SPACEMAKER SURGICAL BALLOON DISSECTOR, SPACEMAKER II SURGICAL BALLOON DISSECTOR, SAPHTRAK

K973046 · General Surgical Innovations · GCJ · Jan 16, 1998 · Gastroenterology, Urology

Device Facts

Record IDK973046
Device NameSPACEMAKER SURGICAL BALLOON DISSECTOR, SPACEMAKER II SURGICAL BALLOON DISSECTOR, SAPHTRAK
ApplicantGeneral Surgical Innovations
Product CodeGCJ · Gastroenterology, Urology
Decision DateJan 16, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The device is indicated for any area of the body where it is necessary to separate the layers of connective tissue for surgical access. The intended use of the Spacemaker® Surgical Balloon Dissectors is to create an operative space by dissecting layers of connective tissue along natural tissue planes of separation of the extraperitoneal (including preperitoneal and retroperitoneal), or subcutaneous extremity or thoracic space. The indications for use include dissection of tissue for the following surgical procedures: endoscopic; laparoscopic; general surgery; plastic and reconstructive surgery, including aesthetic surgery, vascular surgery for the exposure of superficial vessels of the upper and/or lower extremities (no deep structure would be targeted), including the saphenous vein and femoral vessels; thoracoscopic procedures involving exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels and other tissues of the chest wall; procedures in the extraperitoneal space involving dissection of tissue lavers allowing exposure of structures including: adrenals, bone, spine, lymph nodes, and kidneys. Examples of specific surgical procedures the Spacemaker® Surgical Balloon Dissectors may be used to facilitate tissue dissection include: fascial layers between the skin and peritoneum, hernia repair, varicocele dissection, sympathectomy, lymph node dissection, urethropexy, subfascial perforator vein ligation, access to the saphenous vein and femoral vessels, retroperitoneal access to the anterior spine, access for long bone plating. The cannula portion of the device is intended to provide access for surgical and diagnostic instruments for the duration of the operative procedure.

Device Story

Handheld, manually manipulated surgical balloon dissector; creates operative space by separating connective tissue layers along natural planes. Consists of balloon, center rod (solid or hollow for endoscopic visualization), and optional cannula. Used in OR by surgeons for procedures including hernia repair, spine access, and vessel exposure. Cannula provides port for surgical/diagnostic instruments. Benefits patient by facilitating minimally invasive access to anatomical structures.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Handheld manual balloon dissector; solid or hollow center rod; optional cannula for instrument access. Materials not specified. Non-powered, mechanical dissection.

Indications for Use

Indicated for patients requiring surgical access via connective tissue separation in extraperitoneal, subcutaneous extremity, or thoracic spaces. Used in endoscopic, laparoscopic, general, plastic/reconstructive, vascular, and thoracoscopic procedures. Contraindicated for deep structure targeting in vascular surgery.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K973046 # Appendix 4 ## 510(k) Summary JAN 16 1998 - Cathleen Mantor Contact Person: 1. General Surgical Innovations, Inc. 10460 Bubb Road Cupertino, CA 95014 (408) 863-2531 863-1100 (fax) - Date Summary Prepared : August 8, 1997 2. #### Trade Name(s): 3. Spacemaker® Surgical Balloon Dissector with and without Cannula Spacemaker® II Surgical Balloon Dissector with and without Cannula Spacemaker® Serial Surgical Balloon Dissector w/ or w/o Cannula & w/ or w/o Visualization #### 4. Common Name Surgical Balloon Dissector #### 5. Predicate devices K926010 Surgical Dissector with Cannula K944418 Spacemaker® and Spacemaker II Surgical Balloon Dissector and Cannula K951878 Spacemaker® and Spacemaker® II Surgical Balloon Dissector with and without a Canmila, Class II K952278 Reusable Spacemaker® Surgical Balloon Dissector with Cannula K962702 Spacemaker® Serial Surgical Balloon Dissector with or without Cannula and with or without Visualization K953377 Tapered Tip Balloon Dissection Cannula K935426 Preperitoneal Distention Balloon System K946002 Longitudinal Dissection Balloon System K954174 Duo Balloon System (1and 2) K964171 Vasos View Balloon Dissection System Device Description 6. The Spacemaker® Surgical Balloon Dissector devices are handheld, manually manipulated surgical balloon dissectors. The devices can be supplied with either a solid center rod or a hollow center rod. The hollow center rod allows for the insertion of an endoscopic visualization accessory, if desired. The devices can be supplied with or without a cannula. The cannula {2}------------------------------------------------ provides access for operative and diagnostic instrumentation for the surgical procedure. This submission seeks to gain clearance for the Spacemaker® Surgical Balloon Dissectors with all of the specific indications for use that have been delineated in previous submissions. #### 7. Intended Use The device is indicated for any area of the body where it is necessary to separate the layers of connective tissue for surgical access. The intended use of the Spacemaker® Surgical Balloon Dissectors is to create an operative space by dissecting layers of connective tissue along natural tissue planes of separation of the extraperitoneal (including preperitoneal and retroperitoneal), or subcutaneous extremity or thoracic space. The indications for use include dissection of tissue for the following surgical procedures: endoscopic; laparoscopic; general surgery; plastic and reconstructive surgery, including aesthetic surgery, vascular surgery for the exposure of superficial vessels of the upper and/or lower extremities (no deep structure would be targeted), including the saphenous vein and femoral vessels; thoracoscopic procedures involving exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels and other tissues of the chest wall; procedures in the extraperitoneal space involving dissection of tissue lavers allowing exposure of structures including: adrenals, bone, spine, lymph nodes, and kidneys. Examples of specific surgical procedures the Spacemaker® Surgical Balloon Dissectors may be used to facilitate tissue dissection include: fascial layers between the skin and peritoneum, hernia repair, varicocele dissection, sympathectomy, lymph node dissection, urethropexy, subfascial perforator vein ligation, access to the saphenous vein and femoral vessels, retroperitoneal access to the anterior spine, access for long bone plating. The cannula portion of the device is intended to provide access for surgical and diagnostic instruments for the duration of the operative procedure. #### 8. Comparison to Predicate Devices Substantial equivalence is based on the fact that all of the Spacemaker® Surgical Balloon Dissectors have the same principle of operation as the predicate devices (there are no changes to any of the Spacemaker® devices). There are no new indications for use. All of the indications for 22 {3}------------------------------------------------ use have been included in one or more of the predicate device submissions and including these indications for use in all of the Spacemaker® Surgical Balloon Dissectors labeling does not raise different questions regarding safety or efficacy or have any impact on safety and efficacy of the Spacemaker® Surgical Balloon Dissectors compared to the predicate devices. Therefore, it can be concluded that the Spacemaker® Surgical Balloon Dissectors with the inclusion of the indications for use (from the predicate devices) are substantially equivalent to the predicate devices. #### Testing in Support of Substantial Equivalence Determination 9. No functional bench testing was performed for this submission because no tests were necessary to substantiate the safety, performance or substantial equivalence. {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or other bird-like figure, with stylized wings and tail feathers. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 16 1998 Ms. Cathleen Mantor Regulatory Affairs/Ouality Assurance Manager General Surgical Innovations 10460 Bubb Road Cupertino, California 95014 Re: K973046 > Trade Name: Spacemaker® Surgical Balloon Dissector, Spacemaker® II Surgical Balloon Dissector Regulatory Class: II Product Code: GCJ Dated: October 31, 1997 Received: November 5, 1997 Dear Ms. Mantor: We have reviewed your Section 510(k) notification of intent to market the devices referenced above and we have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major regulations affecting your devices oan be found in the Code of Federal Regulations, Title 21: " Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification {5}------------------------------------------------ Page 2 - Ms. Mantor submissions does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market. If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### Statement of Indications for Use 8.1 **510(k) Number (If Known):** 973046 ### Device Name: Spacemaker® Surgical Balloon Dissector with and without Cannula Spacemaker® II Surgical Balloon Dissector with and without Cannula Spacemaker® Serial Surgical Balloon Dissector w/ or w/o Cannula & w/ or w/o Visualization Indications For Use: The device is indicated for any area of the body where it is necessary to separate the layers of connective tissue for surgical access. The intended use of the Spacemaker® Surgical Balloon Dissectors is to create a operative space by dissecting layers of connective tissue along natural tissue planes of separation of the extraperitoneal (including preperitoneal), or subcutaneous extremity or thoracic space. The indications for use include dissection of tissue for the following surgical procedures: endoscopic; laparoscopic; general surgery; plastic and reconstructive surgery, including aesthetic surgery; vascular surgery for the exposure of superficial vessels of the upper and/or lower extremities (no deep structure would be targeted), including the saphenous vein and femoral vessels; thoracoscopic procedures involving exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels and other tissues of the chest wall; procedures in the extraperitoneal space involving dissection of tissue layers allowing exposure of structures including: adrenals, bone, spine, lymph nodes, and kidneys. Examples of specific surgical procedures the Spacemaker® Surgical Balloon Dissectors may be used to facilitate tissue dissection include: fascial lavers between the skin and peritoneum. hernia repair, varicocele dissection, sympathectomy, lymph node dissection, urethropexy, subfascial perforator vein ligation, access to the saphenous vein and femoral vessels, retroperitoneal access to the anterior spine, access for long bone plating. The cannula portion of the device is intended to provide access for surgical and diagnostic. ..... - - -- --instruments for the duration of the operative procedure. Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use X OR Over-The-Counter Use (Per 21 CFR 801.109) (Optional Format 1-2-96) 5 (Division Sion Off) Dive General Restorative Devices 510(k) Number K973046
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...