GE HEALTHCARE SENOGRAPHE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
P990066 · GE Healthcare · MUE · Jan 28, 2000 · Radiology
Device Facts
Record ID
P990066
Device Name
GE HEALTHCARE SENOGRAPHE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Applicant
GE Healthcare
Product Code
MUE · Radiology
Decision Date
Jan 28, 2000
Decision
APRL
Regulation
21 CFR 892.1715
Device Class
Class 2
Intended Use
The Senographe 2000D system generates digital mammographic images that can be used for screening and in the diagnosis of breast cancer. The Senographe 2000D is intended to be used in the same clinical applications as traditional film-based mammographic systems.
Device Story
Senographe 2000D is a full-field digital mammography (FFDM) system; utilizes dual-track (Mo/Rh) X-ray tube and amorphous silicon flat-panel detector with cesium iodide scintillator. System includes acquisition workstation (AWS) for image acquisition, processing, and display; hard-copy laser-film writer for final interpretation. Operated by radiologists/technicians in clinical settings. AWS allows image manipulation (window/leveling, thickness equalization); images sent to archive or review workstation. Provides linear signal response and quantum-limited noise performance; improves DQE over screen-film systems. Benefits include wider exposure latitude, electronic archival, and reduced examination time. Clinical decision-making supported by digital images comparable to film-screen; enables detection of breast cancer abnormalities.
Clinical Evidence
Two prospective reader studies (N=646 and N=625) compared Senographe 2000D to SFM. Primary endpoints: recall rates, sensitivity, and ROC area. Results: Senographe 2000D recall rates were lower than SFM (p<0.001). Sensitivity was comparable (78% vs 74% in Study 1; 68% vs 70% in Study 2). ROC areas were statistically similar (e.g., 0.722 vs 0.723 in Study 2). Side-by-side analysis (40 cancer cases) showed non-inferiority in lesion conspicuity and tissue visibility.
Technological Characteristics
Dual-track (Mo/Rh) X-ray tube; amorphous silicon flat-panel detector with cesium iodide scintillator. Linear signal response; quantum-limited noise performance. Connectivity: AWS, review workstation, laser camera, CD-ROM, networking. Software includes thickness equalization algorithm. Dimensions: 19x23 cm paddle (AOP mode). Sterilization: N/A (surface disinfection).
Indications for Use
Indicated for screening and diagnostic mammography in women over 40 years of age. Contraindications: None known.
Regulatory Classification
Identification
A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.
Related Devices
K210151 — Digimamo D · Vmi Tecnologias Ltda · Mar 1, 2022
K110572 — MAMMODIAGNOST DR · Philips Medical Systems · Sep 23, 2011
K163328 — Planmed Clarity · Planmed OY · Dec 28, 2017
K120255 — PHILIPS MICRODOSE · Philips Digital Mammography Sweden AB · Apr 19, 2012
K121674 — ASPIRE HD PLUS, ASPIRE HD-S · Fujifilm Medical System U.S.A., Inc. · Sep 21, 2012
Submission Summary (Full Text)
{0}
P990066
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## I. GENERAL INFORMATION
| Device Generic Name | Digital Mammographic X-ray System |
| --- | --- |
| Device Trade Name | Senographe 2000D |
| Applicant's Name and Address | GE Medical Systems
3000 N. Grandview Blvd. W-709
Waukesha, WI 53188 |
| PMA Number: | P990066 |
| Date of Panel Recommendation: | December 16, 1999 |
| Date of Good Manufacturing Practices Inspection: | December 6-9, 1999 |
| Date of Notice of Approval to the Applicant: | January 28, 2000 |
## II. INDICATIONS FOR USE
The Senographe 2000D system generates digital mammographic images that can be used for screening and in the diagnosis of breast cancer. The Senographe 2000D is intended to be used in the same clinical applications as traditional film-based mammographic systems.
## III. DEVICE DESCRIPTION
The Senographe 2000D is equipped with a dual track X-ray tube (molybdenum/rhodium) and a digital detector. The digital detector is a flat panel of amorphous silicon on which the cesium iodide is deposited to maximize detection of X-rays. Positioning operations and X-ray exposure are controlled by the Control Panel, which also controls power to all parts of the Senographe 2000D system. The Senographe 2000D includes an acquisition work station (“AWS”) monitor, keyboard and mouse, computer, electronics, accessory storage, and uninterruptable power supply. The AWS is used for image acquisition, processing, and display. The AWS can also be used for database management and can send images to archive, review, or filming. The device utilizes a hard-copy laser-film writer for image presentation. The printer window widths and window level are set automatically or can be user adjusted prior to printing. Images are displayed per film 1x1.
Several options are available for use with the Senographe 2000D system. These options include a FFDM review workstation, a mass archiving system, a laser camera, networking capabilities and CD-ROM interchange media. The Senographe 2000D review workstation is a stand-alone workstation with its own dedicated computer and image database. The review workstation supports image display, manipulation, and film composing functions.
## IV. CONTRAINDICATIONS
None known
{1}
P990066
# V. WARNINGS and PRECAUTIONS
## For User
### Senographe 2000D
- For U.S. only, until such time as an FDA approved accreditation process for full-field digital mammography has been developed, the Senographe 2000D Full Field Digital Mammography System must only be used in MQSA screen-film accredited/certified facilities.
- The light box (on the AWS cart) must not be used for final interpretation of examinations. The ambient light conditions in the examination room, and the resulting light level of the light box, are incompatible with its use for final interpretations.
- The (AWS) workstation monitor must not be used for final interpretation of examinations. It is set up for optimum visualization with an ambient light level of 50 lux. Leaving the AWS light box illuminated without a film in place may degrade reviewing quality of images displayed on the monitor.
- Only images produced by GE-recommended laser cameras can be used for final interpretation of examinations. For compatible printers, see the latest product data sheets for this system, which you can obtain from your local sales representative.
- The three-section table is not designed to hold items in excess of 20 kg weight.
- Never make a clinical examination without either the Bucky or the Breast Holder or the Magnification stand fitted; the unprotected edge of the Digital Detector may damage sensitive skin when compression is applied.
- Breast compression of at least 3 daN (30 Newtons or 6.7 pounds) is essential in AOP mode.
- The AWS Cart is mounted on wheels so that it can be easily positioned for maximum convenience in the examination room. It should NOT be considered a “mobile” unit. Take great care if you must move it in the vicinity of any person or equipment. Do not move it during an examination.
- The monitor should be used in a suitably dark environment when reviewing a digital image. The optimum ambient light level is 50 lux.
- NEVER switch off at the Uninterruptible Power Supply (UPS) except in emergency (risk of data loss).
- To assure continued high level operation of the Senographe 2000D, the recommended quality control procedures should be followed.
- All measurement values refer to measurements made at the image plane. When magnification mode is used, the zoom factor is NOT taken into account. All processed images in Medical Application are in log format. All raw images in RWS are in linear format. The displayed measurements represent the lengths and areas of the graphic annotations made by the user, not the real length or surface area of the pathologies displayed on the screen.
- Annotations added by the operator on the Acquisition Workstation will be lost during image transfer to the Review Workstation.
- In AOP mode, only use a 19 x 23 cm paddle. The use of all other paddles is only permitted in Manual mode. In AOP mode, the automatic calculation of administered dose will only be accurate if the 19 x
8
{2}
P990066
23 cm paddle (specific to the Senographe 2000D) is used. Paddles that are NOT to be used in AOP mode are labeled with the following symbols:

- Markers larger than 2 mm² should not be present in the ROI. Large markers will affect the calculation of tissue density, which may lead to a degraded image.
- After exposure press (located on the right of the Control Console) for decompression if the automatic decompression is not set in the program.
- In the absence of the compression paddle, leave the space free between the bottom of the paddle arm and the top of the image receptor assembly.
- To avoid image loss, never touch the recordable surface of a recordable CD (CD-R). Handle the disk only by the outer edge. Do not place it face down on a hard surface. Fingerprints or scratches will make the disk unusable. Before usage, verify that CD-R surface has no visible scratches. If there are any scratches, do NOT use the CD-R.
## Senographe 2000D Review Workstation
- The Senographe 2000D Review Workstation has not been cleared by the FDA for final interpretation of examinations. Final interpretations should be done from hard copy films only.
- The monitor should be used in a sufficiently dark environment (optimum ambient light level of 50 lux) when reading a digital image.
- Should it ever be necessary to remove all power from the workstation (for maintenance, or to move the workstation), workstation database corruption may occur if the following procedure is not used: You MUST first bring the workstation to the Shutdown state as described in Section 3 of the Operator’s Manual. Once the system is completely shut down, switch off the equipment in this order: workstation computer unit; monitors; external CD-ROM device; external hard disk unit (option); filming interface unit (option); and keypad.
- When saving images on a CD, it is strongly recommended that no other operation should be performed. For a full CD-R the save operation can take up to 45 minutes.
## For Device
### Senographe 2000D
- If the digital detector casing is punctured, it must be removed by authorized GE Service personnel wearing protective gloves and dust masks; send the protective items for disposal along with the defective detector.
- Only Senographe 2000D recommended accessories should be used with this equipment. Failure to heed this warning may cause unexpected functions and possible data loss.
- Software programs other than those supplied by General Electric Medical Systems specifically for use with this system must NOT be loaded onto the system.
3
9
{3}
P990066
- The Interchange Media option is NOT recommended for permanent archiving. GE does not guarantee the suitability of the media for such purposes.
## Senographe 2000D Review Workstation
- The Interchange Media option is NOT recommended for permanent archiving. GE does not guarantee the suitability of the media for such purposes.
- The Review Workstation monitors are unshielded. Placing them in a high magnetic field (e.g., near an MR system magnet) can cause permanent damage, and may void the warranty.
## For Cleaning And Disinfection
### Senographe 2000D
- CIDEX (a cleaning solution) contains glutaraldehyde. Direct contact is corrosive to exposed tissue, causing eye damage and skin irritation/damage. Do not get into eyes, on skin or on clothing. Use in well ventilated area and store in closed containers.
- Adequate cleaning and disinfection is necessary to prevent disease transmission. Be sure to thoroughly clean and disinfect equipment surfaces that contact the patient and all equipment surfaces likely to become soiled during use.
- Improper cleaning methods or the use of certain cleaning and disinfecting agents can damage the equipment, cause poor imaging performance or increase the risk of electric shock. To avoid possible injury or equipment damage:
- Do not use harsh detergents, abrasive cleaners, high alcohol concentration or Methanol at any concentration. If skin preparations contain high alcohol concentrations, allow sufficient drying time before applying compression;
- Do not expose equipment parts to steam or high temperature sterilization;
- Never allow liquids to enter the internal parts of the equipment. Do not apply cleaning sprays or liquids directly to the equipment; always use a clean cloth dampened with the spray or liquid. If you become aware of liquid entry, disconnect the electrical supply and have the equipment checked by qualified service personnel before returning it to use.
- Always follow the germicide manufacturer’s instructions and precautions for mixing, storage, method of application, contact time, rinsing requirements, protective clothing, shelf life and disposal to help assure effective and safe use of the product.
## VI. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
No serious adverse events were reported for the patients enrolled in the clinical study. However excessive breast compression, excessive X-ray exposure, electric shock, infection and skin irritation, abrasion or puncture wounds are potential adverse effects of mammography.
10
{4}
P990066
# VII. ALTERNATIVE PRACTICES AND PROCEDURES
There are several methods available for screening and diagnosing cancer in the breast. These include the following: clinical breast examination; film-screen mammography; ultrasound examination; magnetic resonance imaging; and xeromammography. After an abnormality is detected, a biopsy may be performed to diagnose the cancer.
# VIII. MARKETING HISTORY
The Senographe 2000D is currently marketed in all Europeans countries, Canada, all Latin American countries (except Mexico), and all Asian countries. No Senographe 2000D has been withdrawn from marketing for reasons related to the safety or effectiveness of the device.
# IX. SUMMARY OF NONCLINICAL STUDIES
The sponsor performed several tests to document image and detector parameters and, in some cases, compare them to the results of screen film systems.
A. Comparison of Dynamic Range and Sensitometric Response for Screen Film Mammography (SFM) and Senographe 2000D

Figure 1. Response curve of the Digital Detector
5
{5}
P990066

Figure 2. Response curve for film screen
Contract
The solid-state detector exhibits a response that is linear over a range of 0 to $153\mathrm{mR}$ at the breast support surface. The range of exposure over which linear response is achieved is at least 3 times larger than the linear portion of the SFM sensitometric curve as seen in Figures 1 and 2.
## B. Comparison of Senographe 2000D and SFM DQE
### Definition of Detective Quantum Efficiency (DQE)
DQE measures the efficiency by which a system transfers both signal and noise. It is defined as the ratio of the squares of signal-to-noise ratios (SNRs). Since the output SNR is a measure of the displayed image quality and the input SNR can be related to patient dose, DQE can be considered as a ratio of image quality to patient dose. Doubling DQE means that Senographe 2000D has the same image quality at half the dose. Therefore Senographe 2000D yields a $40\%$ improvement in image quality.
$$
\mathrm{DQE} = \frac{\mathrm{SNR}^2 \text{ at detector output}}{\mathrm{SNR}^2 \text{ at detector input}} \propto \frac{\text{Image Quality}}{\text{Patient Dose}}
$$
where $SNR =$ signal-to-noise ratio
12
{6}
P990066
## Comparison
The sponsor measured the DQE for the Senograph 2000D and for a screen-film system. These measurements are shown in Figure 3.
Figure 3. DQE of the Senographe 2000D and a Screen-Film System

All Senographe 2000D DQE measurements shown here were made using 30 kVp, Rh anode track, Rh filter, and added filtration of 1.5 cm acrylic plus 33 mm Al. Exposures at the detector entrance plane are included with the data.
At detector exposure levels characteristic of those used in screen-film mammography (10 to 15 mR), the DQE of the Senographe 2000D exceeds that of common screen-film systems throughout the 0 to 5 lp/mm range. Even at 1 mR, the low-frequency DQE of the Senographe 2000D exceeds that of screen-film systems when used at ten times the exposure level. While the DQE of screen-film systems extends to higher spatial frequencies, film-grain noise and not quantum noise often dominate the output SNR. Because of the essentially linear response of the Senographe 2000D detector, quantum-limited performance and substantially constant DQE are achieved over a wide range of exposure levels.
## C. DQE versus Exposure Level - Quantum-Limited Performance
The detector of the Senographe 2000D not only exhibits a linear signal response over a wide exposure range, it also exhibits a quantum-limited noise response. Quantum-limited response is evidenced by image noise variance (the square of the standard deviation of the pixel values within a region of interest) that increases linearly with the exposure to the detector (Fig. 4).
13
{7}
P990066
Figure 4. Image Noise vs. Exposure for the Senographe 2000D

At low exposure levels, the plot of image noise variance vs. exposure commonly deviates from a straight line due to the influence of noise in the imaging system. For the Senographe 2000D, this deviation is not evident until the exposure at the detector is less than about 3 mR (Fig. 5).
Figure 5. Image Noise vs. Exposure for the Senographe 2000D

8
14
{8}
P990066
The combination of linear signal response and quantum-limited noise response leads to a DQE that is independent of exposure level over a broad range of exposures. Screen-film systems tend to have low DQE at high film densities (near the skin line) and at low film densities (in the image of glandular tissue or near the chest wall), the Senographe 2000D provides uniformly high DQE throughout the mammogram (Fig. 6). This provides a probability of detection of an abnormality that is less dependent on its location than is the case for screen-film detectors.

Figure 6. DQE vs. Exposure at Three Spatial Frequencies
DQE is essentially independent of exposure for exposures at the detector plane greater than about $3\mathrm{mR}$. Therefore, Senographe 2000D DQE has little dependence on the exposure level for the clinical exposure range.
## D. Summary of findings from the Preclinical Studies
The range of linear signal response for the Senographe 2000D is significantly better than that of screen film.
The detective quantum efficiency (DQE) for the Senographe 2000D is comparable to or higher than screen film up to its Nyquist frequency (5 lp/mm), though beyond that frequency the Senographe 2000D gives no meaningful information.
Below the Nyquist frequency, for exposures that are typical near the skin line or in a dense area of the breast, the Senographe 2000D has a DQE that is higher than that of screen-film by a factor of 2 to 10. Furthermore, at these spatial frequencies the Senographe 2000D has nearly constant DQE over the range of exposures commonly found in a mammogram, while that of screen-film varies by a factor of 3 to 5 between the mid- and high-density regions.
While the higher DQE for the Senographe 2000D at spatial frequencies less than or equal to the Nyquist frequency suggests the possibility that mammograms may be taken at a lower radiation dose, particularly in dense breasted women, the accuracy of mammographic interpretation at lower doses has not been studied to determine whether the absence of information above the Nyquist frequency is clinically important.
The combination of linear signal response and quantum-limited noise response leads to a DQE that is independent of exposure level over a broad range of exposures. Unlike screen-film systems, which tend to have low DQE at high film densities (near the skin line) and at low film densities (in the image of glandular tissue or near the chest wall), the Senographe 2000D provides uniformly high DQE throughout the mammogram. This provides a probability of detection of an abnormality that is less dependent on its
9
15
{9}
P990066
location than is the case for screen-film detectors. This was shown to be true by the side-by-side feature comparison (see section I of CLINICAL STUDIES).
The Senographe 2000D has wider exposure latitudes, higher DQE, allows for electronic archival transmission and image manipulation, and involves shorter examination time. The ability to manipulate images has the potential to eliminate the need for additional magnification of views. Similarly, windowing and leveling can compensate for exposure variations eliminating the need for repeated exposures or exposures adjusted to specifically view the skin line or the chest wall.
## X. SUMMARY OF CLINICAL STUDIES
## READER STUDY DATA
Four clinical sites enrolled a single cohort and the clinical data were used in two reading studies.
## A. Study Inclusion/Exclusion Criteria for Both Reader Studies
### 1. Inclusion Criteria
Women over 40 years of age who were entering a participating facility for diagnostic mammography due to a recommendation for diagnostic mammography based on screening mammography results; palpable findings by clinical breast exam; skin thickening; spontaneous nipple discharge; or focal unilateral breast pain were included in the study.
### 2. Exclusion Criteria
Women who were under the age of 40, pregnant or suspicious of being pregnant; had breast implants; had breasts too large to be adequately positioned on 24x30 cm screen-film image receptor; had non-focal or bilateral breast pain; or who were unable or unwilling to understand or execute the patient consent form were excluded from the study.
Entry bias was minimized by entering all women attending for diagnostic mammography and willing to consent to the study by both Senographe 2000D and SFM. Entry bias was further minimized by imaging all women entering the study with both SFM and Senographe 2000D. Even if a woman entered the study because of a positive screening mammogram, a second SFM as well as a Senographe 2000D was taken to minimize any possibility of entry bias.
## B. Study Population
Six hundred forty-one (641) patients were enrolled in the study at the four study sites. These women underwent both SFM and Senographe 2000D of one or both breasts.
To enrich the number of cancer cases in the study cohort, 21 additional cancer cases were obtained from a complementary study of Senographe 2000D currently being conducted at the University of Colorado Health Sciences Center (UCHSC) and University of Massachusetts Medical Center (UMMC) on a screening population using the GE-Senographe 2000D system (hereafter, “Army screening study”).
Informed consent was obtained from 21 patients enrolled in the Army screening study. This provided 19 additional cancer cases for Reader Study #1 (11 from UCHSC and 8 from UMMC) and 20 additional cancer cases for Reader Study #2 (12 from UCHSC and 8 from UMMC) from a total of approximately 4,000 women screened by both SFM and Senographe 2000D in the Army screening study up to the end of 1998. One (1) patient, however, had to be excluded from both reader studies because her Senographe 2000D images were not available for interpretation in the study. A second case had to be excluded from Reader Study #1 because the additional informed consent had not been obtained at the time of the conclusion of the reader study.
10
16
{10}
P990066
# 1. Reader Study #1 Cohort
For Reader Study #1, 16 subjects (14 diagnostic study non-cancer cases and 2 UCHSC-UMMC screening study cancer cases) could not be included in the study analysis for the following reasons:
- 5 diagnostic cohort patients did not meet the inclusion/exclusion criteria.
- 11 cases (9 diagnostic cohort and 2 screening cohort cases) could not be included because their SFM images (5) or Senographe 2000D images (6) were not available for interpretation during the study (i.e., SFM films checked out for patient care, additional consent form for screening study patients not obtained prior to conclusion of the reader study, or lost films).
Excluding these 16 women brought the total number of patients who contributed both SFM and Senographe 2000D images to the Reader Study #1 to 646, 641 enrolled in the diagnostic study plus 21 screening study cancer cases minus 16 excluded cases. No additional selection of study participants occurred. These 646 women are referred to as the “Reader Study #1 cohort”. The number of known cancers included in the Reader Study #1 cohort totaled 47 (28 cancer cases from the diagnostic cohort and 19 cancer cases from the Army screening study cohort). Of the 47 cancers, 22 were in the left breast, 25 were in the right breast. There were no known cases of bilateral breast cancer.
No serious adverse events occurred among the women in the study.
# 2. Reader Study #2 Cohort
The second reader study was performed on essentially the same set of acquired cases. The purpose of the second reader study was to reduce as many of the sources of variability as possible and to provide a balanced study design for efficient estimation of outcomes.
For Reader Study #2, 37 patients (32 non-cancer patients and 5 cancer patients) could not be included in the study analysis. The reasons for exclusion from the study analysis were as follows:
- 5 patients did not meet the inclusion/exclusion criteria.
- 13 patients could not be included because their SFM images (8) or Senographe 2000D images (5) were not available for interpretation during the study (i.e., SFM films checked out for patient care, lost films, etc.).
- 19 patients could not be included due to inconsistency in lesion markers between SFM and Senographe 2000D acquisitions.
Excluding these 37 cases brought the total number of patients who contributed both SFM and Senographe 2000D images to the Reader Study #2 to 625, 641 enrolled in the diagnostic study plus 21 Army screening study cancer cases minus 37 excluded cases. No additional selection of study participants occurred. These 625 women are referred to as the “Reader Study #2 cohort”. The number of known cancers included in the Reader Study #2 cohort totaled 44 (24 cancer cases from the diagnostic cohort and 20 cancer cases from the screening cohort). Of the 44 cancers, 19 were in the left breast, 25 were in the right breast. There were no known cases of bilateral breast cancer.
No serious adverse events occurred among the women in the study.
# C. Demographics
Table 1 describes the study cohort for Reader Study #1 and Reader Study #2 with respect to age, ethnicity, history of breast disease, and hormone replacement use.
17
{11}
P990066
Table 1: Age, Ethnicity, and History
| | | Study #1 | Study #2 |
| --- | --- | --- | --- |
| Age Data: | Mean: | 55 | 55 |
| | Median: | 56 | 53 |
| | Range: | 40-86 | 40-86 |
| Ethnicity: | White (non-Hispanic) | 85% | 87% |
| | African American | 8% | 6% |
| | Hispanic | 2% | 2% |
| | Asian | 1% | 1% |
| | Other | 2% | 2% |
| | Not reported | 3% | 3% |
| Reported history of breast disease | | 34% | 35% |
| Reported history of hormone replacement therapy | | 33% | 34% |
## IMAGE INTERPRETATION
### A. Reader Study #1
In the first reader study, both screen-film and digital mammograms were independently interpreted for each case by at least two MQSA-qualified interpreting physicians. To minimize recall bias, an interval of at least 30 days between interpretations of the same case on the two different modalities was used. To ensure that interpreting physicians had no prior knowledge of each case, both SFM and Senographe 2000D images were interpreted by primary readers located at a different site than the site of image acquisition.
No prior films, patient histories or any other demographic information accompanied the interpretation of either modality. These reading conditions are similar to those encountered in a screening environment. Original screen-film images included the patient name; digital images did not include the patient name, nor did they include any other identifier that could be linked to the screen-film images. Each interpreting physician read approximately one-half of the screen-film mammograms first and approximately one-half of the digital mammograms first.
ACR BIRADS categories shown in Table 2 were used to assess findings for each modality.
Table 2: ACR BIRADS Categories
| ACR BIRADS Category | Finding |
| --- | --- |
| 0 | Needs further evaluation |
| 1 | Normal |
| 2 | Abnormal – benign |
| 3 | Abnormal – probably benign |
| 4 | Suspicious for cancer |
| 5 | Highly suspicious for cancer |
ACR BIRADS categories 1 (normal), 2 (benign), and 3 (probably benign) with greater than 6 month follow-up were considered negative for breast cancer for both SFM and Senographe 2000D interpretations. ACR BIRADS categories 0 (needs further evaluation), 3 (probably benign) with less than or equal to 6 month follow-up, 4 (suspicious for cancer) and 5 (highly suspicious for cancer) were considered positive for breast cancer for both SFM and Senographe 2000D interpretations.
In an attempt to reduce interpretation variability, when the two independent primary interpretations for a given modality disagreed in terms of positive or negative for breast cancer, a third independent interpretation was performed by one of the secondary readers. The final interpretation for that modality
12
{12}
P990066
was that of the majority (2 out of 3). The same adjudication procedure was applied to each modality in cases where the primary readers for that modality disagreed. Reading of the second modality by each interpreting physician was spaced at least 30 days from the interpretation of the previous modality. If a third independent interpretation was required for both SFM and Senographe 2000D, those readings were also spaced at least 30 days apart. Data for both recall rates and sensitivities of the two modalities were analyzed comparing reader to reader (to eliminate inter-reader variability) and by comparing adjudicated (best 2 of 2 or 3) interpretations for each modality.
## B. Reader Study #2
The purpose of the second reader study was to reduce as many sources of study variability as possible and to provide a balanced design for efficient estimation of outcomes. To reduce reader variability, five MQSA-qualified interpreting physicians each read all Senographe 2000D and all SFM images included in the study cohort. Analysis compared each reader's interpretation of Senographe 2000D to their own interpretation of SFM, so that inter-reader variability could be eliminated. Additional quality control measures were taken to ensure that lesion markers, when present, were identical in both Senographe 2000D and SFM images and the printing of Senographe 2000D images was adequate. All interpreting physicians participating in Reader Study #2 received uniform instructions, as described below.
The reading team for this reader study consisted of five MQSA-qualified radiologists. None of these interpreting physicians had any involvement in image acquisition or in Reader Study #1. Each interpreting physician independently read all SFM and all Senographe 2000D cases, separated by at least 30 days. Each interpreting physician read half of the SFM cases first and half of the Senographe 2000D cases first.
Both SFM and Senographe 2000D images were presented without any history data or marks on the mammograms to each the five readers in random order. No prior films, patient histories or any other demographic information accompanied the interpretation of either modality. These reading conditions are similar to those encountered in a screening environment. All five readers interpreted all films with each modality (separated by at least 30 days), a total of 1250 assessments per reader.
ACR BIRADS categories shown in Table 5 were used to assess findings for each modality.¹ ACR BIRADS categories 1 (normal) and 2 (benign) were considered negative for breast cancer for both SFM and Senographe 2000D interpretations. ACR BIRADS categories 0 (needs further evaluation), 3 (probably benign), 4 (suspicious for cancer) and 5 (highly suspicious for cancer) were considered positive for breast cancer for both SFM and Senographe 2000D interpretations. In an actual screening situation or a purely screening study, the use of the BIRADS 3 category is relatively uncommon.
## C. Side-By-Side Feature Analysis Comparison Study
A side-by-side feature analysis comparison of lesions in Senographe 2000D and SFM images, was conducted again using hardcopy film output of Senographe 2000D images and original SFM films. This side-by-side feature analysis comparison included 40 cancer cases within the study population for which the previous reading study had been completed for all primary readers. Five experienced mammography radiologists (all of whom were primary readers in Reader Study #1) independently examined all films from each modality on these cases and ranked each view (MLO or CC) of each of the 40 cases in terms of 1) the conspicuity of cancers, 2) visibility (inclusion) of tissue at the chest wall, and 3) visibility of tissue at or near the skin line of the breast. Each criterion was scored for relative lesion conspicuity or tissue visibility between Senographe 2000D and SFM using an 11-point Likert scale. In addition, radiologists participating in the side-by-side analysis recorded the primary and, if relevant, the secondary signs of cancer for the four choices of calcifications, mass, focal asymmetry, or architectural distortion. Lesion conspicuity scores were analyzed separately for each lesion type.
19
{13}
P990066
# STAGE AND SIZE DISTRIBUTION OF CANCERS IN THE STUDY COHORT
The distribution of cancer stages in the study cohort for the Reader Study #1 is listed in Table 3. The table is further broken down into the stage distribution of cancers in the diagnostic cohort and the stage distribution of cancers in the screening study cohort. The Agency for Health Care Policy and Research (AHCPR) Clinical Practice Guidelines #13: Quality Determinants of Mammography lists the marker of a good mammography program that at least 50% of detected cancers are Stage 0 and 1 cancers. The study group satisfies that requirement by having 58% of the total number of cancers in the stage 0 and 1 category.
Table 3: Stage Distribution of Cancers in Reader Study #1
| Cancer Stage | 0
In-Situ | I | II | III | IV | Total |
| --- | --- | --- | --- | --- | --- | --- |
| Entire Study Cancer Cases | 12
26% | 15
32% | 15
32% | 3
6% | 2
4% | 47
100% |
| Diagnostic Cohort Cancer Cases | 7
25% | 8
29% | 9
32% | 3
11% | 1
3% | 28
100% |
| Screening Study Cancer Cases | 5
26% | 7
37% | 6
32% | 0
0% | 1
3% | 19
100% |
Table 4 gives the size distribution of cancer cases in Reader Study #1. The AHCPR guidelines list the marker of a good mammography program that at least 30% of detected cancers are less than or equal to 1 cm in size. The entire study cohort for Reader Study #1 had 38% of cancers less than or equal to 1 cm in size.
Table 4: Size Distribution of Cancers in Reader Study #1
| Tumor Size | All Patients (N=47) | Diagnostic Cohort Cancers Only (N=28) | Screening Cohort Cancers Only (N=19) |
| --- | --- | --- | --- |
| Number of Cancer Cases
≤1 cm | 18
38% | 8
29% | 10
53% |
| Number of Cancer Cases >1 cm | 29
62% | 20
71% | 9
47% |
The stage distribution of cancers in the study cohort for Reader Study #2 is listed in Table 5. The sum of Stage 0 and 1 cancers is 61%, well above the AHCPR guideline recommended limit of 50%.
Table 5: Stage Distribution of Cancers in Reader Study #2
| Cancer Stage | 0
In-Situ | I | II | III | IV | Total |
| --- | --- | --- | --- | --- | --- | --- |
| Entire Study Cancer Cases | 12
27% | 15
34% | 13
29% | 2
5% | 2
5% | 44
100% |
| Diagnostic Cohort Cancer Cases | 7
29% | 7
29% | 7
29% | 2
8% | 1
4% | 24
100% |
| Screening Study Cancer Cases | 5
25% | 8
40% | 6
30% | 0
0% | 1
5% | 20
100% |
Table 6 gives the size distribution of cancer cases in Reader Study #2, showing that the study distribution of minimal cancers (≤1 cm in size) of 43% is well within the AHCPR recommended limit of 30%.
14
2°
{14}
P990066
Table 6: Size Distribution of Cancers in Reader Study #2
| Tumor Size | All Patients (N=44) | Diagnostic Cohort Cancers Only (N=24) | Screening Study Cohort Only (N=20) |
| --- | --- | --- | --- |
| Number of Cancer Cases
≤1 cm | 19 | 8 | 11 |
| | 43% | 33% | 55% |
| Number of Cancer Cases > 1 cm | 25 | 16 | 9 |
| | 57% | 67% | 45% |
The results, along with the absence of any case selection in the methods of accrual of study subjects, indicate that the subjects were representative of the general mammography population in the United States.
## RESULTS AND DATA ANALYSES
Results are presented below for Reader Study #1, Reader Study #2, and the Side-by-Side Feature Analysis Study. Three separate analyses were performed for each reader study to evaluate the non-inferiority of Senographe 2000D to SFM. Those analyses were: comparison of recall rates for Senographe 2000D and SFM; a comparison of sensitivities of Senographe 2000D and SFM for cancers found in the cohort evaluated; and receiver operating characteristics (ROC) analysis.
## A. Test 1: Recall Rate Comparison Data Analysis
A primary concern in the clinical use of Senographe 2000D is whether the modality results in a significant excess of false positives (patients recommended for additional imaging studies or for biopsy) compared to SFM. The entire study cohort and the non-cancer cohort were analyzed to compare recall rates using a paired comparison analysis of cases based on comparisons of overall recall rates and their Senographe 2000D and SFM final assessments. The determination of non-inferiority was based on the cases that disagreed concerning recall (i.e., the final interpretation using one modality was positive, the final interpretation using the other modality was negative).
### 1. Recall Rate Results – Reader Study #1
Primary interpretation recall rate results for SFM and Digital are provided in Table 7.
Table 7: Study #1 - Primary Interpretation Recall Rate Results
Reader A vs. Reader A Plus Reader B vs. Reader B
| | All Cases | Non-Cancer Cases Only |
| --- | --- | --- |
| Number of Readings of Each Modality | 1,292 | 1198 |
| Digital Recall Rate | 45% | 43% |
| Screen-film Recall Rate | 47% | 45% |
| p-value for Rejection of the Null Hypothesis | <0.001 | <0.001 |
The recall rate test assumes a tolerable difference of $\delta = 0.05$ between the recall rate of Senographe 2000D and SFM (that is, that the recall rate of Senographe 2000D is not greater than $5\%$ more than the recall rate of SFM). The observed results indicate that the recall rate of Senographe 2000D was lower than the recall rate of SFM. Moreover, the recall rate of Senographe 2000D was sufficiently lower than that of SFM (and the number of study participants was sufficiently large) to reject the
21
{15}
P990066
null hypothesis with a high degree of statistical significance (p<0.001), whether all cases or only non-cancer cases were considered.
These unadjusted combined p-values were confirmed by modeling the differences in recall rates using a generalization of the standard linear model. This enables the analysis of data generated from several sources of variation, multiple random effects and repeated measures. The estimation was made using the SAS/STAT PROC Mixed software. The estimated mean difference between the recall rates was –1.78%, with a 95% confidence interval of [-5.66, 2.10]. The null hypothesis (digital recall rate greater by more than 5% than film-screen recall rate) can thus be rejected and the corresponding adjusted p-value is less than 0.001. The value of π was also estimated and tested using Generalized Estimating Equations (GEE models) with PRO GENMOD. The estimated value was 0.5322, with 0.4167 as the lower value of the 95% confidence interval. This result allows also to reject the null hypothesis and the corresponding adjusted p-value 0.029% is larger as the size of the sample is limited to the discordant pairs.
Table 8 shows the adjusted p-values for both models applied to all cases and non-cancer cases only. The null hypothesis can be rejected in all cases.
Table 8: Study #1 – Recall Rate Results Adjusted For Multiple Readers Reading Same Cases
| | All Cases | Non-Cancer Cases Only |
| --- | --- | --- |
| PROC MIXED Results | | |
| Estimated Digital Recall Rate (Mean) | 49.22% | 47.13% |
| Estimated Screen-film Recall Rate (Mean) | 51.00% | 49.31% |
| Difference in Means (SE) | -1.78 (1.94) | -2.17 (2.01) |
| 95% Confidence Intervals(CI) | [-5.66, 2.10] | [-6.11, 1.77] |
| To Reject Null Hypothesis the Difference in Means Must Be | <5% | <5% |
| Adjusted p-value for Rejection of the Null Hypothesis using PROC MIXED | <0.001 | <0.001 |
| PROC GENMOD Results | | |
| Model Estimated □ value | 0.5322 | 0.5241 |
| Lower Bound of 95% CI for □ | 0.4167 | 0.4048 |
| To Pass Recall Rate Test, □ Must Be ≥ | 0.40 | 0.40 |
| Adjusted p-value for Rejection of the Null Hypothesis using GENMOD | 0.0294 | 0.0433 |
Note: Test in PROC MIXED uses all observations. GENMOD uses only the discordant pairs, thus has a much smaller sample size.
## 2. Adjudicated Results for Recall Rate in Study #1
Comparing Senographe 2000D and SFM interpretations for adjudicated reading results for non-cancer cases only yields the results shown in Table 9.
16
22
{16}
P990066
Table 9: Study #1 - Adjudicated Mammography Recall Rate Results
| Entire Cohort Included | All Cases | Non-Cancer Cases Only |
| --- | --- | --- |
| Number of Readings of Each Modality | 646 | 599 |
| Digital Recall Rate | 44% | 41% |
| Screen-film Recall Rate | 46% | 42% |
| p-value for Rejection of the Null Hypothesis | <0.001 | <0.001 |
The recall rate test again assumes a tolerable difference of $\delta = 0.05$ between the recall rate of Senographe 2000D and SFM (that is, that the recall rate of Senographe 2000D is not more than $5\%$ greater than the recall rate of SFM). The observed results based on adjudicated readings indicate that the recall rate of Senographe 2000D is lower than the recall rate of SFM. Moreover, the recall rate of Senographe 2000D is sufficiently lower than that of SFM (and the number of study participants is sufficiently large) to reject the null hypothesis with a high degree of statistical significance $(p < 0.001)$, whether the adjudicated readings for all cases or only non-cancer cases are considered.
The recall rate test results indicate, with a high degree of statistical confidence, that Senographe 2000D does not recall a higher proportion of women than SFM. This conclusion holds whether we compare primary reader interpretations of the two modalities (which includes only intra-reader variability along with possible positioning and modality differences) or adjudicated interpretations of the two modalities (which also includes inter-reader variability).
## 3. Reader Study #2 Recall Rate Results
The recall rate comparison for Reader Study #2 used the same null hypothesis and the same methods of comparison as that given above for Reader Study #1. Comparing Senographe 2000D and SFM interpretations by readers for all cases, and for non-cancer cases only, yields the results in Table 10.
Table 10: Study #2 - Interpretation Recall Rate Results
| | All Cases | Non-Cancer Cases Only |
| --- | --- | --- |
| Number of Readings of Each Modality | 3125 | 2905 |
| Digital Recall Rate | 47% | 45% |
| Screen-film Recall Rate | 49% | 47% |
| Unadjusted p-value for Rejection of the Null Hypothesis | <0.001 | <0.001 |
The recall rate test assumes a tolerable difference of $\delta = 0.05$ between the recall rate of Senographe 2000D and SFM (that is, that the recall rate of Senographe 2000D is not greater than $5\%$ more than the recall rate of SFM). The observed results indicate that the recall rate of Senographe 2000D was lower than the recall rate of SFM. Moreover, the recall rate of Senographe 2000D was sufficiently lower than that of SFM (and the number of study participants was sufficiently large) to reject the
23
{17}
P990066
null hypothesis with a high degree of statistical significance (unadjusted p<0.001), whether all cases or only non-cancer cases were considered.
These unadjusted combined p-values were confirmed using a generalized linear model PROC MIXED and Table 11 shows the adjusted p-values (adjusted for multiple readers reading the same cases) obtained by modeling the differences in recall rates using a generalization of the standard linear model as previously described in the Study #1. This enables the analysis of data generated from several sources of variation, multiple random effects and repeated measures.
Table 11: Study #2 – Recall Rate Results Adjusted for Multiple Readers Reading Same Cases
| | All Cases | Non-Cancer Cases Only |
| --- | --- | --- |
| PROC MIXED Results | | |
| Estimated Digital Recall Rate (Mean) | 46.98% | 45.37% |
| Estimated Screen-film Recall Rate (Mean) | 48.83% | 47.26% |
| Difference in Means (SE) | 1.86 (1.22) | 1.89 (1.26) |
| 95% Confidence Intervals(CI) | [-4.25, 0.54] | [-4.36, 0.58] |
| To Reject Null Hypothesis the Difference in Means Must Be | <5% | <5% |
| Adjusted p-value for Rejection of the Null Hypothesis using PROC MIXED | <0.001 | <0.001 |
| PROC GENMOD Results | | |
| Model Estimated □ value | 0.6737 | 0.4782 |
| Lower Bound of 95% CI for □ | 0.4164 | 0.4200 |
| To Pass Recall Rate Test, □ Must Be ≥ | 0.412 | 0.411 |
| Adjusted p-value for Rejection of the Null Hypothesis using GENMOD | 0.0382 | 0.0286 |
Note: Test in PROC MIXED uses all observations. GENMOD uses only the discordant pairs, thus has a much smaller sample size.
The difference in means of recall rates = 1.86 with a corresponding 95% confidence interval [-4.25, 0.54] allows rejection of the null hypothesis with an adjusted p-value <0.001. Similarly, the lower bound of 5% confidence interval for π is estimated to be 0.4164 which allows rejection of the null hypothesis with an adjusted p-value = 0.0382.
## B. Test 2: Sensitivity Comparison
Every effort was made to determine breast cancers occurring among the study cohort, by obtaining biopsy results in all cases going to biopsy and through follow-up at each acquisition site. All confirmed breast cancers found by any methods among the study cohort were used in this analysis. A comparison was made of the fraction of total cancers among the study cohort that were detected by Senographe 2000D and SFM, giving an estimated sensitivity for each modality. The fraction of cases correctly classified by a modality (true positives among the cancer cases by that modality divided by all known cancers) was compared using an exact binomial calculation, taking into account the fact that the true positive fractions were correlated.
The null hypothesis was that Senographe 2000D was inferior to SFM in terms of sensitivity by having a true positive fraction at least 10% lower than that of SFM. The rejection of the null hypothesis would provide further evidence for the non-inferiority of Senographe 2000D.
24
{18}
P990066
# 1. Sensitivity Results – Study #1
Comparing Senographe 2000D and SFM interpretations in terms of sensitivity to breast cancer, both by primary reader interpretations and by adjudicated readings, using the cancer case data yields the results in Table 12:
Table 12: Study #1 - Modality Sensitivity Results
| Cancer Case Readings Only | A vs. A
+ B vs. B | Adjudicated Readings |
| --- | --- | --- |
| Number of Readings of Each Modality | 94 | 47 |
| Senographe 2000D Sensitivity | 78% | 85% |
| SFM Sensitivity | 74% | 85% |
| p-value for Rejection of the Null Hypothesis | <0.01 | = 0.062 |
The alternative hypothesis of Senographe 2000D not having a significantly lower sensitivity than SFM assumed a tolerable difference of $\delta = 0.10$. In that case, the alternative hypothesis stated above would be supported by $\pi \geq 0.28$ for comparing primary readings and $\pi \geq 0.21$ for adjudicated readings.
The results stated above for primary readings yielded $\pi = b / (b + c) = 13 / (13 + 10) = 0.57$, which is consistent with the non-inferiority of Senographe 2000D in terms of sensitivity. This yielded a measured sensitivity of $78\%$ for Senographe 2000D and $74\%$ for SFM, supporting the claim that Senographe 2000D does not have a significantly lower sensitivity to breast cancer than SFM.
Statistical analysis of these measured results showed that primary readings are consistent with rejection of the null hypothesis (that Senographe 2000D has a lower sensitivity to breast cancer than SFM) with an unadjusted p-value of less than 0.01. Modeling the difference between the sensitivities led to a mean value of 3.19 with a $95\%$ confidence interval: [-8.94, 15.32] which allows to reject the null hypothesis with an adjusted p-value = 0.0166. Generalized estimating equation models for $\pi$ were unable to converge and estimate the off-diagonal elements due to the reasonably high agreement between the modalities by observer. For Reader A1, there were only 6 of 17 cancers discordant. For Reader B1 there were 2, for Reader C1 there were 7, for Reader D1 there were 4, and for Reader E1 there were 3 discordant cancers. Three of the readers found more cancers with Senographe 2000D and two found more with SFM.
Table 13 shows the estimated $3\%$ difference in sensitivity for Senographe 2000D and a p-value of 0.0166 for the rejection of the inferiority of Senographe 2000D by $10\%$ or more.
19
25
{19}
P990066
Table 13: Study #1 - Sensitivity Results Adjusted For Multiple Readers Reading Same Cases
| | Estimated Sensitivity |
| --- | --- |
| PROC MIXED Results | |
| Estimated Senographe 2000D Sensitivity (Mean) | 77.44% |
| Estimated SFM Sensitivity (Mean) | 74.25% |
| Difference in Means (SE) | 3.19 (6.19) |
| 95% Confidence Intervals(CI) | [-8.94, 15.32] |
| To Reject Null Hypothesis the Difference in Means Must Be | > -10% |
| Adjusted p-Value for Rejection of the Null Hypothesis using PROC MIXED | 0.0166 |
| PROC GENMOD Results | |
| Adjusted p-value for Rejection of the Null Hypothesis using GENMOD was not estimable due to small sample size on off-diagonal elements | |
Note: Test in PROC MIXED uses all observations. GENMOD uses only the discordant pairs, thus has a much smaller sample size.
## 2. Adjudicated Results for Sensitivity – Study #1
Within our study cohort using adjudicated readings, the measured sensitivity of both Senographe 2000D and SFM was 85%, supporting the claim that Senographe 2000D does not have a significantly lower sensitivity to breast cancer than SFM. Statistical analysis of these adjudicated results showed that they are consistent with rejection of the null hypothesis (that Senographe 2000D has a lower sensitivity to breast cancer than SFM) with a p-value of 0.062. The small number of discordant breast cancers with adjudicated interpretations (a total of 8), means that there are only 9 possibilities for the discordant cases, summarized in Table 14 below. The table gives the p-values for each discrete possibility for b and c among 8 discordant cases, indicating that the p-values jump from 0.013 for 5 out of 8 correctly called positive by digital to 0.062 for 4 out of 8 correctly called positive by digital, to 0.23 for 3 correctly called positive by digital. Thus, a p-value of 0.062 with only 8 discordant findings is consistent with rejection of the null hypothesis (that Senographe 2000D has a lower sensitivity in detecting breast cancer than SFM).
20
{20}
P990066
Table14: Discrete p-values with 8 Discordant Cancer Cases
| b | C | p-value |
| --- | --- | --- |
| 8 | 0 | --- |
| 7 | 1 | .00012 |
| 6 | 2 | .0016 |
| 5 | 3 | .013 |
| 4 | 4 | .062 |
| 3 | 5 | .23 |
| 2 | 6 | .53 |
| 1 | 7 | .85 |
| 0 | 8 | 1.0 |
Restricting the sensitivity comparison in Reader Study #1 to Stage 0 and 1 cancers (Table 3), Senographe 2000D had a sensitivity of 85% compared to 74% for SFM. Restricting the comparison of sensitivity to minimal cancers (≤ 1 cm is maximum size) (Table 4), Senographe 2000D had a sensitivity of 83% compared to a sensitivity of 70% for SFM. These results indicate that Senographe 2000D was not shown to have equal sensitivity to SFM based on the detection of larger, later stage cancers; Senographe 2000D had higher sensitivity than SFM in the detection of smaller, earlier stage cancers.
## 3. Sensitivity Comparison Results Study #2
The sensitivity comparison for Reader Study #2 used the same null hypothesis and the same method of comparison as that given above for Reader Study #1. Comparing Senographe 2000D and SFM interpretations in terms of sensitivity to breast cancer using the cancer case data yields the results shown in Table 15.
Table 15: Study #2 - Modality Sensitivity Results
| Cancer Case Readings Only | |
| --- | --- |
| Number of Readings of Each Modality | 220 |
| Senographe 2000D Sensitivity | 68% |
| SFM Sensitivity | 70% |
| Unadjusted p-value for Rejection of the Null Hypothesis | 0.006 |
The alternative hypothesis of Senographe 2000D not having a significantly lower sensitivity than SFM assumed a tolerable difference of $\delta = 0.10$. In that case, the alternative hypothesis stated above would be supported by $\pi \geq 0.331$. The results stated above for primary readings yielded $\pi = b / (b + c) = 31 / (31 + 34) = 0.477$, which is consistent with the non-inferiority of Senographe 2000D in terms of sensitivity. This yielded a measured sensitivity of 68% for Senographe 2000D and 70% for SFM, supporting the claim that Senographe 2000D does not have a significantly lower sensitivity to breast cancer than SFM.
27
{21}
P990066
These unadjusted combined p-values were confirmed using a generalized linear model using PROC MIXED and PROC GENMOD available in the SAS/STAT Software package. These models enable the adjustment for reader and subjects, while estimating the off diagonal (non-concordant) results.
Table 16 shows the estimated sensitivities and 2-sided confidence intervals for the difference in sensitivity.
Table 16: Study #2 - Sensitivity Results Adjusted For Multiple Readers Reading Same Cases
| | Estimated Sensitivity |
| --- | --- |
| PROC MIXED Results | |
| Estimated Senographe 2000D Sensitivity (Mean) | 68.18% |
| Estimated SFM Sensitivity (Mean) | 69.55% |
| Difference in Means (SE) | -1.36 (4.39) |
| 95% Confidence Intervals(CI) | [-9.96, 7.24] |
| To Reject Null Hypothesis the Difference in Means Must Be | > -10% |
| Adjusted p-Value for Rejection of the Null Hypothesis using PROC MIXED | 0.0245 |
| PROC GENMOD Results | |
| Model Estimated □ value | 0.4636 |
| Lower Bound of 95% CI for π | 0.2994 |
| To Pass Alternative Hypothesis Sensitivity Criterion, π Must Be Greater Than: | 0.331 |
| Adjusted p-value for Rejection of the Null Hypothesis using GENMOD | 0.0963 |
Note: Test in PROC MIXED uses all observations. GENMOD uses only the discordant pairs, thus has a much smaller sample size.
It is important to emphasize that the PROC MIXED test uses all observations. GENMOD uses only discordant pairs. In this particular study, the number of discordant pairs was very low. Thus, the GENMOD analysis has a much smaller sample size. More importantly, the PROC MIXED had a larger sample size and thus led to more statistically meaningful results. The difference in means of sensitivities: -1.36 with a corresponding 95% confidence interval [-9.96, 7.24] allows rejection of the null hypothesis with an adjusted p-value = 0.0245. The estimated value for π is 0.4636, with a lower bound of the 95% confidence interval equal to 0.2994. This value is not sufficient to reject the null hypothesis. This result is driven by the reduction of the sample size.
Considering ACR BIRADS category 3 cases as negative shifts the SFM call of one reader's assessment of one case from positive to negative, lowering the overall sensitivity of SFM to 69%, compared to 68% for Senographe 2000D. None of the statistical analysis conclusions with regard to sensitivity are changed by shifting BIRADS category 3 cases from positive to negative.
Restricting the sensitivity comparison in Reader Study #2 to Stage 0 and 1 cancers (Table 5), Senographe 2000D had a sensitivity of 65% compared to 64% for SFM. Restricting the comparison of sensitivity to minimal cancers (≤ 1 cm is maximum size) (Table 6), Senographe 2000D had a sensitivity of 63% compared to 57% for SFM. These results indicate that Senographe 2000D was not shown equal in sensitivity to SFM based on the detection of larger, later stage cancers. Senographe 2000D had slightly higher sensitivity than SFM in the detection of smaller, earlier stage cancers.
22
{22}
P990066
## C. Test 3: ROC Analysis
The independent interpretation of Senographe 2000D and SFM, along with the collection of probabilities of cancer from each interpretation given a BIRADS code of 0, 3, 4 or 5 (on a 0-100% scale), allowed the construction of ROC curves for each modality.
ROC curves have been constructed using probabilities of cancer based on Hanley-McNeil ROC analysis methods²,³ under three different conditions: for all individual primary readers in the study (other than one primary reader who interpreted only 9 non-cancer cases), the collective results for all primary readers for each case, and for adjudicated readings based on the mean probability of cancer for the two readers making the adjudicated decision.
Since a probability of cancer was estimated for each breast, for all patients having both breasts imaged, the maximum probability of the two non-cancer breasts was used; for breasts with cancer, the probability of breast cancer specific to the breast with cancer was used. The ROC curves for each case, along with the integrated areas under the ROC curves using the Hanley-McNeil method (which is consistent with a linear interpolation between ROC data points and usually results in areas lower than the binormal ROC curves because fewer assumptions are made) are presented below.
## 1. ROC Results – Study #1
Figure 7 shows ROC results combining all initial readings (corresponding to A vs. A plus B vs. B, where A and B represent the two primary readers of Senographe 2000D and SFM for each case, regardless of the actual reader). Figure 8 shows ROC results for the adjudicated readings, averaging the probabilities of cancer for the two readers that determined the adjudicated result (positive or negative). ROC curves based on BIRADS categories rather than probabilities of cancer gave similar results.

Figure 7: ROC Results for All Primary Readers Combined in Reader Study #1
29
{23}
P990066

Figure 8: Adjudicated ROC Results in Reader Study#1
The null hypothesis for ROC areas is analogous to the null hypothesis for recall rate and sensitivity test. The null hypothesis is that the ROC area for Senographe 2000D is significantly lower than the ROC area of SFM (by more than 0.10). In each case in which collective ROC analysis results were compared for the two modalities, the alternative hypothesis was easily met. Moreover, in each case of collective ROC analysis, statistical tests based on a z-statistic indicated that the null hypothesis could be rejected with reasonable statistical significance. These ROC analysis results and the p-values associated with rejection of the null hypothesis in each case are summarized in Table 17.
Table 17: Study #1 - Summary of ROC Results and Associated p-values
| ROC Curve Comparison | A_{SFM} | A_{Senographe 2000D} | SE_{SFM} | SE_{Senographe 2000D} | Adjusted Correlation r | p-value |
| --- | --- | --- | --- | --- | --- | --- |
| All primary readings Combined | 0.735 | 0.726 | 0.032 | 0.030 | 0.49 | 0.0015 |
| Adjudicated readings, taking mean cancer probability of readers | 0.801 | 0.779 | 0.037 | 0.038 | 0.52 | 0.017 |
Notes: $SE_{SFM} =$ standard error of the SFM ROC curve area
$SE_{Senographe 2000D} =$ standard error of the Senographe 2000D ROC curve area
$r =$ the correlation between ROC curve areas for Senographe 2000D and SFM (ref: Hanley and McNeil)$^{2,3}$
Recognizing that there are correlated observations in the case of all primary readings combined, an attempt was made to analyze for the distinction between ROC curves that accounts for the overlap in cases interpreted. In this first study, the concern for potential bias or additional information constrained the design of the readings (i.e., each reader did not read the same cases). This led to overlap in cases read that was unbalanced among readers. To attempt to estimate an adjusted standard error of the area under the ROC curve adjusted for some correlation, we created a dataset of first readers and second readers. Using this dataset, we performed an ROC Analysis using the Multi-Reader, Multiple Case Software LABMRMC available from the University of Chicago. The results from this analysis are shown in Table18.
30
{24}
P990066
Table 18: Study #1 - Summary of ROC Results
| ROC Curve Comparison | A_{SFM} | A_{Senegraphe 2000D} | SE_{SFM} | SE_{Senegraphe 2000D} | Mean Difference (95% Confidence Interval) |
| --- | --- | --- | --- | --- | --- |
| First Reader | 0.774 | 0.771 | 0.043 | 0.041 | Mean=0.004 (SE=0.0538)
95% CI= [-0.1017,0.1096] |
| Second Reader | 0.705 | 0.693 | 0.050 | 0.048 | Mean=0.012 (SE=0.0419)
95% CI= [-0.0699,0.0946] |
| Estimated Difference | 0.741 | 0.733 | 0.049 | 0.047 | Mean=0.0082 (0.0362) |
| Confidence Interval | | | | | 95% CI=[-0.0629, 0.0792] |
Notes: $SE_{SFM}$ = standard error of the SFM ROC curve area<br/>SE<sub>Senegraphe 2000D</sub> = standard error of the Senographe 2000D ROC curve area
The lower bound of the one sided confidence interval for testing the null hypothesis is -0.0513 significantly rejecting the inferiority hypothesis (p<.003).
These ROC results indicate that collective ROC curve areas are similar for the two modalities regardless of the method of analysis. Senographe 2000D had ROC curve areas well within 0.1 of SFM areas for all composite results. Analysis of the statistical significance of differences between ROC curve areas demonstrated that the null hypothesis could be rejected with statistical power (at least p< 0.02) in each case. Thus, ROC analysis results support the conclusion that Senographe 2000D is not inferior to SFM in the detection of breast cancer.
## 2. ROC Results – Study #2
ROC curves for Reader Study #2 have been constructed using probabilities of cancer (on a 0-100% scale) based on Hanley-McNeil ROC analysis methods$^{2,3}$: for all individual readers in the study, and the collective results for all readers for each case.
Since a probability of cancer was estimated for each breast, this ROC analysis included each breast as a “case”. Therefore, 44 “cases” had cancer and 953 “cases” (non-cancerous breasts) were without cancer. The ROC curves for all 5 readers combined are presented in Figure 3.

Figure 9: ROC Curves for All 5 Readers Combined in Reader Study #2
{25}
P990066
Table 19 summarizes ROC curve area results for each reader and all readers combined.
Table 19: Study #2 - Summary of ROC Analyses by Reader and For All Readers Combined and Associated P-values
| Reader | A_{x} Senographe 2000D | A_{z} SFM | SE_{Senographe 2000D} On A_{x} | SE_{Senographe 2000D} On A_{z} | Adjusted Correlation (Hanley, McNeil) | H_{0}: Inferiority P-Value |
| --- | --- | --- | --- | --- | --- | --- |
| A2 | .729 | .685 | .044 | .047 | 0.60 | 0.0002 |
| B2 | .720 | .771 | .046 | .042 | 0.51 | 0.132 |
| C2 | .729 | .768 | .045 | .041 | 0.56 | 0.066 |
| D2 | .720 | .738 | .046 | .047 | 0.46 | 0.043 |
| E2 | .729 | .708 | .046 | .047 | 0.62 | 0.0014 |
| Total | .722 | .723 | .020 | .020 | .56 | (unadjusted) <0.0001 |
The null hypothesis for ROC areas is analogous to the null hypothesis for recall rate and sensitivity test. The null hypothesis is that the ROC area for Senographe 2000D is significantly lower than the ROC area of SFM (by more than 0.1). The p-values listed in the last column of Table 20 indicate the confidence level for rejection of the null hypothesis. Rejection of the null hypothesis was indicated with three of the five individual readers and by the combined results of all 5 readers with high statistical significance (unadjusted p<0.0001).
## 3. Study #2 – ROC Curves Adjusted for Correlation of Observations
Recognizing that there are correlated observations, we analyzed ROC curves accounting for the overlap in patients and estimated a standard error of the area under the ROC curve adjusted for this correlation. We performed these ROC analyses using the Multi-Reader, Multiple Case Software LABMRMC available from the University of Chicago. We utilized all readings (breast by breast) and have checked the independence of the results by breast. There was virtually no correlation between the probability of cancer between the breast. In each modality-reader combination, the correlation never exceeded 0.06 and was generally 0.03 or lower. The effect of including all breasts was also checked by randomly selecting one breast per women (except when a cancer was present) and these results were virtually identical to those with the breast by breast analysis. This is likely due to the fact that the majority of breasts utilized had concordant probabilities of cancer of zero. Thus, a breast by breast analysis is reported. The results from this analysis are shown in Table 20.
32
{26}
P990066
Table 20: Study #2 - Summary of ROC Results and Associated p-values
| ROC Curve Comparison | A_{SFM} | A_{Senographe 2000D} | SE_{SFM} | SE_{Senographe 2000D} | Mean Difference (95% Confidence Interval) |
| --- | --- | --- | --- | --- | --- |
| Reader #A2 | 0.785 | 0.797 | 0.043 | 0.042 | Mean=-.011 (SE=.0469)
95% CI= (-0.1032,0.0809) |
| Reader #B2 | 0.686 | 0.739 | 0.051 | 0.043 | Mean=-.053 (SE=.0502)
95% CI= (-0.1517,0.0454) |
| Reader #C2 | 0.756 | 0.756 | 0.057 | 0.056 | Mean=-.000 (SE=.0646)
95% CI= (-0.1268,0.1266) |
| Reader #D2 | 0.794 | 0.749 | 0.037 | 0.048 | Mean=0.047 (SE=.0588)
95% CI= (-0.0686,0.1621) |
| Reader #E2 | 0.796 | 0.731 | 0.040 | 0.053 | Mean=0.063 (SE=0.0623)
95% CI= (-0.0592,0.1853) |
| Overall Results | 0.767 | 0.758 | 0.0368 | 0.0344 | Mean=0.009 (0.0324) |
| Confidence Interval | | | | | 95% CI=(-0.0545, 0.0727) |
Notes: $SE_{SFM}$ = standard error of the SFM ROC curve area<br/>SE<sub>Senographe 2000D</sub> = standard error of the Senographe 2000D ROC curve area
These ROC results indicate that collective ROC curve areas are within 0.01 of each other for the two modalities. Senographe 2000D had ROC curve areas well within 0.1 of SFM areas for composite results. Analysis of the statistical significance of differences between ROC curve areas demonstrated that the null hypothesis could be rejected and the confidence interval for the difference in areas shows that the ROC curve areas of the two modalities are not different. Thus, ROC analysis results support the conclusion that Senographe 2000D is not inferior to SFM in the detection of breast cancer.
## C. Side-by-Side Feature Analysis
The criterion for case inclusion in the side-by-side analysis was all cancer cases for which all primary readings had been completed by the time of the side-by-side analysis. This criterion was used to prevent the possibility that readers in the first reader study might see a case in the side-by-side analysis and then be asked to perform an independent interpretation on that case later for the recall rate/sensitivity study. Therefore, 40 cancer cases of the total 47 identified in the study cohort were used for the side-by-side comparison. All comparisons involved original SFM (except for two cases for which only copy films of the SFM were available) and Senographe 2000D images printed on film.
All 40 cases had MLO comparisons of Senographe 2000D and SFM and 39 had CC comparisons (one case had an exaggerated CC supplied on SFM instead of a routine CC, so the CC comparison was eliminated prior to scoring by reviewers). This resulted in 40 MLO views and 39 CC views, or a total of 79 views, that were each scored independently by each of the 5 primary readers.
All side-by-side comparisons were performed by placing both SFM and Senographe 2000D on the same mammography alternator or bank of viewboxes. Film masking was used for each comparison. A mammography hot light was available at each reading station for each reader. Each case was scored independently by 5 board-certified, MQSA-qualified radiologists. Prior to the side-by-side comparison, the cancerous area was marked in each view on each modality. Each radiologist was asked to note the type of finding, including all that applied, selecting from the following list: mass, calcifications, architectural distortion, and focal asymmetry. Each view (MLO and CC) was compared separately using three different evaluation criteria: lesion conspicuity, inclusion of tissue along the chest wall, and visibility of tissue at or near the skin line of the breast.
33
{27}
P990066
The analysis was done by averaging the score in each of the three categories across readers for all views (5 readers x 79 views = 395 scores). An overall mean value was calculated across all readers and views. Scores for each reader were averaged across all views, giving a mean value for that reader. Scores for each view were averaged across all readers giving a mean value for that view. Table 21 shows the overall mean and the range of mean values for readers and for views.
The average score for lesion conspicuity, $S_C$, was used to represent the lesion conspicuity result. As stated in the study protocol, the null hypothesis was that the average score $S_C$, averaged over all cancer cases for the 5 readers, is significantly worse for Senographe 2000D than for SFM.
$H_0: S_C > 6$, implying that visibility is significantly inferior on Senographe 2000D
The alternative hypothesis was:
$H_0: S_C \leq 6$, implying non-inferiority of Senographe 2000D.
The results of the side-by-side analysis are shown in Table 21. The overall mean is averaged over all readers and all views. The mean range by reader shows the range of mean scores across the five readers, averaged over all cases and views for each reader. The mean range by view shows the range of mean scores across all views, averaged over the five readers. The test requirement was that the overall mean score in each test category be less than 6.0.
Table 21: Summary of Side-by-Side Feature Analysis
| Test | Overall Mean | Mean Range (by Reader) | Mean Range (by View) | Meets Test Requirement? |
| --- | --- | --- | --- | --- |
| Lesion Conspicuity | 5.17 | 4.91-5.34 | 0-9.8 | Yes |
| Tissue Visibility at Chest Wall | 5.21 | 4.80-5.42 | 3.4-8.4 | Yes |
| Tissue Visibility at Skin Line | 2.95 | 0.39-4.30 | 1.7-5.2 | Yes |
The mean score for lesion conspicuity, $S_C = 5.17$, reflects the non-inferiority of Senographe 2000D for lesion conspicuity and the ability to detect low-contrast lesions. Averaged over all 79 views, the range of $S_C$ for the 5 readers was 4.91 to 5.34. Averaged over all five readers, the range of $S_C$ for the 79 cancer case views was 0 to 9.8. T-tests indicate that the mean lesion conspicuity score of 5.17 is consistent with rejection of the null hypothesis (that $S_C > 6$) with a high degree of statistical significance ($p < .007$).
The mean score for visibility of tissue at the chest wall, $S_{CW} = 5.21$, reflects the non-inferiority of Senographe 2000D for inclusion of tissue at the chest wall. Averaged over the 79 views, the range of $S_{CW}$ for the 5 consistent readers was from 4.80 to 5.42. Averaged over the 5 readers, the range of $S_{CW}$ for the 79 views was 3.4 to 8.4. T-tests indicate that the mean chest wall visibility score of 5.21 is consistent with rejection of the null hypothesis (that $S_{CW} > 6$) with a high degree of statistical significance ($p < .001$).
SAS PRO mixed was again utilized to adjust for multiple readers assessing the same cases. The results are confirmed by this analysis, not only overall but for each radiologist separately.
The mean score for tissue visibility at the skin line of the breast was $S_{SL} = 2.95$, reflecting the non-inferiority of Senographe 2000D for inclusion of tissue at the skin line. This score in favor of Senographe 2000D was due to the use of a thickness equalization algorithm applied to all images of all cases included in the recall-rate, sensitivity, and side-by-side comparisons. Averaged over the 79 cases, the range of $S_{CW}$ for the 5 readers was from 0.39 To 4.30. Averaged over the five readers, the range of $S_{CW}$ for the 79 views was 1.7 to 5.2. T-tests indicate that the mean skin line visibility score of 2.95 is consistent with rejection of the null hypothesis (that $S_{SL} > 6$) with a high degree of statistical significance ($p < 0.00001$).
34
{28}
P990066
Table 22 summarizes mean lesion conspicuity scores and the score range (across views, averaged over the 5 readers) for each specific lesion type. The five interpreting physicians performing the side-by-side analysis were asked to specify the primary and secondary indicators of the lesion from the four categories listed in Table 22. The total number of views listed exceeds 79 because of multiple listings of lesion type for some cancers. The score range (averaged over all 5 reviewers) for each lesion type indicates that reviewers used a wide range of lesion conspicuity scores in each lesion category. The mean lesion conspicuity scores, however, are similar to the overall mean lesion conspicuity score of 5.17. There is no trend toward a particular lesion type having lower conspicuity on Senographe 2000D than on SFM.
Table 22: Summary of Side-by-Side Lesion Conspicuity for Specific Lesion Types Within the 40 Cancer Cases
| Lesion Type | Number of Views | Reader Score | |
| --- | --- | --- | --- |
| | | Mean | Range |
| Calcifications Present | 42 | 5.21 | 0-8.0 |
| Masses Present | 53 | 5.17 | 1.8-9.8 |
| Architectural Distortion Present | 34 | 5.24 | 1.8-9.8 |
| Focal Asymmetry Present | 27 | 5.26 | 3.2-9.8 |
| Calcifications Primary | 24 | 5.33 | 0-8.0 |
| Masses Primary | 51 | 5.10 | 1.8-9.8 |
| Architectural Distortion Primary | 4 | 5.05 | 4.2-6.2 |
Note: No cases listed focal asymmetry as the primary lesion type.
These results demonstrate that Senographe 2000D is substantially equivalent (i.e., non-inferior) to SFM in lesion conspicuity, chest wall visibility, and visibility of tissue at the skin line.
## STUDY CONCLUSIONS
Three different reader studies have been conducted to test the non-inferiority of Senographe 2000D relative to SFM. Two reader studies were analyzed to compare ROC results, sensitivities, and recall rates for the two modalities. A side-by-side feature analysis was performed to compare lesion conspicuity (including analysis of conspicuity for specific lesion types), visibility of tissue at the chest wall, and visibility of tissue at the skin line.
The following are the study results:
- The ROC curve areas for the Senographe 2000D and SFM systems are virtually identical.
- The sensitivity analysis demonstrated that the Senographe 2000D has a sensitivity comparable to that of SFM in the screening and detection of breast cancer.
- The specificity analysis demonstrated that the Senographe 2000D system results in fewer women being recalled than SFM.
- The side-by-side feature comparison data demonstrated that the Senographe 2000D system allows better visibility of tissue at the skin line than SFM and that the Senographe 2000D system is comparable to SFM for lesion conspicuity and visibility of tissue at the chest wall.
35
{29}
P990066
## XI. CONCLUSIONS DRAWN FROM NONCLINICAL AND CLINICAL STUDIES
The results of the nonclinical and clinical studies conducted by the sponsor and described above provide a reasonable assurance of the safety and effectiveness of GE Medical Systems’ full-field digital mammography for screening and diagnostic breast imaging. These findings therefore support FDA approval of the GE Senographe 2000D Full-Field Digital Mammography system for clinical use in screening and diagnostic mammography.
## XII. PANEL RECOMMENDATION
A meeting of the FDA Radiological Devices Panel took place on December 16, 1999 to review the sponsor’s submission, PMA P990066, as amended. The panel voted for approval and made the following recommendations:
1. The sponsor should deploy a soft copy workstation to serve as an adjunct to hard copy.
2. FDA should expedite approval of the soft copy modality.
3. Amend the labeling to include an executive summary section in the front of the labeling that emphasizes the difference between the study population and true screening and diagnostic populations.
The sponsor responded with revised labeling to meet condition 3) above. The other two conditions, 1) and 2), are advice to the sponsor and to FDA and will be considered.
## XIII. FDA DECISION
CDRH concurred with the Radiological Devices Panel recommendation of December 16, 1999. The applicant’s manufacturing facility was inspected on December 6-9, 1999 and was found to be in compliance with the Quality System regulations. FDA issued an approval order on January 28, 2000.
## XIV. APPROVAL SPECIFICATIONS
Directions for use: See the attached labeling
The sale, distribution, and use of this device are restricted to prescription use in accordance with 21 CFR 801.109 within the meaning of section 520(e) of the Federal Food, Drug, and Cosmetic Act (the act) under the authority of section 515(d)(1)(B)(ii) of the act. Until such time as an FDA approved accreditation process for full-field digital mammography has been developed the Senographe 2000D Full Field Digital Mammography System must only be sold to screen-film accredited/certified facilities. FDA has also determined that, to ensure the safe and effective use of the device, the device is further restricted within the meaning of section 520(e) under the authority of section 515(d)(1)(B)(ii), (1) insofar as the labeling specify the requirements that apply to the training of practitioners who may use the device as approved in this order and (2) insofar as the sale, distribution, and use must not violate sections 502(q) and (r) of the act.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions and Adverse Events in the attached labeling.
30 36
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.