable devices intended to be used to achieve a controlled skin puncture, typically on the fingertip, to obtain a capillary blood specimen. | Unchanged from the predicate device. | | Use environment | Home and clinical. | Unchanged from the predicate device. | | Operating principle | Manually activated (side fire button) spring-powered automatic lancing and needle retraction. | Unchanged from the predicate device. | | Design/ construction | Stainless steel lancet needle moulded into a plastic lancet holder component, which in turn is assembled into a moulded plastic outer housing with a pre-loaded steel lancet firing spring. The external plastic outer housing incorporates a plastic tab which functions as the firing button on the side of the device. | Unchanged from the predicate device with the exception of an additional spring for lancet retraction and a trigger component which internally engages with the lancet holder and externally functions as the firing button. | Section 5-5 {10} K231124 Table 5.2 (continued) - Characteristics of the Unistik® 3 new design variant safety lancets (submission device) in comparison to the predicate device | Device Characteristic | Predicate Device: Unistik® 3 Safety Lancets - K221126 | Submission Device - Unistik® 3 Safety Lancets – 21G, 23G, 28G & 30G (new design variants) | | --- | --- | --- | | Integral sharps injury prevention feature? | Yes | Unchanged from the predicate device | | Single-use? | Yes | Unchanged from the predicate device | | Sterility | Sterile | Unchanged from the predicate device | | Components and Materials | Plastic external and internal components, stainless steel needle and steel spring. | Unchanged from the predicate device, though some changes in precise material specifications of the plastics and steels. | | Package | Laminate pulp board cartons | Unchanged from the predicate device. | Section 5-6 {11} K231124 Table 5.2 (continued) - Characteristics of the Unistik® 3 new design variant safety lancets (submission device) in comparison to the predicate device | Device Characteristic | | Predicate Device: Unistik® 3 Safety Lancets - K221126 | Submission Device - Unistik® 3 Safety Lancets – 21G, 23G, 28G & 30G (new design variants) | | --- | --- | --- | --- | | Needle Specifications | Needle Gauges (SWG) | Unistik® 3 existing design: 18G, 21G, 23G, 28G, 30G | Unistik® 3 new design variants: 21G, 23G, 28G, 30G The Unistik® 3 new design variant gauge range of 21G to 30G is within the range of Unistik® 3 existing design gauges (18G – 30G) | | | Lancing Depths (mm) | Unistik® 3 existing design: 18G variant: 1.8 mm depth ± 0.35 mm 21G variant: 2.0 mm depth ± 0.35 mm 23G variant – 1.8 mm depth ± 0.35 mm 28G variant – 1.8 mm depth ± 0.35 mm 30G variant – 1.5 mm depth ± 0.35 mm | Unistik® 3 new design variants: 21G new variant: 2.0 mm depth (Min. 1.49 mm / Max. 3.01mm) 23G new variant: 1.8 mm depth (Min. 1.29 mm / Max. 3.01 mm) 28G variant: 1.8 mm depth (Min. 1.29 mm / Max. 3.01 mm) 30G variant: 1.5 mm depth (Min. 0.99 mm / Max. 3.01 mm) The range of nominal lancing depths for the Unistik® 3 new design variants (1.5 mm to 2.0 mm depth) is the same as the depth range offered by the existing Unistik® 3 design. | | | Needle Tip Configuration | 2-facet chisel (18G) 3-facet (21G) 3-facet -(23G) 3-facet (28G) 3-facet (30G) | 3-facet (21G) 3-facet -(23G) 3-facet (28G) 3-facet (30G) The Unistik® 3 new design variants have the same 3-facet needle tips as the corresponding gauges of the existing Unistik® 3 design. | | | Sterilisation method | 20-40 kGy Cobalt 60 gamma radiation validated to achieve 10^{-6} sterility assurance level (SAL). | 25-50 kGy Cobalt 60 gamma radiation validated to achieve 10^{-6}. sterility assurance level (SAL) | Section 5-7 {12} K231124 # 7. Performance Data # Non-clinical performance data - bench testing: Design verification testing of the Unistik® 3 new design variant safety lancets has been carried out to evaluate the performance of the devices against defined acceptance criteria. Table 5.3 below provides a summary of the relevant design verification testing. Table 5.3 - Design verification tests carried out on Unistik® 3 new design variants | Test Description | Result | | --- | --- | | Visual inspection and numbering | Pass | | Standard pre-conditioning | Pass | | Premature activation | Pass | | Torque to rotate cap | Pass | | Force to actuate | Pass | | Needle retention force | Pass | | Needle safety test | Pass | | Depth of Penetration | Pass | | Needle safety after drop testing | Pass | | Simulated clinical use testing | Pass | The design verification testing has established that the Unistik® 3 new design variant safety lancets comply with the acceptance criteria established based on the specifications of the devices. The results from these tests demonstrate that the Unistik® 3 new design variant safety lancets are safe and effective when used as intended. # Simulated clinical use testing: Simulated clinical use testing has been carried for the Unistik® 3 new design variant safety lancets according to the relevant FDA Guidance – (Medical Devices with Sharps Injury Prevention Features, August 9, 2005). The Unistik® 3 new design variant safety lancets achieved the performance criteria specified for the simulated clinical use testing. Section 5-8 {13} K231124 ## Human factors analysis: Human factors analysis has been carried out for the Unistik® 3 new design variant safety lancets which concluded that they are safe and effective for their intended users. ## Biocompatibility: Biocompatibility evaluations have been carried out for the Unistik® 3 new design variant safety lancets and have established the compliance of the devices to ISO 10993-1 requirements. ## Sterilisation: The sterility of the devices is assured using a sterilisation method validated in accordance with ISO 11137-1- “Sterilization of health care products — Radiation — Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices”. Through the sterilisation methods used, all devices are sterilised to provide a Sterility Assurance Level (SAL) of 10⁻⁶. ## 8. Conclusion In summary, the differences between the Unistik® 3 new design variant safety lancets and the predicate device have no impact on safety and effectiveness and the products are therefore substantially equivalent to the predicate device. Section 5-9
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...