K042412 · United States Surgical, A Division of Tyco Healthc · GCJ · Sep 24, 2004 · Gastroenterology, Urology
Device Facts
Record ID
K042412
Device Name
MODIFIED SPACEMAKER SYSTEM
Applicant
United States Surgical, A Division of Tyco Healthc
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Sep 24, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SPACEMAKER* Structural Balloon Trocar is primarily indicated for patients undergoing laparoscopic surgical procedures requiring a sealed port of access and/or tissue retraction. This is also indicated in patients undergoing laparoscopic surgery requiring a sealed port of access and/or tissue separation in extraperitoneal procedures, such as in hernia repair, lymphadenectomy or bladder neck suspension procedures. The SPACEMAKER* Blunt Tip Trocar is intended for use in establishing a port of access for insertion of endoscopic instruments into the abdominal cavity or extraperitoneal space in abdominal and extraperitoneal surgery. The SPACEMAKER* Dissection Balloon is primarily indicated for patients undergoing laparoscopic surgery requiring tissue separation of the extraperitoneal space.
Device Story
Modified SPACEMAKER System consists of balloon dissector and trocar; used during laparoscopic surgery. Provides port of access and tissue separation in abdominal or extraperitoneal space. Operated by surgeons in clinical/OR settings. Balloon dissector creates space; trocar maintains access for endoscopic instruments. Benefits include facilitated surgical access and tissue separation. No complex electronics or software.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
System comprises balloon dissector and trocar. Materials comply with ISO 10993-1. Mechanical device; no energy source, software, or connectivity.
Indications for Use
Indicated for patients undergoing laparoscopic surgical procedures requiring sealed port access, tissue retraction, or tissue separation in the extraperitoneal space, including hernia repair, lymphadenectomy, or bladder neck suspension.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
Auto Suture* Dissection Balloon
Auto Suture* Blunt Tip Trocar
Auto Suture* Structural Balloon Trocar
Related Devices
K151356 — Spacemaker Pro Access and Dissector System · Covidien, LLC · Jun 18, 2015
K182024 — Dissecting Balloon System · Applied Medical Resources Corp. · Aug 30, 2018
K973046 — SPACEMAKER SURGICAL BALLOON DISSECTOR, SPACEMAKER II SURGICAL BALLOON DISSECTOR, SAPHTRAK · General Surgical Innovations · Jan 16, 1998
K090631 — PAJUNK DISPOSABLE BALLOONS AND BALLOON SYSTEMS · PAJUNK GmbH Medizintechnologie · Apr 16, 2009
K962104 — AIRLIFT BALLOON RETRACTON SYSTEM · Origin Medsystems, Inc. · Sep 11, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
## Attachment 4
SEP 2 4 2004
косячіг
# 510(k) Summary
## Modified SPACEMAKER* System
United States Surgical 150 Glover Avenue Norwalk, CT 06856 USA
### DEVICE DESCRIPTION
The Modified SPACEMAKER* System consists of a balloon dissector and trocar which are used in conjunction to provide a port of access and tissue separation during laparoscopic surgery.
### CLASSIFICATION NAME
Balloon Dissectors/ Balloon Trocars
#### INDICATIONS FOR USE
#### SPACEMAKER* Structural Balloon Trocar
The SPACEMAKER * Structural Balloon Trocar is primarily indicated for patients undergoing laparoscopic surgical procedures requiring a sealed port of access and/or tissue retraction. This is also indicated in patients undergoing laparoscopic surgery requiring a sealed port of access and/or tissue separation in extraperitoneal procedures, such as in hernia repair, lymphadenectomy or bladder neck suspension procedures.
#### SPACEMAKER* Blunt Tip Trocar
The SPACEMAKER* Blunt Tip Trocar is intended for use in establishing a port of access for insertion of endoscopic instruments into the abdominal cavity or extraperitoneal space in abdominal and extraperitoneal surgery.
### SPACEMAKER* Dissection Balloon
The SPACEMAKER* Dissection Balloon is primarily indicated for patients undergoing laparoscopic surgery requiring tissue separation of the extraperitoneal space.
#### MATERIALS:
All material components of the Modified SPACEMAKER* System comply with ISO Standard #10993-1.
## PREDICATE DEVICES
Auto Suture* Dissection Balloon Auto Suture* Blunt Tip Trocar Auto Suture* Structural Balloon Trocar
Special 510(k) August 31, 2004 *Trademark
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 4 2004
Ms. Reneé Alfano Associate, Regulatory Affairs United States Surgical 150 Glover Avenue Norwalk, Connecticut 06856
Re: K042412
Trade/Device Name: Modified SPACEMAKER System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: September 2, 2004 Received: September 7, 2004
Dear Ms. Alfano:
:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your becamed the device is substantially equivalent (for the indications felerenced above and nave determined and marketed predicate devices marketed in interstate for use stated in the encrosule) to regars actment date of the Medical Device Art on the closed Force commerce prior to May 20, 1978, the excordance with the provisions of the Federal Food, Drug, devices that have been require approval of a premarket approval application (PMA). and Costlience Act (Act) that to not require to the general controls provisions of the Act. The 1 ou may, mercere, market the act include requirements for annual registration, listing of general controls provisions of the tice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (600 as controls. Existing major regulations affecting your device can inay be subject to sublications, Title 21, Parts 800 to 898. In addition, FDA may be found in the over ments concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I least be advised that I Drivesan that your device complies with other requirements of the Act that I DA has made a acterialized administered by other Federal agencies. You must of any I cacal statutes and registements, including, but not limited to: registration and listing (21 Comply with an the Fee Brequirements, while and manufacturing practice requirements as set CI K Fat 607); adoming (21 CFR Part 820); and if applicable, the electronic forth in the qualis) 350cc provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
## Page 2 - Ms. Reneé Alfano
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will unow your o begin mailing of substantial equivalence of your device to a legally premailted predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire specific as noliance at (301) 594-4659. Also, please note the regulation entitled, Connot the Office of Court Court Cation" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
for
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Attachment 2
## Indications for Use Statement
#### Device Name
Modified SPACEMAKER* System
#### Indications For Use
The Modified SPACEMAKER* System:
## SPACEMAKER* Structural Balloon Trocar
The SPACEMAKER* Structural Balloon Trocar is primarily indicated for patients undergoing laparoscopic surgical procedures requiring a sealed port of access and/or tissue retraction. This is also indicated in patients undergoing laparoscopic surgery requiring a sealed port of access and/or tissue separation in extraperitoneal procedures, such as in hernia repair, lymphadenectomy or bladder neck suspension procedures.
## SPACEMAKER* Blunt Tip Trocar
The SPACEMAKER* Blunt Tip Trocar is intended for use in establishing a port of access for insertion of endoscopic instruments into the abdominal cavity or extraperitoneal space in abdominal and extraperitoneal surgery.
#### SPACEMAKER* Dissection Balloon
The SPACEMAKER* Dissection Balloon is primarily indicated for patients undergoing laparoscopic surgery requiring tissue separation of the extraperitoneal space.
(Please do not write below this line - continue on another page if needed)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|----------------------------------------------------------------|
| Prescription Use:<br>(Per 21 CFR 801.109) | <div> <span style="text-decoration: overline;">✓</span> </div> |
| | OR Over-The-Counter Use: |
Miriam C. Provat
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| Special 510 (k)<br>August 31, 2004<br>*Trademark | Attachment 2.1 |
|--------------------------------------------------|----------------------------------------------------------|
| 510(k) Number | <span style="text-decoration: underline;">K042412</span> |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches