SOFTXIL

K171851 · Bistool · MIB · Mar 16, 2018 · Ear, Nose, Throat

Device Facts

Record IDK171851
Device NameSOFTXIL
ApplicantBistool
Product CodeMIB · Ear, Nose, Throat
Decision DateMar 16, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3620
Device ClassClass 2
AttributesTherapeutic

Indications for Use

SOFTXIL is intended for the augmentation or reconstruction of the nasal, malar, and or chin contour.

Device Story

SOFTXIL consists of injection-molded silicone elastomer implants designed for facial surgery. Implants are available in various pre-formed shapes and sizes for nasal, paranasal, and chin augmentation. The device is intended for use by surgeons in clinical settings. The surgeon may carve or cut the silicone material to customize the fit for individual patient anatomy. The implants provide structural support for soft tissue augmentation. The device is provided sterile, single-use, and is intended to improve facial contouring.

Clinical Evidence

Bench testing only. Testing included sterilization validation (ISO 11737-1), shelf life (ASTM F1980), biocompatibility (ISO 10993-5, -6, -10, -11; USP <51>), and performance testing (appearance, dimensions, extractables, hardness, tensile strength, elongation, and tear strength per ASTM F881).

Technological Characteristics

Material: Silicone elastomer. Form factor: Injection-molded facial implants with smooth surfaces and tapered ends. Sterilization: Ethylene oxide. Standards: ASTM F1980 (shelf life), ASTM F881 (mechanical properties), ISO 10993 series (biocompatibility), ISO 11737-1 (sterilization).

Indications for Use

Indicated for patients requiring augmentation or reconstruction of nasal, malar, or chin contours.

Regulatory Classification

Identification

Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box. To the right of the blue box is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 16, 2018 Bistool % Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 690 Roosevelt Irvine, California 92620 Re: K171851 Trade/Device Name: SOFTXIL Regulation Number: 21 CFR 874.3620 Regulation Name: Ear, Nose, And Throat Synthetic Polymer Material Regulatory Class: Class II Product Code: MIB, LZK, FWP Dated: February 12, 2018 Received: February 13, 2018 Dear Ms. Chung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {2}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # David Krause -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # ATTACHMENT 1 Revised IFU Statement {4}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171581 Device Name SOFTXIL Indications for Use (Describe) SOFTXIL is intended for the augmentation or reconstruction of the nasal, malar, and or chin contour. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {5}------------------------------------------------ # 510(k) Summary (K171581) This 510(k) summary is being submitted in accordance with requirements of 21 CFR Part 807.92 Date: Mar 14, 2018 ### 1. 510K Applicant / Submitter: BISTOOL 9, Gwangnaru-ro 6-gil, Seongdong-gu, Seoul, Republic of Korea, 04796 Contact Person: Eunhyun Bae, CEO Tel : +82-2-3446-7688 #### 2. Submission Contact Person LK Consulting Group USA, Inc. 690 Roosevelt, Irvine CA 92620 Phone: 714.202.5789 Fax: 714.409.3357 Contact: Priscilla Chung Email: juhee.c@lkconsultinggroup.com #### 3. Device - 트 Trade Name : SOFTXIL - Common Name: Facial Silicone Implant - I Classification: Elastomer. Silicone Block - 874.3620. Primary Product Code: MIB Additional product codes: LZK, FWP #### 4. Predicate Device AART Silicone Carving Block by Aesthetic and Reconstructive Technologies, Inc. (AART) (K021820) #### 5. Device Description SOFTXIL is injection molded silicone implants used in facial surgery as nasal, paranasal, and chin implants. The SOFTXIL offers various shapes to meet the needs in different locations. The devices are also provided in various sizes. The SOFTXIL is individually packaged and sterilized by ethylene oxide, and is labeled for single use. It has smooth surface with tapered ends. SOFTXIL is ideal for use in soft tissue augmentation where the use of a soft silicone elastomer is appropriate. {6}------------------------------------------------ # 8. Indications for Use SOFTXIL is intended for the augmentation or reconstruction of the nasal, malar, and or chin contour. # 9. Substantial Equivalence Discussion: The SOFTXIL is substantially equivalent in material, design, function, and performance to the AART Silicone Carving Block marketed by Aesthetic and Reconstructive Technologies, Inc. (AART). All products have identical intended uses and are offered in similar shapes and sizes. The size might be different between the devices, however, since the physician can carve or cut the material to fit each patient so that this difference does not raise of safety and performance. # 10. Performance Tests (Non-clinical) Non-clinical bench tests were performed as followings: - Sterilization Validation in accordance with ISO11737-1 - Shelf Life Test in accordance with ASTM F 1980; Appearance, Sterility, Tensile Strength, Elongation Percentage, Shore Hardness, Tearing Strength, Seal Strength, and Dye Penetration - Biocompatibility Test in accordance with ISO 10993-5. ISO 10993-6. ISO 10993-10. ISO 10993-11, USP<51> - · Performance Tests; Appearance, Dimensions, Extractable substances, Hardness (ASTM F881), Tensile Strength (ASTM F881), Elongation at Break (ASTM F881), Tear Strength (ASTM F881) The test results of the tests support that the subject device is substantially equivalent to the predicate devices. ## 11. Conclusions: Based on the information provided in this premarket notification. BISTOOL concludes that SOFTXIL is substantially equivalent to the predicate device as described herein in safety and effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...